Microscopy ListServer Archives  


File Requested = 0701.txt
Retrival Software Version=NJZ07060908

From: zaluzec-at-microscopy.com
Date: Mon, 1 Jan 2007 10:48:51 -0600
Subject: [Microscopy] Adminsitrivia: 2006 Listserver Archives on-line

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Happy New Year Colleagues

Welcome to the 14th year of the Microscopy Listserver operation.

The complete searchable Microscopy Listserver archive for 2006
(~11.8 Mb) is now on-line at:

http://www.microscopy.com

Last year we delivered nearly 4.8 Million messages to subscribers
around the world (~198.4 Gbits of traffic). I hate to say it but as
in preceeding years spam attacks continue to increase. In 2006,
the custom filters on the Listserver blocked 127,504 suspect postings for
an average of 349.3 suspect spam messages/day. Just imagine
what that would have done to your in-boxes as well as to the utility of
this server.

As you probably realize, a infinitesimal number of those suspect Emails
were actually from subscribers. The major problem continues to
be hidden attachments.

Please if your message is rejected, read the entire rejection message for
the reason and remember check the settings on your Email client program
to insure that it is set to send only plain text messages. Most Email clients
send formatted text as a hidden attachment to your mail, and this will be
stopped by the Email filters. As always, there are no exceptions allowed.
If you can't change the settings use the WWW based posting form at
http://www.microscopy.com, this will strip any attachments
from your message.

Cheers,

Nestor
Your Friendly Neighborhood SysOp

==============================Original Headers==============================
9, 11 -- From zaluzec-at-microscopy.com Mon Jan 1 10:48:51 2007
9, 11 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
9, 11 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l01GmpmK003674
9, 11 -- for {microscopy-at-microscopy.com} ; Mon, 1 Jan 2007 10:48:51 -0600
9, 11 -- Mime-Version: 1.0
9, 11 -- Message-Id: {p06110400c1bee40981cb-at-[206.69.208.22]}
9, 11 -- Date: Mon, 1 Jan 2007 10:48:50 -0600
9, 11 -- To: microscopy-at-microscopy.com
9, 11 -- From: "Nestor J. Zaluzec" {zaluzec-at-microscopy.com}
9, 11 -- Subject: Adminsitrivia: 2006 Listserver Archives on-line
9, 11 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: microscopytoday-at-tampabay.rr.com
Date: Mon, 1 Jan 2007 11:32:58 -0600
Subject: [Microscopy] Microscopy Today January 2007 Table of Contents

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Listers,

Here is the January 2007 Microscopy Today table of contents. I will
close the subscription list for this issue on Monday, Jan. 8th, 2007.

Microscopists in North America and MSA members anywhere can subscribe
for free. Anyone else may subscribe for US$50 per year (to PARTIALLY
cover postage). All subscriptions at http://www.microscopy-today.com
Thank you.

Ron Anderson, Editor
================================

Why Penguin Beaks are Sexy!
Stephen W. Carmichael, ,Mayo Clinic

A ‘Different’ Kind of Microscopy
Fred Schamber and Kai van Beek, Aspex Corporation

Microwave Myths and Tissue Processing
Phillip McArdle, Energy Beam Sciences, East Granby, CT

Recent Developments in CrossBeam® Technology A. Thesen, H. Hoffmeister,
M. Schumann, P. Gnauck, Carl Zeiss SMT Oberkochen, Germany

Heated-Tip AFM: Applications in Nanocomposite Polymer Membranes and
Energetic Materials
Jason P. Killgore1, William King2, Kevin Kjoller3 and René M. Overney1,
1 U. of Washington, Seattle, WA, 2 U. of Illinois at Urbana-Champaign,
IL, 3 Anasys Instruments, Santa Barbara, CA

Automated S/TEM Sample Preparation for Semiconductor Process Support
Greg Cuti* and Taha Jabbar**, *Sela USA, Inc. Sunnyvale, CA, and
**Athenian Institute, Danville, CA

Serial Sectioning via Microtomy (or, How To Get Over 100 Consecutive
Serial Sections On One TEM Gird)
David Elliott, University of Arizona, Tucson, AZ

A Combined In-situ and Electron Tomography Holder for (S)TEM
C. Mitterbauer*, N.D. Browning*,** and P. V. Deshmukh***,*U. of
California, Davis, CA, **Lawrence Livermore National Lab., CA, ***E.A.
Fischione Instruments, Inc., Export, PA

Infrared Laser Confocal Microscopy: Fast, Flexible, Cost-Effective
Inspection, and Metrology Tool for Microelectronic Manufacturing
David Rideout, Olympus Micro-Imaging Orangeburg, NY

New Approaches to Managing, Marketing, and Money for Maintaining a Core
Facility (D. Sherman, Organizer)
Part Ia: Case Study: Strategic plan for an EM Facility
Elaine Humphrey

Part Ib: How to Make a Business Plan for Facility Maintenance and Growth
Donald A. Blewett, Purdue University

A Note on Storing and Testing Gold Conjugates
Jan Leunissen, Aurion, Costerweg, The Netherlands

Cross-sectional TEM Sample Preparation for Nanowires or Porous Films
Grown on a Substrate
Chengyu Song, NCEM, Lawrence Berkeley National Laboratory

New & Interesting at Cell Biology and Industry News

Netnotes
SAMPLE PREPARATION – buffers for fixation
SAMPLE PREPARATION – high pH buffer for fixation
SAMPLE PREPARATION – cells grown on collagen gels
SAMPLE PREPARATION - preservation of microbes in external mucous
SAMPLE PREPARATION – charcoal and wood
SAMPLE PREPARATION - embedding wood
SAMPLE PREPARATION - propylene oxide vs. acetone
SAMPLE PREPARATION - LR White flat embedding
SAMPLE PREPARATION - LR White contrast
SAMPLE PREPARATION – previously frozen tissues
SAMPLE PREPARATION - embedding acrylamide gel
SAMPLE PREPARATION – oil-in-water nanoemulsion
SAMPLE PREPARATION - cross section of multilayers on stainless substrate
SAMPLE PREPARATION – negative staining with ammonium molybdate
LM - dark field microscopy
LM - phase contrast vs. dark field
TEM – digital cameras
Crystallographic indices
EBSD - sharpness
EBIC vs. SE

Index of Advertisers


==============================Original Headers==============================
21, 18 -- From microscopytoday-at-tampabay.rr.com Mon Jan 1 11:32:58 2007
21, 18 -- Received: from ms-smtp-01.tampabay.rr.com (ms-smtp-01.tampabay.rr.com [65.32.5.131])
21, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l01HWvZp014982
21, 18 -- for {Microscopy-at-Microscopy.Com} ; Mon, 1 Jan 2007 11:32:57 -0600
21, 18 -- Received: from [127.0.0.1] (rrcs-24-73-73-214.se.biz.rr.com [24.73.73.214])
21, 18 -- by ms-smtp-01.tampabay.rr.com (8.13.6/8.13.6) with ESMTP id l01HWrsE002406;
21, 18 -- Mon, 1 Jan 2007 12:32:55 -0500 (EST)
21, 18 -- Message-ID: {459945C2.1020703-at-tampabay.rr.com}
21, 18 -- Date: Mon, 01 Jan 2007 12:32:50 -0500
21, 18 -- From: Microscopy Today {microscopytoday-at-tampabay.rr.com}
21, 18 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
21, 18 -- MIME-Version: 1.0
21, 18 -- To: Listserver {Microscopy-at-Microscopy.Com} ,
21, 18 -- Confocal Listserver {confocal-at-listserv.buffalo.edu}
21, 18 -- Subject: Microscopy Today January 2007 Table of Contents
21, 18 -- Content-Type: text/plain; charset=windows-1252; format=flowed
21, 18 -- Content-Transfer-Encoding: 8bit
21, 18 -- X-Virus-Scanned: Symantec AntiVirus Scan Engine
==============================End of - Headers==============================




From: carnahan-at-edison-labs.com
Date: Tue, 2 Jan 2007 08:28:38 -0600
Subject: [Microscopy] viaWWW: Tracor-Northern Xray detector

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both carnahan-at-edison-labs.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: carnahan-at-edison-labs.com
Name: Jim Carnahan

Organization: Edison Analytical Laboratories, Inc.

Title-Subject: [Filtered] Tracor-Northern Xray detector.

Question: The computer, including the 8" floppy drives, on our ancient Tracor-Northern (Noran) TN-5500 detector died some time ago but now we would like to rejuvenate the instrument. I recall some years ago a company was selling a PC upgrade that used the existing HV, pulse processor and other boards but with a modern PC interface and software. Does anyone know if that product still exists and know the source?

An alternative for us would be if anyone has written a Labview (National Instruments) driver for these old instruments that we could implement.

Thanks in advance,

Jim Carnahan

---------------------------------------------------------------------------

==============================Original Headers==============================
9, 12 -- From zaluzec-at-microscopy.com Tue Jan 2 08:28:38 2007
9, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
9, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l02ESaNV018922
9, 12 -- for {microscopy-at-microscopy.com} ; Tue, 2 Jan 2007 08:28:37 -0600
9, 12 -- Mime-Version: 1.0
9, 12 -- X-Sender: (Unverified)
9, 12 -- Message-Id: {p06110400c1c01c85214b-at-[206.69.208.22]}
9, 12 -- Date: Tue, 2 Jan 2007 08:28:36 -0600
9, 12 -- To: microscopy-at-microscopy.com
9, 12 -- From: carnahan-at-edison-labs.com (by way of MicroscopyListserver)
9, 12 -- Subject: viaWWW: Tracor-Northern Xray detector
9, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: kenconverse-at-qualityimages.biz
Date: Tue, 2 Jan 2007 09:41:15 -0600
Subject: [Microscopy] viaWWW: Tracor-Northern Xray detector

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Jim,
One company that several of my customers have been happy with is IXRF
http://www.ixrfsystems.com/
You can include a digital imaging option with it.
No financial interest just good experience with them.

Ken Converse
owner


QUALITY IMAGES
Servicing Scanning Electron Microscopes
Since 1981
474 So. Bridgton Rd.
Bridgton, ME 04009
207-647-4348
Fax 207-647-2688
kenconverse-at-qualityimages.biz
qualityimages.biz



-----Original Message-----
X-from: carnahan-at-edison-labs.com [mailto:carnahan-at-edison-labs.com]
Sent: Tuesday, January 02, 2007 9:32 AM
To: kenconverse-at-qualityimages.biz

This Question/Comment was submitted to the Microscopy Listserver using the
WWW based Form at
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying

please copy both carnahan-at-edison-labs.com as well as the MIcroscopy
Listserver
---------------------------------------------------------------------------

Email: carnahan-at-edison-labs.com
Name: Jim Carnahan

Organization: Edison Analytical Laboratories, Inc.

Title-Subject: [Filtered] Tracor-Northern Xray detector.

Question: The computer, including the 8" floppy drives, on our ancient
Tracor-Northern (Noran) TN-5500 detector died some time ago but now we would
like to rejuvenate the instrument. I recall some years ago a company was
selling a PC upgrade that used the existing HV, pulse processor and other
boards but with a modern PC interface and software. Does anyone know if
that product still exists and know the source?

An alternative for us would be if anyone has written a Labview (National
Instruments) driver for these old instruments that we could implement.

Thanks in advance,

Jim Carnahan

---------------------------------------------------------------------------

==============================Original Headers==============================
9, 12 -- From zaluzec-at-microscopy.com Tue Jan 2 08:28:38 2007 9, 12 --
Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
9, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l02ESaNV018922
9, 12 -- for {microscopy-at-microscopy.com} ; Tue, 2 Jan 2007 08:28:37
-0600
9, 12 -- Mime-Version: 1.0
9, 12 -- X-Sender: (Unverified)
9, 12 -- Message-Id: {p06110400c1c01c85214b-at-[206.69.208.22]}
9, 12 -- Date: Tue, 2 Jan 2007 08:28:36 -0600
9, 12 -- To: microscopy-at-microscopy.com
9, 12 -- From: carnahan-at-edison-labs.com (by way of MicroscopyListserver) 9,
12 -- Subject: viaWWW: Tracor-Northern Xray detector 9, 12 -- Content-Type:
text/plain; charset="us-ascii" ==============================End of -
Headers==============================





==============================Original Headers==============================
26, 28 -- From kenconverse-at-qualityimages.biz Tue Jan 2 09:41:15 2007
26, 28 -- Received: from dpmailmta02.doteasy.com (dpmailmta02.doteasy.com [65.61.209.81])
26, 28 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l02FfEuF031066
26, 28 -- for {Microscopy-at-MSA.Microscopy.Com} ; Tue, 2 Jan 2007 09:41:15 -0600
26, 28 -- Received: from qualityimages.biz (unverified [192.168.101.16])
26, 28 -- by dpmailmta02.doteasy.com (DEO) with ESMTP id 43262079-1814644
26, 28 -- for multiple; Tue, 02 Jan 2007 07:53:44 -0800
26, 28 -- Received: from Ken [76.179.237.214] by qualityimages.biz with ESMTP
26, 28 -- (SMTPD32-8.05) id ABD077C30114; Tue, 02 Jan 2007 07:35:44 -0800
26, 28 -- From: "Ken Converse" {kenconverse-at-qualityimages.biz}
26, 28 -- To: {carnahan-at-edison-labs.com} ,
26, 28 -- "MSA Listserver" {Microscopy-at-MSA.Microscopy.Com}
26, 28 -- Subject: RE: [Microscopy] viaWWW: Tracor-Northern Xray detector
26, 28 -- Date: Tue, 2 Jan 2007 10:37:25 -0500
26, 28 -- Message-ID: {00ec01c72e83$e9146590$6401a8c0-at-Ken}
26, 28 -- MIME-Version: 1.0
26, 28 -- Content-Type: text/plain;
26, 28 -- charset="us-ascii"
26, 28 -- X-Priority: 3 (Normal)
26, 28 -- X-MSMail-Priority: Normal
26, 28 -- X-Mailer: Microsoft Outlook, Build 10.0.6626
26, 28 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.2962
26, 28 -- Importance: Normal
26, 28 -- In-Reply-To: {200701021432.l02EWHfg023728-at-ns.microscopy.com}
26, 28 -- X-IP-stats: Incoming Last 0, First 96, in=2370213, out=0, spam=0 ip=192.168.101.16
26, 28 -- X-External-IP: 192.168.101.16
26, 28 -- Content-Transfer-Encoding: 8bit
26, 28 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l02FfEuF031066
==============================End of - Headers==============================




From: NWWhite-at-bwxt.com
Date: Tue, 2 Jan 2007 12:19:41 -0600
Subject: [Microscopy] Photoshop glitch - Followup

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Happy New Year!

Original problem: File type omitted when saving images(E.G. picture
instead of picture.tif). Also some menu windows had all the graphics,
but omitted the text describing the "radio buttons" function. No other
programs were malfunctioning. A photoshop reinstall did not fix the
issue.

The problem was finally traced to something odd in the Windows 2000 user
file. Have not figured out what or why, however. I installed an older
hard drive clone, updated the windows install and now PS works fine.

Plan to clone the now functional (was backup clone) drive onto the
problematic one. Looks like I will never find out what exactly happened
- but it works so...

Thanks for all the suggestions,

Woody White
BWXT Services



==============================Original Headers==============================
8, 26 -- From nwwhite-at-bwxt.com Tue Jan 2 12:19:19 2007
8, 26 -- Received: from bwxt.com (lynsmtp1.bwxt.com [131.184.146.8])
8, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l02IJJgY013057
8, 26 -- for {microscopy-at-msa.microscopy.com} ; Tue, 2 Jan 2007 12:19:19 -0600
8, 26 -- Received: from ([131.184.13.224])
8, 26 -- by lynsmtp1.bwxt.com with ESMTP id 5202488.3446807;
8, 26 -- Tue, 02 Jan 2007 13:18:50 -0500
8, 26 -- Received: from BWXSPO01.BWXS.BWXTECH.NET ([131.184.13.31]) by bwxslynbh01.BWXS.BWXTECH.NET with Microsoft SMTPSVC(6.0.3790.1830);
8, 26 -- Tue, 2 Jan 2007 13:18:50 -0500
8, 26 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
8, 26 -- Content-class: urn:content-classes:message
8, 26 -- MIME-Version: 1.0
8, 26 -- Content-Type: text/plain;
8, 26 -- charset="us-ascii"
8, 26 -- Subject: Photoshop glitch - Followup
8, 26 -- Date: Tue, 2 Jan 2007 13:18:49 -0500
8, 26 -- Message-ID: {360DDAABED436B49BA397357CCEE8BB703D0FE-at-BWXSPO01.BWXS.BWXTECH.NET}
8, 26 -- X-MS-Has-Attach:
8, 26 -- X-MS-TNEF-Correlator:
8, 26 -- Thread-Topic: Photoshop glitch - Followup
8, 26 -- Thread-Index: AccumnNqN1dvz9CRRFOy9vWqnW5ZCg==
8, 26 -- From: "White, Woody N." {NWWhite-at-bwxt.com}
8, 26 -- To: "MicroscopyListServer" {Microscopy-at-msa.microscopy.com}
8, 26 -- X-OriginalArrivalTime: 02 Jan 2007 18:18:50.0495 (UTC) FILETIME=[73BCE8F0:01C72E9A]
8, 26 -- Content-Transfer-Encoding: 8bit
8, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l02IJJgY013057
==============================End of - Headers==============================




From: pbozzano-at-cnea.gov.ar
Date: Tue, 2 Jan 2007 12:26:16 -0600
Subject: [Microscopy] EMS on line

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi listers!
I used to run the free EMS on line (http://cimesg1.epfl.ch/CIOL) to draw
SAD and Kikuchi patterns, (hkl) distances, indexing SAD patterns, etc.,
but I am having problems now. The legend "This calcul has failed for an
undefined reason" always appeared when I try to run a routine. Can
anybody help me with it?
Thanks in advanced!
Patricia



---------------------------------------------
Dra. Patricia B.Bozzano
Lab. de Microscopia Electrónica
Centro Atómico Constituyentes
Comisión Nacional de Energía Atómica
Buenos Aires, Argentina
pbozzano-at-cnea.gov.ar
Tel: 54 11 6772 7395
Fax: 54 11 6772 7362





==============================Original Headers==============================
7, 26 -- From pbozzano-at-cnea.gov.ar Tue Jan 2 12:26:16 2007
7, 26 -- Received: from cnea.gov.ar (cnea.gov.ar [168.96.65.229])
7, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l02IQEb8019562
7, 26 -- for {Microscopy-at-Microscopy.Com} ; Tue, 2 Jan 2007 12:26:15 -0600
7, 26 -- Received: from cnea-mail.cnea.gov.ar (cnea-mail.cnea.gov.ar [168.96.68.244])
7, 26 -- by cnea.gov.ar (8.11.2/8.11.2) with ESMTP id l02IRPh19720
7, 26 -- for {Microscopy-at-Microscopy.Com} ; Tue, 2 Jan 2007 15:27:25 -0300
7, 26 -- Received: from uam61 (uam61.cnea.gov.ar [168.96.65.126])
7, 26 -- by cnea-mail.cnea.gov.ar (8.12.10/8.12.10) with ESMTP id l02IU3Fn008495
7, 26 -- for {Microscopy-at-Microscopy.Com} ; Tue, 2 Jan 2007 15:30:03 -0300
7, 26 -- From: "Patricia Bozzano" {pbozzano-at-cnea.gov.ar}
7, 26 -- To: {Microscopy-at-Microscopy.Com}
7, 26 -- Subject: EMS on line
7, 26 -- Date: Tue, 2 Jan 2007 15:26:46 -0300
7, 26 -- Message-ID: {000001c72e9b$8fd4e050$7e4160a8-at-uam61}
7, 26 -- MIME-Version: 1.0
7, 26 -- Content-Type: text/plain;
7, 26 -- charset="iso-8859-1"
7, 26 -- X-Priority: 3 (Normal)
7, 26 -- X-MSMail-Priority: Normal
7, 26 -- X-Mailer: Microsoft Outlook, Build 10.0.2627
7, 26 -- Importance: Normal
7, 26 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.2180
7, 26 -- X-RAVMilter-Version: 8.4.4(snapshot 20030410) (cnea-mail)
7, 26 -- Content-Transfer-Encoding: 8bit
7, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l02IQEb8019562
==============================End of - Headers==============================




From: johnf-at-geology.wisc.edu
Date: Tue, 2 Jan 2007 13:20:16 -0600
Subject: [Microscopy] Re: viaWWW: Tracor-Northern Xray detector

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

}
} Email: carnahan-at-edison-labs.com
} Name: Jim Carnahan
}
} Organization: Edison Analytical Laboratories, Inc.
}
} Title-Subject: [Filtered] Tracor-Northern Xray detector.
}
} Question: The computer, including the 8" floppy drives, on our
} ancient Tracor-Northern (Noran) TN-5500 detector died some time ago
} but now we would like to rejuvenate the instrument. I recall some
} years ago a company was selling a PC upgrade that used the existing
} HV, pulse processor and other boards but with a modern PC interface
} and software. Does anyone know if that product still exists and
} know the source?
}
} An alternative for us would be if anyone has written a Labview
} (National Instruments) driver for these old instruments that we
} could implement.
}
} Thanks in advance,
}
} Jim Carnahan

Jim

One company that provides the service you mentioned is TN Analyzer,
www.tnas.net. (run by a former Tracor-Northern engineer). We upgraded
to their WinEDS system several years ago, at a very reasonable cost,
and are happy with the product.

John
--
========================================================
John Fournelle, Ph.D. office: (608) 262-7964 cell: (608) 438-7480
Cameron Electron Microprobe Lab lab: (608) 265-4798
Dept of Geology & Geophysics fax: (608) 262-0693
University of Wisconsin home: (608) 274-2245
1215 West Dayton St. email: johnf-at-geology.wisc.edu
Madison, WI 53706 amateur radio: WA3BTA
Personal http://www.geology.wisc.edu/~johnf/
Probe lab http://www.geology.wisc.edu/~johnf/sx51.html
Probe Sign Up Calender:
http://www.microscopy.wisc.edu/cgi-bin/calendar/microprobe/calendar.cgi

"The first rule of all intelligent tinkering is to save every cog and
wheel." -- Aldo Leopold

"For a successful technology, reality must take precedence over
public relations, for Nature cannot be fooled." -- Richard P.
Feynman

==============================Original Headers==============================
6, 26 -- From johnf-at-geology.wisc.edu Tue Jan 2 13:20:16 2007
6, 26 -- Received: from ice.geology.wisc.edu (ice.geology.wisc.edu [144.92.206.14])
6, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l02JKGwm003181
6, 26 -- for {microscopy-at-microscopy.com} ; Tue, 2 Jan 2007 13:20:16 -0600
6, 26 -- Received: from localhost (localhost [127.0.0.1])
6, 26 -- by localhost (Postfix) with ESMTP id 2D83320D1F;
6, 26 -- Tue, 2 Jan 2007 13:20:16 -0600 (CST)
6, 26 -- Received: from ice.geology.wisc.edu ([127.0.0.1])
6, 26 -- by localhost (geology.wisc.edu [127.0.0.1]) (amavisd-new, port 10024)
6, 26 -- with ESMTP id 09133-06; Tue, 2 Jan 2007 13:20:02 -0600 (CST)
6, 26 -- Received: from [144.92.206.57] (beamer.geology.wisc.edu [144.92.206.57])
6, 26 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
6, 26 -- (No client certificate requested)
6, 26 -- by ice.geology.wisc.edu (Postfix) with ESMTP id 23EA020D05;
6, 26 -- Tue, 2 Jan 2007 13:20:02 -0600 (CST)
6, 26 -- Mime-Version: 1.0
6, 26 -- Message-Id: {p06230908c1c05f72f433-at-[144.92.206.57]}
6, 26 -- In-Reply-To: {200701021437.l02EbDLL027026-at-ns.microscopy.com}
6, 26 -- References: {200701021437.l02EbDLL027026-at-ns.microscopy.com}
6, 26 -- Date: Tue, 2 Jan 2007 13:16:25 -0600
6, 26 -- To: carnahan-at-edison-labs.com
6, 26 -- From: John Fournelle {johnf-at-geology.wisc.edu}
6, 26 -- Subject: Re: [Microscopy] viaWWW: Tracor-Northern Xray detector
6, 26 -- Cc: microscopy-at-microscopy.com
6, 26 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
6, 26 -- X-Virus-Scanned: amavisd-new at geology.wisc.edu
==============================End of - Headers==============================




From: wesaia-at-iastate.edu
Date: Tue, 2 Jan 2007 15:30:13 -0600
Subject: [Microscopy] AskAMicroscopist: solid state versus a scintilater

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

We have both types of detector on one of our SEMs. While I can offer you some distinctions, others may have newer models and care to join the discussion.
 
Our SEM was originally supplied with a Robinson scintillator style detector. Its strengths were fast response for rapid scan rates up to and including TV-rate. It is good for general imaging and gives a nice topographic effect.
 
We later purchased a solid-sate detector for particular use for image analysis. Our application required lower voltage (6kV) than normal. We also required an even response across the field of view. For that, the solid-state detector was considerably better, but it has a slower response than the Robinson detector.
 
So the better detector will depend on your requirements. As it is, we run the solid state most of the time.
 
Warren Straszheim

________________________________________
X-from: wawennekes-at-woh.rr.com [mailto:wawennekes-at-woh.rr.com]
Sent: Sat 12/30/2006 8:59 AM
To: wesaia-at-iastate.edu

Below is the result of your feedback form (NJZFM-ultra-55).  It was submitted by  (wawennekes-at-woh.rr.com) from http://www.microscopy.com/Ask-A-Microscopist/Ask-A-Microscopist.html on Friday, December 29, 2006 at 20:19:28
---------------------------------------------------------------------------

Email: wawennekes-at-woh.rr.com
Name: Willem Wennekes

Organization: UES

Education: Graduate College

Location: Dayton, Ohio, USA

Question: What is the difference in performance between a solid state & a scintilater (Robinson) backscatter detector?  I am very interested in a side by side comparison.

---------------------------------------------------------------------------

==============================Original Headers==============================
7, 12 -- From zaluzec-at-ultra5.microscopy.com Sat Dec 30 08:58:18 2006
7, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
7, 12 --        by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id kBUEwHsU001371
7, 12 --        for {microscopy-at-microscopy.com} ; Sat, 30 Dec 2006 08:58:18 -0600
7, 12 -- Mime-Version: 1.0
7, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
7, 12 -- Message-Id: {p06110402c1bc2ec80375-at-[206.69.208.22]}
7, 12 -- Date: Sat, 30 Dec 2006 08:58:16 -0600
7, 12 -- To: microscopy-at-microscopy.com
7, 12 -- From: wawennekes-at-woh.rr.com (by way of Ask-A-Microscopist)
7, 12 -- Subject: AskAMicroscopist: solid state versus a scintilater
7, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================


==============================Original Headers==============================
13, 31 -- From wesaia-at-iastate.edu Tue Jan 2 15:30:12 2007
13, 31 -- Received: from mailhub-5.iastate.edu (mailhub-5.iastate.edu [129.186.140.15])
13, 31 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l02LUCwu017107
13, 31 -- for {Microscopy-at-msa.microscopy.com} ; Tue, 2 Jan 2007 15:30:12 -0600
13, 31 -- Received: from mailout-1.iastate.edu (mailout-1.iastate.edu [129.186.140.1])
13, 31 -- by mailhub-5.iastate.edu (8.12.11.20060614/8.12.10) with SMTP id l02LUBNi006001;
13, 31 -- Tue, 2 Jan 2007 15:30:11 -0600
13, 31 -- Received: from (owa.eng.iastate.edu [129.186.23.85]) by mailout-1.iastate.edu with smtp
13, 31 -- id 3fae_d4c74c54_9a7d_11db_92cf_00137253420a;
13, 31 -- Tue, 02 Jan 2007 16:25:16 +0000
13, 31 -- Received: from maire.eng.iastate.edu ([10.10.196.69]) by owa.eng.iastate.edu with Microsoft SMTPSVC(6.0.3790.1830);
13, 31 -- Tue, 2 Jan 2007 15:30:12 -0600
13, 31 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
13, 31 -- Content-class: urn:content-classes:message
13, 31 -- MIME-Version: 1.0
13, 31 -- Content-Type: text/plain;
13, 31 -- charset="iso-8859-1"
13, 31 -- Subject: RE: [Microscopy] AskAMicroscopist: solid state versus a scintilater
13, 31 -- Date: Tue, 2 Jan 2007 15:31:56 -0600
13, 31 -- Message-ID: {16A330AC32056A40B32842EC4BB8D727014B1DF8-at-maire.eng.iastate.edu}
13, 31 -- X-MS-Has-Attach:
13, 31 -- X-MS-TNEF-Correlator:
13, 31 -- Thread-Topic: [Microscopy] AskAMicroscopist: solid state versus a scintilater
13, 31 -- Thread-Index: AccsIye5TlCvghdhSD+O6ICCJH3GHgBRULo3
13, 31 -- References: {200612301459.kBUExmFU002731-at-ns.microscopy.com}
13, 31 -- From: "Straszheim, Warren E [CCE E]" {wesaia-at-iastate.edu}
13, 31 -- To: "MSA" {Microscopy-at-msa.microscopy.com}
13, 31 -- Cc: {wawennekes-at-woh.rr.com}
13, 31 -- X-OriginalArrivalTime: 02 Jan 2007 21:30:12.0459 (UTC) FILETIME=[2F8593B0:01C72EB5]
13, 31 -- Content-Transfer-Encoding: 8bit
13, 31 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l02LUCwu017107
==============================End of - Headers==============================




From: sales-at-tradezone.net
Date: Tue, 2 Jan 2007 19:57:39 -0600
Subject: [Microscopy] viaWWW: Tracor-Northern X-ray detector

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both sales-at-tradezone.net as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: sales-at-tradezone.net
Name: Michael S

Organization: TradeZone.net

Title-Subject: [Filtered] Tracor-Northern X-ray detector

Question: At one time Evex (www.evex.com) of Princeton, NJ supplied x-ray microanalysis systems that interfaced to older systems like your Tracor 5500.

FYI: Though we currently don't have an inventory of tracor parts, you may want to check back with us in a few weeks to see if that changes.


Michael S.
Used Laboratory Equipment Broker
sales-at-TradeZone.net

---------------------------------------------------------------------------

==============================Original Headers==============================
9, 12 -- From zaluzec-at-microscopy.com Tue Jan 2 19:57:39 2007
9, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
9, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l031vciY000739
9, 12 -- for {microscopy-at-microscopy.com} ; Tue, 2 Jan 2007 19:57:39 -0600
9, 12 -- Mime-Version: 1.0
9, 12 -- X-Sender: (Unverified)
9, 12 -- Message-Id: {p06110401c1c0bdf2f7c2-at-[206.69.208.22]}
9, 12 -- Date: Tue, 2 Jan 2007 19:57:37 -0600
9, 12 -- To: microscopy-at-microscopy.com
9, 12 -- From: sales-at-tradezone.net (by way of MicroscopyListserver)
9, 12 -- Subject: viaWWW: Tracor-Northern X-ray detector
9, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: Martin.Hoppe-at-leica-microsystems.com
Date: Wed, 3 Jan 2007 08:08:33 -0600
Subject: [Microscopy] viaWWW: Re: Understanding STED

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both Martin.Hoppe-at-leica-microsystems.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: Martin.Hoppe-at-leica-microsystems.com
Name: Martin Hoppe

Organization: Leica Microsystems

Title-Subject: [Filtered] Re: Understanding STED

Question: In order to increase resolution with STED, it is not necessary to have a diffraction-unlimited detection point spread function (PSF). In STED, the diffraction-unlimited sharp excitation PSF is superimposed by the larger diffraction-limited detection PSF. In order to separate (=resolve) 2 adjacent points or molecules, they can both be located within the larger detection PSF. In STED, these points are sequentially excited due to the sharp excitation PSF in conjunction with scanning the beam across the specimen.

Regards
Martin

--------------------------------------------------------
Martin Hoppe, Ph.D.
Global Market Manager
Advanced Fluorescence Systems
Leica Microsystems CMS GmbH
Am Friedensplatz 3
D 68165 Mannheim / Germany

Phone +49 (0)621-7028-0
Cell phone: +49 (0) 172-623-0409
Fax: + 49 (0) 621-7028-1180
Email. Martin.Hoppe-at-leica-microsystems.com



-----------------------------------------------------
Email: onkel98-at-yahoo.com
Name: pipetman

Title-Subject: [Filtered] Q: understanding STED?

Question: I have a problem conceptually understanding STED.

I can see, how a diffraction-limited spot of excited fluorophores is shaped to narrow the PSF. As a result, I get something akin to a point source of emission. But as the emitted light travels back through the microscope's optics, it is subject to diffraction again (and should thus destroy the narrowed PSF).

Is there somewhere a Gaussian fit applied at the other end (similar to PALM) to trace back the original spot? Or am I missing something basic here?

Any explanation would be greatly appreciated. Thanks - p



---------------------------------------------------------------------------

==============================Original Headers==============================
19, 12 -- From zaluzec-at-microscopy.com Wed Jan 3 08:08:32 2007
19, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
19, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l03E8Vbx031721
19, 12 -- for {microscopy-at-microscopy.com} ; Wed, 3 Jan 2007 08:08:32 -0600
19, 12 -- Mime-Version: 1.0
19, 12 -- X-Sender: (Unverified)
19, 12 -- Message-Id: {p06110407c1c169331bcf-at-[206.69.208.22]}
19, 12 -- Date: Wed, 3 Jan 2007 08:08:30 -0600
19, 12 -- To: microscopy-at-microscopy.com
19, 12 -- From: Martin.Hoppe-at-leica-microsystems.com (by way of MicroscopyListserver)
19, 12 -- Subject: viaWWW: Re: Understanding STED
19, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: denniscward-at-earthlink.net
Date: Wed, 3 Jan 2007 14:17:27 -0600
Subject: [Microscopy] follow-up: Appeal for Spectra

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Recently I made an "appeal for spectra" to this listserve. To those who responded, we sincerely appreciate your contributions, and enthusiasm for the project.
To those who did not respond because of my use of a personal email address, I have included my "FBI domain" address in the signature below.
At the risk of redundancy I will repeat my original appeal (OK, Nestor?).
Again, thank you.

} } }
SEM/EDS has been important analytical instrumentation for the FBI for decades. We have recently expanded its utility in investigative roles, however, and therefore would like to ask the community for assistance.
Whereas most SEM/EDS users need answers provided by quantitation and structural characterization, forensic inquiries generally necessitate "identifications"of questioned materials. This is generally possible only if the analyst is able to search his analytical results (spectra, primarily) against data from reference materials, as is the case with most other mature spectroscopies. We therefore have developed tools to archive and query the spectra and associated metadata of reference materials, and have produced a data set of thousands of materials in a variety of materials categories. Spectral and data queries against this database are consistently providing significant investigative direction. Of course quantitative analysis and other analytical techniques are additionally used where appropriate.
Because the FBI has limited analytical capability and limited access to specific materials, I am appealing to you for either reference spectra that can be uploaded directly into our database, or materials that we can borrow and analyze at our facility (on a very limited, very specific basis). Literally all materials are viable candidates for inclusion, from commercial products to biologicals. Spectra must be exportable in EMSA format.
Any data contributed will be restricted to FBI use only. If you would be willing to consider participating in this project, please contact me directly for details.
We sincerely appreciate your consideration of this appeal!

Dennis Ward, FBI



Dennis Ward
FBI Laboratory
Chemistry Unit
2501 Investigation Parkway
Quantico, VA 22135
v 703.632.7424

denniscward-at-earthlink.net
Dennis.Ward-at-ic.fbi.gov


==============================Original Headers==============================
8, 25 -- From denniscward-at-earthlink.net Wed Jan 3 14:17:26 2007
8, 25 -- Received: from elasmtp-mealy.atl.sa.earthlink.net (elasmtp-mealy.atl.sa.earthlink.net [209.86.89.69])
8, 25 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l03KHQJF020515
8, 25 -- for {Microscopy-at-microscopy.com} ; Wed, 3 Jan 2007 14:17:26 -0600
8, 25 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
8, 25 -- s=dk20050327; d=earthlink.net;
8, 25 -- b=gWnC1ci8x6BhDFPNr90YdxG7h49cfR28bi0RqXX7ojHUdp/ZN5vfJA2WmxbAx68B;
8, 25 -- h=Message-ID:Date:From:Reply-To:To:Subject:Cc:Mime-Version:Content-Type:Content-Transfer-Encoding:X-Mailer:X-ELNK-Trace:X-Originating-IP;
8, 25 -- Received: from [209.86.224.32] (helo=elwamui-cypress.atl.sa.earthlink.net)
8, 25 -- by elasmtp-mealy.atl.sa.earthlink.net with asmtp (Exim 4.34)
8, 25 -- id 1H2CYA-0002ID-5P; Wed, 03 Jan 2007 15:17:26 -0500
8, 25 -- Received: from 204.249.77.1 by webmail.pas.earthlink.net with HTTP; Wed, 3 Jan 2007 15:17:25 -0500
8, 25 -- Message-ID: {16699149.1167855446173.JavaMail.root-at-elwamui-cypress.atl.sa.earthlink.net}
8, 25 -- Date: Wed, 3 Jan 2007 15:17:25 -0500 (GMT-05:00)
8, 25 -- From: DWard {denniscward-at-earthlink.net}
8, 25 -- Reply-To: DWard {denniscward-at-earthlink.net}
8, 25 -- To: Microscopy-at-microscopy.com
8, 25 -- Subject: follow-up: Appeal for Spectra
8, 25 -- Cc: Michael.Rickenbach-at-ic.fbi.gov
8, 25 -- Mime-Version: 1.0
8, 25 -- Content-Type: text/plain; charset=UTF-8
8, 25 -- Content-Transfer-Encoding: 7bit
8, 25 -- X-Mailer: EarthLink Zoo Mail 1.0
8, 25 -- X-ELNK-Trace: 0f5e20a2ad636252da76355bff6752029ef193a6bfc3dd48ca82da3eed552b7421761a92b8dc031c93caf27dac41a8fd350badd9bab72f9c350badd9bab72f9c
8, 25 -- X-Originating-IP: 209.86.224.32
==============================End of - Headers==============================




From: mrsean-at-u.washington.edu
Date: Wed, 3 Jan 2007 19:18:51 -0600
Subject: [Microscopy] viaWWW: Can I process frozen tissue for TEM?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both mrsean-at-u.washington.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: mrsean-at-u.washington.edu
Name: Sean Kelly

Organization: University of Washington

Title-Subject: [Filtered] Can I process frozen tissue for TEM?

Question: I have some tissue that is embedded in OCT and stored at
-70C and want to know if it can be processed for TEM.

Any experience with this or relevent publications would be greatly appreciated.

Thanks,
Sean

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-microscopy.com Wed Jan 3 19:18:51 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l041IoFA004761
8, 12 -- for {microscopy-at-microscopy.com} ; Wed, 3 Jan 2007 19:18:51 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: (Unverified)
8, 12 -- Message-Id: {p06110408c1c20668f4ee-at-[206.69.208.22]}
8, 12 -- Date: Wed, 3 Jan 2007 19:18:49 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: mrsean-at-u.washington.edu (by way of MicroscopyListserver)
8, 12 -- Subject: viaWWW: Can I process frozen tissue for TEM?
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: stuart-at-foto45.co.uk
Date: Wed, 3 Jan 2007 19:20:24 -0600
Subject: [Microscopy] viaWWW: Photography Masters Project

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both stuart-at-foto45.co.uk as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: stuart-at-foto45.co.uk
Name: Stuart Handley

Title-Subject: [Filtered] Photography

Question: Hi,the reason for ask a massive favour is, I am starting a masters degree in photography and related media in a few days time and need to put my proposal in for my research project. I am going to propose images and related images on the subject of mines and minerals. This will involve photography underground and of samples that I collect, my hope is that I can find an electron microscopists who can help me to get down to the micro level of the minerals. The photography has to be different or done in a way that has not been done before so by combining the conventional and the electro scanning microscope images plus studio images of minerals I will be able to achive my set goal. Can any one out there help in any way or put me on to some one who maybe able to. Finding access to such equipment is going to be hard, any help most appreciated, I'm doing the MA part time over three years.
Stuart Handley

---------------------------------------------------------------------------

==============================Original Headers==============================
5, 12 -- From zaluzec-at-microscopy.com Wed Jan 3 19:20:23 2007
5, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
5, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l041KMh6006384
5, 12 -- for {microscopy-at-microscopy.com} ; Wed, 3 Jan 2007 19:20:23 -0600
5, 12 -- Mime-Version: 1.0
5, 12 -- X-Sender: (Unverified)
5, 12 -- Message-Id: {p06110409c1c206bb0849-at-[206.69.208.22]}
5, 12 -- Date: Wed, 3 Jan 2007 19:20:21 -0600
5, 12 -- To: microscopy-at-microscopy.com
5, 12 -- From: stuart-at-foto45.co.uk (by way of MicroscopyListserver)
5, 12 -- Subject: viaWWW: Photography Masters Project
5, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: jbs-at-temple.edu
Date: Wed, 3 Jan 2007 20:43:20 -0600
Subject: [Microscopy] AO Ultracut

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Greetings and Happy New Year.

Does anyone know the current price of the latest version of the old
AO Ultracut microtome?

Joel



Joel B. Sheffield, Ph.D
Department of Biology
Temple University
Philadelphia, PA 19122
Voice: 215 204 8839
e-mail: jbs-at-temple.edu


==============================Original Headers==============================
7, 26 -- From jbs-at-temple.edu Wed Jan 3 20:43:18 2007
7, 26 -- Received: from po-smtp3.temple.edu (po-smtp3.temple.edu [155.247.166.231])
7, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l042hI7f027756
7, 26 -- for {microscopy-at-microscopy.com} ; Wed, 3 Jan 2007 20:43:18 -0600
7, 26 -- Received: from [192.168.0.199] (c-68-80-94-164.hsd1.nj.comcast.net [68.80.94.164])
7, 26 -- by po-smtp3.temple.edu (MOS 3.8.2-GA)
7, 26 -- with ESMTP id ECL66709 (AUTH jbs);
7, 26 -- Wed, 3 Jan 2007 21:43:17 -0500 (EST)
7, 26 -- From: "Joel Sheffield" {jbs-at-temple.edu}
7, 26 -- To: microscopy-at-microscopy.com
7, 26 -- Date: Wed, 03 Jan 2007 21:43:18 -0500
7, 26 -- MIME-Version: 1.0
7, 26 -- Subject: AO Ultracut
7, 26 -- Reply-to: jbs-at-temple.edu
7, 26 -- Message-ID: {459C2376.26392.2EEF117B-at-jbs.temple.edu}
7, 26 -- Priority: normal
7, 26 -- X-mailer: Pegasus Mail for Windows (4.41)
7, 26 -- Content-type: text/plain; charset=US-ASCII
7, 26 -- Content-transfer-encoding: 7BIT
7, 26 -- Content-description: Mail message body
7, 26 -- X-Junkmail-Status: score=10/50, host=po-smtp3.temple.edu
7, 26 -- X-Junkmail-SD-Raw: score=unknown,
7, 26 -- refid=str=0001.0A09020A.459C6819.0035,ss=1,fgs=0,
7, 26 -- ip=192.168.0.199,
7, 26 -- so=2006-09-22 03:48:54,
7, 26 -- dmn=5.2.125/2006-10-10
==============================End of - Headers==============================




From: lcgould-at-med.cornell.edu
Date: Thu, 4 Jan 2007 08:06:52 -0600
Subject: [Microscopy] Re: Can I process frozen tissue for TEM?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Sean,
Can you process it? Yes.
Should you process it? Maybe.
Will you get good ultrastrucure? No.

A lot will depend on how the tissue was treated before it was frozen:
was it flash frozen without any fixation or cryo-protection? It will
have a lot of freezing artefacts at the EM level...ice crystal damage
to membranes, etc.
If it was lightly fixed (eg:paraformaldehyde), it might fair a bit better.
If it was fixed AND cryo-protected (eg: treated with sucrose), you
might get some decent structure.

You'll have to cut away as much of the OCT as you can and then thaw
the tissue. I would advise thawing it in a buffered glut. solution
so that it fixes as it thaws. Then process for TEM as usual and keep
you fingers crossed.
Over the years, I've had to process tissues that were not originally
intended for EM and I've been surprised by what may be rescued.
Good luck,
Lee
--
Leona Cohen-Gould, M.S.
Sr. Staff Associate
Director, Electron Microscopy & Histology Core Facility
Manager, Optical Microscopy Core Facility
Joan & Sanford I. Weill Medical College
of Cornell University
voice (212)746-6146
fax (212)746-8175
http://www.cornellcelldevbiology.org
http://www.cornellbiochem.org

==============================Original Headers==============================
3, 24 -- From lcgould-at-med.cornell.edu Thu Jan 4 08:06:52 2007
3, 24 -- Received: from smtp-gw2.med.cornell.edu (smtp-gw2.med.cornell.edu [157.139.3.45])
3, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l04E6qHt017830
3, 24 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 08:06:52 -0600
3, 24 -- Received: from mpx1.med.cornell.edu (biglb-vlan511vip.med.cornell.edu [140.251.11.120])
3, 24 -- by smtp-gw2.med.cornell.edu (Switch-3.1.8/Switch-3.1.7) with ESMTP id l04E6dIe025222;
3, 24 -- Thu, 4 Jan 2007 09:06:50 -0500 (EST)
3, 24 -- Received: from [140.251.48.23] (mac110773.med.cornell.edu [140.251.48.23])
3, 24 -- by mpx1.med.cornell.edu
3, 24 -- (Sun Java System Messaging Server 6.1 HotFix 0.11 (built Jan 28 2005))
3, 24 -- with ESMTPA id {0JBC009AJKJ2MF10-at-mpx1.med.cornell.edu} ; Thu,
3, 24 -- 04 Jan 2007 09:06:39 -0500 (EST)
3, 24 -- Date: Thu, 04 Jan 2007 09:06:36 -0500
3, 24 -- From: "Leona Cohen-Gould" {lcgould-at-med.cornell.edu}
3, 24 -- Subject: Re: [Microscopy] Can I process frozen tissue for TEM?
3, 24 -- In-reply-to: {200701040120.l041K3Md005942-at-ns.microscopy.com}
3, 24 -- Sender: "Leona Cohen-Gould" {lcgould-at-med.cornell.edu}
3, 24 -- To: Microscopy Listserver {microscopy-at-microscopy.com} , mrsean-at-u.washington.edu
3, 24 -- Message-id: {p0623092ac1c2b8f49172-at-[140.251.48.23]}
3, 24 -- MIME-version: 1.0
3, 24 -- Content-type: text/plain; charset=us-ascii; format=flowed
3, 24 -- Content-transfer-encoding: 7BIT
3, 24 -- References: {200701040120.l041K3Md005942-at-ns.microscopy.com}
3, 24 -- X-PMX-Version: 5.2.2.285561, Antispam-Engine: 2.5.0.283055, Antispam-Data: 2007.1.4.55433
==============================End of - Headers==============================




From: bfoster-at-mme1.com
Date: Thu, 4 Jan 2007 11:33:04 -0600
Subject: [Microscopy] Re: viaWWW: Photography Masters Project

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi, Stuart

If you are looking for art, I'd really suggest that you consider polarized light (an optical microscopy technique). The sample prep is about the same and the results are incredibly beautiful.

I have a colleague in the UK who can probably help. Contact me off-line for further info.

Good hunting!
Barbara Foster

Microscopy/Microscopy Education
313 S Jupiter Rd, Suite 100
Allen, TX 75002
P: 972-954-8011
W: www.MicroscopyEducation.com


MME is now scheduling customized, on-site courses through April. Call us today for details.

P. S.
Need a good general reference or light microscopy text for the Spring semester? Call us today to learn more about "Optimizing LIght Microscopy". Copies still available through MME... even for class-room lots ... and we give quantity discounts. Call Ken Piel at (972)954-8011.


At 11:26 AM 1/4/2007, stuart-at-foto45.co.uk wrote:



} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
14, 17 -- From bfoster-at-mme1.com Thu Jan 4 11:33:04 2007
14, 17 -- Received: from 5starpro.com (enterprise.5starpro.com [207.44.136.95])
14, 17 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l04HX4cO000988
14, 17 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 11:33:04 -0600
14, 17 -- Received: (qmail 14375 invoked by uid 2020); 4 Jan 2007 11:58:16 -0600
14, 17 -- Received: from host57-99.rancor.birch.net (HELO barbsd505.mme1.com) (65.17.57.99)
14, 17 -- by enterprise.5starpro.com with (DHE-RSA-AES256-SHA encrypted) SMTP; 4 Jan 2007 11:58:16 -0600
14, 17 -- Message-Id: {7.0.1.0.0.20070104113024.01d6f680-at-mme1.com}
14, 17 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
14, 17 -- Date: Thu, 04 Jan 2007 11:33:01 -0600
14, 17 -- To: stuart-at-foto45.co.uk, microscopy-at-microscopy.com
14, 17 -- From: Barbara Foster {bfoster-at-mme1.com}
14, 17 -- Subject: Re: [Microscopy] viaWWW: Photography Masters Project
14, 17 -- In-Reply-To: {200701040122.l041MWho012563-at-ns.microscopy.com}
14, 17 -- References: {200701040122.l041MWho012563-at-ns.microscopy.com}
14, 17 -- Mime-Version: 1.0
14, 17 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: bob-at-rockisland.com
Date: Thu, 4 Jan 2007 15:16:17 -0600
Subject: [Microscopy] viaWWW: Photography Masters Project

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Stuart,

There is a neat book: "Photography Through The Microscope" By J. G.
Delly and Eastman Kodak. It has many examples of photos taken through light
microscopes along with easy to follow 'how to...' instructions. My copy was
published in 1988. I don't know if the book is still in print, but your
University library can most likely has it or find a copy for you.

Take care, Bob Carter
2000 Bayshore Road
Lopez Island, WA 98261-8595

} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
14, 17 -- From bfoster-at-mme1.com Thu Jan 4 11:33:04 2007
14, 17 -- Received: from 5starpro.com (enterprise.5starpro.com
[207.44.136.95])
14, 17 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l04HX4cO000988
14, 17 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 11:33:04 -0600
14, 17 -- Received: (qmail 14375 invoked by uid 2020); 4 Jan 2007
11:58:16 -0600
14, 17 -- Received: from host57-99.rancor.birch.net (HELO
barbsd505.mme1.com) (65.17.57.99)
14, 17 -- by enterprise.5starpro.com with (DHE-RSA-AES256-SHA encrypted)
SMTP; 4 Jan 2007 11:58:16 -0600
14, 17 -- Message-Id: {7.0.1.0.0.20070104113024.01d6f680-at-mme1.com}
14, 17 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
14, 17 -- Date: Thu, 04 Jan 2007 11:33:01 -0600
14, 17 -- To: stuart-at-foto45.co.uk, microscopy-at-microscopy.com
14, 17 -- From: Barbara Foster {bfoster-at-mme1.com}
14, 17 -- Subject: Re: [Microscopy] viaWWW: Photography Masters Project
14, 17 -- In-Reply-To: {200701040122.l041MWho012563-at-ns.microscopy.com}
14, 17 -- References: {200701040122.l041MWho012563-at-ns.microscopy.com}
14, 17 -- Mime-Version: 1.0
14, 17 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================


--
No virus found in this incoming message.
Checked by AVG Free Edition.
Version: 7.5.432 / Virus Database: 268.16.5/616 - Release Date: 1/4/2007
1:34 PM


==============================Original Headers==============================
9, 33 -- From bob-at-rockisland.com Thu Jan 4 15:16:17 2007
9, 33 -- Received: from mars.rockisland.com (dnscache2.rockisland.com [64.119.0.12])
9, 33 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l04LGHhO016614
9, 33 -- for {Microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 15:16:17 -0600
9, 33 -- Received: from localhost (localhost [127.0.0.1])
9, 33 -- by localhost.rockisland.com (Postfix) with ESMTP id 87CC95ABA;
9, 33 -- Thu, 4 Jan 2007 13:16:16 -0800 (PST)
9, 33 -- (envelope-from bob-at-rockisland.com)
9, 33 -- X-Virus-Scanned: amavisd-new at rockisland.com
9, 33 -- Received: from mars.rockisland.com ([127.0.0.1])
9, 33 -- by localhost (smtp.rockisland.com [127.0.0.1]) (amavisd-new, port 10024)
9, 33 -- with ESMTP id osxVkIDZtANC; Thu, 4 Jan 2007 13:16:16 -0800 (PST)
9, 33 -- Received: from housecat (adsl-sj-11-138.rockisland.net [64.119.11.138])
9, 33 -- by mars.rockisland.com (Postfix) with SMTP id 0199B5A43;
9, 33 -- Thu, 4 Jan 2007 13:16:15 -0800 (PST)
9, 33 -- (envelope-from bob-at-rockisland.com)
9, 33 -- Message-ID: {005f01c73045$944401b0$0302a8c0-at-housecat}
9, 33 -- From: "Bob Carter" {bob-at-rockisland.com}
9, 33 -- To: {Microscopy-at-microscopy.com}
9, 33 -- Cc: {stuart-at-foto45.co.uk}
9, 33 -- References: {200701041734.l04HYhoT002801-at-ns.microscopy.com}
9, 33 -- Subject: Re: [Microscopy] Re: viaWWW: Photography Masters Project
9, 33 -- Date: Thu, 4 Jan 2007 13:16:19 -0800
9, 33 -- MIME-Version: 1.0
9, 33 -- Content-Type: text/plain;
9, 33 -- format=flowed;
9, 33 -- charset="iso-8859-1";
9, 33 -- reply-type=original
9, 33 -- Content-Transfer-Encoding: 7bit
9, 33 -- X-Priority: 3
9, 33 -- X-MSMail-Priority: Normal
9, 33 -- X-Mailer: Microsoft Outlook Express 6.00.2900.3028
9, 33 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: steve.klingaman-at-normandale.edu
Date: Thu, 4 Jan 2007 17:55:03 -0600
Subject: [Microscopy] AskAMicroscopist: entry-level SEM for teaching

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question was submitted to Ask-A-Microscopist by (steve.klingaman-at-normandale.edu)
from http://www.msa.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on Thursday, January 4, 2007 at 14:07:06
Remember to consider the Grade/Age of the student when considering the Question
---------------------------------------------------------------------------
Please reply to both steve.klingaman-at-normandale.edu as well as to the Microscopy Listserver
---------------------------------------------------------------------------

Email: steve.klingaman-at-normandale.edu
Name: Steve Klingaman

Organization: Normandale Community College

Education: Undergraduate College

Location: Bloomington, MN USA

Title: SEM

Question: Can you make a recommendation as to an entry-level SEM for teaching purposes to be used in a life sciences lab? Would a "table top" model suffice?

Price range might be the $50 - $100K range

Thank you,


---------------------------------------------------------------------------

==============================Original Headers==============================
11, 12 -- From zaluzec-at-ultra5.microscopy.com Thu Jan 4 17:55:03 2007
11, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
11, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l04Nt24S030988
11, 12 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 17:55:03 -0600
11, 12 -- Mime-Version: 1.0
11, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
11, 12 -- Message-Id: {p06110400c1c3443e7485-at-[206.69.208.22]}
11, 12 -- Date: Thu, 4 Jan 2007 17:55:02 -0600
11, 12 -- To: microscopy-at-microscopy.com
11, 12 -- From: steve.klingaman-at-normandale.edu (by way of Ask-A-Microscopist)
11, 12 -- Subject: AskAMicroscopist: entry-level SEM for teaching
11, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: swatkins-at-pitt.edu
Date: Thu, 4 Jan 2007 17:55:41 -0600
Subject: [Microscopy] viaWWW: cryo-ultramicrotome wanted

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both swatkins-at-pitt.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: swatkins-at-pitt.edu
Name: simon watkins

Organization: University of Pittsburgh

Title-Subject: [Filtered] cryo-ultramicrotome wanted

Question: Hi folks, I am trying to buy a used Reichert/Leica Cryo-ultramicrotome. I really only need the cryokit if you have one lying around or in storage. Both of our units have failed and we cant get the parts to fix them!!!! Built in obsolescence I guess. If you have a system, or know someone who would like to sell one to my group, please contact me offline
Simon

Simon C. Watkins Ph.D. FRC Path.
Professor, Cell Biology and Physiology and Immunology
Director Graduate Program in Cell Biology and Physiology
Vice Chair, Cell Biology and Physiology
Director Center for Biologic Imaging
BSTS 225
University of Pittsburgh
3500 Terrace St.
Pittsburgh PA 15261
Tel:412-648-3051
Fax:412-648-2797
URL: http://www.cbi.pitt.edu


---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-microscopy.com Thu Jan 4 17:55:41 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l04Ntelc031729
8, 12 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 17:55:40 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: (Unverified)
8, 12 -- Message-Id: {p06110401c1c3446a7eea-at-[206.69.208.22]}
8, 12 -- Date: Thu, 4 Jan 2007 17:55:40 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: swatkins-at-pitt.edu (by way of MicroscopyListserver)
8, 12 -- Subject: viaWWW: cryo-ultramicrotome wanted
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: jzheng-at-uci.edu
Date: Thu, 4 Jan 2007 19:39:20 -0600
Subject: [Microscopy] question: value of CM20 TEM

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Question: Hi, Folks,
I have a Philips CM20 traditional TEM with an EDS system. It is about 15
years old, but has a good working condition. Could you please tell me
how much it is worth now?
Thanks, Jian-Guo

------------------------------------------------------------------
Jian-Guo Zheng, PhD
Director, Nanomaterials Characterization and Fabrication Facility (NCF2)
University of California, Irvine
3415 Calit2 Building
Irvine, CA 92697-2800
phone: 949-824-0441 (office), 949-468-9980 (cell)
fax: 949-824-8197
email: jzheng-at-calit2.uci.edu
----------------------------------------------------------------




==============================Original Headers==============================
5, 20 -- From jzheng-at-uci.edu Thu Jan 4 19:39:19 2007
5, 20 -- Received: from relay2.es.uci.edu (relay2.es.uci.edu [128.200.80.28])
5, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l051dIBg022095
5, 20 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 19:39:19 -0600
5, 20 -- Received: from webmail.uci.edu (webmail2.es.uci.edu [128.200.80.37])
5, 20 -- by relay2.es.uci.edu (8.13.1/8.13.1) with ESMTP id l051dHai009969
5, 20 -- for {microscopy-at-microscopy.com} ; Thu, 4 Jan 2007 17:39:18 -0800
5, 20 -- Received: from 128.195.177.193
5, 20 -- (SquirrelMail authenticated user jzheng)
5, 20 -- by webmail.uci.edu with HTTP;
5, 20 -- Thu, 4 Jan 2007 17:39:18 -0800 (PST)
5, 20 -- Message-ID: {4616.128.195.177.193.1167961158.squirrel-at-webmail.uci.edu}
5, 20 -- Date: Thu, 4 Jan 2007 17:39:18 -0800 (PST)
5, 20 -- Subject: question: value of CM20 TEM
5, 20 -- From: jzheng-at-uci.edu
5, 20 -- To: microscopy-at-microscopy.com
5, 20 -- User-Agent: SquirrelMail/1.4.9a
5, 20 -- MIME-Version: 1.0
5, 20 -- Content-Type: text/plain;charset=iso-8859-1
5, 20 -- Content-Transfer-Encoding: 8bit
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Fri, 5 Jan 2007 04:04:12 -0600
Subject: [Microscopy] low vac SEM questions

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi!

Back again with my basic questions about SEM ;-)
I took pictures using both low vac mode (1 Pa) and
high vac mode (8x10^-3 Pa) at magnifications from 3kX
up to 25kX with my tescan SEM (at 10kV with quartz). I
can't see the slightest improvement in image quality!
The only difference is a slight increase in contrast
at higher vacuum (which is not necessarily better
because low-contrasted parts are not visible).

Is it normal? What would be the use of high vacuum? Is
it useful for EDX analysis but not normal scanning?

Stephane


__________________________________________________
Do You Yahoo!?
Tired of spam? Yahoo! Mail has the best spam protection around
http://mail.yahoo.com

==============================Original Headers==============================
6, 19 -- From nizets2-at-yahoo.com Fri Jan 5 04:04:12 2007
6, 19 -- Received: from web37415.mail.mud.yahoo.com (web37415.mail.mud.yahoo.com [209.191.91.147])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l05A4BRJ008576
6, 19 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 04:04:12 -0600
6, 19 -- Received: (qmail 71687 invoked by uid 60001); 5 Jan 2007 10:04:11 -0000
6, 19 -- Message-ID: {20070105100411.71685.qmail-at-web37415.mail.mud.yahoo.com}
6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 -- h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
6, 19 -- b=6Gx+ph2PmuD2WzkOGIN+ZlNa+5yDUpTAndMMum4S8pYohGncsSeRk3Y7gMi6xMHk2L8saJZQYR9Ua9syw5S6OPvXMIxaLqqTfcgXnIml3pIjv4hRBM8f/yFB8L0wO40wgWkLiX5hO0T1umkDdEV/F9lswkOv0zraYu2BOPpL2kM=;
6, 19 -- X-YMail-OSG: Dytz84gVM1n9PsZNUfrMuvo5W9jrKdFdIN.p9A8I14W61PGS.hkHntDikHM.T.jW9sjQILdZH0BQYhaajWnaO7Gq3f6jKqQbihJLFOF4BgLBiSDijNJINair5E1.bkxQcrxj9g4_hWHDowA-
6, 19 -- Received: from [80.122.101.102] by web37415.mail.mud.yahoo.com via HTTP; Fri, 05 Jan 2007 02:04:11 PST
6, 19 -- Date: Fri, 5 Jan 2007 02:04:11 -0800 (PST)
6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: low vac SEM questions
6, 19 -- To: microscopy-at-microscopy.com
6, 19 -- MIME-Version: 1.0
6, 19 -- Content-Type: text/plain; charset=iso-8859-1
6, 19 -- Content-Transfer-Encoding: 8bit
==============================End of - Headers==============================




From: W.Muss-at-salk.at
Date: Fri, 5 Jan 2007 04:13:00 -0600
Subject: [Microscopy] STAINING: Zeil Neelson, Zeil-Neilson,Zeil Neelsen, Zeil Neilsen, Ziehl-Nielson: which one is correct

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

To all:
} Happy, Healthy and Successful New Year"

Conc:
} Zeil Neelson {, } Zeil-Neilson {, } Zeil Neelsen {, } Zeil-Neilsen {,
} Ziehl-Nielson {:
which "name"/"term" one is correct


Dear all,
I would like to ask a question which came into minds when ToC-reading the
title of apreviousely published article on:
} Zeil Neelson [sic!]and Wade-Fite stains to demonstrate medlar bodies of
chromoblastomycosis {...

Can anybody of the group could tell or inform me wether a
} Zeil-Neelson {(SIC!) stain (for TB and mycobacterium, acid fast bacilli)
really does exist or is an other, classical, histochemical, specific stain
- I always thought to correctly be called "Ziehl-Nielson"....

When googleing shortly for } Zeil-Neelson { I found out that there exist
some other "terms" sounding very similar....see above....

Have I missed something??

Any information gladly is appreciated,

regards and THANK YOU,


Wolfgang Muss
Salzburg-AUSTRIA


OR Dr. Wolfgang Muss
Head of EM-Lab
{=with pride: 25 years in operation by 2nd of Feb.2006=}
Institute of Pathology, SALK
(Salzburger Landeskliniken gemeinnuetzige GesmbH)
(privatized General Hospital Fed. State of Salzburg)
Muellner Hauptstrasse 48
A-5020 SALZBURG AUSTRIA/EUROPE

and/or/alternatively (same Lab, same address)

Paracelsus Medical Private University (PMU)
Institute of Pathology
Electron Microscopy Lab
Muellner Hauptstrasse 48
A-5020 SALZBURG, Austria/Europe
Phone work: +43+662+4482+4720
Mobile phone work:+43+662+4482-57704
Fax-No. at work: ++43+662+4482-882 ext (please, only by indicating: "c/o
W.Muss")
E-Mail work: W.Muss-at-SALK.at

Mobile-phone private: ++43+676+5 369-456
E-Mail private: wij.Muss-at-aon.at


Ankuendigung namens der (Information on behalf of)
Society for Cutaneous Ultrastructure Research (SCUR)
PLEASE VISIT THE UPDATED WEBSITE of SCUR at
} http://www.scur.org {
-------------------------------------------------------------------------
post-congress information:
33rd Annual Meeting of the SCUR, 8th-10th June, 2006, WARSZAW, Poland
For information visit the WEBSITE, http://www.scur.org/
Click: Meetings ==} Previous Meetings ==} 33rd. Ann. Meeting
or, directly:
http://orgs.dermis.net/content/e04scur/e03meetings/e775/e896/index_ger.html


Forthcoming Meetings:
34th Annual Meeting of the SCUR, June 14-16, 2007,PRAGUE
Czech Republic
Local Host: Prof. Dr. Peter ARENBERGER
e-mail: pa-at-verum.cz

correct dates of meeting to be updated ASAP
35th Annual Meeting of the SCUR, ?11th -12th? May, 2008, OTHSU near KYOTO,
Japan
Joint Meeting with the
JSUCB, the Japanese Society for Ultrastructural Cutaneous Biology



==============================Original Headers==============================
23, 40 -- From W.Muss-at-salk.at Fri Jan 5 04:13:00 2007
23, 40 -- Received: from hermes.salk.at (hermes.lks.at [193.170.167.9])
23, 40 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05ACxrH017009
23, 40 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 04:13:00 -0600
23, 40 -- Received: from localhost (localhost [127.0.0.1])
23, 40 -- by hermes.salk.at (Postfix) with ESMTP id 5EE56C386F
23, 40 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 11:12:58 +0100 (CET)
23, 40 -- X-Virii-Scanned: Kaspersky Antivirus at salk.at
23, 40 -- Received: from hermes.salk.at ([127.0.0.1])
23, 40 -- by localhost (n1ex218.lks.local [127.0.0.1]) (amavisd-new, port 10024)
23, 40 -- with ESMTP id 0BokYcsAqnZH for {microscopy-at-microscopy.com} ;
23, 40 -- Fri, 5 Jan 2007 11:12:58 +0100 (CET)
23, 40 -- Received: from hermes.lks.at (hermes.salk.at [192.168.13.210])
23, 40 -- by hermes.salk.at (Postfix) with ESMTP id F39AAC3868
23, 40 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 11:12:57 +0100 (CET)
23, 40 -- Received: from localhost (localhost [127.0.0.1])
23, 40 -- by hermes.lks.at (Postfix) with ESMTP id DC0445A9022
23, 40 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 11:12:57 +0100 (CET)
23, 40 -- Received: from hermes.lks.at ([127.0.0.1])
23, 40 -- by localhost (hermes.lks.local [127.0.0.1]) (amavisd-new, port 10024)
23, 40 -- with ESMTP id 13523-01 for {microscopy-at-microscopy.com} ;
23, 40 -- Fri, 5 Jan 2007 11:12:57 +0100 (CET)
23, 40 -- Received: from c1pa003 (unknown [192.168.42.3])
23, 40 -- by hermes.lks.at (Postfix) with SMTP id 8F8075A901F
23, 40 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 11:12:57 +0100 (CET)
23, 40 -- Received: by localhost with Microsoft MAPI; Fri, 5 Jan 2007 11:12:56 +0100
23, 40 -- Message-ID: {01C730BA.73B46140.W.Muss-at-salk.at}
23, 40 -- From: MUSS Wolfgang PhD {W.Muss-at-salk.at}
23, 40 -- Reply-To: "W.Muss-at-salk.at" {W.Muss-at-salk.at}
23, 40 -- To: "'microscopy-at-microscopy.com'" {microscopy-at-microscopy.com}
23, 40 -- Subject: [Microscopy] STAINING: Zeil Neelson, Zeil-Neilson,Zeil Neelsen, Zeil Neilsen, Ziehl-Nielson: which one is correct
23, 40 -- Date: Fri, 5 Jan 2007 11:12:55 +0100
23, 40 -- Return-Receipt-To: MUSS Wolfgang PhD {W.Muss-at-salk.at}
23, 40 -- Organization: SALK,Salzburger Landeskliniken, Pathologie, EM-Labor
23, 40 -- X-Mailer: Microsoft Internet E-Mail/MAPI - 8.0.0.4211
23, 40 -- MIME-Version: 1.0
23, 40 -- Content-Type: text/plain; charset="us-ascii"
23, 40 -- Content-Transfer-Encoding: 7bit
23, 40 -- X-Virii-Scanned: by Sophos/ClamAV at salk.at
23, 40 -- X-Scanned-By: SALK-Content-Filter
==============================End of - Headers==============================




From: michael-at-shaffer.net
Date: Fri, 5 Jan 2007 07:06:48 -0600
Subject: [Microscopy] RE: low vac SEM questions

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Stephane writes ...

} Back again with my basic questions about SEM ;-) I took
} pictures using both low vac mode (1 Pa) and high vac mode
} (8x10^-3 Pa) at magnifications from 3kX up to 25kX with my
} tescan SEM (at 10kV with quartz). I can't see the slightest
} improvement in image quality!
} The only difference is a slight increase in contrast at
} higher vacuum (which is not necessarily better because
} low-contrasted parts are not visible).

You do not give any indication of the beam current you are using (possibly
the 'spot size' adjustment for your SEM). I suspect you will not realize
the real advantages of employing high vacuum unless the beam scuurent (and
spot size) is small. Better vacuum will also provide better contrast ...
But exactly what you see is specimen and preparation dependent. I suspect
the samples are coated ... But with ??

cheerios :o)
michael shaffer

SEM/MLA Research Coordinator
{http://www.mun.ca/creait/maf/}
Inco Innovation Centre
Memorial University
St. John's, Newfoundland



==============================Original Headers==============================
6, 20 -- From michael-at-shaffer.net Fri Jan 5 07:06:48 2007
6, 20 -- Received: from n034.sc1.he.tucows.com (fh1021.dia.cp.net [64.97.168.31] (may be forged))
6, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05D6mgZ001124
6, 20 -- for {Microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 07:06:48 -0600
6, 20 -- Received: from roamingwolf (134.153.130.141) by n034.sc1.he.tucows.com (7.2.069.1) (authenticated as Michael-at-Shaffer.net)
6, 20 -- id 458A9BAC0021124E; Fri, 5 Jan 2007 13:06:47 +0000
6, 20 -- From: "michael shaffer" {michael-at-shaffer.net}
6, 20 -- To: {nizets2-at-yahoo.com}
6, 20 -- Cc: "MSA Microscopy list" {Microscopy-at-microscopy.com}
6, 20 -- Subject: RE: [Microscopy] low vac SEM questions
6, 20 -- Date: Fri, 5 Jan 2007 09:36:34 -0330
6, 20 -- Message-ID: {001501c730ca$5b1b47b0$8d829986-at-roamingwolf}
6, 20 -- MIME-Version: 1.0
6, 20 -- Content-Type: text/plain;
6, 20 -- charset="US-ASCII"
6, 20 -- Content-Transfer-Encoding: 7bit
6, 20 -- X-Mailer: Microsoft Office Outlook 11
6, 20 -- In-Reply-To: {200701051004.l05A4oPC009282-at-ns.microscopy.com}
6, 20 -- Thread-Index: AccwsLAWLLyNoD6YR6+msTLKPCGAKgAGNgFA
6, 20 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: pburcke-at-utnet.utoledo.edu
Date: Fri, 5 Jan 2007 07:48:29 -0600
Subject: [Microscopy] viaWWW: EDS: Link exL had drive

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both pburcke-at-utnet.utoledo.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: pburcke-at-utnet.utoledo.edu
Name: Pannee Burckel

Organization: Chemistry, University of Toledo

Title-Subject: [Filtered] EDS: Link exL had drive

Question:
Dear Listers: The Link exL hard drive crashed a month ago. We need an inexpensive temporarily solution. If anyone have a it somewhere in the lab, please let me know. It is a 40S 40MB Quantum Prodrive. I could purchase an HD online but I am not sure if it can be installed with Link software. Thank you.

Pannee

---------------------------------------------------------------------------

==============================Original Headers==============================
7, 12 -- From zaluzec-at-microscopy.com Fri Jan 5 07:48:29 2007
7, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
7, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05DmS95012792
7, 12 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 07:48:29 -0600
7, 12 -- Mime-Version: 1.0
7, 12 -- X-Sender: (Unverified)
7, 12 -- Message-Id: {p06110400c1c407999c02-at-[206.69.208.22]}
7, 12 -- Date: Fri, 5 Jan 2007 07:48:27 -0600
7, 12 -- To: microscopy-at-microscopy.com
7, 12 -- From: pburcke-at-utnet.utoledo.edu (by way of MicroscopyListserver)
7, 12 -- Subject: viaWWW: EDS: Link exL had drive
7, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: Anjeanette.Ormonde-at-unilever.com
Date: Fri, 5 Jan 2007 07:58:28 -0600
Subject: [Microscopy] Supplies for film processing to give away

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


In cleaning out the lab I came across a drawer full of small cylinders
that have "Polaroid Print Coater" embossed on one side and "For Black &
White Pictures Only". Inside is a fluid drenched material with a handy
little holder along one side. I've never used them and we no longer
have need to process film in the dark room but I believe the material is
used to add a protective coating to the processed photographs. Is there
anyone that would be interested in having some or all of these
materials? In addition we have some powdered Kodak Developer D-19 and
while I can't find an expiration date I'm sure it is several years old
(I've been here over 5 years and it was purchased before my arrival.)
There is also some EM film I'm sure has expired as well as some polaroid
film and even a role of film called "Kodachrome 64". I also came across
some liquid activator and fixer that was never opened. If you are
interested in any or all of these items please contact me directly for
more details.

Thanks!
Angie


==============================Original Headers==============================
2, 23 -- From Anjeanette.Ormonde-at-unilever.com Fri Jan 5 07:58:28 2007
2, 23 -- Received: from ntrsegw30001.s3.ms.unilever.com (mailout02.unilever.com [204.110.170.4])
2, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05DwSSp023886
2, 23 -- for {Microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 07:58:28 -0600
2, 23 -- Received: from NTRSEVS30001.s3.ms.unilever.com ([162.87.40.60]) by ntrsegw30001.s3.ms.unilever.com with Microsoft SMTPSVC(6.0.3790.1830);
2, 23 -- Fri, 5 Jan 2007 08:58:28 -0500
2, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
2, 23 -- Content-class: urn:content-classes:message
2, 23 -- MIME-Version: 1.0
2, 23 -- Content-Type: text/plain;
2, 23 -- charset="US-ASCII"
2, 23 -- Subject: Supplies for film processing to give away
2, 23 -- Date: Fri, 5 Jan 2007 08:58:26 -0500
2, 23 -- Message-ID: {DBE4410766F3FE4991E69A1CD3CA13AA01240A51-at-NTRSEVS30001.s3.ms.unilever.com}
2, 23 -- X-MS-Has-Attach:
2, 23 -- X-MS-TNEF-Correlator:
2, 23 -- Thread-Topic: Supplies for film processing to give away
2, 23 -- Thread-Index: AccwINGrFgMmmlrVRx28VMx/O/3OsAAsCWdQ
2, 23 -- From: "Ormonde, Anjeanette" {Anjeanette.Ormonde-at-unilever.com}
2, 23 -- To: {Microscopy-at-microscopy.com}
2, 23 -- X-OriginalArrivalTime: 05 Jan 2007 13:58:28.0302 (UTC) FILETIME=[936CAEE0:01C730D1]
2, 23 -- Content-Transfer-Encoding: 8bit
2, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l05DwSSp023886
==============================End of - Headers==============================




From: Sven.Terclavers-at-med.kuleuven.be
Date: Fri, 5 Jan 2007 08:10:56 -0600
Subject: [Microscopy] Stereology analysis

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

First of all, my best wishes for 2007 - May the Light shine through your
Microscope!

As often in this lab however, the beginning of a new year brings new
challenges. We are looking for software to perform stereology analysis
(Cavalieri, Fractionator, ...). At the moment we are using already one
packge from Zeiss, incorporated into their KS300 software, however, we would
like to expand the analysis to more pc's. Therefore the following question:
is someone out there using the stereology modules for Image J? And if so,
is there a manual available or could you pass some tips?
Many thanks in advance,

Sven

°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°
Sven Terclavers
Bio-Imaging Facility
Center for Transgene Technology & Gene Therapy
Campus Gasthuisberg O&N3
Herestraat 49
3000 Leuven
Belgium
Tel: +32.(0)16.34.71.97
Fax: +32.(0)16.34.59.90
Email: Sven.Terclavers-at-med.kuleuven.be
Web: www.vib.be - www.kuleuven.be/mcm
°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°°


Disclaimer: http://www.kuleuven.be/cwis/email_disclaimer.htm



==============================Original Headers==============================
8, 33 -- From Sven.Terclavers-at-med.kuleuven.be Fri Jan 5 08:10:55 2007
8, 33 -- Received: from thumbler.kulnet.kuleuven.ac.be (thumbler.kulnet.kuleuven.ac.be [134.58.240.45])
8, 33 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05EAtNo002553
8, 33 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 08:10:55 -0600
8, 33 -- Received: from localhost (localhost [127.0.0.1])
8, 33 -- by thumbler.kulnet.kuleuven.ac.be (Postfix) with ESMTP id D4DCF138606
8, 33 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 15:10:54 +0100 (CET)
8, 33 -- Received: from smtp02.kuleuven.be (lepidus.kulnet.kuleuven.ac.be [134.58.240.72])
8, 33 -- by thumbler.kulnet.kuleuven.ac.be (Postfix) with ESMTP id 01A1813855E
8, 33 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 15:10:54 +0100 (CET)
8, 33 -- Received: from smtp02.kuleuven.be (localhost.localdomain [127.0.0.1])
8, 33 -- by smtp02.kuleuven.be (Postfix) with ESMTP id C95372CAAF3;
8, 33 -- Fri, 5 Jan 2007 15:10:53 +0100 (CET)
8, 33 -- Received: from raven (unknown [134.58.245.175])
8, 33 -- by smtp02.kuleuven.be (Postfix) with ESMTP id ADE522CAAE2
8, 33 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 15:10:53 +0100 (CET)
8, 33 -- Reply-To: {Sven.Terclavers-at-med.kuleuven.be}
8, 33 -- From: "Sven Terclavers" {Sven.Terclavers-at-med.kuleuven.be}
8, 33 -- To: "Microscopy" {microscopy-at-microscopy.com}
8, 33 -- Subject: Stereology analysis
8, 33 -- Date: Fri, 5 Jan 2007 15:10:46 +0100
8, 33 -- Organization: KULeuven
8, 33 -- MIME-Version: 1.0
8, 33 -- Content-Type: text/plain;
8, 33 -- charset="iso-8859-1"
8, 33 -- X-Mailer: Microsoft Office Outlook, Build 11.0.5510
8, 33 -- In-Reply-To: {200606121312.k5CDChT4020354-at-ns.microscopy.com}
8, 33 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
8, 33 -- Thread-Index: AcaOIeVgnvnnEli/TiWPxThcEgrVeiirzJOA
8, 33 -- Message-Id: {20070105141053.ADE522CAAE2-at-smtp02.kuleuven.be}
8, 33 -- X-Virus-Scanned: by KULeuven Antivirus Cluster
8, 33 -- Content-Transfer-Encoding: 8bit
8, 33 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l05EAtNo002553
==============================End of - Headers==============================




From: renaudgeological-at-execulink.com
Date: Fri, 5 Jan 2007 14:47:06 -0600
Subject: [Microscopy] viaWWW: WIN 30 PGH Board

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi

One of the problems with assessing SEM performance is often the specimen you
select. I travel around the world looking at customers instruments training
the people to obtain more from their instrument. In order to judge the
performance of the instruments I spent a number of months experimenting with
test specimens until a came up with one where the image always changed in
some way when I changed parameters.

I remember a ListServer posting some time ago when a scientist commented
that the new FEG instruments are amazing as they are just as good (with
their specimen) at 1kV as 30kV! In my mind the test specimen was not
testing and I feel this could be the reason that you do not see changes when
you vary the parameters in your instrument.

I poor quality vacuum is great to reduce charge. But a poor vacuum causes
beam spread, a problem that will increase the spot size and, if sufficient
performance is demanded, ultimately degrade the image quality.

Steve Chapman
Senior Consultant Protrain
For electron microscopy consultancy and training world wide
Tel +44 1280 816512 Fax +44 1280 814007
Mobile +44 7711 606967 Web www.emcourses.com

----- Original Message -----
X-from: {nizets2-at-yahoo.com}
To: {protrain-at-emcourses.com}
Sent: Friday, January 05, 2007 10:05 AM

Dear Stephane,
There are several fundamental differences between high vac and low vac that
make each unique in application.
1. low vac can only use BSE or special low vac SE-type detectors that make
traditional, high-resolution SE images impossible. Low vac is usually 10 to
200 Pa. Usually these detectors require higher beam voltage (10 to 20 kV)
and beam current than high-vac SE imaging.
2. High vac requires that a sample be conductive. Try your experiment with a
freshly-plucked flower, looking at the stamen and pollen grains and you will
see that it is impossible in high vac mode but quite possible in low vac.
You may have to increase the gas pressure to 20 to 100 Pa to eliminate
charging. If you need to do high resolution imaging of the small features on
a sample surface, only high vac will allow you to use {5 kV, small aperture,
short working distance and low beam current to get detailed SE images of the
sample.
3. The scattering of the electron beam in low vac mode gives EDS
contributions from areas away from where you may want to analyse. It is nice
to be able to do EDS on anything without coating, but if you need to analyse
one small spot without contribution from neighbouring areas, you need high
vac.
Both modes have their place and strengths and it is nice to have the choice.
Regards,
Mary Mager
Electron Microscopist
Materials Eng. UBC
#419 - 6350 Stores Rd.
Vancouver, B.C. V6T 1Z4
CANADA
Phone: 604-822-5648
Fax: 604-822-3619
email: mager-at-interchange.ubc.ca

-----Original Message-----
X-from: nizets2-at-yahoo.com [mailto:nizets2-at-yahoo.com]
Sent: Friday, January 05, 2007 2:14 AM
To: mager-at-interchange.ubc.ca

Hi!

Back again with my basic questions about SEM ;-) I took pictures using both
low vac mode (1 Pa) and high vac mode (8x10^-3 Pa) at magnifications from
3kX up to 25kX with my tescan SEM (at 10kV with quartz). I can't see the
slightest improvement in image quality!
The only difference is a slight increase in contrast at higher vacuum (which
is not necessarily better because low-contrasted parts are not visible).

Is it normal? What would be the use of high vacuum? Is it useful for EDX
analysis but not normal scanning?

Stephane


__________________________________________________
Do You Yahoo!?
Tired of spam? Yahoo! Mail has the best spam protection around
http://mail.yahoo.com

==============================Original Headers==============================
6, 19 -- From nizets2-at-yahoo.com Fri Jan 5 04:04:12 2007 6, 19 -- Received:
from web37415.mail.mud.yahoo.com (web37415.mail.mud.yahoo.com
[209.191.91.147])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id
l05A4BRJ008576
6, 19 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 04:04:12
-0600
6, 19 -- Received: (qmail 71687 invoked by uid 60001); 5 Jan 2007 10:04:11
-0000 6, 19 -- Message-ID:
{20070105100411.71685.qmail-at-web37415.mail.mud.yahoo.com}
6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 --
h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Conten
t-Transfer-Encoding:Message-ID;
6, 19 --
b=6Gx+ph2PmuD2WzkOGIN+ZlNa+5yDUpTAndMMum4S8pYohGncsSeRk3Y7gMi6xMHk2L8saJZQYR
9Ua9syw5S6OPvXMIxaLqqTfcgXnIml3pIjv4hRBM8f/yFB8L0wO40wgWkLiX5hO0T1umkDdEV/F9
lswkOv0zraYu2BOPpL2kM=;
6, 19 -- X-YMail-OSG:
Dytz84gVM1n9PsZNUfrMuvo5W9jrKdFdIN.p9A8I14W61PGS.hkHntDikHM.T.jW9sjQILdZH0BQ
YhaajWnaO7Gq3f6jKqQbihJLFOF4BgLBiSDijNJINair5E1.bkxQcrxj9g4_hWHDowA-
6, 19 -- Received: from [80.122.101.102] by web37415.mail.mud.yahoo.com via
HTTP; Fri, 05 Jan 2007 02:04:11 PST 6, 19 -- Date: Fri, 5 Jan 2007 02:04:11
-0800 (PST) 6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com} 6, 19 --

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both renaudgeological-at-execulink.com as well as the
MIcroscopy Listserver
---------------------------------------------------------------------------

Email: renaudgeological-at-execulink.com
Name: Jim Renaud

Organization: Renaud Geological Consulting Ltd.

Title-Subject: [Filtered] WIN 30 PGH Board

Question: I am looking for a WIN 30 PGH board for the digital imaging
system on my electron microprobe. I have checked with the
manufacturer and this is a discontinued product. If you have one of
these boards and would like to sell it or know here I may be able to
find one, I would appreciate any information you could share.

Best Regards,

Jim Renaud.

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-microscopy.com Fri Jan 5 14:47:06 2007
8, 12 -- Received: from [130.202.10.196] (msdvpn8.msd.anl.gov [130.202.238.72])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05Kl52W015543
8, 12 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 14:47:06 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: (Unverified)
8, 12 -- Message-Id: {p06020401c1c46b0f4e84-at-[130.202.10.196]}
8, 12 -- Date: Fri, 5 Jan 2007 14:52:55 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: renaudgeological-at-execulink.com (by way of MicroscopyListserver)
8, 12 -- Subject: viaWWW: WIN 30 PGH Board
8, 12 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
==============================End of - Headers==============================




From: eoptics-at-mcmaster.ca
Date: Fri, 5 Jan 2007 16:53:37 -0600
Subject: [Microscopy] viaWWW: Conical Dark Field Imaging

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both eoptics-at-mcmaster.ca as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: eoptics-at-mcmaster.ca
Name: Fred Pearson

Organization: McMaster Unviersity/Canadian Center for Electron Microscopy

Title-Subject: [Filtered] Conical Dark Field Imaging

Question: Any there any descriptive articles available for Conical
Dark Field imaging and its advantages?



---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-microscopy.com Fri Jan 5 16:53:36 2007
8, 12 -- Received: from [130.202.10.196] (visitor196.et.anl-external.org [130.202.10.196])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l05Mra0F028820
8, 12 -- for {microscopy-at-microscopy.com} ; Fri, 5 Jan 2007 16:53:36 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: (Unverified)
8, 12 -- Message-Id: {p06020401c1c488b9c10c-at-[130.202.10.196]}
8, 12 -- Date: Fri, 5 Jan 2007 16:59:25 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: eoptics-at-mcmaster.ca (by way of MicroscopyListserver)
8, 12 -- Subject: viaWWW: Conical Dark Field Imaging
8, 12 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
==============================End of - Headers==============================




From: Lori_mahaffey-at-yahoo.com
Date: Mon, 8 Jan 2007 08:05:07 -0600
Subject: [Microscopy] [Filtered] AskAMicroscopist: dust mites

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question was submitted to Ask-A-Microscopist by (Lori_mahaffey-at-yahoo.com)
from http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on Sunday, January 7, 2007 at 17:14:47
Remember to consider the Grade/Age of the student when considering the Question
---------------------------------------------------------------------------
Please reply to both Lori_mahaffey-at-yahoo.com as well as to the Microscopy Listserver
---------------------------------------------------------------------------

Email: Lori_mahaffey-at-yahoo.com
Name: Keaton Mahaffey

Organization: Noddin Elementary

Education: K-8 Grade Grammar School

Location: San Jose Ca

Title: Dust Mites

Question: i have tried to look at an old pillow under my microscope hoping to see dust mites but i did not.
Is there something i can do to see them?

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Mon Jan 8 08:05:07 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08E56Ax023549
8, 12 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 08:05:06 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
8, 12 -- Message-Id: {p06110400c1c7fffc2a72-at-[206.69.208.22]}
8, 12 -- Date: Mon, 8 Jan 2007 08:05:05 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: Lori_mahaffey-at-yahoo.com (by way of Ask-A-Microscopist)
8, 12 -- Subject: [Filtered] AskAMicroscopist: dust mites
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: cammer-at-aecom.yu.edu
Date: Mon, 8 Jan 2007 09:04:27 -0600
Subject: [Microscopy] Nikon to Olympus objective converter

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

We have a brass coupler with threads to put a Nikon objective on an
Olympus nosepiece. Don't know where it came from, but it looks homemade.

We need a few more.

Before asking our machine shop to duplicate it, we were wondering if
anybody knows whether we can purchase them.

Thanks.
____________________________________________________________________________
Michael Cammer Analytical Imaging Facility Albert Einstein Coll. of Med.
URL: http://www.aecom.yu.edu/aif/


==============================Original Headers==============================
5, 30 -- From cammer-at-aecom.yu.edu Mon Jan 8 09:04:27 2007
5, 30 -- Received: from mx1.aecom.yu.edu (mx1.aecom.yu.edu [129.98.1.51])
5, 30 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08F4R3n003670
5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:04:27 -0600
5, 30 -- Received: from draco.aecom.yu.edu (draco.aecom.yu.edu [129.98.1.160])
5, 30 -- by mx1.aecom.yu.edu (Postfix) with ESMTP id CABCA9F0015
5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 10:04:26 -0500 (EST)
5, 30 -- Received: from draco.aecom.yu.edu (unknown [127.0.0.1])
5, 30 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id EB05D8B4003
5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:53:57 -0500 (EST)
5, 30 -- X-AuditID: 816201a0-9f8c8bb000000505-bb-45a25b05cc87
5, 30 -- Received: from post.aecom.yu.edu (post.aecom.yu.edu [129.98.1.100])
5, 30 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id B48FE718002
5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:53:57 -0500 (EST)
5, 30 -- Received: from AIF3.aecom.yu.edu (aif3.aif.aecom.yu.edu [129.98.80.70])
5, 30 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
5, 30 -- (No client certificate requested)
5, 30 -- by post.aecom.yu.edu (Postfix) with ESMTP id 6038026
5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 10:04:26 -0500 (EST)
5, 30 -- Message-Id: {7.0.1.0.2.20070108100226.0532b168-at-aecom.yu.edu}
5, 30 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
5, 30 -- Date: Mon, 08 Jan 2007 10:04:28 -0500
5, 30 -- To: microscopy-at-microscopy.com
5, 30 -- From: Michael Cammer {cammer-at-aecom.yu.edu}
5, 30 -- Subject: Nikon to Olympus objective converter
5, 30 -- In-Reply-To: {200701052049.l05KnBOD018011-at-ns.microscopy.com}
5, 30 -- References: {200701052049.l05KnBOD018011-at-ns.microscopy.com}
5, 30 -- Mime-Version: 1.0
5, 30 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
5, 30 -- X-Brightmail-Tracker: AAAAAA==
==============================End of - Headers==============================




From: keith.morris-at-ucl.ac.uk
Date: Mon, 8 Jan 2007 09:20:50 -0600
Subject: [Microscopy] [Filtered] AskAMicroscopist: dust mites

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html



Dear Keaton,

You should be able to see them fairly easily at 10-40x, if they are there,
as they are often over 200um long - you need a decent microscope though. You
can chemically test for their presence using a chemical card. They need 70%
humidity and food to thrive - so an old unused airing cupboard pillow might
not provide any living specimens - perhaps try a mattress that's in use.
Other than allergenic dust problems with their faeces & dead mites (which
can be severe) they are harmless.

A 4 to 120x stereo microscope should also work well, although they are a
little translucent, and they are odd looking beasties. Rather than repeat
other on-line info, have a look at:

http://ohioline.osu.edu/hyg-fact/2000/2157.html

http://www.ehso.com/ehshome/dustmites.php

http://www.dustless.com/alpha/info/mites.html

http://hgic.clemson.edu/factsheets/HGIC2551.htm

http://www.funsci.com/fun3_en/dust/dust.htm

http://www.madsci.org/posts/archives/aug98/900036716.Gb.r.html

http://www.badspiderbites.com/dust-mites.php

There an optical microscope image at

http://strano16.interfree.it/micro03.htm

Keith

_________________________

Dr Keith J Morris
Imaging Facility Manager
Institute of Ophthalmology
University College London
11-43 Bath Street
London EC1V 9EL

email: keith.Morris-at-ucl.ac.uk


-----Original Message-----
X-from: Lori_mahaffey-at-yahoo.com [mailto:Lori_mahaffey-at-yahoo.com]
Sent: 08 January 2007 14:12
To: keith.morris-at-ucl.ac.uk

This Question was submitted to Ask-A-Microscopist by
(Lori_mahaffey-at-yahoo.com)
from http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on
Sunday, January 7, 2007 at 17:14:47
Remember to consider the Grade/Age of the student when considering the
Question
---------------------------------------------------------------------------
Please reply to both Lori_mahaffey-at-yahoo.com as well as to the Microscopy
Listserver
---------------------------------------------------------------------------

Email: Lori_mahaffey-at-yahoo.com
Name: Keaton Mahaffey

Organization: Noddin Elementary

Education: K-8 Grade Grammar School

Location: San Jose Ca

Title: Dust Mites

Question: i have tried to look at an old pillow under my microscope hoping
to see dust mites but i did not.
Is there something i can do to see them?

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Mon Jan 8 08:05:07 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l08E56Ax023549
8, 12 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 08:05:06
-0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
8, 12 -- Message-Id: {p06110400c1c7fffc2a72-at-[206.69.208.22]}
8, 12 -- Date: Mon, 8 Jan 2007 08:05:05 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: Lori_mahaffey-at-yahoo.com (by way of Ask-A-Microscopist)
8, 12 -- Subject: [Filtered] AskAMicroscopist: dust mites
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================



==============================Original Headers==============================
34, 22 -- From keith.morris-at-ucl.ac.uk Mon Jan 8 09:20:49 2007
34, 22 -- Received: from smtp3.global.net.uk (smtp3.global.net.uk [80.189.92.91])
34, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08FKnHQ014967
34, 22 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:20:49 -0600
34, 22 -- Received: from 124.126.gr5.adsl.brightview.com ([80.189.126.124] helo=loungepc)
34, 22 -- by smtp3.global.net.uk with esmtp (Exim 4.42)
34, 22 -- id 1H3wIq-000DhT-Dz; Mon, 08 Jan 2007 15:20:48 +0000
34, 22 -- From: "Keith Morris" {keith.morris-at-ucl.ac.uk}
34, 22 -- To: {Lori_mahaffey-at-yahoo.com}
34, 22 -- Cc: {microscopy-at-microscopy.com}
34, 22 -- Subject: RE: [Microscopy] [Filtered] AskAMicroscopist: dust mites
34, 22 -- Date: Mon, 8 Jan 2007 15:20:50 -0000
34, 22 -- Message-ID: {000901c73338$94f07440$0401a8c0-at-loungepc}
34, 22 -- MIME-Version: 1.0
34, 22 -- Content-Type: text/plain;
34, 22 -- charset="us-ascii"
34, 22 -- Content-Transfer-Encoding: 7bit
34, 22 -- X-Mailer: Microsoft Office Outlook 11
34, 22 -- Thread-Index: AcczL5Kj2rdmM517QuSugGHZKI5O1wABQUUg
34, 22 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
34, 22 -- In-Reply-To: {200701081411.l08EBgGq030454-at-ns.microscopy.com}
34, 22 -- Authenticated-Sender:
==============================End of - Headers==============================




From: mcauliff-at-umdnj.edu
Date: Mon, 8 Jan 2007 09:35:21 -0600
Subject: [Microscopy] decalcification of Zebrafish

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Alan,

We also decomissioned our darkroom several years ago, since we had only
one remaining client that wanted prints, and she only wanted one or two
8x10's per year. Our Durst Labrator (sp?) went to another building and
is still in use, I'm told.

I purchased an inkjet, "photo-quality" printer, in case anyone really
needed prints, but it was never used and the ink cartridges eventually
went bad.

I also paid my darkroom dues over many years. I think every piece of
work clothing I owned had developer stains and I always squinted in
normal light. I will always have a fond place in my heart for
silver-based photography, and may even go back to it someday for the
artsy-fartsy part of my life, but I will not shed a tear for it in
research work. We will soon replace our old TEM with a film- and
digital-capable new one, but I expect that we will go from thousands of
sheets of 4489 per year to tens----if that.

Cheers,
Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


-----Original Message-----
X-from: Debby Sherman [mailto:dsherman-at-purdue.edu]
Sent: Thursday, January 04, 2007 4:13 PM
To: Alan Nicholls
Cc: Christopher J. Gilpin; Angela Klaus; Anlee Krupp; Ann Lehman;
Anne-Marie Girard; Annette Andrews; Anthony McCormick; Bobbie Schneider;
Brigitte Shaw; Bryony James; Chere Petty; Clive Wells; David Dorward;
David Hall; Dee Breger; Donald Gantz; Douglas Cromey; Edward Seijo;
Elaine Humphrey; Elison Blancaflor; Evelyn York; Miller, F. Scott; Frank
Macaluso; Gary Chandler; Geoffrey Williams; Halina Witkiewicz; Hendrik
Colijn; Jacob Mey; Jaime Dant; Janice Pennington; Jeffrey Hanson; Jim
Conner; John Bozzola; John Curulli; John Gardner; Jon Mulholland;
Jonathan Krupp; Joseph Kulik; Judith Drazba; Julie Getz; Karen Kelley;
Karen Stevenson; Karl Wendt; Laura Raymond; Leona Cohen-Gould; Longzhou
Ma; Louis Kerr; Luisa Dempere; Margaret Bisher; Margaret Sherwood;
Marilyn Dunlap; Mark Walters; Melanie Barfels; Michael Goheen; Michelle
Ocana; Missy Hazen; Mitchell McCartney; Neelima Shah; Norman Olson;
Patricia Connelly; Patricia Jansma; Paula Allan-Wojtas; Philip Oshel;
Phillip Russell; Rakesh Bhatnagar; Rebecca Stearns; Richard Gursky;
Robin Griffin; Rogelio Pescador; Sardiaa Plaud; Steven Barlow; Stuart
McKernan; Thomas Freeman; Thomas Murray; Timothy Maugel; Tina Carvalho;
Tsuey-Chen Long; Wendy Salmon; William L'Amoreaux; William Russin; Zsolt
Lazar; Anita K. McCauley PhD; Chunfei Li Ph.D.; Earnest W. Truby Ph. D.;
Edwina W. Westbrook (Winnie); Gary M. Brown; James Hayden (Jamie);
Kenneth L. Tiekotter; Leslie M. Eibest; Mayandi Sivaguru Ph.D. (Shiv);
Qian-Chun Yu MB Ph.D.; R. Holland Cheng; S. Kelly Sears Ph.D. B.F.A.;
Sousan Abolhassani Ph.D.; Warren Straszheim Ph.D.; Barbara maloney; Ben
Fowler; Beverly Giammara; Bill Tivol; Caroline Miller; Chaoying Ni PhD;
Christian Mellen; Christine Davis; Donald Robertson; Eduardo
Rosa-Molinar; Fred Monson PhD; Greg Ning; Hilary Holloway; Jacqueline
Mills; Jan Redick; Jeff Braswell; Jeff Horn; Judy Murphy; Kent McDonald;
Kim Christensen; Lois Anderson; Ross Jr, Louis M.; Marc Pypaert; Marilee
Sellers; Michael Dunlap; Mike Gemble; Randy Nessler; Tindall, Randy D.;
Reinhard Rachel; Richard Schalek; Ron Anderson; Rosemary White; Sara
Miller; Steve Chapman; Phillips, Thomas E.

Dear colleeagues:

Anyone have a protocol for the decalcification of adult zebrafish?

Geoff

--
--
**********************************************
Geoff McAuliffe, Ph.D.
Neuroscience and Cell Biology
Robert Wood Johnson Medical School
675 Hoes Lane, Piscataway, NJ 08854
voice: (732)-235-4583; fax: -4029
mcauliff-at-umdnj.edu
**********************************************



==============================Original Headers==============================
6, 33 -- From mcauliff-at-umdnj.edu Mon Jan 8 09:35:21 2007
6, 33 -- Received: from zix02.umdnj.edu (zix02.UMDNJ.EDU [130.219.34.125])
6, 33 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08FZLfS004547
6, 33 -- for {microscopy-at-msa.microscopy.com} ; Mon, 8 Jan 2007 09:35:21 -0600
6, 33 -- Received: from zix02.umdnj.edu (ZixVPM [127.0.0.1])
6, 33 -- by Outbound.umdnj.edu (Proprietary) with ESMTP id BEE984BE2E
6, 33 -- for {microscopy-at-msa.microscopy.com} ; Mon, 8 Jan 2007 10:35:20 -0500 (EST)
6, 33 -- Received: from umdnj.edu (imail.UMDNJ.EDU [130.219.34.129])
6, 33 -- by zix02.umdnj.edu (Proprietary) with ESMTP id 855C44BE76
6, 33 -- for {microscopy-at-msa.microscopy.com} ; Mon, 8 Jan 2007 10:35:19 -0500 (EST)
6, 33 -- Received: from ([130.219.34.131])
6, 33 -- by imail.umdnj.edu with ESMTP id KP-BRACD.123713466;
6, 33 -- Mon, 08 Jan 2007 10:35:00 -0500
6, 33 -- Received: from conversion-daemon.Polaris.umdnj.edu by Polaris.umdnj.edu
6, 33 -- (iPlanet Messaging Server 5.2 HotFix 2.02 (built Oct 21 2004))
6, 33 -- id {0JBK005012VZK1-at-Polaris.umdnj.edu} (original mail from mcauliff-at-umdnj.edu)
6, 33 -- for microscopy-at-msa.microscopy.com; Mon, 08 Jan 2007 10:35:00 -0500 (EST)
6, 33 -- Received: from [127.0.0.1] ([10.138.2.123])
6, 33 -- by Polaris.umdnj.edu (iPlanet Messaging Server 5.2 HotFix 2.02 (built Oct 21
6, 33 -- 2004)) with ESMTP id {0JBK003FJ3A7Y9-at-Polaris.umdnj.edu} ; Mon,
6, 33 -- 08 Jan 2007 10:34:56 -0500 (EST)
6, 33 -- Date: Mon, 08 Jan 2007 10:36:17 -0500
6, 33 -- From: Geoff McAuliffe {mcauliff-at-umdnj.edu}
6, 33 -- Subject: decalcification of Zebrafish
6, 33 -- To: MicroscopyListserver {microscopy-at-msa.microscopy.com} ,
6, 33 -- Histonet {histonet-at-pathology.swmed.edu}
6, 33 -- Message-id: {45A264F1.4000904-at-umdnj.edu}
6, 33 -- MIME-version: 1.0
6, 33 -- Content-type: text/plain; format=flowed; charset=ISO-8859-1
6, 33 -- Content-transfer-encoding: 7BIT
6, 33 -- X-Accept-Language: en-us, en
6, 33 -- User-Agent: Mozilla/5.0 (Windows; U; Windows NT 5.1; en-US; rv:1.7.2)
6, 33 -- Gecko/20040804 Netscape/7.2 (ax)
==============================End of - Headers==============================




From: jureller-at-uchicago.edu
Date: Mon, 8 Jan 2007 11:17:03 -0600
Subject: [Microscopy] Re: Nikon to Olympus objective converter

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Alan, et al.,

Having been trained in classic darkroom B&W procedures back in late
70's/early 80's (TEM 4x5, 35mm (Plus-X, Tri-X), -- and Polaroid Type
55), we decommissioned our darkroom ~8 years ago, having moved to
digital TEM and LM imaging two years prior to that. We stocked color
film (Kodachrome, Ektachrome, Kodacolor) but that was always sent out
for processing.

Basic photographic theory and practice still applies. The only thing
that has changed is the workflow, with more emphasis at the point of
capture, rather than after-the-fact processing/printing. In fact, we
perform very little printing, as 99% of needs are in digital format. Any
prints are done with a high end Fuji Pictrography printer (silver halide
based digital print system that rivals a classical photograph!)

Also, as long as you understand the technology, an inkjet can
equal/exceed a photograph. You must carefully choose the printer &
corresponding paper, and ink media. For B&W photography, one should
examine quad tone inks (multi-tone black inks).

With digital, there is a learning curve to understand concepts of camera
sensors (what the camera is actually recording), image pixels,
resolution, how to match the proper amount of digital information for a
particular output, and file storage formats (TIF, JPG, etc) and
advantages/disadvantages of each. Everything you ever did in the
darkroom is right there in your image editor (like the consumer
Photoshop Elements or even professional strength Photoshop CS).

Best regards,

Walter F. Bobrowski
Sr. Scientist
Pfizer Global R&D
2800 Plymouth Rd.
Ann Arbor, MI 48105

TEL: 734-622-7814
FAX: 734-622-5718

"The ultimate human freedom is the ability to choose one's attitude in a
given set of circumstances." -Viktor Frankl





-----Original Message-----
X-from: TindallR-at-missouri.edu [mailto:TindallR-at-missouri.edu]
Sent: Monday, January 08, 2007 10:33 AM
To: Bobrowski, Walter

Alan,

We also decomissioned our darkroom several years ago, since we had only
one remaining client that wanted prints, and she only wanted one or two
8x10's per year. Our Durst Labrator (sp?) went to another building and
is still in use, I'm told.

I purchased an inkjet, "photo-quality" printer, in case anyone really
needed prints, but it was never used and the ink cartridges eventually
went bad.

I also paid my darkroom dues over many years. I think every piece of
work clothing I owned had developer stains and I always squinted in
normal light. I will always have a fond place in my heart for
silver-based photography, and may even go back to it someday for the
artsy-fartsy part of my life, but I will not shed a tear for it in
research work. We will soon replace our old TEM with a film- and
digital-capable new one, but I expect that we will go from thousands of
sheets of 4489 per year to tens----if that.

Cheers,
Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan



Hi,

We had to make a few of these up in our in-shouse shop too. However, Thorlabs
now sells RMS to M25 (and reverse) commercially for cheap

http://www.thorlabs.com/newgrouppage9.cfm?objectGroup_ID=1747

regards
justin

cammer-at-aecom.yu.edu wrote:
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} We have a brass coupler with threads to put a Nikon objective on an
} Olympus nosepiece. Don't know where it came from, but it looks homemade.
}
} We need a few more.
}
} Before asking our machine shop to duplicate it, we were wondering if
} anybody knows whether we can purchase them.
}
} Thanks.
} ____________________________________________________________________________
} Michael Cammer Analytical Imaging Facility Albert Einstein Coll. of Med.
} URL: http://www.aecom.yu.edu/aif/
}
}
} ==============================Original Headers==============================
} 5, 30 -- From cammer-at-aecom.yu.edu Mon Jan 8 09:04:27 2007
} 5, 30 -- Received: from mx1.aecom.yu.edu (mx1.aecom.yu.edu [129.98.1.51])
} 5, 30 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08F4R3n003670
} 5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:04:27 -0600
} 5, 30 -- Received: from draco.aecom.yu.edu (draco.aecom.yu.edu [129.98.1.160])
} 5, 30 -- by mx1.aecom.yu.edu (Postfix) with ESMTP id CABCA9F0015
} 5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 10:04:26 -0500 (EST)
} 5, 30 -- Received: from draco.aecom.yu.edu (unknown [127.0.0.1])
} 5, 30 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id EB05D8B4003
} 5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:53:57 -0500 (EST)
} 5, 30 -- X-AuditID: 816201a0-9f8c8bb000000505-bb-45a25b05cc87
} 5, 30 -- Received: from post.aecom.yu.edu (post.aecom.yu.edu [129.98.1.100])
} 5, 30 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id B48FE718002
} 5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 09:53:57 -0500 (EST)
} 5, 30 -- Received: from AIF3.aecom.yu.edu (aif3.aif.aecom.yu.edu [129.98.80.70])
} 5, 30 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
} 5, 30 -- (No client certificate requested)
} 5, 30 -- by post.aecom.yu.edu (Postfix) with ESMTP id 6038026
} 5, 30 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 10:04:26 -0500 (EST)
} 5, 30 -- Message-Id: {7.0.1.0.2.20070108100226.0532b168-at-aecom.yu.edu}
} 5, 30 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
} 5, 30 -- Date: Mon, 08 Jan 2007 10:04:28 -0500
} 5, 30 -- To: microscopy-at-microscopy.com
} 5, 30 -- From: Michael Cammer {cammer-at-aecom.yu.edu}
} 5, 30 -- Subject: Nikon to Olympus objective converter
} 5, 30 -- In-Reply-To: {200701052049.l05KnBOD018011-at-ns.microscopy.com}
} 5, 30 -- References: {200701052049.l05KnBOD018011-at-ns.microscopy.com}
} 5, 30 -- Mime-Version: 1.0
} 5, 30 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
} 5, 30 -- X-Brightmail-Tracker: AAAAAA==
} ==============================End of - Headers==============================
}

--
---------------------------------------------------------------------------
Justin E. Jureller, Ph.D.
Technical Director, IBD NanoBiology Facility, The University of Chicago
Gordon Center for Integrative Science, 929 E 57th Street, Chicago, IL 60637
office: GCIS ESB06B lab: GCIS ESB18
phone: (773) 834-3864 fax: (773) 834-1917
email: jureller-at-uchicago.edu web: http://nanobio.uchicago.edu/
---------------------------------------------------------------------------

==============================Original Headers==============================
6, 18 -- From jureller-at-uchicago.edu Mon Jan 8 11:17:03 2007
6, 18 -- Received: from relay01.uchicago.edu (relay01.uchicago.edu [128.135.12.136])
6, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08HH2GA026129
6, 18 -- for {Microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 11:17:03 -0600
6, 18 -- Received: from [128.135.35.206] (godard.uchicago.edu [128.135.35.206])
6, 18 -- by relay01.uchicago.edu (8.13.6.20060614/8.12.9) with ESMTP id l08HH2TI011530
6, 18 -- for {Microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 11:17:02 -0600 (CST)
6, 18 -- Message-ID: {45A27C94.3090200-at-uchicago.edu}
6, 18 -- Date: Mon, 08 Jan 2007 11:17:08 -0600
6, 18 -- From: j jureller {jureller-at-uchicago.edu}
6, 18 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
6, 18 -- MIME-Version: 1.0
6, 18 -- To: Microscopy-at-microscopy.com
6, 18 -- Subject: Re: [Microscopy] Nikon to Olympus objective converter
6, 18 -- References: {200701081509.l08F97Vd011986-at-ns.microscopy.com}
6, 18 -- In-Reply-To: {200701081509.l08F97Vd011986-at-ns.microscopy.com}
6, 18 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
6, 18 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: cammer-at-aecom.yu.edu
Date: Mon, 8 Jan 2007 15:14:55 -0600
Subject: [Microscopy] Olympus IX70 technical question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

try B&H Photo-Video
http://www.bhphotovideo.com/bnh/controller/home?O=NavBar&A=FetchChildren&Q=&ci=9824 ,
check "Lens to Body Adapters" and "Lens Adapters and Mounts" under "General
Lens Accessories".

Even if they don't have Olympus body to Nikon lens adapter, you may combine
necessary parts from 2 adapters, these are cheap.

Vitaly Feingold
SIA
2773 Heath Line
Duluth GA 30096
Ph. 770-232-7785
Fax 770-232-1791
www.sia-cam.com
----- Original Message -----
X-from: {cammer-at-aecom.yu.edu}
To: {vitalylazar-at-att.net}
Sent: Monday, January 08, 2007 10:05 AM

never mind, I miss-understood the posting.

Vitaly Feingold
SIA
2773 Heath Line
Duluth GA 30096
Ph. 770-232-7785
Fax 770-232-1791
www.sia-cam.com
----- Original Message -----
X-from: {vitalylazar-at-att.net}
To: {vitalylazar-at-att.net}
Sent: Monday, January 08, 2007 12:23 PM

We have an Olympus IX 70 that is about six years old that we want to
rebuild for a custom laser application.

There are two optics in the light path we need to identify. We would
greatly appreciate if anybody knows the answer to please let us
know. (The alternative it to disassemble the microscope completely
which we'd rather not do.)

The light path from the sample to the detector through the side port
is as follows:

Objective lens to window to lens to prism to window. We need to
identify the lens (B) and the prism (C). B and C are diagrammed at
http://cammer.net/temp/IX70question/

Is the lens what Olympus calls the "tube lens"?
What focal length is the lens?
How far into the UV and into the IR does the lens pass } 70%?

Similarly, for the prism, how far into the UV and into the IR does
the glass pass } 70%?

Thanks!

-Michael
____________________________________________________________________________
Michael Cammer Analytical Imaging Facility Albert Einstein Coll. of Med.
URL: http://www.aecom.yu.edu/aif/


==============================Original Headers==============================
9, 28 -- From cammer-at-aecom.yu.edu Mon Jan 8 15:14:55 2007
9, 28 -- Received: from mx1.aecom.yu.edu (mx1.aecom.yu.edu [129.98.1.51])
9, 28 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08LEsvP002842
9, 28 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 15:14:54 -0600
9, 28 -- Received: from draco.aecom.yu.edu (draco.aecom.yu.edu [129.98.1.160])
9, 28 -- by mx1.aecom.yu.edu (Postfix) with ESMTP id 621A39F0072
9, 28 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 16:14:54 -0500 (EST)
9, 28 -- Received: from draco.aecom.yu.edu (unknown [127.0.0.1])
9, 28 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id 5531D8B4002
9, 28 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 16:04:24 -0500 (EST)
9, 28 -- X-AuditID: 816201a0-9da69bb000000505-eb-45a2b1d82c74
9, 28 -- Received: from post.aecom.yu.edu (post.aecom.yu.edu [129.98.1.100])
9, 28 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id 1C3E7718002
9, 28 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 16:04:24 -0500 (EST)
9, 28 -- Received: from AIF3.aecom.yu.edu (aif3.aif.aecom.yu.edu [129.98.80.70])
9, 28 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
9, 28 -- (No client certificate requested)
9, 28 -- by post.aecom.yu.edu (Postfix) with ESMTP id DBFA326;
9, 28 -- Mon, 8 Jan 2007 16:14:53 -0500 (EST)
9, 28 -- Message-Id: {7.0.1.0.2.20070108161224.026baad0-at-aecom.yu.edu}
9, 28 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
9, 28 -- Date: Mon, 08 Jan 2007 16:14:37 -0500
9, 28 -- To: microscopy-at-microscopy.com
9, 28 -- From: Michael Cammer {cammer-at-aecom.yu.edu}
9, 28 -- Subject: Olympus IX70 technical question
9, 28 -- Mime-Version: 1.0
9, 28 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
9, 28 -- X-Brightmail-Tracker: AAAAAA==
==============================End of - Headers==============================




From: jmkrupp-at-ucsc.edu
Date: Mon, 8 Jan 2007 16:12:04 -0600
Subject: [Microscopy] digital camera withdrawals

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Greetings for the New Year:

Have I become a wimp? I cannot face the new year without my TEM digital camera.

Here is the story:

About 4 years ago a PI on campus bought a digital camera for our
central campus lab's TEM. She said anyone could use it, no problem.
She got a lot of use from it and so did everyone else.

Now she is leaving and wants to take the camera with her, leaving me
without one. None of the users who got used to the camera want to go
to film, I don't miss the chemical mixing, disposal, and darkroom
time. Users really got used to taking their pictures with them at the
end of a session on a memory stick instead of waiting half a day to
see if anything turned out and then having to figure out a filing
system for all their negatives.

What would you do if someone pulled the rug, er, uh, camera, out from
under you? I need to convince our Dean that it is really the only way
to go now a days. Is that true or have I become too lazy?

There is a chance that we might be able to cut a deal with the owner
of the camera if we offer her a fair price for it. It is a Gatan 1K x
1K Bioscan mounted at the 35 mm port. Any guesses on current value? I
know it is kind of old and might be obsolete in some circles, but
it's the only one I have and I would like to keep it until we can
afford something newer. What would be a fair offer to the owner?

Jon
--


Jonathan Krupp
Microscopy & Imaging Lab
C230 Earth & Marine Science
University of California
Santa Cruz, CA 95064
(831) 459-2477
jmkrupp-at-ucsc.edu

==============================Original Headers==============================
10, 21 -- From jmkrupp-at-ucsc.edu Mon Jan 8 16:12:03 2007
10, 21 -- Received: from smtp-prod-mx3.ucsc.edu (smtp-prod-mx3.ucsc.edu [128.114.125.45])
10, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l08MC3PU014681
10, 21 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 16:12:03 -0600
10, 21 -- Received: from ucsc.edu (cruzmail-fe2.ucsc.edu [128.114.125.2])
10, 21 -- by smtp-prod-mx3.ucsc.edu (8.13.8/8.13.8) with ESMTP id l08MBng5007837
10, 21 -- for {microscopy-at-microscopy.com} ; Mon, 8 Jan 2007 14:12:02 -0800 (PST)
10, 21 -- Received: from [128.114.25.227] (account jmkrupp-at-ucsc.edu HELO [128.114.25.227])
10, 21 -- by silver.ucsc.edu (CommuniGate Pro SMTP 4.3.7)
10, 21 -- with ESMTPA id 95721644 for microscopy-at-microscopy.com; Mon, 08 Jan 2007 14:11:54 -0800
10, 21 -- Mime-Version: 1.0
10, 21 -- Message-Id: {p06230909c1c86f391353-at-[128.114.25.227]}
10, 21 -- Date: Mon, 8 Jan 2007 14:11:53 -0800
10, 21 -- To: microscopy-at-microscopy.com
10, 21 -- From: Jon Krupp {jmkrupp-at-ucsc.edu}
10, 21 -- Subject: digital camera withdrawals
10, 21 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
10, 21 -- X-UCSC-EDU-MIMEDefang: Checked
10, 21 -- X-UCSC-EDU-Sender: {jmkrupp-at-ucsc.edu}
10, 21 -- X-UCSC-CATS-MailScanner-From: {jmkrupp-at-ucsc.edu}
10, 21 -- X-Scanned-By: MIMEDefang 2.58 on 128.114.125.45
==============================End of - Headers==============================




From: cervantes-at-bendres.com
Date: Tue, 9 Jan 2007 10:51:07 -0600
Subject: [Microscopy] Cryo TEM Imaging of Nanoparticles

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This is somewhat of a solicitation for information from the cryo TEM community. I'm imaging small (small = less than 20 nm) polymeric nanoparticles with cryo TEM, and as always, am striving for the best image possible. I'd like to obtain good (with respect to characteristics such as resolution and signal/noise ratio) images of small, low contrast, spherical particles, to serve as comparisons for images I generate . Journal and book references would be gladly appreciated. Conventional TEM images, with or without staining, would also be very helpful.

Jessica Cervantes
Research Chemist
Bend Research, Inc
64550 Research Rd
Bend, OR 97701
(541) 382-4100
(541) 382-0212 x240
(541) 382-6177 fax

LEGAL NOTICE: This message (and/or any attachments accompanying it) is confidential and proprietary. It is intended for the addressee(s) only. Access to this e-mail by anyone else is unauthorized. If you are not an addressee any disclosure, copying, or distribution of the contents of this e-mail (and any attachments) or any action taken (or not taken) in reliance upon this message is unauthorized and may be unlawful. If you are not an addressee, please contact the sender immediately by calling (541) 382-4100.



==============================Original Headers==============================
5, 14 -- From cervantes-at-bendres.com Tue Jan 9 10:51:07 2007
5, 14 -- Received: from BRIEX04A.bri.local (bc161112.bendcable.com [216.228.161.112])
5, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l09Gp5Qd030963
5, 14 -- for {microscopy-at-microscopy.org} ; Tue, 9 Jan 2007 10:51:06 -0600
5, 14 -- MIME-Version: 1.0
5, 14 -- Content-Type: text/plain;
5, 14 -- charset="iso-8859-1"
5, 14 -- Subject: Cryo TEM Imaging of Nanoparticles
5, 14 -- Date: Tue, 9 Jan 2007 08:51:03 -0800
5, 14 -- Message-ID: {8943D65F9AD70E4488AD6DE09F15088768C25E-at-BRIEX04A}
5, 14 -- From: "Cervantes, Jessica" {cervantes-at-bendres.com}
5, 14 -- To: {microscopy-at-microscopy.org}
5, 14 -- Content-Transfer-Encoding: 8bit
5, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l09Gp5Qd030963
==============================End of - Headers==============================




From: m.aindow-at-uconn.edu
Date: Tue, 9 Jan 2007 14:31:49 -0600
Subject: [Microscopy] Postdoctoral Position

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

The following is a re-posting of a vacancy that was advertised originally in
September, but was not filled due to an administrative re-organization.

****************************************************************************

University of Connecticut
Institute of Materials Science

Position in Scanning Electron Microscopy and Optical Microscopy


The Institute of Materials Science (IMS) at UConn is an interdisciplinary
center with the threefold mission of fostering education, research and
outreach in all areas of the materials sciences. The Microscopy Laboratory
is a user facility which houses the main research microscopes (optical, SEM
and TEM) for the IMS. There is an opening in the Laboratory for a Scanning
Electron Microscopy / Optical Microscopy specialist. The appointee will be
involved in a range of academic and industrial projects, and will assist in
the operation of the Laboratory including performing routine maintenance and
training/assisting users of the microscopes.

Candidates should hold a higher degree (MS or PhD) in Materials Science or a
related discipline and must have extensive hands-on SEM and optical
microscopy experience. Experience in maintenance of electron microscopes,
use of transmission electron microscopy, microscopy of soft materials and/or
microtomy would also be beneficial. This is a fixed-term appointment and is
available from February 1st. Screening of the applications will begin
immediately and will continue until the post is filled. Applications from
under-represented groups, including minorities, women and persons with
disabilities are encouraged.

Interested candidates should send a curriculum vitae and the names of at
least three referees with postal addresses, telephone numbers and Email
addresses to: Prof. Mark Aindow, Institute of Materials Science, University
of Connecticut, 97 North Eagleville Road, Unit 3136, Storrs, CT 06269-3136
USA. Email: m.aindow-at-uconn.edu





==============================Original Headers==============================
5, 22 -- From m.aindow-at-uconn.edu Tue Jan 9 14:31:49 2007
5, 22 -- Received: from mail2.uits.uconn.edu (mail2.uits.uconn.edu [137.99.25.204])
5, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l09KVnTP014869
5, 22 -- for {Microscopy-at-MSA.Microscopy.Com} ; Tue, 9 Jan 2007 14:31:49 -0600
5, 22 -- Received: from [137.99.20.179] (d20h179.public.uconn.edu [137.99.20.179])
5, 22 -- by mail2.uits.uconn.edu (8.13.6/8.11.6) with ESMTP id l09KViqW010917;
5, 22 -- Tue, 9 Jan 2007 15:31:44 -0500
5, 22 -- User-Agent: Microsoft-Entourage/11.3.2.061213
5, 22 -- Date: Tue, 09 Jan 2007 15:31:48 -0500
5, 22 -- Subject: Postdoctoral Position
5, 22 -- From: Mark Aindow {m.aindow-at-uconn.edu}
5, 22 -- To: {Microscopy-at-MSA.Microscopy.Com}
5, 22 -- Message-ID: {C1C965E4.1623B%m.aindow-at-uconn.edu}
5, 22 -- Thread-Topic: Postdoctoral Position
5, 22 -- Thread-Index: Acc0LS+XbfUTZaAgEdu8nQARJEXN3A==
5, 22 -- Mime-version: 1.0
5, 22 -- Content-type: text/plain;
5, 22 -- charset="US-ASCII"
5, 22 -- Content-transfer-encoding: 7bit
5, 22 -- X-UConn-MailScanner-Information: Contact UConn Help Desk 860-486-4357 for more information.
5, 22 -- X-UConn-MailScanner: Found to be clean
5, 22 -- X-UConn-MailScanner-SpamCheck:
==============================End of - Headers==============================




From: ani.issaian-at-csun.edu
Date: Tue, 9 Jan 2007 18:13:46 -0600
Subject: [Microscopy] Re: Cryo TEM Imaging of Nanoparticles

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Jessica,

Try staining the polymeric sections/particles with OsO4 at different times, ie., 5, 7, 10min, look at them with the TEM and select the best results.

Ani

---- Original message ----
} Date: Tue, 9 Jan 2007 11:03:32 -0600
} From: cervantes-at-bendres.com
} Subject: [Microscopy] Cryo TEM Imaging of Nanoparticles
} To: ani.issaian-at-csun.edu
}
}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America

==============================Original Headers==============================
4, 29 -- From ani.issaian-at-csun.edu Tue Jan 9 18:13:45 2007
4, 29 -- Received: from plover.csun.edu (plover.csun.edu [130.166.1.24])
4, 29 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0A0DjED002579
4, 29 -- for {microscopy-at-microscopy.org} ; Tue, 9 Jan 2007 18:13:45 -0600
4, 29 -- Received: from puffin.csun.edu (puffin.csun.edu [130.166.1.21])
4, 29 -- by plover.csun.edu (MOS 3.8.2-GA)
4, 29 -- with ESMTP id DLF31054
4, 29 -- for {microscopy-at-microscopy.org} ;
4, 29 -- Tue, 9 Jan 2007 16:13:44 -0800 (PST)
4, 29 -- Received: from cuckoo.csun.edu (cuckoo.csun.edu [130.166.1.230])
4, 29 -- by puffin.csun.edu (MOS 3.7.5a-GA)
4, 29 -- with ESMTP id FIS39177
4, 29 -- for {microscopy-at-microscopy.org} ;
4, 29 -- Tue, 9 Jan 2007 16:13:43 -0800 (PST)
4, 29 -- Received: (from cuckoo.csun.edu [130.166.114.202])
4, 29 -- by cuckoo.csun.edu (MOS 3.8.2-GA)
4, 29 -- with HTTPS/1.1 id AQU44765 (AUTH ami24015);
4, 29 -- Tue, 9 Jan 2007 16:13:43 -0800 (PST)
4, 29 -- From: Ani M Issaian {ani.issaian-at-csun.edu}
4, 29 -- Subject: Re: [Microscopy] Cryo TEM Imaging of Nanoparticles
4, 29 -- To: microscopy-at-microscopy.org
4, 29 -- Reply-To: ani.issaian-at-csun.edu
4, 29 -- X-Mailer: Mirapoint Webmail Direct 3.8.2-GA
4, 29 -- MIME-Version: 1.0
4, 29 -- Content-Type: text/plain; charset=us-ascii
4, 29 -- Content-Transfer-Encoding: 7bit
4, 29 -- Message-Id: {20070109161343.AQU44765-at-cuckoo.csun.edu}
4, 29 -- Date: Tue, 9 Jan 2007 16:13:43 -0800 (PST)
4, 29 -- X-Junkmail-Whitelist: YES (by domain whitelist at plover.csun.edu)
==============================End of - Headers==============================




From: connellyps-at-mail.nih.gov
Date: Tue, 9 Jan 2007 19:35:00 -0600
Subject: [Microscopy] AskAMicroscopist: fixation for mitochondria

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question was submitted to Ask-A-Microscopist by (connellyps-at-mail.nih.gov)
from http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on Tuesday, January 9, 2007 at 17:33:37
Remember to consider the Grade/Age of the student when considering the Question
---------------------------------------------------------------------------
Please reply to both connellyps-at-mail.nih.gov as well as to the Microscopy Listserver
---------------------------------------------------------------------------

Email: connellyps-at-mail.nih.gov
Name: Pat Connelly

Organization: NIH

Education: Graduate College

Location: Bethesda, MD

Title: fixation for mitochondria

Question: Dear Fellow Experts,

I am in need of the "best" fixation for mitochondria in mouse and rat muscle and kidney. Past experiments have yeilded OK but not great results using 2.5%glut with and without paraformaldehyde in sodium cacodylate buffer followed by 1% OsO4 and UA block stain. I'd like to be saying "Wow, Look at these!"
Any suggestions or references? Today I did a Pub Med search and most of what I found was the same as I have been doing.

Thanks,
Pat Connelly
connellyps-at-mail.nih.gov

---------------------------------------------------------------------------

==============================Original Headers==============================
10, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 9 19:35:00 2007
10, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
10, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0A1Yxi3014862
10, 12 -- for {microscopy-at-microscopy.com} ; Tue, 9 Jan 2007 19:35:00 -0600
10, 12 -- Mime-Version: 1.0
10, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
10, 12 -- Message-Id: {p06110401c1c9f33021fc-at-[206.69.208.22]}
10, 12 -- Date: Tue, 9 Jan 2007 19:34:58 -0600
10, 12 -- To: microscopy-at-microscopy.com
10, 12 -- From: connellyps-at-mail.nih.gov (by way of Ask-A-Microscopist)
10, 12 -- Subject: AskAMicroscopist: fixation for mitochondria
10, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: WAHeeschen-at-dow.com
Date: Wed, 10 Jan 2007 07:53:49 -0600
Subject: [Microscopy] Job Posting: Microscopy technologist for Dow Chemical in Freeport

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Microscopy Technologist Position at The Dow Chemical Company in Freeport, Texas

The Dow Chemical Company, an equal opportunity employer, is seeking applications for an immediate opening for a microscopy technologist. Candidates should have training in light and electron microscopies and related sample preparation methods. The applicant should have a minimum of an Associate Degree in Science. The successful applicant will work under the direction of a Dow Researcher to deliver microscopy related solutions to real world industrial problems. The job focus will be primarily related to microscopy of polymers including; light microscopy, scanning electron microscopy and transmission electron microscopy and soft material sample preparation to include; ultramicrotomy and cryo-ultramicrotomy. The position will be at Dow's Freeport, Texas Analytical Laboratory, which houses modern LM, SEM, TEM, SPM and surface science equipment.

The salary is commensurate with experience. Please send hardcopy resume' to:
Dr. John Blackson
Building 1897 E29C
The Dow Chemical Company
Midland, MI 48667

Electronic resume' can be sent to
tdominowski-at-dow.com


Best Regards,

Bill
William A. Heeschen, Ph.D.
Building 1897 E84
The Dow Chemical Company
Midland, MI 48667
waheeschen-at-dow.com

Dow is a leader in science and technology, providing innovative chemical, plastic and agricultural products and services to many essential consumer markets. With annual sales of $40 billion, Dow serves customers in 175 countries and a wide range of markets that are vital to human progress: food, transportation, health and medicine, personal and home care, and building and construction, among others. Committed to the principles of sustainable development, Dow and its 43,000 employees seek to balance economic, environmental and social responsibilities. References to "Dow" or the "Company" mean The Dow Chemical Company and its consolidated subsidiaries unless otherwise expressly noted.

==============================Original Headers==============================
8, 23 -- From WAHeeschen-at-dow.com Wed Jan 10 07:53:49 2007
8, 23 -- Received: from mail86.messagelabs.com (mail86.messagelabs.com [216.82.244.115])
8, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0ADrmHG005876
8, 23 -- for {Microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 07:53:49 -0600
8, 23 -- X-VirusChecked: Checked
8, 23 -- X-Env-Sender: WAHeeschen-at-dow.com
8, 23 -- X-Msg-Ref: server-13.tower-86.messagelabs.com!1168437227!28046730!1
8, 23 -- X-StarScan-Version: 5.5.10.7; banners=-,-,-
8, 23 -- X-Originating-IP: [216.99.65.27]
8, 23 -- Received: (qmail 19531 invoked from network); 10 Jan 2007 13:53:47 -0000
8, 23 -- Received: from mail7.dow.com (HELO mante88.nam.dow.com) (216.99.65.27)
8, 23 -- by server-13.tower-86.messagelabs.com with SMTP; 10 Jan 2007 13:53:47 -0000
8, 23 -- Received: by mante88.nam.dow.com with Internet Mail Service (5.5.2658.3)
8, 23 -- id {C4JXN6G9} ; Wed, 10 Jan 2007 08:53:46 -0500
8, 23 -- Message-ID: {9FAC91DDE67EF3448238E5E33D3E5AEF05383A-at-USMDLMDOWX026.dow.com}
8, 23 -- From: "Heeschen, Bill (WA)" {WAHeeschen-at-dow.com}
8, 23 -- To: "'Microscopy-at-microscopy.com'" {Microscopy-at-microscopy.com}
8, 23 -- Subject: Job Posting: Microscopy technologist for Dow Chemical in Freeport
8, 23 -- , TX
8, 23 -- Date: Wed, 10 Jan 2007 08:53:37 -0500
8, 23 -- MIME-Version: 1.0
8, 23 -- X-Mailer: Internet Mail Service (5.5.2658.3)
8, 23 -- Content-Type: text/plain
==============================End of - Headers==============================




From: zhangj16-at-msu.edu
Date: Wed, 10 Jan 2007 14:04:17 -0600
Subject: [Microscopy] Hitachi H800 vacuum problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear everyone

I had problems with Hitachi H800 TEM vacuum sequences now. From start up of
Evac, three rough pumps(RP) start to work. RP1 and 2 pump the valve of two
diffusion pump to open. RP3 cause the pre evac and a valve of column to
open. Then the vacuum level of column and pre evac are just stop at around
8*10^-1 pa, no further progress can be reach.

Please share your idea what may be the problems. Thanks.


Jiaming Zhang

Research Assistant
Department of Chem. Eng. & Mater. Sci. Michigan State University
Tel: 517-231-8013
Email: zhangj16-at-egr.msu.edu


==============================Original Headers==============================
7, 16 -- From zhangj16-at-msu.edu Wed Jan 10 14:04:17 2007
7, 16 -- Received: from sys30.mail.msu.edu (sys30.mail.msu.edu [35.9.75.130])
7, 16 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0AK4H5S026401
7, 16 -- for {Microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 14:04:17 -0600
7, 16 -- Received: from zhangj16 by sys30.mail.msu.edu with local (Exim 4.52 #1)
7, 16 -- id 1H4jgG-0007Ed-VX
7, 16 -- for Microscopy-at-microscopy.com; Wed, 10 Jan 2007 15:04:17 -0500
7, 16 -- From: "Jiaming Zhang" {zhangj16-at-msu.edu}
7, 16 -- To: Microscopy-at-microscopy.com
7, 16 -- Subject: Hitachi H800 vacuum problem
7, 16 -- Date: Wed, 10 Jan 2007 15:04:07 -0500
7, 16 -- Mime-Version: 1.0
7, 16 -- Content-Type: text/plain; format=flowed; charset="utf-8"
7, 16 -- Content-Transfer-Encoding: 7bit
7, 16 -- Message-Id: {E1H4jgG-0007Ed-VX-at-sys30.mail.msu.edu}
7, 16 -- X-Virus: None found by Clam AV
==============================End of - Headers==============================




From: bigelow-at-engin.umich.edu
Date: Wed, 10 Jan 2007 15:13:37 -0600
Subject: [Microscopy] RE: H800 vacuum problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

It sounds as though your vacuum system has basically the same
configuration as the one shown in Figure 9.5 of my book, Vacuum
Methods in Electron Microscopy (available from SPI Supplies, M. E.
Taylor, Ladd, etc.), except that you have diffusion pumps rather than
the turbomolecular pumps shown in this figure. The discussion given
of the operation of the system in this figure should give you a
pretty good understanding of how your system should operate.

In particular, I suspect that RP3 on your system might serve two
functions: 1. it might be used to rough pump the column during an
initial pump down operation, and 2. it probably serves to evacuate
the specimen chamber during specimen exchange operations. If these
assumptions are correct then the valves between RP3 and the column
should open during the initial stages of pumpdown, but once a
suitable vacuum (about 10-1 Pa) is reached in the column the valves
between RP3 and the column should close and remain closed, except
during specimen change operations. You might want to check the
operation of these valves on your system to be sure that they are
operating properly. If they do not close when they should they could
cause the column pressure to remain too high, as you describe. If
you check with your department's electrician he will have a gadget
that he can simply clamp around the wires leading to these valves
that will read the current flowing to them, and thus be able to
determine when they are activated and when they are not.

If the problem does not involve the RP3 system, then there must be
something wrong with the diffusion pumps. Either the valves between
them and the column are not opening as they should or the diffusion
pumps themselves are not working properly. There are many potential
problems with diffusion pumps, too many to discuss here; however,
most of them are discussed in considerable detail in Section 5.8 (p.
214) of my book. First check to be sure the valves in these systems
are operating properly, next check to be sure the DPs are getting
proper electrical power , then check that the flow of cooling water
is as it should be. Beyond that you have a really serious problem and
probably should call in the Hitachi service representative.

Good luck,

--
Wilbur C. Bigelow, Professor Emeritus
Materials Sci. & Engr., Univ. of Michigan
Ann Arbor, Michigan 48109-2136
e-mail: bigelow-at-engin.umich.edu;
Fx:734-763-4788; Ph:734-975-0858
Address mail to: 2911 Whittier Court
Ann Arbor, MI 48104-6731

==============================Original Headers==============================
5, 14 -- From bigelow-at-engin.umich.edu Wed Jan 10 15:13:36 2007
5, 14 -- Received: from srvr22.engin.umich.edu (srvr22.engin.umich.edu [141.213.75.21])
5, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0ALDZug006727
5, 14 -- for {microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 15:13:36 -0600
5, 14 -- Received: from [141.212.131.221] (bigelow-g4.engin.umich.edu [141.212.131.221])
5, 14 -- by srvr22.engin.umich.edu (8.13.6/8.13.6) with ESMTP id l0ALDYrn016264
5, 14 -- for {microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 16:13:34 -0500 (EST)
5, 14 -- Mime-Version: 1.0
5, 14 -- Message-Id: {p06210200c1cafe775158-at-[141.212.131.221]}
5, 14 -- Date: Wed, 10 Jan 2007 16:13:33 -0500
5, 14 -- To: Microscopy Listserver {microscopy-at-microscopy.com}
5, 14 -- From: Wil Bigelow {bigelow-at-engin.umich.edu}
5, 14 -- Subject: RE: H800 vacuum problem
5, 14 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
==============================End of - Headers==============================




From: emlabservices-at-cox.net
Date: Wed, 10 Jan 2007 16:03:42 -0600
Subject: [Microscopy] Hitachi H800 vacuum problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hello Jiaming,

The vacuum level stated in your message indicates only roughing pressure. I
would check to see if both diffusion pump heaters are hot (don't touch these
directly with your hand; see if you can sense warmth by holding your hand an
inch or so away from the base of the pump), the sides of the diffusion pumps
are cool (confirming you have adequate cooling water; again be cautious),
and if all this seems ok, clean your penning head as the vacuum system
sequence relies on this reading to continue on to high vacuum pumping and
appropriate valving. I think your microscope instruction manual details a
way to do this cleaning.

Good luck !

Bob Roberts
EM Lab Services, Inc.
449 NW 62nd St.
Topeka, Kansas 66617-1780
785.246.1232 voice
785.246.0168 fax
www.emlabservices.com



==============================Original Headers==============================
6, 26 -- From emlabservices-at-cox.net Wed Jan 10 16:03:41 2007
6, 26 -- Received: from centrmmtao02.cox.net (centrmmtao02.cox.net [70.168.83.82])
6, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0AM3eWM018806
6, 26 -- for {microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 16:03:41 -0600
6, 26 -- Received: from eastrmimpo01.cox.net ([68.1.16.119]) by centrmmtao02.cox.net
6, 26 -- (InterMail vM.6.01.06.03 201-2131-130-104-20060516) with ESMTP
6, 26 -- id {20070110220340.MQPF25837.centrmmtao02.cox.net-at-eastrmimpo01.cox.net}
6, 26 -- for {microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 17:03:40 -0500
6, 26 -- Received: from EMLabServices ([24.255.213.54])
6, 26 -- by eastrmimpo01.cox.net with bizsmtp
6, 26 -- id 9a2A1W0081AzDvc0000000; Wed, 10 Jan 2007 17:02:17 -0500
6, 26 -- Message-ID: {006c01c73503$2f0836f0$6400a8c0-at-EMLabServices}
6, 26 -- From: "EM Lab Services" {emlabservices-at-cox.net}
6, 26 -- To: {Microscopy-at-microscopy.com}
6, 26 -- Subject: [Microscopy] Hitachi H800 vacuum problem
6, 26 -- Date: Wed, 10 Jan 2007 16:03:36 -0600
6, 26 -- MIME-Version: 1.0
6, 26 -- Content-Type: text/plain;
6, 26 -- format=flowed;
6, 26 -- charset="iso-8859-1";
6, 26 -- reply-type=original
6, 26 -- Content-Transfer-Encoding: 7bit
6, 26 -- X-Priority: 3
6, 26 -- X-MSMail-Priority: Normal
6, 26 -- X-Mailer: Microsoft Outlook Express 6.00.2900.3028
6, 26 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: smiller-at-umr.edu
Date: Wed, 10 Jan 2007 17:13:27 -0600
Subject: [Microscopy] TEM postdoctoral position

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Post-Doctoral Fellowship in TEM and Materials Characterization
Materials Science and Engineering Department
University of Missouri-Rolla

The University of Missouri-Rolla, Department of Materials Science and Engineering, is seeking candidates for a postdoctoral fellowship position in transmission electron microscopy of thin films and hard materials. In addition to experience in TEM sample preparation and analysis, the ability to utilize other materials characterization techniques such as Auger and x-ray photoelectron spectroscopies and scanning electron microscopy is also desired. Experience in the characterization of interfaces is also beneficial, though not required. Candidate must be able to work independently and in a group, have excellent communication skills, and provide guidance to team members. The project is a part of a multi-university research initiative on diamond coatings.

A Ph.D. degree in materials science or a related field is preferred; persons with a MS degree may be acceptable with extensive background in TEM. A minimum of 2 years of experience operating a TEM, analyzing images and electron diffraction results, and preparation of TEM samples is required.

Interested candidates should send a Resume/Curriculum Vitae, Letter of Intent, and References to: Prof. Matt O'Keefe, 302 Materials Research Center, University of Missouri-Rolla, Rolla, MO 65409 USA. Email: mjokeefe-at-umr.edu




==============================Original Headers==============================
6, 32 -- From smiller-at-umr.edu Wed Jan 10 17:13:26 2007
6, 32 -- Received: from smtpout2.cc.umr.edu (smtpout2.cc.umr.edu [131.151.0.94])
6, 32 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0ANDNov031485
6, 32 -- for {Microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 17:13:25 -0600
6, 32 -- Received: (qmail 20051 invoked from network); 10 Jan 2007 23:13:20 -0000
6, 32 -- Received: from scansrv2.cc.umr.edu (131.151.1.114)
6, 32 -- by smtpout-ipvs.cc.umr.edu with SMTP; 10 Jan 2007 23:13:20 -0000
6, 32 -- Received: (qmail 25870 invoked from network); 10 Jan 2007 23:13:19 -0000
6, 32 -- Received: from smtp1.cc.umr.edu (131.151.1.43)
6, 32 -- by scanout-ipvs.cc.umr.edu with SMTP; 10 Jan 2007 23:13:19 -0000
6, 32 -- Received: from umr-exproto2.cc.umr.edu (umr-exproto2.cc.umr.edu [131.151.0.192])
6, 32 -- by smtp1.cc.umr.edu (8.13.1/8.13.1) with ESMTP id l0ANDJ2V008604
6, 32 -- for {Microscopy-at-microscopy.com} ; Wed, 10 Jan 2007 17:13:19 -0600
6, 32 -- Received: from UMR-CMAIL3.umr.edu ([131.151.1.77]) by umr-exproto2.cc.umr.edu with Microsoft SMTPSVC(6.0.3790.1830);
6, 32 -- Wed, 10 Jan 2007 17:13:19 -0600
6, 32 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
6, 32 -- Content-class: urn:content-classes:message
6, 32 -- MIME-Version: 1.0
6, 32 -- Content-Type: text/plain;
6, 32 -- charset="iso-8859-1"
6, 32 -- Subject: TEM postdoctoral position
6, 32 -- Date: Wed, 10 Jan 2007 17:13:18 -0600
6, 32 -- Message-ID: {0FE4BF38C44B644F8E12B86EFC392F86031EAA-at-UMR-CMAIL3.umr.edu}
6, 32 -- X-MS-Has-Attach:
6, 32 -- X-MS-TNEF-Correlator:
6, 32 -- Thread-Topic: TEM postdoctoral position
6, 32 -- Thread-Index: Acc1DOmyuPrjYjTFEce63WJENA16mA==
6, 32 -- From: "Miller, F. Scott" {smiller-at-umr.edu}
6, 32 -- To: {Microscopy-at-microscopy.com}
6, 32 -- X-OriginalArrivalTime: 10 Jan 2007 23:13:19.0166 (UTC) FILETIME=[EA6425E0:01C7350C]
6, 32 -- Content-Transfer-Encoding: 8bit
6, 32 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0ANDNov031485
==============================End of - Headers==============================




From: johnf-at-geology.wisc.edu
Date: Thu, 11 Jan 2007 05:16:37 -0600
Subject: [Microscopy] HKL ebsd image question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

If anyone on the list is using the HKL ebsd system, please contact me
at johnf-at-geology.wisc.edu
I have a question regarding the quality of backscatter vs forescatter
image display from the detectors sitting above and below the camera.

Thanks.

John

==============================Original Headers==============================
3, 24 -- From johnf-at-geology.wisc.edu Thu Jan 11 05:16:37 2007
3, 24 -- Received: from ice.geology.wisc.edu (ice.geology.wisc.edu [144.92.206.14])
3, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0BBGbkp022802
3, 24 -- for {microscopy-at-microscopy.com} ; Thu, 11 Jan 2007 05:16:37 -0600
3, 24 -- Received: from localhost (localhost [127.0.0.1])
3, 24 -- by localhost (Postfix) with ESMTP id BCBF520D1F
3, 24 -- for {microscopy-at-microscopy.com} ; Thu, 11 Jan 2007 05:16:36 -0600 (CST)
3, 24 -- Received: from ice.geology.wisc.edu ([127.0.0.1])
3, 24 -- by localhost (geology.wisc.edu [127.0.0.1]) (amavisd-new, port 10024)
3, 24 -- with ESMTP id 02001-04 for {microscopy-at-microscopy.com} ;
3, 24 -- Thu, 11 Jan 2007 05:16:31 -0600 (CST)
3, 24 -- Received: from [192.168.0.122] (dyn-192-16.vpn.wisc.edu [146.151.192.16])
3, 24 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
3, 24 -- (No client certificate requested)
3, 24 -- by ice.geology.wisc.edu (Postfix) with ESMTP id 9664720D07
3, 24 -- for {microscopy-at-microscopy.com} ; Thu, 11 Jan 2007 05:16:31 -0600 (CST)
3, 24 -- Mime-Version: 1.0
3, 24 -- Message-Id: {p06210202c1cbcc7bd7f3-at-[192.168.0.122]}
3, 24 -- Date: Thu, 11 Jan 2007 07:46:27 -0330
3, 24 -- To: microscopy-at-microscopy.com
3, 24 -- From: John Fournelle {johnf-at-geology.wisc.edu}
3, 24 -- Subject: HKL ebsd image question
3, 24 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
3, 24 -- X-Virus-Scanned: amavisd-new at geology.wisc.edu
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Thu, 11 Jan 2007 09:05:43 -0600
Subject: [Microscopy] Re: AskAMicroscopist: fixation for mitochondria

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Pat,

If what you mean by "best fixation" is a fixation
which does not allow artifacts or difformations of the
structure, apart from cryo-fixation I think that your
fixation is optimal (and is pretty classical).
Personally I have a preference for the addition of a
low concentration of formaldehyde (2%) because it
penetrates faster than glutar in the tissue. It acts
as a sort of "pre-fixative" before the glutar comes
and stabilizes the structures.

If you mean "increasing contrasting of the membranes",
you could consider using ferrocyanide (which I tried
successfully) and/or tannic acid (which I didn't try
but is well documented).

I found this reference for you but there are probably
others:
Berryman MA, et al (1992) Effects of tannic acid on
antigenicity and membrane contrast in ultrastructural
immunocytochemistry. J Histochem Cytochem 40, 6,
845-857

Regards or LG as they say here in Austria

Stephane


--- connellyps-at-mail.nih.gov wrote:

}
}
}
}
----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The
} Microscopy Society of America
} To Subscribe/Unsubscribe --
} http://www.microscopy.com/MicroscopyListserver
} On-Line Help
}
http://www.microscopy.com/MicroscopyListserver/FAQ.html
}
----------------------------------------------------------------------------
}
} This Question was submitted to Ask-A-Microscopist by
} (connellyps-at-mail.nih.gov)
} from
}
http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html
} on Tuesday, January 9, 2007 at 17:33:37
} Remember to consider the Grade/Age of the student
} when considering the Question
}
---------------------------------------------------------------------------
} Please reply to both connellyps-at-mail.nih.gov as
} well as to the Microscopy Listserver
}
---------------------------------------------------------------------------
}
} Email: connellyps-at-mail.nih.gov
} Name: Pat Connelly
}
} Organization: NIH
}
} Education: Graduate College
}
} Location: Bethesda, MD
}
} Title: fixation for mitochondria
}
} Question: Dear Fellow Experts,
}
} I am in need of the "best" fixation for mitochondria
} in mouse and rat muscle and kidney. Past experiments
} have yeilded OK but not great results using 2.5%glut
} with and without paraformaldehyde in sodium
} cacodylate buffer followed by 1% OsO4 and UA block
} stain. I'd like to be saying "Wow, Look at these!"
} Any suggestions or references? Today I did a Pub
} Med search and most of what I found was the same as
} I have been doing.
}
} Thanks,
} Pat Connelly
} connellyps-at-mail.nih.gov
}
}
---------------------------------------------------------------------------
}
} ==============================Original
} Headers==============================
} 10, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan
} 9 19:35:00 2007
} 10, 12 -- Received: from [206.69.208.22]
} (mac22.zaluzec.com [206.69.208.22])
} 10, 12 -- by ns.microscopy.com
} (8.12.11.20060308/8.12.8) with ESMTP id
} l0A1Yxi3014862
} 10, 12 -- for {microscopy-at-microscopy.com} ; Tue, 9
} Jan 2007 19:35:00 -0600
} 10, 12 -- Mime-Version: 1.0
} 10, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com
} (Unverified)
} 10, 12 -- Message-Id:
} {p06110401c1c9f33021fc-at-[206.69.208.22]}
} 10, 12 -- Date: Tue, 9 Jan 2007 19:34:58 -0600
} 10, 12 -- To: microscopy-at-microscopy.com
} 10, 12 -- From: connellyps-at-mail.nih.gov (by way of
} Ask-A-Microscopist)
} 10, 12 -- Subject: AskAMicroscopist: fixation for
} mitochondria
} 10, 12 -- Content-Type: text/plain;
} charset="us-ascii"
} ==============================End of -
} Headers==============================
}




____________________________________________________________________________________
Any questions? Get answers on any topic at
www.Answers.yahoo.com. Try it now.



____________________________________________________________________________________
Do you Yahoo!?
Everyone is raving about the all-new Yahoo! Mail beta.
http://new.mail.yahoo.com

==============================Original Headers==============================
14, 20 -- From nizets2-at-yahoo.com Thu Jan 11 09:05:42 2007
14, 20 -- Received: from web37404.mail.mud.yahoo.com (web37404.mail.mud.yahoo.com [209.191.91.136])
14, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0BF5fSZ005801
14, 20 -- for {microscopy-at-microscopy.com} ; Thu, 11 Jan 2007 09:05:42 -0600
14, 20 -- Received: (qmail 63482 invoked by uid 60001); 11 Jan 2007 15:05:39 -0000
14, 20 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
14, 20 -- s=s1024; d=yahoo.com;
14, 20 -- h=Received:Date:From:Subject:To:Cc:In-Reply-To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
14, 20 -- b=o5tVbGn23hSH3z6CRVLjoyWzV9top0IRPXe4qip8oENP+LnVHbJP57XOFCof212p1Im11YjpytjOf23oojveAba5ip3SQjwKsxfioVmRLAcvjgQDw0Aj+eg8kRUpq3REL9enFbZNtV/pQr0HWktUlPrCiohcY16p5ZHr+VAo6ic=;
14, 20 -- Received: from [80.122.101.102] by web37404.mail.mud.yahoo.com via HTTP; Thu, 11 Jan 2007 07:05:39 PST
14, 20 -- Date: Thu, 11 Jan 2007 07:05:39 -0800 (PST)
14, 20 -- From: Stephane Nizet {nizets2-at-yahoo.com}
14, 20 -- Subject: Re: [Microscopy] AskAMicroscopist: fixation for mitochondria
14, 20 -- To: connellyps-at-mail.nih.gov
14, 20 -- Cc: microscopy-at-microscopy.com
14, 20 -- In-Reply-To: {200701100139.l0A1dAlE022824-at-ns.microscopy.com}
14, 20 -- MIME-Version: 1.0
14, 20 -- Content-Type: text/plain; charset=iso-8859-1
14, 20 -- Content-Transfer-Encoding: 8bit
14, 20 -- Message-ID: {729801.62970.qm-at-web37404.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: mcauliff-at-umdnj.edu
Date: Thu, 11 Jan 2007 09:29:07 -0600
Subject: [Microscopy] Re: fixation for mitochondria

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Pat:

I agree with Stephane, there is nothing wrong with the fixative you
are using. If more contrast is what you seek you could also try a
post-fix or mordant in potassium dichromate. Substituting Dalton's
chrome-osmium fix for your regular osmium post-fix might be worth a try.
Alternatively, you could try a "light" fix followed by
histochemistry to demo mitochondrial enzymes, they post fix with osmium.
This would be more work.

Geoff

connellyps-at-mail.nih.gov wrote:

}
} Email: connellyps-at-mail.nih.gov
} Name: Pat Connelly
}
} Organization: NIH
}
} Education: Graduate College
}
} Location: Bethesda, MD
}
} Title: fixation for mitochondria
}
} Question: Dear Fellow Experts,
}
} I am in need of the "best" fixation for mitochondria in mouse and rat muscle and kidney. Past experiments have yeilded OK but not great results using 2.5%glut with and without paraformaldehyde in sodium cacodylate buffer followed by 1% OsO4 and UA block stain. I'd like to be saying "Wow, Look at these!"
} Any suggestions or references? Today I did a Pub Med search and most of what I found was the same as I have been doing.
}
} Thanks,
} Pat Connelly
} connellyps-at-mail.nih.gov
}
} ---------------------------------------------------------------------------
}
} ==============================Original Headers==============================
} 10, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 9 19:35:00 2007
} 10, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
} 10, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0A1Yxi3014862
} 10, 12 -- for {microscopy-at-microscopy.com} ; Tue, 9 Jan 2007 19:35:00 -0600
} 10, 12 -- Mime-Version: 1.0
} 10, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
} 10, 12 -- Message-Id: {p06110401c1c9f33021fc-at-[206.69.208.22]}
} 10, 12 -- Date: Tue, 9 Jan 2007 19:34:58 -0600
} 10, 12 -- To: microscopy-at-microscopy.com
} 10, 12 -- From: connellyps-at-mail.nih.gov (by way of Ask-A-Microscopist)
} 10, 12 -- Subject: AskAMicroscopist: fixation for mitochondria
} 10, 12 -- Content-Type: text/plain; charset="us-ascii"
} ==============================End of - Headers==============================
}
}
}
}


--
--
**********************************************
Geoff McAuliffe, Ph.D.
Neuroscience and Cell Biology
Robert Wood Johnson Medical School
675 Hoes Lane, Piscataway, NJ 08854
voice: (732)-235-4583; fax: -4029
mcauliff-at-umdnj.edu
**********************************************



==============================Original Headers==============================
9, 35 -- From mcauliff-at-umdnj.edu Thu Jan 11 09:29:07 2007
9, 35 -- Received: from zix04.umdnj.edu (zix04.UMDNJ.EDU [130.219.34.127])
9, 35 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0BFT6Ri017223
9, 35 -- for {microscopy-at-msa.microscopy.com} ; Thu, 11 Jan 2007 09:29:06 -0600
9, 35 -- Received: from zix04.umdnj.edu (ZixVPM [127.0.0.1])
9, 35 -- by Outbound.umdnj.edu (Proprietary) with ESMTP id 395204BEB8
9, 35 -- for {microscopy-at-msa.microscopy.com} ; Thu, 11 Jan 2007 10:29:06 -0500 (EST)
9, 35 -- Received: from umdnj.edu (imail.UMDNJ.EDU [130.219.34.129])
9, 35 -- by zix04.umdnj.edu (Proprietary) with ESMTP id 8AF234BECD
9, 35 -- for {microscopy-at-msa.microscopy.com} ; Thu, 11 Jan 2007 10:29:01 -0500 (EST)
9, 35 -- Received: from ([130.219.34.131])
9, 35 -- by imail.umdnj.edu with ESMTP id KP-BRACD.124903795;
9, 35 -- Thu, 11 Jan 2007 10:27:54 -0500
9, 35 -- Received: from conversion-daemon.Polaris.umdnj.edu by Polaris.umdnj.edu
9, 35 -- (iPlanet Messaging Server 5.2 HotFix 2.02 (built Oct 21 2004))
9, 35 -- id {0JBP00F01MQTBQ-at-Polaris.umdnj.edu} (original mail from mcauliff-at-umdnj.edu)
9, 35 -- for microscopy-at-msa.microscopy.com; Thu, 11 Jan 2007 10:27:53 -0500 (EST)
9, 35 -- Received: from [127.0.0.1] ([10.138.2.123])
9, 35 -- by Polaris.umdnj.edu (iPlanet Messaging Server 5.2 HotFix 2.02 (built Oct 21
9, 35 -- 2004)) with ESMTP id {0JBP00C2JMYA9J-at-Polaris.umdnj.edu} ; Thu,
9, 35 -- 11 Jan 2007 10:27:47 -0500 (EST)
9, 35 -- Date: Thu, 11 Jan 2007 10:29:07 -0500
9, 35 -- From: Geoff McAuliffe {mcauliff-at-umdnj.edu}
9, 35 -- Subject: Re: fixation for mitochondria
9, 35 -- In-reply-to: {200701100136.l0A1aDAo016699-at-ns.microscopy.com}
9, 35 -- To: connellyps-at-mail.nih.gov,
9, 35 -- MicroscopyListserver {microscopy-at-msa.microscopy.com}
9, 35 -- Message-id: {45A657C3.4040907-at-umdnj.edu}
9, 35 -- MIME-version: 1.0
9, 35 -- Content-type: text/plain; format=flowed; charset=ISO-8859-1
9, 35 -- Content-transfer-encoding: 7BIT
9, 35 -- X-Accept-Language: en-us, en
9, 35 -- User-Agent: Mozilla/5.0 (Windows; U; Windows NT 5.1; en-US; rv:1.7.2)
9, 35 -- Gecko/20040804 Netscape/7.2 (ax)
9, 35 -- References: {200701100136.l0A1aDAo016699-at-ns.microscopy.com}
==============================End of - Headers==============================




From: sshroff-at-bmrn.com
Date: Thu, 11 Jan 2007 18:16:32 -0600
Subject: [Microscopy] viaWWW: peptide diffusion

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both sshroff-at-bmrn.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: sshroff-at-bmrn.com
Name: Shilpa

Organization: BioMarin

Title-Subject: [Filtered] peptide diffusion

Question: Hi -

I am looking at antigen presentation on the surface of a cell line. I am first incubating the cell with a protein (for about 2hrs). The cells are fixed after the incubation in 4% paraformaldehyde, permeabilized in TritonX.

If my protein were fragmented, would I have to worry about loosing the small peptides even though I fixed in paraformaldehyde? I am staining the protien with a fluorescent polyclonal to check its internalization

thanks
-shilpa

---------------------------------------------------------------------------

==============================Original Headers==============================
9, 12 -- From zaluzec-at-microscopy.com Thu Jan 11 18:16:31 2007
9, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
9, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0C0GUsX008394
9, 12 -- for {microscopy-at-microscopy.com} ; Thu, 11 Jan 2007 18:16:31 -0600
9, 12 -- Mime-Version: 1.0
9, 12 -- X-Sender: (Unverified)
9, 12 -- Message-Id: {p06110402c1cc83ce9c1b-at-[206.69.208.22]}
9, 12 -- Date: Thu, 11 Jan 2007 18:16:29 -0600
9, 12 -- To: microscopy-at-microscopy.com
9, 12 -- From: sshroff-at-bmrn.com (by way of MicroscopyListserver)
9, 12 -- Subject: viaWWW: peptide diffusion
9, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: phillipst-at-missouri.edu
Date: Thu, 11 Jan 2007 18:23:19 -0600
Subject: [Microscopy] Re: viaWWW: peptide diffusion

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

One other concern would be that your protein was cross-linked to another
protein that was then extracted by the Triton X-100.

At 06:17 PM 01/11/07, you wrote:



} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America

Thomas E. Phillips, PhD
Professor of Biological Sciences
Director, Molecular Cytology Core
2 Tucker Hall
University of Missouri
Columbia, MO 65211-7400

573-882-4712 (office)
573-882-0123 (fax)
PhillipsT-at-missouri.edu
http://www.biology.missouri.edu/faculty/phillips.html
http://www.biotech.missouri.edu/mcc/


==============================Original Headers==============================
8, 21 -- From PhillipsT-at-missouri.edu Thu Jan 11 18:23:19 2007
8, 21 -- Received: from um-nsmtpout1.um.umsystem.edu (um-nsmtpout1.um.umsystem.edu [209.106.228.53])
8, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0C0NJK5015971
8, 21 -- for {Microscopy-at-msa.microscopy.com} ; Thu, 11 Jan 2007 18:23:19 -0600
8, 21 -- Received: from um-tsmtpout1.um.umsystem.edu ([209.106.228.23]) by um-nsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
8, 21 -- Thu, 11 Jan 2007 18:23:19 -0600
8, 21 -- Received: from phillips-dell.missouri.edu ([128.206.81.132]) by um-tsmtpout1.um.umsystem.edu over TLS secured channel with Microsoft SMTPSVC(6.0.3790.1830);
8, 21 -- Thu, 11 Jan 2007 18:23:18 -0600
8, 21 -- Message-Id: {6.0.0.22.2.20070111182352.048d5710-at-pop.missouri.edu}
8, 21 -- X-Sender: phillipst-at-pop.missouri.edu
8, 21 -- X-Mailer: QUALCOMM Windows Eudora Version 6.0.0.22
8, 21 -- Date: Thu, 11 Jan 2007 18:25:17 -0600
8, 21 -- To: {sshroff-at-bmrn.com}
8, 21 -- From: Tom Phillips {phillipst-at-missouri.edu}
8, 21 -- Subject: Re: [Microscopy] viaWWW: peptide diffusion
8, 21 -- Cc: Microscopy-at-msa.microscopy.com
8, 21 -- In-Reply-To: {200701120017.l0C0Hr8t009809-at-ns.microscopy.com}
8, 21 -- References: {200701120017.l0C0Hr8t009809-at-ns.microscopy.com}
8, 21 -- Mime-Version: 1.0
8, 21 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
8, 21 -- X-OriginalArrivalTime: 12 Jan 2007 00:23:18.0908 (UTC) FILETIME=[DC0BABC0:01C735DF]
==============================End of - Headers==============================




From: almecid-at-tcd.ie
Date: Fri, 12 Jan 2007 08:08:48 -0600
Subject: [Microscopy] viaWWW: Carbon Grids and Furnace

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both almecid-at-tcd.ie as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: almecid-at-tcd.ie
Name: Dorothee

Organization: Trinity College Dublin

Title-Subject: [Filtered] TEM grids

Question: Can carbon TEM grids withstand 450 degrees in a furnace?


---------------------------------------------------------------------------

==============================Original Headers==============================
7, 12 -- From zaluzec-at-microscopy.com Fri Jan 12 08:08:48 2007
7, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
7, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CE8kL7016513
7, 12 -- for {microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 08:08:47 -0600
7, 12 -- Mime-Version: 1.0
7, 12 -- X-Sender: (Unverified)
7, 12 -- Message-Id: {p06110403c1cd46cd50dc-at-[206.69.208.22]}
7, 12 -- Date: Fri, 12 Jan 2007 08:08:46 -0600
7, 12 -- To: microscopy-at-microscopy.com
7, 12 -- From: almecid-at-tcd.ie (by way of MicroscopyListserver)
7, 12 -- Subject: viaWWW: Carbon Grids and Furnace
7, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: phillipst-at-missouri.edu
Date: Fri, 12 Jan 2007 10:00:55 -0600
Subject: [Microscopy] Photoshop question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

I consider myself pretty skilled in Photoshop but there is one simple thing
I don't know how to do and it drives me crazy. I have searched the
Photoshop online help and other websites to no avail. I want to keep the
images anchored to the Photoshop screen frame. The problem is that I start
to zoom an image and it grows to cover the menu bar. This makes subsequent
manipulation tough. I am sure there is just some simple thing to toggle but
I can't find it. I use Photoshop CS2 (version 9). thanks for any tips you
have. tom



Thomas E. Phillips, PhD
Professor of Biological Sciences
Director, Molecular Cytology Core
2 Tucker Hall
University of Missouri
Columbia, MO 65211-7400

573-882-4712 (office)
573-882-0123 (fax)
PhillipsT-at-missouri.edu
http://www.biology.missouri.edu/faculty/phillips.html
http://www.biotech.missouri.edu/mcc/


==============================Original Headers==============================
6, 18 -- From PhillipsT-at-missouri.edu Fri Jan 12 10:00:55 2007
6, 18 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
6, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CG0trJ029917
6, 18 -- for {Microscopy-at-msa.microscopy.com} ; Fri, 12 Jan 2007 10:00:55 -0600
6, 18 -- Received: from um-tsmtpout1.um.umsystem.edu ([209.106.228.23]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
6, 18 -- Fri, 12 Jan 2007 10:00:54 -0600
6, 18 -- Received: from phillips-dell.missouri.edu ([128.206.81.132]) by um-tsmtpout1.um.umsystem.edu over TLS secured channel with Microsoft SMTPSVC(6.0.3790.1830);
6, 18 -- Fri, 12 Jan 2007 10:00:54 -0600
6, 18 -- Message-Id: {6.0.0.22.2.20070112095904.04914ad0-at-pop.missouri.edu}
6, 18 -- X-Sender: phillipst-at-pop.missouri.edu
6, 18 -- X-Mailer: QUALCOMM Windows Eudora Version 6.0.0.22
6, 18 -- Date: Fri, 12 Jan 2007 10:02:53 -0600
6, 18 -- To: Microscopy-at-msa.microscopy.com
6, 18 -- From: Tom Phillips {phillipst-at-missouri.edu}
6, 18 -- Subject: Photoshop question
6, 18 -- Mime-Version: 1.0
6, 18 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
6, 18 -- X-OriginalArrivalTime: 12 Jan 2007 16:00:54.0593 (UTC) FILETIME=[D70BFB10:01C73662]
==============================End of - Headers==============================




From: DrJohnRuss-at-aol.com
Date: Fri, 12 Jan 2007 10:05:40 -0600
Subject: [Microscopy] Re: Photoshop question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Select the magnifying glass icon. In the options bar that appears
across the top of your screen, uncheck the "resize images to fit"
item. When you zoom the image the window in which it appears will not
change size or position.


On Jan 12, 2007, at 11:01 AM, phillipst-at-missouri.edu wrote:

}
}
}
} ----------------------------------------------------------------------
} ------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of
} America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/
} MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------
} ------
}
} I consider myself pretty skilled in Photoshop but there is one
} simple thing
} I don't know how to do and it drives me crazy. I have searched the
} Photoshop online help and other websites to no avail. I want to
} keep the
} images anchored to the Photoshop screen frame. The problem is that
} I start
} to zoom an image and it grows to cover the menu bar. This makes
} subsequent
} manipulation tough. I am sure there is just some simple thing to
} toggle but
} I can't find it. I use Photoshop CS2 (version 9). thanks for any
} tips you
} have. tom
}
}
}
} Thomas E. Phillips, PhD
} Professor of Biological Sciences
} Director, Molecular Cytology Core
} 2 Tucker Hall
} University of Missouri
} Columbia, MO 65211-7400
}
} 573-882-4712 (office)
} 573-882-0123 (fax)
} PhillipsT-at-missouri.edu
} http://www.biology.missouri.edu/faculty/phillips.html
} http://www.biotech.missouri.edu/mcc/
}
}
}




==============================Original Headers==============================
7, 22 -- From DrJohnRuss-at-aol.com Fri Jan 12 10:05:40 2007
7, 22 -- Received: from imo-d05.mx.aol.com (imo-d05.mx.aol.com [205.188.157.37])
7, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CG5dBk004437
7, 22 -- for {microscopy-at-msa.microscopy.com} ; Fri, 12 Jan 2007 10:05:39 -0600
7, 22 -- Received: from DrJohnRuss-at-aol.com
7, 22 -- by imo-d05.mx.aol.com (mail_out_v38_r7.6.) id f.c14.e14d6c9 (52468);
7, 22 -- Fri, 12 Jan 2007 11:05:32 -0500 (EST)
7, 22 -- Received: from [192.168.123.187] (cpe-065-190-140-239.nc.res.rr.com [65.190.140.239]) by ciaaol-d09.mx.aol.com (v114.2) with ESMTP id MAILCIAAOLD097-ccf445a7b1c931f; Fri, 12 Jan 2007 11:05:30 -0500
7, 22 -- In-Reply-To: {200701121601.l0CG1LCo030612-at-ns.microscopy.com}
7, 22 -- References: {200701121601.l0CG1LCo030612-at-ns.microscopy.com}
7, 22 -- Mime-Version: 1.0 (Apple Message framework v752.2)
7, 22 -- Content-Type: text/plain; charset=US-ASCII; delsp=yes; format=flowed
7, 22 -- Message-Id: {3EF10D56-909A-4B3D-91CA-0CAD4B91154B-at-AOL.com}
7, 22 -- Cc: microscopy-at-msa.microscopy.com
7, 22 -- Content-Transfer-Encoding: 7bit
7, 22 -- From: John Russ {DrJohnRuss-at-aol.com}
7, 22 -- Subject: Re: [Microscopy] Photoshop question
7, 22 -- Date: Fri, 12 Jan 2007 11:05:25 -0500
7, 22 -- To: phillipst-at-missouri.edu
7, 22 -- X-Mailer: Apple Mail (2.752.2)
7, 22 -- X-AOL-IP: 65.190.140.239
7, 22 -- X-Spam-Flag: NO
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Fri, 12 Jan 2007 10:18:55 -0600
Subject: [Microscopy] TEM: Infiltration of lens

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Listers,

Does anyone have a protocol for effective infiltration of lenses encased
in retinas? Pretty please?? We have tried extended infiltration times
(up to three days) and microwave-assisted infiltrations with Epon/Spurrs
and pure Spurrs and have still not achieved good infiltration. There
are holes on either side of the retinas to help solutions move in and
out, but the lenses are still crumbling during sectioning. The retinas
are okay.

I just know somebody's out there who can help with this......

Thanks much.

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original Headers==============================
7, 23 -- From TindallR-at-missouri.edu Fri Jan 12 10:18:55 2007
7, 23 -- Received: from um-nsmtpout1.um.umsystem.edu (um-nsmtpout1.um.umsystem.edu [209.106.228.53])
7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CGItPZ019583
7, 23 -- for {microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 10:18:55 -0600
7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-nsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
7, 23 -- Fri, 12 Jan 2007 10:18:54 -0600
7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
7, 23 -- Content-class: urn:content-classes:message
7, 23 -- MIME-Version: 1.0
7, 23 -- Content-Type: text/plain;
7, 23 -- charset="us-ascii"
7, 23 -- Subject: TEM: Infiltration of lens
7, 23 -- Date: Fri, 12 Jan 2007 10:18:54 -0600
7, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68CD1-at-UM-XMAIL08.um.umsystem.edu}
7, 23 -- X-MS-Has-Attach:
7, 23 -- X-MS-TNEF-Correlator:
7, 23 -- Thread-Topic: TEM: Infiltration of lens
7, 23 -- Thread-Index: Acc2ZVpHGt3f/kCtSteTmqnj0MoVsw==
7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
7, 23 -- To: {microscopy-at-microscopy.com}
7, 23 -- X-OriginalArrivalTime: 12 Jan 2007 16:18:54.0545 (UTC) FILETIME=[5ABF9410:01C73665]
7, 23 -- Content-Transfer-Encoding: 8bit
7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0CGItPZ019583
==============================End of - Headers==============================




From: RFleck-at-nibsc.ac.uk
Date: Fri, 12 Jan 2007 10:36:24 -0600
Subject: [Microscopy] Postdoctoral Research Opportunity at NIBSC

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

National Institute for Biological Standards and Control

Hertfordshire, UK

Post-doctoral Researcher

Division of Cell Biology and Imaging

As part of a DTI initiative in Regenerative Medicine Technologies a three year post-doctoral research position has become available. The project is part of a consortium bringing together strengths in advanced imaging, robotic microwell-based engineering and combinatorial techniques. The programme aims to create a leading technology for quickly and efficiently examining large numbers of linked variables, addressing a central issue of regenerative medicine; the capacity to reproducibly control the differentiation of a high proportion of stem cells.
The successful candidate will undertake studies to investigate differentiation of cells and propose solutions to the technical challenges for scaling the patented concept. Work will require development and study of differentiation systems . Interaction between cells, the cells and carrier beads, effects of culture conditions, environment, differentiation media and the functional capacity of cells in biological assays will be considered.
We are seeking an enthusiastic post-doctoral scientist capable of pursuing productive research in molecular cell biology. You will have a Life Science PhD or appropriate subject with at least 3 years experience in microscopy in a research environment. It is essential you have experience of confocal and/or electron microscopy. In addition, you will require experience of one or more: immuno-gold/fluorescent, confocal, live cell, cryo-SEM or cryo-TEM techniques. You will be based in the recently refurbished imaging facility equipped with new high resolution electron microscopes and supporting specimen preparation equipment.

Salary will be within Pay Band E on a scale from £24,815 up to £28,478, dependent on experience.

Please see our website www.nibsc.ac.uk for a full job description.

To apply, please submit a full Curriculum Vitae (2 copies), including the names and addresses of two referees, to the Human Resources Department, National Institute for Biological Standards and Control, Blanche Lane, South Mimms, Potters Bar, Hertfordshire, EN6 3QG , or email HR-at-nibsc.ac.uk
Closing date for applications is 1 February 2007.
Please quote reference SU.CB.567 clearly on your application.




==============================Original Headers==============================
11, 25 -- From RFleck-at-nibsc.ac.uk Fri Jan 12 10:36:24 2007
11, 25 -- Received: from jack.nibsc.ac.uk (jack.nibsc.ac.uk [212.219.216.54])
11, 25 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CGaOHi030862
11, 25 -- for {Microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 10:36:24 -0600
11, 25 -- Received: from postbox.nibsc.ac.uk ([212.219.216.15] helo=postbox.ad.nibsc.ac.uk)
11, 25 -- by jack.nibsc.ac.uk with esmtp (Exim 4.63)
11, 25 -- (envelope-from {RFleck-at-nibsc.ac.uk} )
11, 25 -- id 1H5PO3-0002OS-Pp
11, 25 -- for Microscopy-at-microscopy.com; Fri, 12 Jan 2007 16:36:15 +0000
11, 25 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
11, 25 -- Content-class: urn:content-classes:message
11, 25 -- MIME-Version: 1.0
11, 25 -- Content-Type: text/plain;
11, 25 -- charset="iso-8859-1"
11, 25 -- Subject: Postdoctoral Research Opportunity at NIBSC
11, 25 -- Date: Fri, 12 Jan 2007 16:37:18 -0000
11, 25 -- Message-ID: {0DBB443159C8DA4C83E7B8B6546FC4370520AA68-at-postbox.ad.nibsc.ac.uk}
11, 25 -- X-MS-Has-Attach:
11, 25 -- X-MS-TNEF-Correlator:
11, 25 -- Thread-Topic: Postdoctoral Research Opportunity at NIBSC
11, 25 -- Thread-Index: Acc2Z+ynoEP81rtpTj6tLL+f0lJ3bA==
11, 25 -- From: "Roland Fleck" {RFleck-at-nibsc.ac.uk}
11, 25 -- To: {Microscopy-at-microscopy.com}
11, 25 -- Content-Transfer-Encoding: 8bit
11, 25 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0CGaOHi030862
==============================End of - Headers==============================




From: PAGE.BALUCH-at-asu.edu
Date: Fri, 12 Jan 2007 11:41:28 -0600
Subject: [Microscopy] TEM artifact ?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


Hi, I am new at using the TEM and preparing my adherent culture cells for
this type of imaging so am hoping that someone may have experience with this
unusual artifact that showed up in some of my cells. I had 2 sets of cells.
One set was untreated and the others were treated with membrane permeate
pseudosubstrate peptides and or nocodazole. The cells that were not treated
in any way appeared normal upon imaging but the cells that had been treated
had fiber-like artifacts throughout the cytoplasm and nuclei. The "fibers"
were localized mainly within the cells and appeared to be consistently
0.1umin length. I am assuming the artifact is originating from the
peptides but was wondering if anyone else has had this experience. I also
repeated the post staining process with new sections to rule out any
problems with lead
citrate or uranyl acetate precipitants but had the same problem. Thanks for
any advice.
Page

Page Baluch
Arizona State University/SoLS
P.O. Box 4601
Tempe, AZ 85287-4601
Lab: 480-965-7011
Fax: 480-965-7599


==============================Original Headers==============================
3, 22 -- From pbaluch-at-imap3.asu.edu Fri Jan 12 11:41:28 2007
3, 22 -- Received: from epo-int1.asu.edu (epo-int1.asu.edu [129.219.187.20])
3, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CHfR3c010746
3, 22 -- for {microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 11:41:28 -0600
3, 22 -- Received: from webmail2.asu.edu (webmail2.asu.edu [129.219.117.231])
3, 22 -- by epo-int1.asu.edu (Switch-3.1.8/Switch-3.1.7/asu-postoffice-prod) with ESMTP id l0CHewwh028440
3, 22 -- for {microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 10:40:58 -0700
3, 22 -- Received: (from emma-at-localhost)
3, 22 -- by webmail2.asu.edu (8.12.9/8.12.9/Submit) id l0CHfMsW011067
3, 22 -- for microscopy-at-microscopy.com; Fri, 12 Jan 2007 10:41:23 -0700
3, 22 -- X-Authentication-Warning: webmail2.asu.edu: emma set sender to pbaluch-at-imap3.asu.edu using -f
3, 22 -- To: microscopy-at-microscopy.com
3, 22 -- Subject: TEM artifact ?
3, 22 -- Message-ID: {1168623682.45a7c842ea514-at-webmail.asu.edu}
3, 22 -- Date: Fri, 12 Jan 2007 10:41:22 -0700 (MST)
3, 22 -- From: PAGE.BALUCH-at-asu.edu
3, 22 -- MIME-Version: 1.0
3, 22 -- Content-Type: text/plain; charset=ISO-8859-1
3, 22 -- Content-Transfer-Encoding: 8bit
3, 22 -- User-Agent: IMP/PHP IMAP webmail program 2.2.3
3, 22 -- X-Originating-IP: 149.169.133.32
3, 22 -- X-Virus-Scanned: by amavisd-new
==============================End of - Headers==============================




From: tivol-at-caltech.edu
Date: Fri, 12 Jan 2007 11:45:49 -0600
Subject: [Microscopy] Re: viaWWW: Carbon Grids and Furnace

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


On Jan 12, 2007, at 6:08 AM, almecid-at-tcd.ie wrote:

} Question: Can carbon TEM grids withstand 450 degrees in a furnace?
}
Dear Dorothee,
Carbon can certainly withstand 450 C, so if the grid itself is made of
carbon, the answer should be "yes", assuming that there are no problems
with heating or cooling at a rate such that thermal stresses are not
introduced. Carbon films on metal grids could be a different story,
however. The large difference between the coefficients of expansion of
metal and carbon will very likely cause problems.
Yours,
Bill Tivol, PhD
EM Scientist and Manager
Cryo-Electron Microscopy Facility
Broad Center, Mail Code 114-96
California Institute of Technology
Pasadena CA 91125
(626) 395-8833
tivol-at-caltech.edu


==============================Original Headers==============================
4, 22 -- From tivol-at-caltech.edu Fri Jan 12 11:45:49 2007
4, 22 -- Received: from outgoing-mail.its.caltech.edu (outgoing-mail.its.caltech.edu [131.215.239.19])
4, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CHjnIu016042
4, 22 -- for {microscopy-at-msa.microscopy.com} ; Fri, 12 Jan 2007 11:45:49 -0600
4, 22 -- Received: from water-dog.caltech.edu (water-dog [192.168.1.26])
4, 22 -- by wood-ox-postvirus (Postfix) with ESMTP
4, 22 -- id 86AF82EF5D; Fri, 12 Jan 2007 09:45:43 -0800 (PST)
4, 22 -- Received: from [192.168.159.233] (jpix-01.caltech.edu [131.215.2.133])
4, 22 -- by wood-ox.its.caltech.edu (Postfix) with ESMTP
4, 22 -- id 470BA2EF19; Fri, 12 Jan 2007 09:45:42 -0800 (PST)
4, 22 -- Mime-Version: 1.0 (Apple Message framework v624)
4, 22 -- In-Reply-To: {200701121408.l0CE8ujR016675-at-ns.microscopy.com}
4, 22 -- References: {200701121408.l0CE8ujR016675-at-ns.microscopy.com}
4, 22 -- Content-Type: text/plain; charset=US-ASCII; format=flowed
4, 22 -- Message-Id: {261ad5a527abf386fcf548bbf0241ba5-at-caltech.edu}
4, 22 -- Content-Transfer-Encoding: 7bit
4, 22 -- From: Bill Tivol {tivol-at-caltech.edu}
4, 22 -- Subject: Re: [Microscopy] viaWWW: Carbon Grids and Furnace
4, 22 -- Date: Fri, 12 Jan 2007 09:53:11 -0800
4, 22 -- To: microscopy-at-msa.microscopy.com, almecid-at-tcd.ie
4, 22 -- X-Mailer: Apple Mail (2.624)
4, 22 -- X-Spam-Scanned: at Caltech-ITS on water-dog by amavisd-2.3.3
==============================End of - Headers==============================




From: ian.anderson-at-nist.gov
Date: Fri, 12 Jan 2007 17:59:40 -0600
Subject: [Microscopy] viaWWW: Microbeam Analysis Society (MAS) 2007 Topical Workshop

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both ian.anderson-at-nist.gov as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: ian.anderson-at-nist.gov
Name: Ian Anderson

Organization: NIST

Title-Subject: [Filtered] Microbeam Analysis Society (MAS) 2007 Topical Workshop

Question: Microbeam Analysis Society (MAS)
2007 Topical Workshop
Hyperspectral Imaging II
Advanced Measurement Laboratory (AML)
May 1 ñ 4, 2007 ? NIST, Gaithersburg, MD, USA


Special provisions for students and Special provisions for students and postdocs postdocs!


http://www.microbeamanalysis.org/workshops/HI /workshops/HI-II

---------------------------------------------------------------------------


==============================Original Headers==============================
11, 14 -- From zaluzec-at-microscopy.com Fri Jan 12 17:59:40 2007
11, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
11, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0CNxdCS006984
11, 14 -- for {microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 17:59:39 -0600
11, 14 -- Mime-Version: 1.0
11, 14 -- X-Sender: (Unverified)
11, 14 -- Message-Id: {p06110400c1cdd154c901-at-[206.69.208.22]}
11, 14 -- Date: Fri, 12 Jan 2007 17:59:38 -0600
11, 14 -- To: microscopy-at-microscopy.com
11, 14 -- From: ian.anderson-at-nist.gov (by way of MicroscopyListserver)
11, 14 -- Subject: viaWWW: Microbeam Analysis Society (MAS) 2007 Topical Workshop
11, 14 -- Content-Type: text/plain; charset="iso-8859-1"
11, 14 -- Content-Transfer-Encoding: 8bit
11, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0CNxdCS006984
==============================End of - Headers==============================




From: cosandey-at-rci.rutgers.edu
Date: Fri, 12 Jan 2007 18:02:20 -0600
Subject: [Microscopy] viaWWW: PostDoc Research Associate Position

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both cosandey-at-rci.rutgers.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: cosandey-at-rci.rutgers.edu
Name: Fred Cosandey

Organization: Rutgers University

Title-Subject: [Filtered] PostDoc Research Associate Position

Question: Research Associate Position
Transmission Electron Microscopy

The Department of Materials Science and Engineering at Rutgers University is currently seeking a candidate for a Research Associate position in Transmission Electron Microscopy.
The successful candidate should have a PhD degree in Materials Science or related discipline and have extensive experience in the use of transmission electron microscope including HAADF, GIF and EELS spectroscopy techniques. Prior experience with JEOL 2010F STEM would be preferable.
Specific projects involve STEM analysis of nanostructured electrode materials used in Li-Ion batteries, including EELS determination of valence state and structural as well as chemical analysis during Li-induced phase transformations. Also, chemical and electronic structure determination of interfaces in oxides and carbide systems.

The appointment will be for one year and may be renewable for subsequent years. Interested candidate should send their CV with three references to:

Professor Frederic Cosandey
Department of Materials Science and Engineering
Rutgers University
607 Taylor Rd.
Piscataway, NJ 08854-8065,
cosandey-at-rci.rutgers.edu, 732-445-4942 (phone), 732-445-5977 (fax)

Rutgers University is an equal opportunity/affirmative action employer.



---------------------------------------------------------------------------

==============================Original Headers==============================
12, 12 -- From zaluzec-at-microscopy.com Fri Jan 12 18:02:20 2007
12, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
12, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0D02Jax008731
12, 12 -- for {microscopy-at-microscopy.com} ; Fri, 12 Jan 2007 18:02:20 -0600
12, 12 -- Mime-Version: 1.0
12, 12 -- X-Sender: (Unverified)
12, 12 -- Message-Id: {p06110401c1cdd1e4eae3-at-[206.69.208.22]}
12, 12 -- Date: Fri, 12 Jan 2007 18:02:18 -0600
12, 12 -- To: microscopy-at-microscopy.com
12, 12 -- From: cosandey-at-rci.rutgers.edu (by way of MicroscopyListserver)
12, 12 -- Subject: viaWWW: PostDoc Research Associate Position
12, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: johnf-at-geology.wisc.edu
Date: Sat, 13 Jan 2007 03:33:37 -0600
Subject: [Microscopy] Re: Microbeam Analysis Society (MAS) 2007 Topical Workshop

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Re the MAS 2007 Topical Workshop on Spectral Imaging:
if the previous url did not work, try
http://www.microprobe.org/workshops/HI-II/

sdf

==============================Original Headers==============================
2, 24 -- From johnf-at-geology.wisc.edu Sat Jan 13 03:33:37 2007
2, 24 -- Received: from ice.geology.wisc.edu (ice.geology.wisc.edu [144.92.206.14])
2, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0D9XbMF004596
2, 24 -- for {microscopy-at-microscopy.com} ; Sat, 13 Jan 2007 03:33:37 -0600
2, 24 -- Received: from localhost (localhost [127.0.0.1])
2, 24 -- by localhost (Postfix) with ESMTP id F080120D05
2, 24 -- for {microscopy-at-microscopy.com} ; Sat, 13 Jan 2007 03:33:36 -0600 (CST)
2, 24 -- Received: from ice.geology.wisc.edu ([127.0.0.1])
2, 24 -- by localhost (geology.wisc.edu [127.0.0.1]) (amavisd-new, port 10024)
2, 24 -- with ESMTP id 13966-09 for {microscopy-at-microscopy.com} ;
2, 24 -- Sat, 13 Jan 2007 03:33:34 -0600 (CST)
2, 24 -- Received: from [192.168.0.122] (dyn-192-20.vpn.wisc.edu [146.151.192.20])
2, 24 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
2, 24 -- (No client certificate requested)
2, 24 -- by ice.geology.wisc.edu (Postfix) with ESMTP id 92D9720D02
2, 24 -- for {microscopy-at-microscopy.com} ; Sat, 13 Jan 2007 03:33:33 -0600 (CST)
2, 24 -- Mime-Version: 1.0
2, 24 -- Message-Id: {p06210202c1ce5692f42b-at-[192.168.0.122]}
2, 24 -- Date: Sat, 13 Jan 2007 06:03:30 -0330
2, 24 -- To: microscopy-at-microscopy.com
2, 24 -- From: John Fournelle {johnf-at-geology.wisc.edu}
2, 24 -- Subject: Re: Microbeam Analysis Society (MAS) 2007 Topical Workshop
2, 24 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
2, 24 -- X-Virus-Scanned: amavisd-new at geology.wisc.edu
==============================End of - Headers==============================




From: michael-at-shaffer.net
Date: Sat, 13 Jan 2007 06:09:16 -0600
Subject: [Microscopy] RE: viaWWW: Microbeam Analysis Society (MAS) 2007 Topical Workshop

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

The correct link would be:

http://www.microprobe.org/workshops/HI-II/


} Special provisions for students and Special provisions for
} students and postdocs postdocs!
}
}
} http://www.microbeamanalysis.org/workshops/HI /workshops/HI-II
}


==============================Original Headers==============================
5, 21 -- From michael-at-shaffer.net Sat Jan 13 06:09:16 2007
5, 21 -- Received: from n034.sc1.he.tucows.com (smtpout1433.sc1.he.tucows.com [64.97.157.133])
5, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0DC9GvD018476
5, 21 -- for {Microscopy-at-microscopy.com} ; Sat, 13 Jan 2007 06:09:16 -0600
5, 21 -- Received: from rarewolf (205.251.83.78) by n034.sc1.he.tucows.com (7.2.069.1) (authenticated as michael-at-shaffer.net)
5, 21 -- id 45A8B00F00003B89; Sat, 13 Jan 2007 12:09:06 +0000
5, 21 -- From: "michael shaffer" {michael-at-shaffer.net}
5, 21 -- To: {ian.anderson-at-nist.gov}
5, 21 -- Cc: "MSA Microscopy list" {Microscopy-at-microscopy.com}
5, 21 -- References: {200701130000.l0D00QeE007702-at-ns.microscopy.com}
5, 21 -- Subject: RE: [Microscopy] viaWWW: Microbeam Analysis Society (MAS) 2007 Topical Workshop
5, 21 -- Date: Sat, 13 Jan 2007 08:37:05 -0330
5, 21 -- Message-ID: {005201c7370b$58db3f40$4701a8c0-at-rarewolf}
5, 21 -- MIME-Version: 1.0
5, 21 -- Content-Type: text/plain;
5, 21 -- charset="US-ASCII"
5, 21 -- Content-Transfer-Encoding: 7bit
5, 21 -- X-Mailer: Microsoft Office Outlook 11
5, 21 -- In-Reply-To: {200701130000.l0D00QeE007702-at-ns.microscopy.com}
5, 21 -- Thread-Index: Acc2pZAxXxlbkOJ4SA6cOAtSV283HAAZYHmA
5, 21 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: jl523136-at-gmail.com
Date: Sun, 14 Jan 2007 10:11:42 -0600
Subject: [Microscopy] viaWWW: EXELFS analysis to obtain RDF

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both jl523136-at-gmail.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: jl523136-at-gmail.com
Name: Juntao Li

Organization: NY

Title-Subject: [Filtered] EXELFS analysis to obtain RDF

Question: Hi, everybody,
I have a question about the EXELFS analysis.
I try to use EXELFS analysis to obtain radial distribution function of amorphous materials.
I followed the standard procedure,
1. Acquisition of high quality edge, deconvolution and background removal
2. Isolation of the oscillatory part of the intensity (normalization)
3. Scale conversion —(k) of into k space —(k)
4. Truncation of —(k)
5. Fourier distribution gives raw Radial Distribution Function (RDF)
6. Correction for phase shifts

} From step 4 to step 5 I was confused, I use Origin to the FFT, but in the image which I got, the X-axies is frequency, but in many papers I found that they got the RDF, where the x-axis is radial distance,the unit is angstrom or nm. I can not understand this, is there anybody can tell me how to get the RDF? Thanks a lot.




---------------------------------------------------------------------------


==============================Original Headers==============================
11, 14 -- From zaluzec-at-microscopy.com Sun Jan 14 10:11:42 2007
11, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
11, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0EGBfKO004132
11, 14 -- for {microscopy-at-microscopy.com} ; Sun, 14 Jan 2007 10:11:42 -0600
11, 14 -- Mime-Version: 1.0
11, 14 -- X-Sender: (Unverified)
11, 14 -- Message-Id: {p06110400c1d006ab5c04-at-[206.69.208.22]}
11, 14 -- Date: Sun, 14 Jan 2007 10:11:40 -0600
11, 14 -- To: microscopy-at-microscopy.com
11, 14 -- From: jl523136-at-gmail.com (by way of MicroscopyListserver)
11, 14 -- Subject: viaWWW: EXELFS analysis to obtain RDF
11, 14 -- Content-Type: text/plain; charset="iso-8859-1"
11, 14 -- Content-Transfer-Encoding: 8bit
11, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0EGBfKO004132
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Mon, 15 Jan 2007 06:32:38 -0600
Subject: [Microscopy] Re: viaWWW: peptide diffusion

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Shilpa,

Although I have no direct experience to answer this
question, on the basis of what is known about PFA
fixation I would say there is a chance you lose a
non-negligible part of your peptides.
Do you retain enough to allow immunodetection is a
question you can only answer by trying. A negative
signal will be hard to interpret, though, because it
could indeed come from this technical limitation.
If I were you, I'd include as a control cells
incubated with your protein at 4°C, this would allow
the protein to bind its receptor but would block the
internalization and subsequent processing (if you can
block the processing after internalization it is even
better, but you have to prove that the processing is
actually blocked - with a gel for example). Your
antibody should recognize the peptide sequence in the
context of the full protein (?) so this would allow
you to confirm that the immunodetection actually
works.

Good luck,

Stephane

--- sshroff-at-bmrn.com wrote:

}
}
}
}
----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The
} Microscopy Society of America
} To Subscribe/Unsubscribe --
} http://www.microscopy.com/MicroscopyListserver
} On-Line Help
}
http://www.microscopy.com/MicroscopyListserver/FAQ.html
}
----------------------------------------------------------------------------
}
} This Question/Comment was submitted to the
} Microscopy Listserver
} using the WWW based Form at
}
---------------------------------------------------------------------------
} Remember this posting is most likely not from a
} Subscriber, so when replying
} please copy both sshroff-at-bmrn.com as well as the
} MIcroscopy Listserver
}
---------------------------------------------------------------------------
}
} Email: sshroff-at-bmrn.com
} Name: Shilpa
}
} Organization: BioMarin
}
} Title-Subject: [Filtered] peptide diffusion
}
} Question: Hi -
}
} I am looking at antigen presentation on the surface
} of a cell line. I am first incubating the cell with
} a protein (for about 2hrs). The cells are fixed
} after the incubation in 4% paraformaldehyde,
} permeabilized in TritonX.
}
} If my protein were fragmented, would I have to worry
} about loosing the small peptides even though I fixed
} in paraformaldehyde? I am staining the protien with
} a fluorescent polyclonal to check its
} internalization
}
} thanks
} -shilpa
}
}
---------------------------------------------------------------------------
}
} ==============================Original
} Headers==============================
} 9, 12 -- From zaluzec-at-microscopy.com Thu Jan 11
} 18:16:31 2007
} 9, 12 -- Received: from [206.69.208.22]
} (mac22.zaluzec.com [206.69.208.22])
} 9, 12 -- by ns.microscopy.com
} (8.12.11.20060308/8.12.8) with ESMTP id
} l0C0GUsX008394
} 9, 12 -- for {microscopy-at-microscopy.com} ; Thu, 11
} Jan 2007 18:16:31 -0600
} 9, 12 -- Mime-Version: 1.0
} 9, 12 -- X-Sender: (Unverified)
} 9, 12 -- Message-Id:
} {p06110402c1cc83ce9c1b-at-[206.69.208.22]}
} 9, 12 -- Date: Thu, 11 Jan 2007 18:16:29 -0600
} 9, 12 -- To: microscopy-at-microscopy.com
} 9, 12 -- From: sshroff-at-bmrn.com (by way of
} MicroscopyListserver)
} 9, 12 -- Subject: viaWWW: peptide diffusion
} 9, 12 -- Content-Type: text/plain;
} charset="us-ascii"
} ==============================End of -
} Headers==============================
}




____________________________________________________________________________________
Food fight? Enjoy some healthy debate
in the Yahoo! Answers Food & Drink Q&A.
http://answers.yahoo.com/dir/?link=list&sid=396545367

==============================Original Headers==============================
9, 21 -- From nizets2-at-yahoo.com Mon Jan 15 06:32:38 2007
9, 21 -- Received: from web37404.mail.mud.yahoo.com (web37404.mail.mud.yahoo.com [209.191.91.136])
9, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0FCWbn4006357
9, 21 -- for {microscopy-at-microscopy.com} ; Mon, 15 Jan 2007 06:32:37 -0600
9, 21 -- Received: (qmail 30535 invoked by uid 60001); 15 Jan 2007 12:32:37 -0000
9, 21 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
9, 21 -- s=s1024; d=yahoo.com;
9, 21 -- h=X-YMail-OSG:Received:Date:From:Subject:To:Cc:In-Reply-To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
9, 21 -- b=1QY6nQMhnY626Zkh0FP5cmfWRlgq6SFa023kgy8Lp2m/0GYH6mx/X6I8z9QYh3Gc2jDf5jXwMrR6b3WvNfLqkSYbboPhADryPgBSlXz9OJS4DYIaGRUDKfifmSHx5fYNGCsUa3/ofQb+L1zDrz1aP6iGdLFvaRrU3Ghah4O7l14=;
9, 21 -- X-YMail-OSG: RhmN7WoVM1mmd.rM9qtcIRwllhN7nwPsG5qu.xGbhh2yvctn97Ttg_KaczWezN2myw--
9, 21 -- Received: from [80.122.101.102] by web37404.mail.mud.yahoo.com via HTTP; Mon, 15 Jan 2007 04:32:36 PST
9, 21 -- Date: Mon, 15 Jan 2007 04:32:36 -0800 (PST)
9, 21 -- From: Stephane Nizet {nizets2-at-yahoo.com}
9, 21 -- Subject: Re: [Microscopy] viaWWW: peptide diffusion
9, 21 -- To: sshroff-at-bmrn.com
9, 21 -- Cc: microscopy-at-microscopy.com
9, 21 -- In-Reply-To: {200701120024.l0C0OVwg019903-at-ns.microscopy.com}
9, 21 -- MIME-Version: 1.0
9, 21 -- Content-Type: text/plain; charset=iso-8859-1
9, 21 -- Content-Transfer-Encoding: 8bit
9, 21 -- Message-ID: {949223.29988.qm-at-web37404.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: kim.kiely-at-hrx.com.au
Date: Tue, 16 Jan 2007 01:01:17 -0600
Subject: [Microscopy] Atom Probe Engineer at the University of Sydney

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Atom Probe Engineers - 2 positions available
Electron Microscope Unit at the University of Sydney, Australia

The Electron Microscope Unit (www.emu.usyd.edu.au) is located on the
main campus of the University of Sydney and incorporates the Australian
Key Centre for Microscopy and Microanalysis as its research and training
arm. The Key Centre is a research leader in the field and has some of
the most advanced facilities in the world for microscopy and
microanalysis. As such, it seeks to develop techniques, expertise and
understanding of the relationships between the structure and function of
physical, chemical, biological and other systems. It also aims to
provide leadership in the development of innovation and ingenuity in
Australian science and engineering, and is seeking to recruit two
outstanding Atom Probe Engineers to advance this aim.

The Atom Probe Engineers will provide the national user community with
technical support for research using atom probe tomography, particularly
as the Key Centre becomes the headquarters of a new National Microscopy
and Microanalysis Research Facility during 2007. The Engineers will work
closely with the Director and all unit staff to assure the success of
the centre's research services and programs, and their integration with
the new research facility, including national and international
collaborations. They will provide technical support to users during
their training and operation of the existing advanced atom probe and the
forthcoming laser atom probe platform.

In this position, the Engineers will create solutions to specimen
fabrication, determine sound running conditions for data acquisition,
and conduct data reconstruction and analysis. So far as is possible, we
prefer to train users in the relevant areas. In addition, there is scope
for the appointees to undertake their own research in collaboration with
the Director, and leading edge projects are available for immediate
commencement. These exciting positions will require a commitment to
technical excellence and will expose the appointees to cutting-edge
research and development in nanotechnology. The advanced atom probe
platforms are Imago LEAP(tm) instruments (www.imago.com).

The Engineers will have exceptional communication, interpersonal and
problem solving skills, and the capacity to operate sophisticated
scientific instrumentation. Skills in MS Office, email and internet will
be essential, as will strong organisational and presentation skills and
the ability to work both independently and in a team. In addition, the
appointees will be committed to understanding and servicing the
technical requirements of users, and be willing to undertake national
and international travel.

To succeed, the appointees will have a Bachelor degree with Honours, and
a background in applied physics, computer systems engineering,
electronic engineering, mechanical/mechatronic engineering or materials
engineering. A PhD in these areas will be highly advantageous.

These positions are full-time fixed term for five years, subject to the
completion of a satisfactory probation period for new appointees.
Following the initial five years, longer term appointments may be
possible subject to available funding and facility demand. Membership of
a University-approved superannuation scheme is a condition of employment
for new appointees.

Remuneration package: AU$57,161- $64,050 p.a. (which includes a base
salary Level 5 $48,302 - $54,123 p.a., leave loading and up to 17%
employer's contribution to superannuation)

Depending on expertise, skills and qualifications, the remuneration
package can be potentially higher than Level 5. This will be negotiated
with quality candidates with relevant experience.

Closing: 2 February 2007

All applications must be completed online at the following url:
http://positions.usyd.edu.au, search by position reference number
92079. Specific enquiries about the role can be directed to Kim Kiely on
+61 2 9561 9068 or k.kiely-at-usyd.edu.au.



Kim Kiely
Strategic Sourcing Specialist

E kim.kiely-at-hrx.com.au
T 02 9561 9068
ABN 97 116 399 690
Level 6, 213 Miller Street, North Sydney, NSW 2060





==============================Original Headers==============================
15, 20 -- From kim.kiely-at-hrx.com.au Tue Jan 16 01:01:16 2007
15, 20 -- Received: from mail.hrx.com.au (210-193-133-5.macquarie.net.au [210.193.133.5] (may be forged))
15, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0G71FtJ006159
15, 20 -- for {Microscopy-at-Microscopy.Com} ; Tue, 16 Jan 2007 01:01:16 -0600
15, 20 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
15, 20 -- Content-class: urn:content-classes:message
15, 20 -- MIME-Version: 1.0
15, 20 -- Content-Type: text/plain;
15, 20 -- charset="us-ascii"
15, 20 -- Subject: Atom Probe Engineer at the University of Sydney
15, 20 -- Date: Tue, 16 Jan 2007 18:02:54 +1100
15, 20 -- Message-ID: {310E171DC233684F9BFBEBC32170AFFB7F7534-at-syd01ex0001.hrx.com.au}
15, 20 -- X-MS-Has-Attach:
15, 20 -- X-MS-TNEF-Correlator:
15, 20 -- Thread-Topic: Atom Probe Engineer at the University of Sydney
15, 20 -- Thread-Index: Acc5PFhFtzyt6o1VTvuaYkXKuJ5Bug==
15, 20 -- From: "Kim Kiely" {kim.kiely-at-hrx.com.au}
15, 20 -- To: {Microscopy-at-Microscopy.Com}
15, 20 -- Content-Transfer-Encoding: 8bit
15, 20 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0G71FtJ006159
==============================End of - Headers==============================




From: kim.kiely-at-hrx.com.au
Date: Tue, 16 Jan 2007 01:03:08 -0600
Subject: [Microscopy] Nanostructural Analysis of Steels

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Research Associate - Nanostructural Analysis of Steels
The Australian Key Centre for Microscopy and Microanalysis

The Australian Key Centre for Microscopy and Microanalysis
(www.emu.usyd.edu.au) intends to appoint an outstanding Research
Associate to conduct excellent research on a project to be conducted in
collaboration with BlueScope Steel (www.bluescopesteel.com.au). The
Research Associate will be based at the University of Sydney, and work
closely with the staff at the BlueScope Steel Research Centre in
Wollongong. The project contains scope for international travel in the
form of visits to the United States for production trials and attendance
at international conferences worldwide.

Castrip(r) is a revolutionary new method of producing steel strip
directly from liquid steel that BlueScope Steel plan to commercialise in
Australia. Currently, only plain carbon steels are produced using this
method. This project will investigate microalloying additions known to
be highly effective in conventional steels with an emphasis on nanoscale
microstructure using advanced microscopy techniques including
transmission electron microscopy (TEM) and atom probe tomography (APT).
For the first time, we will explore and optimise the effect of these
alloying additions in strip cast steels where dramatically higher
cooling rates and special deoxidation practices present opportunities
for innovative new design of steel nanostructure. Ultimately we will
perform production trials of the most promising alloys with a view to
developing design concepts for a new class of strip cast steels.

Located on the University's Camperdown campus, the Key Centre for
Microscopy and Microanalysis headquarters the National Microscopy and
Microanalysis Research Facility and is a major node of the ARC Centre of
Excellence for Design in Light Metals. Researchers have access to an
outstanding array of nanostructural analysis equipment, both within the
Unit and at our partner nodes.

The successful applicant will be expected to engage in all aspects of
the academic life of the facility. This includes approved collaboration
with facility users, other facility staff and approved involvement with
training activities. In this way, excellent opportunities for an
academically-rich experience are available through this position. In
addition, the appointee will be expected to write quality manuscripts
that are suitable for publication in peer-reviewed journals.

To succeed, a recent PhD in metallurgy, materials science, engineering,
physics, chemistry or a related discipline is essential. The position is
full-time fixed term for 2 years, subject to the completion of a
satisfactory probation period for new appointees. There is the
possibility of further offers of employment for 12 months, subject to
funding and need. Membership of a University approved superannuation
scheme is a condition of employment for new appointees. The appointment
can only commence after the Collaborating Organisation Agreement is
signed with the BlueScope Steel.

Remuneration package: AU$72,327 - $77,638 p.a. (which includes a base
salary Level A $61,117 - $65,605 p.a., leave loading and up to 17%
employer's contribution to superannuation)

Closing: 2nd February 2007

All applications must be completed online at the following url:
http://positions.usyd.edu.au and search by position reference number,
92070. Specific enquiries about the role can be directed to Kim Kiely on
+61 2 9561 9068 or k.kiely-at-usyd.edu.au.









==============================Original Headers==============================
15, 20 -- From kim.kiely-at-hrx.com.au Tue Jan 16 01:03:08 2007
15, 20 -- Received: from mail.hrx.com.au (ns1.recruitasp.com.au [210.193.133.5])
15, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0G7365f007628
15, 20 -- for {Microscopy-at-Microscopy.Com} ; Tue, 16 Jan 2007 01:03:07 -0600
15, 20 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
15, 20 -- Content-class: urn:content-classes:message
15, 20 -- MIME-Version: 1.0
15, 20 -- Content-Type: text/plain;
15, 20 -- charset="us-ascii"
15, 20 -- Subject: Nanostructural Analysis of Steels
15, 20 -- Date: Tue, 16 Jan 2007 18:04:46 +1100
15, 20 -- Message-ID: {310E171DC233684F9BFBEBC32170AFFB7F7536-at-syd01ex0001.hrx.com.au}
15, 20 -- X-MS-Has-Attach:
15, 20 -- X-MS-TNEF-Correlator:
15, 20 -- Thread-Topic: Nanostructural Analysis of Steels
15, 20 -- Thread-Index: Acc5PJsRgXhhdXR+TAOBpeT6bUDfpQ==
15, 20 -- From: "Kim Kiely" {kim.kiely-at-hrx.com.au}
15, 20 -- To: {Microscopy-at-Microscopy.Com}
15, 20 -- Content-Transfer-Encoding: 8bit
15, 20 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0G7365f007628
==============================End of - Headers==============================




From: kim.kiely-at-hrx.com.au
Date: Tue, 16 Jan 2007 01:05:39 -0600
Subject: [Microscopy] Research Associate in Advanced Microscopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


Research Associate - Advanced Microscopy
The Australian Key Centre for Microscopy and Microanalysis

The Australian Key Centre for Microscopy and Microanalysis
(www.emu.usyd.edu.au) currently has an outstanding opportunity available
for a results-driven Research Associate to conduct specialised research
investigating grain boundary chemistry using state-of-the-art microscopy
and microanalysis techniques.

Located on the University's Camperdown campus, the Key Centre hosts
world-class facilities, and is renowned for its research excellence. The
Centre headquarters the National Microscopy and Microanalysis Research
Facility and is a major node of the ARC Centre of Excellence for Design
in Light Metals. Researchers have access to an outstanding array of
nanostructural analysis equipment, both within the Unit and at our
partner nodes.

The appointee will be involved in the study of the relationship between
solute atmospheres at grain boundaries and the rate of grain boundary
migration during annealing in practically important alloy systems. He or
she will lead in the development of advanced experimental techniques
utilising Focused Ion Beam (FIB) technology, Transmission Electron
Microscopy (TEM) and Atom Probe Tomography (APT). This research project
will be carried out in collaboration with Imago Scientific Instruments
(www.imago.com), and contains scope for international travel in the form
of research visits to the United States and attendance at international
conferences worldwide.

The Key Centre attracts high-profile scientists and research groups, so
an ability to communicate and work with others will be crucial. In
addition, the appointee will need an ability to write quality
manuscripts that are suitable for publication in peer-reviewed journals.

To succeed, a recent PhD in materials science, engineering, physics,
chemistry or a related discipline is essential. This is the perfect
opportunity for a recent graduate to conduct innovative research in a
specialised field that attracts significant research coverage and
funding.

The position is full-time fixed term for 2 years, subject to the
completion of a satisfactory probation period for new appointees. There
is the possibility of further offers of employment for 12 months,
subject to funding and need. Membership of a University approved
superannuation scheme is a condition of employment for new appointees.

Remuneration package: AU$72,327 - $77,638 p.a. (which includes a base
salary Level A $61,117 - $65,605 p.a., leave loading and up to 17%
employer's contribution to superannuation)

Closing: 2nd Febaury 2007

All applications must be completed online at the following url:
http://positions.usyd.edu.au and search by position reference number,
92070. Specific enquiries about the role can be directed to Kim Kiely on
+61 2 9561 9068 or k.kiely-at-usyd.edu.au.



==============================Original Headers==============================
12, 20 -- From kim.kiely-at-hrx.com.au Tue Jan 16 01:05:39 2007
12, 20 -- Received: from mail.hrx.com.au (ns1.recruitasp.com.au [210.193.133.5])
12, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0G75bZP013668
12, 20 -- for {Microscopy-at-Microscopy.Com} ; Tue, 16 Jan 2007 01:05:38 -0600
12, 20 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
12, 20 -- Content-class: urn:content-classes:message
12, 20 -- MIME-Version: 1.0
12, 20 -- Content-Type: text/plain;
12, 20 -- charset="us-ascii"
12, 20 -- Subject: Research Associate in Advanced Microscopy
12, 20 -- Date: Tue, 16 Jan 2007 18:07:17 +1100
12, 20 -- Message-ID: {310E171DC233684F9BFBEBC32170AFFB7F7537-at-syd01ex0001.hrx.com.au}
12, 20 -- X-MS-Has-Attach:
12, 20 -- X-MS-TNEF-Correlator:
12, 20 -- Thread-Topic: Research Associate in Advanced Microscopy
12, 20 -- Thread-Index: Acc5PPUzjV1w16yVQWSc36zx9JRJmw==
12, 20 -- From: "Kim Kiely" {kim.kiely-at-hrx.com.au}
12, 20 -- To: {Microscopy-at-Microscopy.Com}
12, 20 -- Content-Transfer-Encoding: 8bit
12, 20 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0G75bZP013668
==============================End of - Headers==============================




From: kim.kiely-at-hrx.com.au
Date: Tue, 16 Jan 2007 01:08:15 -0600
Subject: [Microscopy] Microscopy and Microanalysis at the University of Sydney

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Electron Microscope Unit (EMU)
Careers in Microscopy and Microanalysis

Established in 1958, The University of Sydney's Electron Microscope Unit (EMU) produces quality research services, leading research programs, and internationally recognised research training in microscopy and microanalysis. It serves as national headquarters for the NANO Major National Research Facility (see: www.nano.org.au). The Unit recognises the substantial role that microscopy and microanalysis is set to play in the next decade, as biotechnology and nanoscience have increasing impact on science and society.
The Unit has recently received significant competitive research grant funding and is undergoing major expansion. As a result, we are now seeking to recruit quality engineers and scientists with backgrounds in materials science and engineering, nanotechnology, physics, chemistry or other relevant engineering disciplines to undertake research in the Australian Key Centre for Microscopy & Microanalysis, the research and training arm of the EMU. We are presently mapping program and project requirements and have a number of positions to fill, including Postdoctoral Research Fellowships, Research Assistantships, Technical Staff Positions and PhD. Scholarships. A range of tenure in these positions is available, up to 5 years.
We are interested to hear from recent graduates at the Bachelor or PhD. level that have a passion for research and are interested in investigating structure-property relationships in advanced materials using state-of-the-art microscopy and microanalysis approaches. A common theme in this research is to understand the structure and dynamics of materials structure towards the atomic level, and to relate to the engineering design and performance of materials. Necessarily, this will involve advances in microscopy research and techniques as well as advances in design in materials. This research builds on national and international collaborations, so we are looking for researchers that are also willing to undertake travel and short secondments in the laboratories of our colleagues.

The projects relate to design in:

*    Light Alloys (alloys of Mg, Ti and Al)
*    Novel Steels Systems
*    Spintronic Systems
*    Other Nanomaterials

Demonstrated research potential is essential Research experience in relevant fields will be required in some of these positions. For more details on possible projects, see www.emu.usyd.edu.au and follow the "career" link. For more details on selection criteria, see the attachment "EOI - Generic Selection Criteria" below.

To register interest for one of these positions, follow the link http://positions.usyd.edu.au and search by position reference number 92067.  For enquiries about these roles please call Fabrice Noël on +61 2 9036 7295 or f.noel-at-usyd.edu.au.
Closing: Interested parties are urged to express interest as soon as possible to feed into the research mapping process, certainly by no later than Friday, 2 February, 2007.








==============================Original Headers==============================
12, 20 -- From kim.kiely-at-hrx.com.au Tue Jan 16 01:08:15 2007
12, 20 -- Received: from mail.hrx.com.au (210-193-133-5.macquarie.net.au [210.193.133.5] (may be forged))
12, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0G78DPC022723
12, 20 -- for {Microscopy-at-Microscopy.Com} ; Tue, 16 Jan 2007 01:08:14 -0600
12, 20 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
12, 20 -- Content-class: urn:content-classes:message
12, 20 -- MIME-Version: 1.0
12, 20 -- Content-Type: text/plain;
12, 20 -- charset="iso-8859-1"
12, 20 -- Subject: Microscopy and Microanalysis at the University of Sydney
12, 20 -- Date: Tue, 16 Jan 2007 18:09:53 +1100
12, 20 -- Message-ID: {310E171DC233684F9BFBEBC32170AFFB7F7538-at-syd01ex0001.hrx.com.au}
12, 20 -- X-MS-Has-Attach:
12, 20 -- X-MS-TNEF-Correlator:
12, 20 -- Thread-Topic: Microscopy and Microanalysis at the University of Sydney
12, 20 -- Thread-Index: Acc5PVIfM3GFH90GSwyP4NvVLCUdHw==
12, 20 -- From: "Kim Kiely" {kim.kiely-at-hrx.com.au}
12, 20 -- To: {Microscopy-at-Microscopy.Com}
12, 20 -- Content-Transfer-Encoding: 8bit
12, 20 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0G78DPC022723
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Tue, 16 Jan 2007 03:14:48 -0600
Subject: [Microscopy] preparation of cells for SEM - question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear listers,

I have a request for experts in SEM who have some
experience with eukaryotic cells.
We have a basic SEM which is normally used to study
minerals, so we are not equipped with critical point
dryers and the like. Now I would like to have a look
at some cells in culture with the SEM but I have no
idea how I could prepare the cells without critical
point dryer. Of course if I simply fix and dehydrate
in acetone, then air dry the cells I expect the
structure to crumble. Well perhaps it won't give nice
pictures but perhaps it will be enough for what I want
to see. On the other hand perhaps I could keep them in
a semi-hydrated state and look at them at 10 Pa vaccum
(whatever "semi-hydrated state means ;-)).

I would be grateful if you cared to share your opinion
on this subjet with me.

Stephane



____________________________________________________________________________________
Bored stiff? Loosen up...
Download and play hundreds of games for free on Yahoo! Games.
http://games.yahoo.com/games/front

==============================Original Headers==============================
6, 19 -- From nizets2-at-yahoo.com Tue Jan 16 03:14:48 2007
6, 19 -- Received: from web37411.mail.mud.yahoo.com (web37411.mail.mud.yahoo.com [209.191.91.143])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0G9Emoo019552
6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 03:14:48 -0600
6, 19 -- Received: (qmail 62303 invoked by uid 60001); 16 Jan 2007 09:14:47 -0000
6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 -- h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
6, 19 -- b=kpkcnoBDUHMcVkk3aIfi3ch5Ffk324ZCkbIAnLfSxMO5yzLGCpa74RgxqXcpb6rCkUqNh6CevqxvS783hV8rc5yRiXjGPlhG+sMnVSwX/qheKVKYH8tg8dOuIcXQ0xvXvOS/v0G8SSlqv8teAEkzdXQKlgJxBBldpWTS2jBvJxk=;
6, 19 -- X-YMail-OSG: 6E_tvvQVM1lhn5yjbcRFZ33VO8xzQy.Y0.qekDzkGSSd9Vs3SDF0WHtJsnif0Jao410DY2csNk2MjHuUwU1.68o6dDetO9BirDlwIulsuZg6gJF0FQgVpzzHlnJ9St8l8kxEzjvOlJUvb3w-
6, 19 -- Received: from [80.122.101.102] by web37411.mail.mud.yahoo.com via HTTP; Tue, 16 Jan 2007 01:14:47 PST
6, 19 -- Date: Tue, 16 Jan 2007 01:14:47 -0800 (PST)
6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: preparation of cells for SEM - question
6, 19 -- To: microscopy-at-microscopy.com
6, 19 -- MIME-Version: 1.0
6, 19 -- Content-Type: text/plain; charset=iso-8859-1
6, 19 -- Content-Transfer-Encoding: 8bit
6, 19 -- Message-ID: {636541.62156.qm-at-web37411.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: David.Patton-at-uwe.ac.uk
Date: Tue, 16 Jan 2007 03:36:06 -0600
Subject: [Microscopy] preparation of cells for SEM - question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Stephane, I have always found your posts interesting so it is nice to be
able to help you. I think critical point drying is best for cell
preparation but we tend to use a hexamethyldisilazane (HMDS) dehydration
protocol as it is much easier.

I think you will find looking at wet cells problematic. With a tungsten
gun XL30 ESEM I find it very difficult to get adequate resolution with
cultured cells.

My HMDS protocol is given below. If anyone would like to suggest
improvements please fire away. Although fixing in PBS goes against EM
principals the results for modest magnification SEM are often good.

Dave


Preparation of dry samples for SEM using HMDS

Safety
Work in fume hood and use gloves
Retain ethanol and HMDS (hexamethyldisilazane) for disposal in
non-chlorinated waste bottle

Fixation

Fix in 4% glutaraldehyde in buffer (usually 0.1M )*
Leave for 1 hr at room temperature or 24hrs in the fridge
Rinse in buffer x3 (total storage time should exceed fixation time by a
factor of 3).

SEM Dehydration

5m in 20% ethanol
5m in 30% ethanol
5m in 50% ethanol
5m in 70% ethanol
5m in 90% ethanol
5m in 100% ethanol
5m in 100% ethanol


5m in 100% ethanol / HMDS (2:1)
5m in 100% ethanol / HMDS (2:2)
5m in 100% ethanol / HMDS (1:2)
5m in 100% HMDS
5m in 100% HMDS
5m in 100% HMDS


Remove specimen from HMDS and place on filter paper in a petri dish and
leave in fume hood until dry.

*For 4% glutaraldehyde in 0.1M buffer
Take, say, 50mls of 0.2M buffer, 34ml d water and 16ml of 25%
glutaraldehyde.

0.1M Phosphate buffer or sodium cacodylate are used.

Fixation Using PBS
8.4ml PBS
1.6ml glutaraldehyde

Gives 10ml of a 4% glutaraldehyde fixative




-----Original Message-----
X-from: nizets2-at-yahoo.com [mailto:nizets2-at-yahoo.com]
Sent: 16 January 2007 09:20
To: David Patton

Dear listers,

I have a request for experts in SEM who have some
experience with eukaryotic cells.
We have a basic SEM which is normally used to study
minerals, so we are not equipped with critical point
dryers and the like. Now I would like to have a look
at some cells in culture with the SEM but I have no
idea how I could prepare the cells without critical
point dryer. Of course if I simply fix and dehydrate
in acetone, then air dry the cells I expect the
structure to crumble. Well perhaps it won't give nice
pictures but perhaps it will be enough for what I want
to see. On the other hand perhaps I could keep them in
a semi-hydrated state and look at them at 10 Pa vaccum
(whatever "semi-hydrated state means ;-)).

I would be grateful if you cared to share your opinion
on this subjet with me.

Stephane



________________________________________________________________________
____________
Bored stiff? Loosen up...
Download and play hundreds of games for free on Yahoo! Games.
http://games.yahoo.com/games/front

==============================Original
Headers==============================
6, 19 -- From nizets2-at-yahoo.com Tue Jan 16 03:14:48 2007
6, 19 -- Received: from web37411.mail.mud.yahoo.com
(web37411.mail.mud.yahoo.com [209.191.91.143])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP
id l0G9Emoo019552
6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007
03:14:48 -0600
6, 19 -- Received: (qmail 62303 invoked by uid 60001); 16 Jan 2007
09:14:47 -0000
6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 --
h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Co
ntent-Transfer-Encoding:Message-ID;
6, 19 --
b=kpkcnoBDUHMcVkk3aIfi3ch5Ffk324ZCkbIAnLfSxMO5yzLGCpa74RgxqXcpb6rCkUqNh6
CevqxvS783hV8rc5yRiXjGPlhG+sMnVSwX/qheKVKYH8tg8dOuIcXQ0xvXvOS/v0G8SSlqv8
teAEkzdXQKlgJxBBldpWTS2jBvJxk=;
6, 19 -- X-YMail-OSG:
6E_tvvQVM1lhn5yjbcRFZ33VO8xzQy.Y0.qekDzkGSSd9Vs3SDF0WHtJsnif0Jao410DY2cs
Nk2MjHuUwU1.68o6dDetO9BirDlwIulsuZg6gJF0FQgVpzzHlnJ9St8l8kxEzjvOlJUvb3w-
6, 19 -- Received: from [80.122.101.102] by web37411.mail.mud.yahoo.com
via HTTP; Tue, 16 Jan 2007 01:14:47 PST
6, 19 -- Date: Tue, 16 Jan 2007 01:14:47 -0800 (PST)
6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: preparation of cells for SEM - question
6, 19 -- To: microscopy-at-microscopy.com
6, 19 -- MIME-Version: 1.0
6, 19 -- Content-Type: text/plain; charset=iso-8859-1
6, 19 -- Content-Transfer-Encoding: 8bit
6, 19 -- Message-ID: {636541.62156.qm-at-web37411.mail.mud.yahoo.com}
==============================End of -
Headers==============================


This incoming email to UWE has been independently scanned for viruses by
McAfee anti-virus software and none were detected


This email was independently scanned for viruses by McAfee anti-virus software and none were found


==============================Original Headers==============================
38, 34 -- From David.Patton-at-uwe.ac.uk Tue Jan 16 03:36:06 2007
38, 34 -- Received: from mailapp03.uwe.ac.uk (mailapp03.uwe.ac.uk [164.11.132.65])
38, 34 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0G9a5Gk030756
38, 34 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 03:36:05 -0600
38, 34 -- Received: from (mta02.uwe.ac.uk [164.11.132.62]) by mailapp03.uwe.ac.uk with smtp
38, 34 -- id 2d0f_31ed9552_a544_11db_97a3_00142221cca9;
38, 34 -- Tue, 16 Jan 2007 09:30:35 +0000
38, 34 -- Received: from egen-uwe01.campus.ads.uwe.ac.uk
38, 34 -- (egen-uwe01.uwe.ac.uk [164.11.249.121])
38, 34 -- by mta02.uwe.ac.uk (iPlanet Messaging Server 5.2 HotFix 2.07 (built Jun 24
38, 34 -- 2005)) with ESMTP id {0JBY00AR2FZU3S-at-mta02.uwe.ac.uk} for
38, 34 -- microscopy-at-microscopy.com; Tue, 16 Jan 2007 09:35:54 +0000 (GMT)
38, 34 -- Date: Tue, 16 Jan 2007 09:35:06 +0000
38, 34 -- From: David Patton {David.Patton-at-uwe.ac.uk}
38, 34 -- Subject: RE: [Microscopy] preparation of cells for SEM - question
38, 34 -- In-reply-to: {200701160920.l0G9KGNq026487-at-ns.microscopy.com}
38, 34 -- To: nizets2-at-yahoo.com
38, 34 -- Cc: microscopy-at-microscopy.com
38, 34 -- Message-id: {F247F674896BE243AD8263C5280E2BDB02B0AECD-at-egen-uwe01}
38, 34 -- MIME-version: 1.0
38, 34 -- X-MIMEOLE: Produced By Microsoft Exchange V6.5
38, 34 -- Content-type: text/plain; charset=us-ascii
38, 34 -- Content-class: urn:content-classes:message
38, 34 -- Thread-topic: [Microscopy] preparation of cells for SEM - question
38, 34 -- Thread-index: Acc5T4/BzWF0tJWbTBOyK0ths3Tq+wAAIe4Q
38, 34 -- X-MS-Has-Attach:
38, 34 -- X-MS-TNEF-Correlator:
38, 34 -- X-NAIMIME-Disclaimer: 1
38, 34 -- X-NAIMIME-Modified: 1
38, 34 -- X-NAI-Spam-Score: -0.3
38, 34 -- X-NAI-Spam-Rules: 1 Rules triggered
38, 34 -- BAYES_30=-0.3
38, 34 -- Content-Transfer-Encoding: 8bit
38, 34 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0G9a5Gk030756
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Tue, 16 Jan 2007 12:22:32 -0600
Subject: [Microscopy] preparation of cells for SEM - question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Stephane,

If your cells are in suspension I would coat cover slips with
poly-l-lysine (1mg/ml in distilled water), place a drop of the
suspension onto the cover slips and allow to settle for up to an hour in
a humidity chamber. Then, after the cells attach, just gently rinse the
cover slips, fix, and dehydrate using
David's HMDS protocol.

I would suggest 2% glutaraldehyde in buffer (0.1M phosphate, cacodylate,
or common buffer for biological EM) for 10-15 minutes, followed by three
or more buffer rinses, post-fixation in 1% osmium tetroxide (buffered or
aqueous) for the same time, followed by water rinses for 5-10 minutes
each, then into your dehydration schedule. Sputter coat and mount.

Good luck.

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


-----Original Message-----
X-from: nizets2-at-yahoo.com [mailto:nizets2-at-yahoo.com]
Sent: Tuesday, January 16, 2007 3:16 AM
To: Tindall, Randy D.

Dear listers,

I have a request for experts in SEM who have some experience with
eukaryotic cells.
We have a basic SEM which is normally used to study minerals, so we are
not equipped with critical point dryers and the like. Now I would like
to have a look at some cells in culture with the SEM but I have no idea
how I could prepare the cells without critical point dryer. Of course if
I simply fix and dehydrate in acetone, then air dry the cells I expect
the structure to crumble. Well perhaps it won't give nice pictures but
perhaps it will be enough for what I want to see. On the other hand
perhaps I could keep them in a semi-hydrated state and look at them at
10 Pa vaccum (whatever "semi-hydrated state means ;-)).

I would be grateful if you cared to share your opinion on this subjet
with me.

Stephane



________________________________________________________________________
____________
Bored stiff? Loosen up...
Download and play hundreds of games for free on Yahoo! Games.
http://games.yahoo.com/games/front

==============================Original
Headers==============================
6, 19 -- From nizets2-at-yahoo.com Tue Jan 16 03:14:48 2007 6, 19 --
Received: from web37411.mail.mud.yahoo.com (web37411.mail.mud.yahoo.com
[209.191.91.143])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP
id l0G9Emoo019552
6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007
03:14:48 -0600
6, 19 -- Received: (qmail 62303 invoked by uid 60001); 16 Jan 2007
09:14:47 -0000 6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 --
h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Co
ntent-Transfer-Encoding:Message-ID;
6, 19 --
b=kpkcnoBDUHMcVkk3aIfi3ch5Ffk324ZCkbIAnLfSxMO5yzLGCpa74RgxqXcpb6rCkUqNh6
CevqxvS783hV8rc5yRiXjGPlhG+sMnVSwX/qheKVKYH8tg8dOuIcXQ0xvXvOS/v0G8SSlqv8
teAEkzdXQKlgJxBBldpWTS2jBvJxk=;
6, 19 -- X-YMail-OSG:
6E_tvvQVM1lhn5yjbcRFZ33VO8xzQy.Y0.qekDzkGSSd9Vs3SDF0WHtJsnif0Jao410DY2cs
Nk2MjHuUwU1.68o6dDetO9BirDlwIulsuZg6gJF0FQgVpzzHlnJ9St8l8kxEzjvOlJUvb3w-
6, 19 -- Received: from [80.122.101.102] by web37411.mail.mud.yahoo.com
via HTTP; Tue, 16 Jan 2007 01:14:47 PST 6, 19 -- Date: Tue, 16 Jan 2007
01:14:47 -0800 (PST) 6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: preparation of cells for SEM - question 6, 19 -- To:
microscopy-at-microscopy.com 6, 19 -- MIME-Version: 1.0 6, 19 --
Content-Type: text/plain; charset=iso-8859-1 6, 19 --
Content-Transfer-Encoding: 8bit 6, 19 -- Message-ID:
{636541.62156.qm-at-web37411.mail.mud.yahoo.com}
==============================End of -
Headers==============================



==============================Original Headers==============================
20, 26 -- From TindallR-at-missouri.edu Tue Jan 16 12:22:32 2007
20, 26 -- Received: from um-nsmtpout1.um.umsystem.edu (um-nsmtpout1.um.umsystem.edu [209.106.228.53])
20, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0GIMWsj023906
20, 26 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 12:22:32 -0600
20, 26 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-nsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
20, 26 -- Tue, 16 Jan 2007 12:22:31 -0600
20, 26 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
20, 26 -- Content-class: urn:content-classes:message
20, 26 -- MIME-Version: 1.0
20, 26 -- Content-Type: text/plain;
20, 26 -- charset="us-ascii"
20, 26 -- Subject: RE: [Microscopy] preparation of cells for SEM - question
20, 26 -- Date: Tue, 16 Jan 2007 12:22:31 -0600
20, 26 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68CE3-at-UM-XMAIL08.um.umsystem.edu}
20, 26 -- in-reply-to: {200701160916.l0G9GKG9021537-at-ns.microscopy.com}
20, 26 -- X-MS-Has-Attach:
20, 26 -- X-MS-TNEF-Correlator:
20, 26 -- Thread-Topic: [Microscopy] preparation of cells for SEM - question
20, 26 -- Thread-Index: Acc5Tv0//VNi6xXfS1e9n6fo+SQCaAAScKjQ
20, 26 -- References: {200701160916.l0G9GKG9021537-at-ns.microscopy.com}
20, 26 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
20, 26 -- To: {nizets2-at-yahoo.com}
20, 26 -- Cc: {microscopy-at-microscopy.com}
20, 26 -- X-OriginalArrivalTime: 16 Jan 2007 18:22:31.0760 (UTC) FILETIME=[4967C500:01C7399B]
20, 26 -- Content-Transfer-Encoding: 8bit
20, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0GIMWsj023906
==============================End of - Headers==============================




From: tivol-at-caltech.edu
Date: Tue, 16 Jan 2007 18:33:28 -0600
Subject: [Microscopy] Re: preparation of cells for SEM - question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


On Jan 16, 2007, at 1:14 AM, nizets2-at-yahoo.com wrote:

} I have a request for experts in SEM who have some
} experience with eukaryotic cells.
} We have a basic SEM which is normally used to study
} minerals, so we are not equipped with critical point
} dryers and the like. Now I would like to have a look
} at some cells in culture with the SEM but I have no
} idea how I could prepare the cells without critical
} point dryer. Of course if I simply fix and dehydrate
} in acetone, then air dry the cells I expect the
} structure to crumble. Well perhaps it won't give nice
} pictures but perhaps it will be enough for what I want
} to see. On the other hand perhaps I could keep them in
} a semi-hydrated state and look at them at 10 Pa vaccum
} (whatever "semi-hydrated state means ;-)).
}
} I would be grateful if you cared to share your opinion
} on this subjet with me.
}
Dear Stephane,
When I was at the high-voltage TEM, I designed a hydration stage,
which allowed me to examine cells that were fully hydrated--as
determined by the fact that upon removal from the scope, at least some
of the cells were still viable. These were prepared by placing
formvar/carbon coated gold grids in the culture dish, so the cells
would grow on the grids, then just removing the grids from the dish,
placing them in a 100% humidity chamber, blotting off the excess fluid,
and transferring the grids to the hydration stage maintaining the
humidity as close to 100% as possible. If you can operate your VPSEM
at the vapor pressure of water for the temperature of the specimen
chamber (~25 torr), then you can look at them fully hydrated. I do not
know, however, whether you will see anything of interest, or just the
surface of the thin film of water that will still coat the cells. Good
luck.
Yours,
Bill Tivol, PhD
EM Scientist and Manager
Cryo-Electron Microscopy Facility
Broad Center, Mail Code 114-96
California Institute of Technology
Pasadena CA 91125
(626) 395-8833
tivol-at-caltech.edu


==============================Original Headers==============================
4, 22 -- From tivol-at-caltech.edu Tue Jan 16 18:33:28 2007
4, 22 -- Received: from outgoing-mail.its.caltech.edu (outgoing-mail.its.caltech.edu [131.215.239.19])
4, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0H0XRtM009854
4, 22 -- for {microscopy-at-msa.microscopy.com} ; Tue, 16 Jan 2007 18:33:28 -0600
4, 22 -- Received: from water-dog.caltech.edu (water-dog [192.168.1.26])
4, 22 -- by earth-ox-postvirus (Postfix) with ESMTP id 386C510A4AD
4, 22 -- for {microscopy-at-msa.microscopy.com} ; Tue, 16 Jan 2007 16:15:49 -0800 (PST)
4, 22 -- Received: from [192.168.159.233] (jpix-01.caltech.edu [131.215.2.133])
4, 22 -- by fire-ox.its.caltech.edu (Postfix) with ESMTP id C24DE35B20
4, 22 -- for {microscopy-at-msa.microscopy.com} ; Tue, 16 Jan 2007 15:55:41 -0800 (PST)
4, 22 -- Mime-Version: 1.0 (Apple Message framework v624)
4, 22 -- In-Reply-To: {200701160914.l0G9EvlZ019715-at-ns.microscopy.com}
4, 22 -- References: {200701160914.l0G9EvlZ019715-at-ns.microscopy.com}
4, 22 -- Content-Type: text/plain; charset=US-ASCII; format=flowed
4, 22 -- Message-Id: {10f5fb65833744fd4afb91860f232bc1-at-caltech.edu}
4, 22 -- Content-Transfer-Encoding: 7bit
4, 22 -- From: Bill Tivol {tivol-at-caltech.edu}
4, 22 -- Subject: Re: [Microscopy] preparation of cells for SEM - question
4, 22 -- Date: Tue, 16 Jan 2007 16:03:20 -0800
4, 22 -- To: microscopy-at-msa.microscopy.com
4, 22 -- X-Mailer: Apple Mail (2.624)
4, 22 -- X-Spam-Scanned: at Caltech-ITS on water-dog by amavisd-2.3.3
==============================End of - Headers==============================




From: chappell.mark-at-epa.gov
Date: Tue, 16 Jan 2007 20:01:02 -0600
Subject: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question was submitted to Ask-A-Microscopist by (chappell.mark-at-epa.gov)
from http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on Tuesday, January 16, 2007 at 13:59:24
Remember to consider the Grade/Age of the student when considering the Question
---------------------------------------------------------------------------
Please reply to both chappell.mark-at-epa.gov as well as to the Microscopy Listserver
---------------------------------------------------------------------------

Email: chappell.mark-at-epa.gov
Name: Mark Chappell

Organization: U.S. EPA

Education: Graduate College

Location: Cincinnati, OH

Title: SEM/EDX of bird feather

Question: I want to map the Hg in a contaminated bird feather using SEM/EDX/WDS. What's the best way to prepare the sample to minimize charging?

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 16 20:01:02 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0H211P7022216
8, 12 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 20:01:02 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
8, 12 -- Message-Id: {p06110401c1d333bcefd0-at-[206.69.208.22]}
8, 12 -- Date: Tue, 16 Jan 2007 20:01:00 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: chappell.mark-at-epa.gov (by way of Ask-A-Microscopist)
8, 12 -- Subject: AskAMicroscopist: SEM/EDX of bird feather
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: gary-at-gaugler.com
Date: Tue, 16 Jan 2007 20:22:49 -0600
Subject: [Microscopy] Re: AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This should not be too difficult.

First look at Hg Z=80. So, high Z.
M alpha=2.191KeV. L alpha=9.986KeV.

So, for a good results you can do 5KV and look
for the Ma or do 20KV and look for Ma and La.

Coat the specimen with Pd, Ir, C, but not Au
(La=9.710KeV). Doing as suggested will or should
put Hg by itself as a EDS peak.

gary g.



At 06:02 PM 1/16/2007, you wrote:




} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
13, 20 -- From gary-at-gaugler.com Tue Jan 16 20:22:49 2007
13, 20 -- Received: from qsmtp1.mc.surewest.net (qsmtp.mc.surewest.net [66.60.130.145])
13, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0H2Mmjm000978
13, 20 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 20:22:48 -0600
13, 20 -- Message-Id: {200701170222.l0H2Mmjm000978-at-ns.microscopy.com}
13, 20 -- Received: (qmail 15696 invoked from network); 16 Jan 2007 18:22:39 -0800
13, 20 -- Received: by simscan 1.1.0 ppid: 15678, pid: 15693, t: 0.1742s
13, 20 -- scanners: regex: 1.1.0 attach: 1.1.0
13, 20 -- Received: from unknown (HELO ivan.gaugler.com) (66.60.171.211)
13, 20 -- by qsmtp1 with SMTP; 16 Jan 2007 18:22:39 -0800
13, 20 -- X-Mailer: QUALCOMM Windows Eudora Version 7.1.0.9
13, 20 -- Date: Tue, 16 Jan 2007 18:24:43 -0800
13, 20 -- To: chappell.mark-at-epa.gov
13, 20 -- From: Gary Gaugler {gary-at-gaugler.com}
13, 20 -- Subject: Re: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
13, 20 -- Cc: MSA listserver {microscopy-at-microscopy.com}
13, 20 -- In-Reply-To: {200701170202.l0H22ZtI025116-at-ns.microscopy.com}
13, 20 -- References: {200701170202.l0H22ZtI025116-at-ns.microscopy.com}
13, 20 -- Mime-Version: 1.0
13, 20 -- Content-Type: text/plain; charset=us-ascii; format=flowed; x-avg-checked=avg-ok-57F5541
==============================End of - Headers==============================




From: gary-at-gaugler.com
Date: Tue, 16 Jan 2007 20:27:47 -0600
Subject: [Microscopy] Re: AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

As I think about this....this may not be possible.

The amount of Hg in the feathers are probably PPM.
EDS will not show this. You would need WDS or FTIR
to show up the small amounts....unless the feathers
are really loaded with Hg. The shape of the feathers
will also affect collection due to difflection of the
x-rays. But you should still get decent results to show
if any TRACE amounts of Hg are present.

gary g.


At 06:02 PM 1/16/2007, you wrote:




} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
10, 20 -- From gary-at-gaugler.com Tue Jan 16 20:27:47 2007
10, 20 -- Received: from qsmtp4.mc.surewest.net (qsmtp.mc.surewest.net [66.60.130.145])
10, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0H2RkcY011922
10, 20 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 20:27:46 -0600
10, 20 -- Message-Id: {200701170227.l0H2RkcY011922-at-ns.microscopy.com}
10, 20 -- Received: (qmail 1348 invoked from network); 16 Jan 2007 18:22:21 -0800
10, 20 -- Received: by simscan 1.1.0 ppid: 1336, pid: 1346, t: 0.1706s
10, 20 -- scanners: regex: 1.1.0 attach: 1.1.0
10, 20 -- Received: from unknown (HELO ivan.gaugler.com) (66.60.171.211)
10, 20 -- by qsmtp4 with SMTP; 16 Jan 2007 18:22:21 -0800
10, 20 -- X-Mailer: QUALCOMM Windows Eudora Version 7.1.0.9
10, 20 -- Date: Tue, 16 Jan 2007 18:29:42 -0800
10, 20 -- To: chappell.mark-at-epa.gov
10, 20 -- From: Gary Gaugler {gary-at-gaugler.com}
10, 20 -- Subject: Re: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
10, 20 -- Cc: MSA listserver {microscopy-at-microscopy.com}
10, 20 -- In-Reply-To: {200701170202.l0H22ZtI025116-at-ns.microscopy.com}
10, 20 -- References: {200701170202.l0H22ZtI025116-at-ns.microscopy.com}
10, 20 -- Mime-Version: 1.0
10, 20 -- Content-Type: text/plain; charset=us-ascii; format=flowed; x-avg-checked=avg-ok-57F5541
==============================End of - Headers==============================




From: hikonishi-at-gmail.com
Date: Tue, 16 Jan 2007 23:20:03 -0600
Subject: [Microscopy] BALTEK - RES-010 Benchtop Rapid Etching System

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear List members:

I am looking for the information on BALTEK - RES-010 Benchtop Rapid Etching
System. If you have any experience with the ion milling such as artifact and
over milling, please provide the information.

We have an old ion milling from BALTEK, but I heard that it always
overmills. Now our instrument has some problems, but no one wants to fix it
because of over milling problem. I looked at the instrument and I found that
the gun is much similar to the gun from Technoorg-Linda IV3. Based on some
experience with IV3, I think that people at our university might not use
proper way to operate it.

If you use BALTEK - RES-010, I would like to know the milling conditions
such as angle, voltage, and current. Also, I would like to know if you have
any experience of serious artifacts such as reaction, decomposition or phase
transformation.
Thank you,

Hiromi Konishi, Ph.D.
UW-Madison


==============================Original Headers==============================
6, 30 -- From hikonishi-at-gmail.com Tue Jan 16 23:20:03 2007
6, 30 -- Received: from py-out-1112.google.com (py-out-1112.google.com [64.233.166.178])
6, 30 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0H5K2dO027924
6, 30 -- for {Microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 23:20:03 -0600
6, 30 -- Received: by py-out-1112.google.com with SMTP id b50so1226286pyh
6, 30 -- for {Microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 21:20:00 -0800 (PST)
6, 30 -- DomainKey-Signature: a=rsa-sha1; c=nofws;
6, 30 -- d=gmail.com; s=beta;
6, 30 -- h=received:message-id:from:to:subject:date:mime-version:content-type:content-transfer-encoding:x-priority:x-msmail-priority:x-mailer:x-mimeole;
6, 30 -- b=IaBQGRKwpfKAkRx8L8wGRHk3RnDqEJpDCCZNNKQSAZdIlxSXe9NMlY+KDGtAFfFdFcRKMHSloGXp0vQZeSKX66l0oB81xuJkUkNI9T+GJnlg+eWHE+xuejwjiX/j/yHPAzTSu/qcbAw7NztRXONuqzPOQItfD5YjXCZx5JNNoNk=
6, 30 -- Received: by 10.35.103.12 with SMTP id f12mr11756749pym.1169011200239;
6, 30 -- Tue, 16 Jan 2007 21:20:00 -0800 (PST)
6, 30 -- Received: from HKONISHI ( [216.165.154.16])
6, 30 -- by mx.google.com with ESMTP id f57sm9087298pyh.2007.01.16.21.19.59;
6, 30 -- Tue, 16 Jan 2007 21:19:59 -0800 (PST)
6, 30 -- Message-ID: {000901c739f7$2d245690$0200a8c0-at-HKONISHI}
6, 30 -- From: "Hiromi Konishi" {hikonishi-at-gmail.com}
6, 30 -- To: {Microscopy-at-microscopy.com}
6, 30 -- Subject: BALTEK - RES-010 Benchtop Rapid Etching System
6, 30 -- Date: Tue, 16 Jan 2007 23:20:10 -0600
6, 30 -- MIME-Version: 1.0
6, 30 -- Content-Type: text/plain;
6, 30 -- format=flowed;
6, 30 -- charset="iso-2022-jp";
6, 30 -- reply-type=original
6, 30 -- Content-Transfer-Encoding: 7bit
6, 30 -- X-Priority: 3
6, 30 -- X-MSMail-Priority: Normal
6, 30 -- X-Mailer: Microsoft Outlook Express 6.00.2900.2180
6, 30 -- X-MIMEOLE: Produced By Microsoft MimeOLE V6.00.2900.2180
==============================End of - Headers==============================




From: johnf-at-geology.wisc.edu
Date: Wed, 17 Jan 2007 07:12:43 -0600
Subject: [Microscopy] Re: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

}
}
} Question: I want to map the Hg in a contaminated bird feather using
} SEM/EDX/WDS. What's the best way to prepare the sample to minimize
} charging?

When you say "contaminated" do you mean massive contamination,
millions of times greater than what a human would get in their hair
(thru metabolism) eating fish that had "high" levels from Hg used in
adjacent gold mining/extraction? Only in that case would an
electron-based microanalytical technique be of any use.

I attempted to measure Hg in hair from native people from Brazil who
lived near gold extraction areas.The idea was to see if we could see
variations in the hair along length (thus, versus time). Using WDS
EPMA. I knew the approximate levels (thru some technique like INAA or
ICPMS or XRF, I forget) and it was high ppb or low ppm (again, I
forget the details). But I could never put enough current (I probably
went up to 10 or 20 nA) to see anything.

John


==============================Original Headers==============================
5, 25 -- From johnf-at-geology.wisc.edu Wed Jan 17 07:12:42 2007
5, 25 -- Received: from ice.geology.wisc.edu (ice.geology.wisc.edu [144.92.206.14])
5, 25 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HDCfcB014198
5, 25 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 07:12:42 -0600
5, 25 -- Received: from localhost (localhost [127.0.0.1])
5, 25 -- by localhost (Postfix) with ESMTP id 134F720D31;
5, 25 -- Wed, 17 Jan 2007 07:12:40 -0600 (CST)
5, 25 -- Received: from ice.geology.wisc.edu ([127.0.0.1])
5, 25 -- by localhost (geology.wisc.edu [127.0.0.1]) (amavisd-new, port 10024)
5, 25 -- with ESMTP id 27330-08; Wed, 17 Jan 2007 07:12:34 -0600 (CST)
5, 25 -- Received: from [192.168.1.103] (24-183-52-127.dhcp.mdsn.wi.charter.com [24.183.52.127])
5, 25 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
5, 25 -- (No client certificate requested)
5, 25 -- by ice.geology.wisc.edu (Postfix) with ESMTP id 5B2A320D08;
5, 25 -- Wed, 17 Jan 2007 07:12:34 -0600 (CST)
5, 25 -- Mime-Version: 1.0
5, 25 -- Message-Id: {p06210201c1d3cfb242f0-at-[192.168.0.122]}
5, 25 -- In-Reply-To: {200701170205.l0H2595J030194-at-ns.microscopy.com}
5, 25 -- References: {200701170205.l0H2595J030194-at-ns.microscopy.com}
5, 25 -- Date: Wed, 17 Jan 2007 07:14:58 -0600
5, 25 -- To: chappell.mark-at-epa.gov, microscopy-at-microscopy.com
5, 25 -- From: John Fournelle {johnf-at-geology.wisc.edu}
5, 25 -- Subject: Re: SEM/EDX of bird feather
5, 25 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
5, 25 -- X-Virus-Scanned: amavisd-new at geology.wisc.edu
==============================End of - Headers==============================




From: frank.karl-at-degussa.com
Date: Wed, 17 Jan 2007 07:19:14 -0600
Subject: [Microscopy] Re: AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Wow, Some really different answers. I seem to remember Hg (2.19) falls
close to S (2.30), so I would use 25 kv because I would want to see both
the 2.19 M line and 9.8 L line. I like to be able to see at least two
lines, when possible, for any element. I would skip coating, having tried
it and always unhappy with my results and go with low pressure.

I suspect the comments about detectable level as are on target, but you
could very well surprise us.

Please let me know how this works out as I played with feathers using light
and SEM and have an interest.



chappell.mark-at-epa
.gov To: frank.karl-at-degussa.com
cc:
01/16/2007 09:02 Subject: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
PM
Please respond to
chappell.mark








----------------------------------------------------------------------------

The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


This Question was submitted to Ask-A-Microscopist by
(chappell.mark-at-epa.gov)
from http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html
on Tuesday, January 16, 2007 at 13:59:24
Remember to consider the Grade/Age of the student when considering the
Question
---------------------------------------------------------------------------
Please reply to both chappell.mark-at-epa.gov as well as to the Microscopy
Listserver
---------------------------------------------------------------------------

Email: chappell.mark-at-epa.gov
Name: Mark Chappell

Organization: U.S. EPA

Education: Graduate College

Location: Cincinnati, OH

Title: SEM/EDX of bird feather

Question: I want to map the Hg in a contaminated bird feather using
SEM/EDX/WDS. What's the best way to prepare the sample to minimize
charging?

---------------------------------------------------------------------------

==============================Original
Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 16 20:01:02 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0H211P7022216
8, 12 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007
20:01:02 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
8, 12 -- Message-Id: {p06110401c1d333bcefd0-at-[206.69.208.22]}
8, 12 -- Date: Tue, 16 Jan 2007 20:01:00 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: chappell.mark-at-epa.gov (by way of Ask-A-Microscopist)
8, 12 -- Subject: AskAMicroscopist: SEM/EDX of bird feather
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of -
Headers==============================




==============================Original Headers==============================
26, 18 -- From frank.karl-at-degussa.com Wed Jan 17 07:19:14 2007
26, 18 -- Received: from framailout1.rz.itson.com (mailout2.degussa.com [149.216.91.173])
26, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HDJDMf020992
26, 18 -- for {microscopy-at-msa.microscopy.com} ; Wed, 17 Jan 2007 07:19:14 -0600
26, 18 -- Received: from mobuscomm01.mail.degussa.com (uss1026.applications.degussanet.com [10.88.88.98])
26, 18 -- by framailout1.rz.itson.com (8.13.4/8.13.4/Debian-3sarge3) with SMTP id l0HDJ6QH021157;
26, 18 -- Wed, 17 Jan 2007 14:19:09 +0100
26, 18 -- In-Reply-To: {200701170202.l0H22Gnm024324-at-ns.microscopy.com}
26, 18 -- Subject: Re: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
26, 18 -- To: chappell.mark-at-epa.gov, microscopy-at-msa.microscopy.com
26, 18 -- X-Mailer: Lotus Notes Release 6.5.5 November 30, 2005
26, 18 -- Message-ID: {OFC2EB2416.D1D6E534-ON86257266.0047FF38-85257266.004926BB-at-degussa.com}
26, 18 -- From: frank.karl-at-degussa.com
26, 18 -- Date: Wed, 17 Jan 2007 08:19:00 -0500
26, 18 -- X-MIMETrack: Serialize by Router on MOBUSComm01/DHexternal/US(Release 6.5.1|January 21, 2004) at
26, 18 -- 01/17/2007 07:19:10 AM
26, 18 -- MIME-Version: 1.0
26, 18 -- Content-type: text/plain; charset=US-ASCII
==============================End of - Headers==============================




From: dljones-at-bestweb.net
Date: Wed, 17 Jan 2007 08:01:51 -0600
Subject: [Microscopy] looking for parts

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

I apologise for taking space for this request, but I have looked many
places without success. Perhaps someone here may be able to give me some
pointers.

I'm looking for a keyboard for a SEM: a Zeiss DSM 9XX A. It could be from
a 940, 950, or 960 as I believe these are all the same keyboards. This is
for a SEM that is being used in my local high school so I'm looking for a
used, functioning keyboard that is not expensive.

If anyone knows where I might find this, please contact me. As I imagine
this is of no interest to the group, please respond to me directly.

I'm located in Cold Spring, New York USA.

Thank you,

dj


==============================Original Headers==============================
7, 16 -- From dljones-at-bestweb.net Wed Jan 17 08:01:51 2007
7, 16 -- Received: from vms044pub.verizon.net (vms044pub.verizon.net [206.46.252.44])
7, 16 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HE1pkq004628
7, 16 -- for {Microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 08:01:51 -0600
7, 16 -- Received: from localhost ([71.247.123.130])
7, 16 -- by vms044.mailsrvcs.net (Sun Java System Messaging Server 6.2-6.01 (built Apr
7, 16 -- 3 2006)) with ESMTPA id {0JC000NCFMYU4B30-at-vms044.mailsrvcs.net} for
7, 16 -- Microscopy-at-microscopy.com; Wed, 17 Jan 2007 08:01:47 -0600 (CST)
7, 16 -- Date: Wed, 17 Jan 2007 09:00:49 -0500 (Eastern Standard Time)
7, 16 -- From: "David L. Jones" {dljones-at-bestweb.net}
7, 16 -- Subject: looking for parts
7, 16 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
7, 16 -- To: Microscopy-at-microscopy.com
7, 16 -- Message-id: {Pine.WNT.4.64.0701170841590.2632-at-H-F1}
7, 16 -- MIME-version: 1.0
7, 16 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Wed, 17 Jan 2007 09:39:47 -0600
Subject: [Microscopy] Polaron E5100 sputter coater question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Listers,

While our main sputter coater is out for a much-needed overhaul, we are
relying on a backup coater, namely a Polaron E5100. The unit is
operating beautifully, except that the voltage dial has worked loose at
some point and I can no longer tell where the zero point, and thus the
proper operating voltage, is located on the scale. This is because the
scale starts at 2.0, not zero, and I have no reference to line up the
indicator with. I tried looking at the set-screw scar on the knob shaft,
but it didn't help. Am I missing something obvious? (Wouldn't be the
first time.....)

f you will go to http://www.emc.missouri.edu/lookatthis.htm to see the
images of this coater hopefully this will make sense.

Can someone with a similar unit kindly let me know about where the zero
point would be on this machine? I can get it to coat by trial and
error, but it would be nice to have a baseline to experiment from.

Many thanks!

Searching for zero,
Randy


Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original Headers==============================
9, 23 -- From TindallR-at-missouri.edu Wed Jan 17 09:39:47 2007
9, 23 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
9, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HFdlGO018946
9, 23 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 09:39:47 -0600
9, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
9, 23 -- Wed, 17 Jan 2007 09:39:47 -0600
9, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
9, 23 -- Content-class: urn:content-classes:message
9, 23 -- MIME-Version: 1.0
9, 23 -- Content-Type: text/plain;
9, 23 -- charset="us-ascii"
9, 23 -- Subject: Polaron E5100 sputter coater question
9, 23 -- Date: Wed, 17 Jan 2007 09:39:46 -0600
9, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68CF1-at-UM-XMAIL08.um.umsystem.edu}
9, 23 -- X-MS-Has-Attach:
9, 23 -- X-MS-TNEF-Correlator:
9, 23 -- Thread-Topic: Polaron E5100 sputter coater question
9, 23 -- Thread-Index: Acc6TbbzCo8D3MqPS+a3y10wR2ngWQ==
9, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
9, 23 -- To: {microscopy-at-microscopy.com}
9, 23 -- X-OriginalArrivalTime: 17 Jan 2007 15:39:47.0209 (UTC) FILETIME=[B7B14790:01C73A4D]
9, 23 -- Content-Transfer-Encoding: 8bit
9, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0HFdlGO018946
==============================End of - Headers==============================




From: fortner-at-cmt.anl.gov
Date: Wed, 17 Jan 2007 09:58:29 -0600
Subject: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

There may be a better way to do this- nondestructively- using x-rays
from the Advanced Photon Source Users' Facility at Argonne National
Laboratory. See this analysis of lead in human hair (a rather famous,
human, at that):

http://www.aps.anl.gov//News/APS_News/2000/20001017.htm

And

http://www.anl.gov/Media_Center/Frontiers/2002/c3facil.html

If you are interested, I can put you in contact with the investigators
on this project.

Jeffrey A. Fortner, Ph.D.
Chemical Engineering Division
Argonne National Laboratory
Argonne, IL 60439
fortner(at)cmt.anl.gov


.
-----Original Message-----
X-from: chappell.mark-at-epa.gov [mailto:chappell.mark-at-epa.gov]
Sent: Tuesday, January 16, 2007 8:02 PM
To: Fortner, Jeffrey A.

This Question was submitted to Ask-A-Microscopist by
(chappell.mark-at-epa.gov) from
http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on
Tuesday, January 16, 2007 at 13:59:24 Remember to consider the Grade/Age
of the student when considering the Question
------------------------------------------------------------------------
---
Please reply to both chappell.mark-at-epa.gov as well as to the
Microscopy Listserver
------------------------------------------------------------------------
---

Email: chappell.mark-at-epa.gov
Name: Mark Chappell

Organization: U.S. EPA

Education: Graduate College

Location: Cincinnati, OH

Title: SEM/EDX of bird feather

Question: I want to map the Hg in a contaminated bird feather using
SEM/EDX/WDS. What's the best way to prepare the sample to minimize
charging?

------------------------------------------------------------------------
---

==============================Original
Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 16 20:01:02 2007 8,
12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0H211P7022216
8, 12 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007
20:01:02 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified) 8, 12 --
Message-Id: {p06110401c1d333bcefd0-at-[206.69.208.22]}
8, 12 -- Date: Tue, 16 Jan 2007 20:01:00 -0600 8, 12 -- To:
microscopy-at-microscopy.com 8, 12 -- From: chappell.mark-at-epa.gov (by way
of Ask-A-Microscopist) 8, 12 -- Subject: AskAMicroscopist: SEM/EDX of
bird feather 8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of -
Headers==============================


==============================Original Headers==============================
21, 20 -- From fortner-at-cmt.anl.gov Wed Jan 17 09:58:29 2007
21, 20 -- Received: from cmtmail.cmt.anl.gov (CMTMAIL.cmt.anl.gov [146.139.148.212])
21, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HFwSV3030417
21, 20 -- for {microscopy-at-sparc5.microscopy.com} ; Wed, 17 Jan 2007 09:58:29 -0600
21, 20 -- X-MimeOLE: Produced By Microsoft Exchange V6.5.7226.0
21, 20 -- Content-class: urn:content-classes:message
21, 20 -- MIME-Version: 1.0
21, 20 -- Content-Type: text/plain;
21, 20 -- charset="us-ascii"
21, 20 -- Subject: FW: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
21, 20 -- Date: Wed, 17 Jan 2007 09:58:28 -0600
21, 20 -- Message-ID: {661C1AA445464649B30D32151A7C34FA043C7C69-at-CMTMAIL.cmt.anl.gov}
21, 20 -- X-MS-Has-Attach:
21, 20 -- X-MS-TNEF-Correlator:
21, 20 -- Thread-Topic: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
21, 20 -- Thread-Index: Acc523r+QWU1vsJPR2+Jb3c2wb69EQAc0dkA
21, 20 -- From: "Fortner, Jeffrey A." {fortner-at-cmt.anl.gov}
21, 20 -- To: {microscopy-at-ns.microscopy.com} , {chappell.mark-at-epa.gov}
21, 20 -- Content-Transfer-Encoding: 8bit
21, 20 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0HFwSV3030417
==============================End of - Headers==============================




From: ph2-at-sprynet.com
Date: Wed, 17 Jan 2007 10:03:46 -0600
Subject: [Microscopy] Re: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Work on elemental composition in hair was done back in the 70s by Dr. Walter
McCrone using neutron activation. This was used to map elements by
concentration along the shaft. The basis was to look at hair
differentiation in forensics.

I recall Walter mentioning it in a short presentation at Inter-Micro ca.
2000? as well as in a conversation I had with him. I believe he also
published a piece in the The Microscope.

Tony


......................................................................
Andrew Anthony "Tony" Havics, CHMM, CIH, PE
pH2, LLC
5250 E US 36, Suite 830
Avon, IN 46123
(317) 718-7020 off
(317) 718-7038 fax
(317) 409-3238 cell

90% of Risk Management is knowing where to place the decimal point...any
consultant can give you the other 10%(SM)

This message is from pH2. This message and any attachments may contain
legally privileged or confidential information, and are intended only for
the individual or entity identified above as the addressee. If you are not
the addressee, or if this message has been addressed to you in error, you
are not authorized to read, copy, or distribute this message and any
attachments, and we ask that you please delete this message and attachments
(including all copies) and notify the sender by return e-mail or by phone at
317-718-7020. Delivery of this message and any attachments to any person
other than the intended recipient(s) is not intended in any way to waive
confidentiality or a privilege. All personal messages express views only of
the sender, which are not to be attributed to pH2 and may not be copied or
distributed without this statement.


-----Original Message-----
X-from: johnf-at-geology.wisc.edu [mailto:johnf-at-geology.wisc.edu]
Sent: Wednesday, January 17, 2007 8:22 AM
To: ph2-at-sprynet.com

}
}
} Question: I want to map the Hg in a contaminated bird feather using
} SEM/EDX/WDS. What's the best way to prepare the sample to minimize
} charging?

When you say "contaminated" do you mean massive contamination,
millions of times greater than what a human would get in their hair
(thru metabolism) eating fish that had "high" levels from Hg used in
adjacent gold mining/extraction? Only in that case would an
electron-based microanalytical technique be of any use.

I attempted to measure Hg in hair from native people from Brazil who
lived near gold extraction areas.The idea was to see if we could see
variations in the hair along length (thus, versus time). Using WDS
EPMA. I knew the approximate levels (thru some technique like INAA or
ICPMS or XRF, I forget) and it was high ppb or low ppm (again, I
forget the details). But I could never put enough current (I probably
went up to 10 or 20 nA) to see anything.

John


==============================Original Headers==============================
5, 25 -- From johnf-at-geology.wisc.edu Wed Jan 17 07:12:42 2007
5, 25 -- Received: from ice.geology.wisc.edu (ice.geology.wisc.edu
[144.92.206.14])
5, 25 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l0HDCfcB014198
5, 25 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 07:12:42
-0600
5, 25 -- Received: from localhost (localhost [127.0.0.1])
5, 25 -- by localhost (Postfix) with ESMTP id 134F720D31;
5, 25 -- Wed, 17 Jan 2007 07:12:40 -0600 (CST)
5, 25 -- Received: from ice.geology.wisc.edu ([127.0.0.1])
5, 25 -- by localhost (geology.wisc.edu [127.0.0.1]) (amavisd-new, port
10024)
5, 25 -- with ESMTP id 27330-08; Wed, 17 Jan 2007 07:12:34 -0600 (CST)
5, 25 -- Received: from [192.168.1.103]
(24-183-52-127.dhcp.mdsn.wi.charter.com [24.183.52.127])
5, 25 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
5, 25 -- (No client certificate requested)
5, 25 -- by ice.geology.wisc.edu (Postfix) with ESMTP id 5B2A320D08;
5, 25 -- Wed, 17 Jan 2007 07:12:34 -0600 (CST)
5, 25 -- Mime-Version: 1.0
5, 25 -- Message-Id: {p06210201c1d3cfb242f0-at-[192.168.0.122]}
5, 25 -- In-Reply-To: {200701170205.l0H2595J030194-at-ns.microscopy.com}
5, 25 -- References: {200701170205.l0H2595J030194-at-ns.microscopy.com}
5, 25 -- Date: Wed, 17 Jan 2007 07:14:58 -0600
5, 25 -- To: chappell.mark-at-epa.gov, microscopy-at-microscopy.com
5, 25 -- From: John Fournelle {johnf-at-geology.wisc.edu}
5, 25 -- Subject: Re: SEM/EDX of bird feather
5, 25 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
5, 25 -- X-Virus-Scanned: amavisd-new at geology.wisc.edu
==============================End of - Headers==============================



==============================Original Headers==============================
18, 26 -- From ph2-at-sprynet.com Wed Jan 17 10:03:46 2007
18, 26 -- Received: from elasmtp-kukur.atl.sa.earthlink.net (elasmtp-kukur.atl.sa.earthlink.net [209.86.89.65])
18, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HG3kNk009005
18, 26 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 10:03:46 -0600
18, 26 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
18, 26 -- s=dk20050327; d=sprynet.com;
18, 26 -- b=tBJkEKzzh1IzjTxhOotRSKxWhIDpta24Q7RcnMnOCMDojh+ZqkmliVFRswecmYB1;
18, 26 -- h=Received:From:To:Subject:Date:MIME-Version:Content-Type:Content-Transfer-Encoding:X-Mailer:X-MimeOLE:Thread-Index:In-Reply-To:Message-ID:X-ELNK-Trace:X-Originating-IP;
18, 26 -- Received: from [75.61.18.94] (helo=user915fa8f284)
18, 26 -- by elasmtp-kukur.atl.sa.earthlink.net with asmtp (Exim 4.34)
18, 26 -- id 1H7DGL-0000o2-J8; Wed, 17 Jan 2007 11:03:45 -0500
18, 26 -- From: "Tony Havics" {ph2-at-sprynet.com}
18, 26 -- To: {johnf-at-geology.wisc.edu} , {microscopy-at-microscopy.com}
18, 26 -- Subject: RE: [Microscopy] Re: SEM/EDX of bird feather
18, 26 -- Date: Wed, 17 Jan 2007 11:03:41 -0500
18, 26 -- MIME-Version: 1.0
18, 26 -- Content-Type: text/plain;
18, 26 -- charset="us-ascii"
18, 26 -- Content-Transfer-Encoding: 7bit
18, 26 -- X-Mailer: Microsoft Office Outlook, Build 11.0.5510
18, 26 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
18, 26 -- Thread-Index: Acc6OnDmWnJkLwdaTQCNI5up6obfkgAFThXg
18, 26 -- In-Reply-To: {200701171321.l0HDLlSR028784-at-ns.microscopy.com}
18, 26 -- Message-ID: {E1H7DGL-0000o2-J8-at-elasmtp-kukur.atl.sa.earthlink.net}
18, 26 -- X-ELNK-Trace: 6888e50b2be9b4fee5331016acda17f9d351a46f1ff38ab561926f01341a0e9c350badd9bab72f9c350badd9bab72f9c350badd9bab72f9c350badd9bab72f9c
18, 26 -- X-Originating-IP: 75.61.18.94
==============================End of - Headers==============================




From: ph2-at-sprynet.com
Date: Wed, 17 Jan 2007 10:19:49 -0600
Subject: [Microscopy] Re: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

It got the better of me:

The article was:

McCrone, Walter, and Michael Bayard: Individualization of Hair. Microscope
47(3):129-133. 1999.

Good article. Ion microprobe.

Walter passed away a few years ago, but Mike Bayard is still around (Bayard
consulting).

Tony

......................................................................
Andrew Anthony "Tony" Havics, CHMM, CIH, PE
pH2, LLC
5250 E US 36, Suite 830
Avon, IN 46123
(317) 718-7020 off
(317) 718-7038 fax
(317) 409-3238 cell

90% of Risk Management is knowing where to place the decimal point...any
consultant can give you the other 10%(SM)

This message is from pH2. This message and any attachments may contain
legally privileged or confidential information, and are intended only for
the individual or entity identified above as the addressee. If you are not
the addressee, or if this message has been addressed to you in error, you
are not authorized to read, copy, or distribute this message and any
attachments, and we ask that you please delete this message and attachments
(including all copies) and notify the sender by return e-mail or by phone at
317-718-7020. Delivery of this message and any attachments to any person
other than the intended recipient(s) is not intended in any way to waive
confidentiality or a privilege. All personal messages express views only of
the sender, which are not to be attributed to pH2 and may not be copied or
distributed without this statement.


-----Original Message-----
X-from: johnf-at-geology.wisc.edu [mailto:johnf-at-geology.wisc.edu]
Sent: Wednesday, January 17, 2007 8:22 AM
To: ph2-at-sprynet.com

}
}
} Question: I want to map the Hg in a contaminated bird feather using
} SEM/EDX/WDS. What's the best way to prepare the sample to minimize
} charging?

When you say "contaminated" do you mean massive contamination,
millions of times greater than what a human would get in their hair
(thru metabolism) eating fish that had "high" levels from Hg used in
adjacent gold mining/extraction? Only in that case would an
electron-based microanalytical technique be of any use.

I attempted to measure Hg in hair from native people from Brazil who
lived near gold extraction areas.The idea was to see if we could see
variations in the hair along length (thus, versus time). Using WDS
EPMA. I knew the approximate levels (thru some technique like INAA or
ICPMS or XRF, I forget) and it was high ppb or low ppm (again, I
forget the details). But I could never put enough current (I probably
went up to 10 or 20 nA) to see anything.

John


==============================Original Headers==============================
5, 25 -- From johnf-at-geology.wisc.edu Wed Jan 17 07:12:42 2007
5, 25 -- Received: from ice.geology.wisc.edu (ice.geology.wisc.edu
[144.92.206.14])
5, 25 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l0HDCfcB014198
5, 25 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 07:12:42
-0600
5, 25 -- Received: from localhost (localhost [127.0.0.1])
5, 25 -- by localhost (Postfix) with ESMTP id 134F720D31;
5, 25 -- Wed, 17 Jan 2007 07:12:40 -0600 (CST)
5, 25 -- Received: from ice.geology.wisc.edu ([127.0.0.1])
5, 25 -- by localhost (geology.wisc.edu [127.0.0.1]) (amavisd-new, port
10024)
5, 25 -- with ESMTP id 27330-08; Wed, 17 Jan 2007 07:12:34 -0600 (CST)
5, 25 -- Received: from [192.168.1.103]
(24-183-52-127.dhcp.mdsn.wi.charter.com [24.183.52.127])
5, 25 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
5, 25 -- (No client certificate requested)
5, 25 -- by ice.geology.wisc.edu (Postfix) with ESMTP id 5B2A320D08;
5, 25 -- Wed, 17 Jan 2007 07:12:34 -0600 (CST)
5, 25 -- Mime-Version: 1.0
5, 25 -- Message-Id: {p06210201c1d3cfb242f0-at-[192.168.0.122]}
5, 25 -- In-Reply-To: {200701170205.l0H2595J030194-at-ns.microscopy.com}
5, 25 -- References: {200701170205.l0H2595J030194-at-ns.microscopy.com}
5, 25 -- Date: Wed, 17 Jan 2007 07:14:58 -0600
5, 25 -- To: chappell.mark-at-epa.gov, microscopy-at-microscopy.com
5, 25 -- From: John Fournelle {johnf-at-geology.wisc.edu}
5, 25 -- Subject: Re: SEM/EDX of bird feather
5, 25 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
5, 25 -- X-Virus-Scanned: amavisd-new at geology.wisc.edu
==============================End of - Headers==============================



==============================Original Headers==============================
20, 26 -- From ph2-at-sprynet.com Wed Jan 17 10:19:49 2007
20, 26 -- Received: from elasmtp-kukur.atl.sa.earthlink.net (elasmtp-kukur.atl.sa.earthlink.net [209.86.89.65])
20, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HGJm4f020222
20, 26 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 10:19:48 -0600
20, 26 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
20, 26 -- s=dk20050327; d=sprynet.com;
20, 26 -- b=j37OA7fofM46Bxvdk/AJjcToLbIYpxUbWPBhBeMlbU3bmT9mNy7DHXey1mELx3Q6;
20, 26 -- h=Received:From:To:Subject:Date:MIME-Version:Content-Type:Content-Transfer-Encoding:X-Mailer:X-MimeOLE:Thread-Index:In-Reply-To:Message-ID:X-ELNK-Trace:X-Originating-IP;
20, 26 -- Received: from [75.61.18.94] (helo=user915fa8f284)
20, 26 -- by elasmtp-kukur.atl.sa.earthlink.net with asmtp (Exim 4.34)
20, 26 -- id 1H7DVr-0000Gl-Ck; Wed, 17 Jan 2007 11:19:47 -0500
20, 26 -- From: "Tony Havics" {ph2-at-sprynet.com}
20, 26 -- To: {johnf-at-geology.wisc.edu} , {microscopy-at-microscopy.com}
20, 26 -- Subject: RE: [Microscopy] Re: SEM/EDX of bird feather
20, 26 -- Date: Wed, 17 Jan 2007 11:19:43 -0500
20, 26 -- MIME-Version: 1.0
20, 26 -- Content-Type: text/plain;
20, 26 -- charset="us-ascii"
20, 26 -- Content-Transfer-Encoding: 7bit
20, 26 -- X-Mailer: Microsoft Office Outlook, Build 11.0.5510
20, 26 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
20, 26 -- Thread-Index: Acc6OnDmWnJkLwdaTQCNI5up6obfkgAGG2UQ
20, 26 -- In-Reply-To: {200701171321.l0HDLlSR028784-at-ns.microscopy.com}
20, 26 -- Message-ID: {E1H7DVr-0000Gl-Ck-at-elasmtp-kukur.atl.sa.earthlink.net}
20, 26 -- X-ELNK-Trace: 6888e50b2be9b4fee5331016acda17f92216719cbc12cfa5f0f1f9848c34fed8350badd9bab72f9c350badd9bab72f9c350badd9bab72f9c350badd9bab72f9c
20, 26 -- X-Originating-IP: 75.61.18.94
==============================End of - Headers==============================




From: mike.wombwell-at-quorumtech.com
Date: Wed, 17 Jan 2007 10:21:14 -0600
Subject: [Microscopy] Polaron E5100 sputter coater question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Randy,

The maximum voltage from the HV transformer is 2800 volts so it would be
safe to set the control knob shaft to the fully clockwise position and
place the knob at the 2.8 kV end of the scale.
For reference, this is a very old coater (discontinued in around 1988)
but some parts are still available. For further support please contact
our US distributor Energy Beam Sciences
(http://www.quorumtech.com/contact_us/usa.htm).

You can also find a download of the E5100 manual on the technical
support pages of our website
(http://www.quorumtech.com/Tech_Support/polaron-range-technical-support.
htm)

I hope this helps.

Best regards,

Mike Wombwell
Quorum Technologies
Newhaven, East Sussex, UK
Tel: +44(0)1273 510535
Fax: +44(0)1273 510536
mike.wombwell-at-quorumtech.com
http://www.quorumtech.com

E & O E


-----Original Message-----
X-from: TindallR-at-missouri.edu [mailto:TindallR-at-missouri.edu]
Sent: 17 January 2007 15:44
To: Mike Wombwell

Dear Listers,

While our main sputter coater is out for a much-needed overhaul, we are
relying on a backup coater, namely a Polaron E5100. The unit is
operating beautifully, except that the voltage dial has worked loose at
some point and I can no longer tell where the zero point, and thus the
proper operating voltage, is located on the scale. This is because the
scale starts at 2.0, not zero, and I have no reference to line up the
indicator with. I tried looking at the set-screw scar on the knob shaft,
but it didn't help. Am I missing something obvious? (Wouldn't be the
first time.....)

f you will go to http://www.emc.missouri.edu/lookatthis.htm to see the
images of this coater hopefully this will make sense.

Can someone with a similar unit kindly let me know about where the zero
point would be on this machine? I can get it to coat by trial and
error, but it would be nice to have a baseline to experiment from.

Many thanks!

Searching for zero,
Randy


Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original
Headers==============================
9, 23 -- From TindallR-at-missouri.edu Wed Jan 17 09:39:47 2007
9, 23 -- Received: from um-tsmtpout1.um.umsystem.edu
(um-tsmtpout1.um.umsystem.edu [209.106.228.23])
9, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0HFdlGO018946
9, 23 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007
09:39:47 -0600
9, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by
um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
9, 23 -- Wed, 17 Jan 2007 09:39:47 -0600
9, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
9, 23 -- Content-class: urn:content-classes:message
9, 23 -- MIME-Version: 1.0
9, 23 -- Content-Type: text/plain;
9, 23 -- charset="us-ascii"
9, 23 -- Subject: Polaron E5100 sputter coater question
9, 23 -- Date: Wed, 17 Jan 2007 09:39:46 -0600
9, 23 -- Message-ID:
{91108EF9255B394CBF8B7E3789814A41C68CF1-at-UM-XMAIL08.um.umsystem.edu}
9, 23 -- X-MS-Has-Attach:
9, 23 -- X-MS-TNEF-Correlator:
9, 23 -- Thread-Topic: Polaron E5100 sputter coater question
9, 23 -- Thread-Index: Acc6TbbzCo8D3MqPS+a3y10wR2ngWQ==
9, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
9, 23 -- To: {microscopy-at-microscopy.com}
9, 23 -- X-OriginalArrivalTime: 17 Jan 2007 15:39:47.0209 (UTC)
FILETIME=[B7B14790:01C73A4D]
9, 23 -- Content-Transfer-Encoding: 8bit
9, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0HFdlGO018946
==============================End of -
Headers==============================


==============================Original Headers==============================
23, 24 -- From mike.wombwell-at-quorumtech.com Wed Jan 17 10:21:13 2007
23, 24 -- Received: from thb-mta-10.emailfiltering.com (thb-mta-10.emailfiltering.com [213.166.4.221])
23, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HGLAhi021429
23, 24 -- for {Microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 10:21:11 -0600
23, 24 -- Received: from host213-123-204-94.in-addr.btopenworld.com ([213.123.204.94])
23, 24 -- by thb-mta-10.emailfiltering.com with emfmta (version 3.2.1.3113.11.rd-3.2.2-libc2.3.1) vanilla id 174099874
23, 24 -- ; Wed, 17 Jan 2007 16:21:06 +0000
23, 24 -- Content-class: urn:content-classes:message
23, 24 -- Subject: RE: [Microscopy] Polaron E5100 sputter coater question
23, 24 -- MIME-Version: 1.0
23, 24 -- Content-Type: text/plain;
23, 24 -- charset="us-ascii"
23, 24 -- Date: Wed, 17 Jan 2007 16:24:41 -0000
23, 24 -- X-MimeOLE: Produced By Microsoft Exchange V6.5.7226.0
23, 24 -- Message-ID: {4287884323A072458FB10232AA8FDB39293848-at-server01.QuorumTechnologiesLtd.local}
23, 24 -- X-MS-Has-Attach:
23, 24 -- X-MS-TNEF-Correlator:
23, 24 -- Thread-Topic: [Microscopy] Polaron E5100 sputter coater question
23, 24 -- Thread-Index: Acc6TsbyCrZmWbmeQWCcQtMDRN+h8wAAlovg
23, 24 -- From: "Mike Wombwell" {mike.wombwell-at-quorumtech.com}
23, 24 -- To: {TindallR-at-missouri.edu}
23, 24 -- Cc: {Microscopy-at-microscopy.com}
23, 24 -- Content-Transfer-Encoding: 8bit
23, 24 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0HGLAhi021429
==============================End of - Headers==============================




From: eggert-at-mikroanalytik.de
Date: Wed, 17 Jan 2007 10:22:16 -0600
Subject: [Microscopy] Re: FW: AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Mark,

I would also try to use laboratory micro-XRF before using a synchrotron
source. The advantages against common SEM excitation are much better
detection limits (because there is no bremsstrahlung in spectra
background) and no need of vacuum. I think, micro-XRF meets the needs of
your analysis goals.

Frank Eggert

===============================
http://www.microanalyst.net
===============================


fortner-at-cmt.anl.gov wrote:

} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
8, 19 -- From eggert-at-mikroanalytik.de Wed Jan 17 10:22:15 2007
8, 19 -- Received: from smtp-out.kontent.com (smtp-out001.kontent.com [81.88.40.215])
8, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HGME90024076
8, 19 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 10:22:14 -0600
8, 19 -- Received: from mikroanalytik.de (p54BDC949.dip0.t-ipconnect.de [84.189.201.73])
8, 19 -- by smtp-out.kontent.com (Postfix) with ESMTP id 7C0A137C197;
8, 19 -- Wed, 17 Jan 2007 17:22:09 +0100 (CET)
8, 19 -- Message-ID: {45AE4CD6.2040607-at-mikroanalytik.de}
8, 19 -- Date: Wed, 17 Jan 2007 17:20:38 +0100
8, 19 -- From: Frank Eggert {eggert-at-mikroanalytik.de}
8, 19 -- User-Agent: Mozilla/5.0 (Windows; U; Windows NT 5.0; de-DE; rv:1.4) Gecko/20030619 Netscape/7.1 (ax)
8, 19 -- X-Accept-Language: de, de-de, en
8, 19 -- MIME-Version: 1.0
8, 19 -- To: microscopy-at-microscopy.com, chappell.mark-at-epa.gov
8, 19 -- Subject: Re: [Microscopy] FW: AskAMicroscopist: SEM/EDX of bird feather
8, 19 -- References: {200701171600.l0HG06OH001278-at-ns.microscopy.com}
8, 19 -- In-Reply-To: {200701171600.l0HG06OH001278-at-ns.microscopy.com}
8, 19 -- Content-Type: text/plain; charset=us-ascii; format=flowed
8, 19 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Wed, 17 Jan 2007 10:44:54 -0600
Subject: [Microscopy] Polaron E5100 sputter coater question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Mike,

Your link didn't work, but I was able to get to the manual with
http://www.quorumtech.com/Manuals/E5100-manual.pdf. Thanks very much
for this. I'm posting this to the listserve, in case others have this
problem also.

I think we've got the problem solved with your suggestion of setting the
maximum setting at 2.8 on the dial. Full counterclockwise with this pot
then comes out at about 3:00, as Richard Kucklinski also noted in a
off-list posting to me. This makes sense now.

I'm suspecting an argon leak, since with the argon turned fully OFF, I'm
still getting } 40 mA coating current with the dial set at 2.0-2.2. I'll
experiment around with it to get it fine-tuned a bit better, but it's
coating reliably now.

Many thanks to all who emailed and phoned. What a resource!!

Cheers,
Randy

-----Original Message-----
X-from: mike.wombwell-at-quorumtech.com [mailto:mike.wombwell-at-quorumtech.com]

Sent: Wednesday, January 17, 2007 10:22 AM
To: Tindall, Randy D.

Dear Randy,

The maximum voltage from the HV transformer is 2800 volts so it would be
safe to set the control knob shaft to the fully clockwise position and
place the knob at the 2.8 kV end of the scale.
For reference, this is a very old coater (discontinued in around 1988)
but some parts are still available. For further support please contact
our US distributor Energy Beam Sciences
(http://www.quorumtech.com/contact_us/usa.htm).

You can also find a download of the E5100 manual on the technical
support pages of our website
(http://www.quorumtech.com/Tech_Support/polaron-range-technical-support.
htm)

I hope this helps.

Best regards,

Mike Wombwell
Quorum Technologies
Newhaven, East Sussex, UK
Tel: +44(0)1273 510535
Fax: +44(0)1273 510536
mike.wombwell-at-quorumtech.com
http://www.quorumtech.com

E & O E


-----Original Message-----
X-from: TindallR-at-missouri.edu [mailto:TindallR-at-missouri.edu]
Sent: 17 January 2007 15:44
To: Mike Wombwell

Dear Listers,

While our main sputter coater is out for a much-needed overhaul, we are
relying on a backup coater, namely a Polaron E5100. The unit is
operating beautifully, except that the voltage dial has worked loose at
some point and I can no longer tell where the zero point, and thus the
proper operating voltage, is located on the scale. This is because the
scale starts at 2.0, not zero, and I have no reference to line up the
indicator with. I tried looking at the set-screw scar on the knob shaft,
but it didn't help. Am I missing something obvious? (Wouldn't be the
first time.....)

f you will go to http://www.emc.missouri.edu/lookatthis.htm to see the
images of this coater hopefully this will make sense.

Can someone with a similar unit kindly let me know about where the zero
point would be on this machine? I can get it to coat by trial and
error, but it would be nice to have a baseline to experiment from.

Many thanks!

Searching for zero,
Randy


Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original
Headers==============================
9, 23 -- From TindallR-at-missouri.edu Wed Jan 17 09:39:47 2007
9, 23 -- Received: from um-tsmtpout1.um.umsystem.edu
(um-tsmtpout1.um.umsystem.edu [209.106.228.23])
9, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0HFdlGO018946
9, 23 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007
09:39:47 -0600
9, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by
um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
9, 23 -- Wed, 17 Jan 2007 09:39:47 -0600
9, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
9, 23 -- Content-class: urn:content-classes:message
9, 23 -- MIME-Version: 1.0
9, 23 -- Content-Type: text/plain;
9, 23 -- charset="us-ascii"
9, 23 -- Subject: Polaron E5100 sputter coater question
9, 23 -- Date: Wed, 17 Jan 2007 09:39:46 -0600
9, 23 -- Message-ID:
{91108EF9255B394CBF8B7E3789814A41C68CF1-at-UM-XMAIL08.um.umsystem.edu}
9, 23 -- X-MS-Has-Attach:
9, 23 -- X-MS-TNEF-Correlator:
9, 23 -- Thread-Topic: Polaron E5100 sputter coater question
9, 23 -- Thread-Index: Acc6TbbzCo8D3MqPS+a3y10wR2ngWQ==
9, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
9, 23 -- To: {microscopy-at-microscopy.com}
9, 23 -- X-OriginalArrivalTime: 17 Jan 2007 15:39:47.0209 (UTC)
FILETIME=[B7B14790:01C73A4D]
9, 23 -- Content-Transfer-Encoding: 8bit
9, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0HFdlGO018946
==============================End of -
Headers==============================


==============================Original
Headers==============================
23, 24 -- From mike.wombwell-at-quorumtech.com Wed Jan 17 10:21:13 2007
23, 24 -- Received: from thb-mta-10.emailfiltering.com
(thb-mta-10.emailfiltering.com [213.166.4.221])
23, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0HGLAhi021429
23, 24 -- for {Microscopy-at-microscopy.com} ; Wed, 17 Jan 2007
10:21:11 -0600
23, 24 -- Received: from host213-123-204-94.in-addr.btopenworld.com
([213.123.204.94])
23, 24 -- by thb-mta-10.emailfiltering.com with emfmta (version
3.2.1.3113.11.rd-3.2.2-libc2.3.1) vanilla id 174099874
23, 24 -- ; Wed, 17 Jan 2007 16:21:06 +0000
23, 24 -- Content-class: urn:content-classes:message
23, 24 -- Subject: RE: [Microscopy] Polaron E5100 sputter coater
question
23, 24 -- MIME-Version: 1.0
23, 24 -- Content-Type: text/plain;
23, 24 -- charset="us-ascii"
23, 24 -- Date: Wed, 17 Jan 2007 16:24:41 -0000
23, 24 -- X-MimeOLE: Produced By Microsoft Exchange V6.5.7226.0
23, 24 -- Message-ID:
{4287884323A072458FB10232AA8FDB39293848-at-server01.QuorumTechnologiesLtd.l
ocal}
23, 24 -- X-MS-Has-Attach:
23, 24 -- X-MS-TNEF-Correlator:
23, 24 -- Thread-Topic: [Microscopy] Polaron E5100 sputter coater
question
23, 24 -- Thread-Index: Acc6TsbyCrZmWbmeQWCcQtMDRN+h8wAAlovg
23, 24 -- From: "Mike Wombwell" {mike.wombwell-at-quorumtech.com}
23, 24 -- To: {TindallR-at-missouri.edu}
23, 24 -- Cc: {Microscopy-at-microscopy.com}
23, 24 -- Content-Transfer-Encoding: 8bit
23, 24 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0HGLAhi021429
==============================End of -
Headers==============================


==============================Original Headers==============================
36, 26 -- From TindallR-at-missouri.edu Wed Jan 17 10:44:53 2007
36, 26 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
36, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HGiquc021008
36, 26 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 10:44:53 -0600
36, 26 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
36, 26 -- Wed, 17 Jan 2007 10:44:52 -0600
36, 26 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
36, 26 -- Content-class: urn:content-classes:message
36, 26 -- MIME-Version: 1.0
36, 26 -- Content-Type: text/plain;
36, 26 -- charset="us-ascii"
36, 26 -- Subject: RE: [Microscopy] RE: Polaron E5100 sputter coater question
36, 26 -- Date: Wed, 17 Jan 2007 10:44:51 -0600
36, 26 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68CF3-at-UM-XMAIL08.um.umsystem.edu}
36, 26 -- in-reply-to: {200701171622.l0HGMF4f024191-at-ns.microscopy.com}
36, 26 -- X-MS-Has-Attach:
36, 26 -- X-MS-TNEF-Correlator:
36, 26 -- Thread-Topic: [Microscopy] RE: Polaron E5100 sputter coater question
36, 26 -- Thread-Index: Acc6U6huRQGSU+asQaSz9+NIox/asQAAclSA
36, 26 -- References: {200701171622.l0HGMF4f024191-at-ns.microscopy.com}
36, 26 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
36, 26 -- To: {mike.wombwell-at-quorumtech.com}
36, 26 -- Cc: {microscopy-at-microscopy.com}
36, 26 -- X-OriginalArrivalTime: 17 Jan 2007 16:44:52.0515 (UTC) FILETIME=[CF6FAF30:01C73A56]
36, 26 -- Content-Transfer-Encoding: 8bit
36, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0HGiquc021008
==============================End of - Headers==============================




From: mager-at-interchange.ubc.ca
Date: Wed, 17 Jan 2007 11:09:31 -0600
Subject: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Mark,
I have done this in the past, for the Royal Museum in Victoria, BC. They
were interested in the contamination of feathers and fur in old stuffed
animals and birds, because mercury and arsenic used to be used to preserve
museum samples many years ago and the museum wanted to know if these
exhibits were contaminated with heavy elements before they were put on
display or exposed them to the public. The heavy elements were expected to
be applied to the feathers, so the amounts were large, if they were there. I
used the SEM in variable pressure mode, because fluffy things are so
difficult to coat effectively, and just scanned around with the
back-scattered detector, looking for bright things. Any I found I took a
picture of and analysed with EDX. I have also detected elevated levels of
copper in someone's hair using SEM+EDX.
Looking for smaller amounts of heavy elements or organically-bound material
in the interior of the sample would require the sample to be mounted in
epoxy, polished, carbon-coated and then careful WDX analysis.
Good luck,
Mary Mager
Electron Microscopist
Materials Eng. UBC
#419 - 6350 Stores Rd.
Vancouver, B.C. V6T 1Z4
CANADA
Phone: 604-822-5648
Fax: 604-822-3619
email: mager-at-interchange.ubc.ca

-----Original Message-----
X-from: chappell.mark-at-epa.gov [mailto:chappell.mark-at-epa.gov]
Sent: Tuesday, January 16, 2007 6:06 PM
To: mager-at-interchange.ubc.ca

This Question was submitted to Ask-A-Microscopist by (chappell.mark-at-epa.gov)
from http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on
Tuesday, January 16, 2007 at 13:59:24 Remember to consider the Grade/Age of
the student when considering the Question
---------------------------------------------------------------------------
Please reply to both chappell.mark-at-epa.gov as well as to the Microscopy
Listserver
---------------------------------------------------------------------------

Email: chappell.mark-at-epa.gov
Name: Mark Chappell

Organization: U.S. EPA

Education: Graduate College

Location: Cincinnati, OH

Title: SEM/EDX of bird feather

Question: I want to map the Hg in a contaminated bird feather using
SEM/EDX/WDS. What's the best way to prepare the sample to minimize
charging?

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 16 20:01:02 2007 8, 12
-- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l0H211P7022216
8, 12 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007 20:01:02
-0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified) 8, 12 --
Message-Id: {p06110401c1d333bcefd0-at-[206.69.208.22]}
8, 12 -- Date: Tue, 16 Jan 2007 20:01:00 -0600 8, 12 -- To:
microscopy-at-microscopy.com 8, 12 -- From: chappell.mark-at-epa.gov (by way of
Ask-A-Microscopist) 8, 12 -- Subject: AskAMicroscopist: SEM/EDX of bird
feather 8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================


==============================Original Headers==============================
15, 33 -- From mager-at-interchange.ubc.ca Wed Jan 17 11:09:30 2007
15, 33 -- Received: from mta1.mail-relay.ubc.ca (mta1.mail-relay.ubc.ca [137.82.45.2])
15, 33 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HH9U65032360
15, 33 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 11:09:30 -0600
15, 33 -- Received: from mta1.interchange.ubc.ca (mta1.interchange.ubc.ca [142.103.145.69])
15, 33 -- by mta1.mail-relay.ubc.ca (8.12.11.20060308/8.12.11) with ESMTP id l0HH9Qa7021791
15, 33 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 09:09:27 -0800 (PST)
15, 33 -- (envelope-from mager-at-interchange.ubc.ca)
15, 33 -- Received: from Mary (desk.staff.mmat.ubc.ca [137.82.16.178])
15, 33 -- by smtp.interchange.ubc.ca
15, 33 -- (iPlanet Messaging Server 5.2 HotFix 1.21 (built Sep 8 2003))
15, 33 -- with ESMTPS id {0JC000LG7VNQGE-at-smtp.interchange.ubc.ca} for
15, 33 -- microscopy-at-microscopy.com; Wed, 17 Jan 2007 09:09:26 -0800 (PST)
15, 33 -- Date: Wed, 17 Jan 2007 09:08:54 -0800
15, 33 -- From: Mary Mager {mager-at-interchange.ubc.ca}
15, 33 -- Subject: RE: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
15, 33 -- In-reply-to: {200701170205.l0H25aJh031269-at-ns.microscopy.com}
15, 33 -- To: chappell.mark-at-epa.gov
15, 33 -- Cc: Microscopy {microscopy-at-microscopy.com}
15, 33 -- Reply-to: mager-at-interchange.ubc.ca
15, 33 -- Message-id: {061901c73a5a$2b638ac0$b2105289-at-staff.mmat.ubc.ca}
15, 33 -- Organization: Materials Eng. UBC
15, 33 -- MIME-version: 1.0
15, 33 -- X-MIMEOLE: Produced By Microsoft MimeOLE V6.00.2800.1896
15, 33 -- X-Mailer: Microsoft Office Outlook 11
15, 33 -- Content-type: text/plain; charset=us-ascii
15, 33 -- Content-transfer-encoding: 7bit
15, 33 -- Thread-index: Acc52/qy+HURPdd7SuGpzEFJ9qGpSwAfNatg
15, 33 -- X-UBC-Scanned: Sophos PureMessage 4.7.1.128075, Antispam-Engine: 2.5.0.283055, Antispam-Data: 2007.1.17.85932
15, 33 -- X-UBC-Relayed: Relayed through mail-relay.ubc.ca
15, 33 -- X-PerlMx-Spam: Probability=7%, Report=__C230066_P5 0, __CP_URI_IN_BODY 0, __CT 0, __CTE 0, __CT_TEXT_PLAIN 0, __HAS_MSGID 0, __HAS_X_MAILER 0, __MIME_TEXT_ONLY 0, __MIME_VERSION 0, __SANE_MSGID 0, __STOCK_PHRASE_24 0, __USER_AGENT_MS_GENERIC 0
15, 33 -- X-Spam-Level:
15, 33 -- X-Spam-Flag: No
==============================End of - Headers==============================




From: eschumacher-at-mccrone.com
Date: Wed, 17 Jan 2007 11:22:34 -0600
Subject: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Posted on behalf of a colleague here at McCrone:

Actually, the work done by Dr. McCrone in the early 1970's on a forensic
hair case from Detroit, used an ion microprobe/secondary ion mass
spectrometry (SIMS) rather than neutron activation analysis (NAA). We
have abandoned the idea of mapping the trace elemental concentrations
along the shaft of the hair due to many difficulties with that approach.
More recently we concentrated trace elements in hairs by low temperature
ashing (LTA) and used SIMS to identify and measure trace element
concentrations in the ash. The quantification isn't very accurate and
the levels need to be compared with good positive and negative controls
because there is no baseline data for comparison. In other words, you
might determine that a questioned hair contains levels of lead, for
example, greater than expected in hairs from normal people where they
have not received significant exposure to lead. We presume the same
approach could be used for feathers. Of course, the other option is to
use a bulk method such as XRF at Argonne. For more information contact
Dick Bisbing off-line at dbisbing-at-mccrone.com.

Regards,

Elaine
*********************************************************************
Elaine F.Schumacher
Senior Research Scientist
McCrone Associates, Inc.
850 Pasquinelli Drive
Westmont, IL 60559-5539 USA
630-887-7100 (tel)
630-887-7417 (fax)
E-mail: eschumacher-at-mccrone.com
Web Site: www.mccrone.com



*********************************************************************
This message and any attachments are solely for the
intended recipient. If you are not the intended recipient,
disclosure, copying, use or distribution of the information
included in this message is prohibited.
*********************************************************************

-----Original Message-----
X-from: chappell.mark-at-epa.gov [mailto:chappell.mark-at-epa.gov]
Sent: Tuesday, January 16, 2007 8:06 PM
To: Elaine F. Schumacher

This Question was submitted to Ask-A-Microscopist by
(chappell.mark-at-epa.gov)
from
http://www.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on
Tuesday, January 16, 2007 at 13:59:24
Remember to consider the Grade/Age of the student when considering the
Question
------------------------------------------------------------------------
---
Please reply to both chappell.mark-at-epa.gov as well as to the
Microscopy Listserver
------------------------------------------------------------------------
---

Email: chappell.mark-at-epa.gov
Name: Mark Chappell

Organization: U.S. EPA

Education: Graduate College

Location: Cincinnati, OH

Title: SEM/EDX of bird feather

Question: I want to map the Hg in a contaminated bird feather using
SEM/EDX/WDS. What's the best way to prepare the sample to minimize
charging?

------------------------------------------------------------------------
---

==============================Original
Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Tue Jan 16 20:01:02 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com
[206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0H211P7022216
8, 12 -- for {microscopy-at-microscopy.com} ; Tue, 16 Jan 2007
20:01:02 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
8, 12 -- Message-Id: {p06110401c1d333bcefd0-at-[206.69.208.22]}
8, 12 -- Date: Tue, 16 Jan 2007 20:01:00 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: chappell.mark-at-epa.gov (by way of Ask-A-Microscopist)
8, 12 -- Subject: AskAMicroscopist: SEM/EDX of bird feather
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of -
Headers==============================



==============================Original Headers==============================
22, 28 -- From eschumacher-at-mccrone.com Wed Jan 17 11:22:34 2007
22, 28 -- Received: from pgp.mccrone.com (McCrone-Associates-1051626.cust-rtr.ameritech.net [68.23.63.122])
22, 28 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0HHMXn9011085
22, 28 -- for {Microscopy-at-MSA.Microscopy.Com} ; Wed, 17 Jan 2007 11:22:33 -0600
22, 28 -- Received: from pgp.mccrone.com (localhost [127.0.0.1])
22, 28 -- by pgp.mccrone.com (Postfix) with ESMTP id EC24A1A800B
22, 28 -- for {Microscopy-at-MSA.Microscopy.Com} ; Wed, 17 Jan 2007 11:22:35 -0600 (CST)
22, 28 -- Received: from MCCRONEMSG.tmg.mccrone.com ([192.168.101.20])
22, 28 -- by pgp.mccrone.com (PGP Universal service);
22, 28 -- Wed, 17 Jan 2007 11:22:36 -0600
22, 28 -- X-PGP-Universal: processed
22, 28 -- Content-class: urn:content-classes:message
22, 28 -- MIME-Version: 1.0
22, 28 -- Content-Type: text/plain;
22, 28 -- charset="US-ASCII"
22, 28 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
22, 28 -- Subject: RE: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
22, 28 -- Date: Wed, 17 Jan 2007 11:22:23 -0600
22, 28 -- Message-ID: {2A27CE0EC2A5C748974EA513DFB285A7B973B8-at-MCCRONEMSG.tmg.mccrone.com}
22, 28 -- X-MS-Has-Attach:
22, 28 -- X-MS-TNEF-Correlator:
22, 28 -- Thread-Topic: [Microscopy] AskAMicroscopist: SEM/EDX of bird feather
22, 28 -- Thread-Index: Acc53Aa8ajunK7QAS5KNQLnjgiOKowAf8ASA
22, 28 -- From: "Elaine F. Schumacher" {eschumacher-at-mccrone.com}
22, 28 -- To: {chappell.mark-at-epa.gov} , {Microscopy-at-MSA.Microscopy.Com}
22, 28 -- Cc: "Richard E. Bisbing" {dbisbing-at-mccrone.com}
22, 28 -- Content-Transfer-Encoding: 8bit
22, 28 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0HHMXn9011085
==============================End of - Headers==============================




From: sue.tyler-at-noaa.gov
Date: Wed, 17 Jan 2007 18:33:45 -0600
Subject: [Microscopy] viaWWW: TEM and ecological assesment

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Randy

If I remember correctly the top end of the E5000 series variac scale was
2.8kV. Best results, if you do not want to cook your specimen, are achieved
at 20mA and around 800 volts in other words about 1/3 to 1/2 scale from
minimum.

Hope this helps?

Steve Chapman
Protrain for EM training & consultancy world wide
www.emcourses.com
Tel +44 1280 816512 Fax +44 1280 814007
Mobile +44 7711 606967

----- Original Message -----
X-from: {TindallR-at-missouri.edu}
To: {protrain-at-emcourses.com}
Sent: Wednesday, January 17, 2007 4:45 PM

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both sue.tyler-at-noaa.gov as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: sue.tyler-at-noaa.gov
Name: Sue Tyler

Organization: Cooperative Oxford Laboratory

Title-Subject: [Filtered] TEM and ecological assesment

Question: My laboratory is changing directions and shifting from disease research to ecosystem health assessment. The past users of the TEM lab have asked me to justify the need for TEM in this new line of work. I am the technician and not aware of current research in the field of ecosystem health. I have received a few papers off the internet. There are a few institutions who use TEM in that field, but I was wondering if there are anymore out there? Can you point me in the right direction?

With the wealth of knowlege in "microscopy land" I am sure someone can help me.

Thanks in advance

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-microscopy.com Wed Jan 17 18:33:45 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0I0XiZM010213
8, 12 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 18:33:45 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: (Unverified)
8, 12 -- Message-Id: {p06110406c1d470c23ea0-at-[206.69.208.22]}
8, 12 -- Date: Wed, 17 Jan 2007 18:33:44 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: sue.tyler-at-noaa.gov (by way of MicroscopyListserver)
8, 12 -- Subject: viaWWW: TEM and ecological assesment
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: klangwor-at-uoregon.edu
Date: Wed, 17 Jan 2007 18:34:23 -0600
Subject: [Microscopy] viaWWW: EBL/ PMMA Mask

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both klangwor-at-uoregon.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: klangwor-at-uoregon.edu
Name: Kurt Langworthy

Organization: University of Oregon

Title-Subject: [Filtered] EBL/ PMMA Mask

Question: All,
I'm looking for a gold etching recipt that will allow me to use PMMA as a mask. KCN looks promising, but is highly toxic. Does anyone have any better ideas?
Thanks,
Kurt

---------------------------------------------------------------------------

==============================Original Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Wed Jan 17 18:34:22 2007
6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0I0YJlb010881
6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 18:34:21 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110407c1d470f44a74-at-[206.69.208.22]}
6, 12 -- Date: Wed, 17 Jan 2007 18:34:19 -0600
6, 12 -- To: microscopy-at-microscopy.com
6, 12 -- From: klangwor-at-uoregon.edu (by way of MicroscopyListserver)
6, 12 -- Subject: viaWWW: EBL/ PMMA Mask
6, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Thu, 18 Jan 2007 10:12:20 -0600
Subject: [Microscopy] Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

And along a completely different line, we've been asked if we can do a
"spark emission test". Since I don't know what that is, and Google
hasn't helped, I don't know if we can or not.

Has anybody heard of this test? Seems to be a way of checking sample
composition at concentrations below those attainable by EDS, but that's
about all I can tell.

Sorry if this isn't microscopy-related, but I don't even know that, at
this point. We'll see what the collective comes up with on this one!

Thanks to all.

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original Headers==============================
7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 10:12:19 2007
7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IGCJAl020222
7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 10:12:19 -0600
7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
7, 23 -- Thu, 18 Jan 2007 10:12:19 -0600
7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
7, 23 -- Content-class: urn:content-classes:message
7, 23 -- MIME-Version: 1.0
7, 23 -- Content-Type: text/plain;
7, 23 -- charset="us-ascii"
7, 23 -- Subject: Spark emission test?
7, 23 -- Date: Thu, 18 Jan 2007 10:12:18 -0600
7, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D0E-at-UM-XMAIL08.um.umsystem.edu}
7, 23 -- X-MS-Has-Attach:
7, 23 -- X-MS-TNEF-Correlator:
7, 23 -- Thread-Topic: Spark emission test?
7, 23 -- Thread-Index: Acc7G207OlT1I53OTyafRC/FuYiYDg==
7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
7, 23 -- To: {microscopy-at-microscopy.com}
7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 16:12:19.0651 (UTC) FILETIME=[6DDA0930:01C73B1B]
7, 23 -- Content-Transfer-Encoding: 8bit
7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0IGCJAl020222
==============================End of - Headers==============================




From: NWWhite-at-bwxt.com
Date: Thu, 18 Jan 2007 10:24:47 -0600
Subject: [Microscopy] Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Randy,

I don't have one, but am familiar with the technique. It is not
uncommon in foundry and other metallurgical areas. The instrument
generates an electric arc to a (typically metallic) specimen and "looks"
at the resulting emission spectrum. It is not useful for microanalysis,
but the larger analysis volume works pretty well for average
compositions. I don't remember the detection limit(s), but it is pretty
low. Another advantage is that it is quick and there is little sample
preparation required.

Hope this brief response is helpful...
Woody White
BWXT Services



-----Original Message-----
X-from: TindallR-at-missouri.edu [mailto:TindallR-at-missouri.edu]
Sent: Thursday, January 18, 2007 11:13 AM
To: White, Woody N.

And along a completely different line, we've been asked if we can do a
"spark emission test". Since I don't know what that is, and Google
hasn't helped, I don't know if we can or not.

Has anybody heard of this test? Seems to be a way of checking sample
composition at concentrations below those attainable by EDS, but that's
about all I can tell.

Sorry if this isn't microscopy-related, but I don't even know that, at
this point. We'll see what the collective comes up with on this one!

Thanks to all.

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original
Headers==============================
7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 10:12:19 2007
7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu
(um-tsmtpout1.um.umsystem.edu [209.106.228.23])
7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0IGCJAl020222
7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
10:12:19 -0600
7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by
um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
7, 23 -- Thu, 18 Jan 2007 10:12:19 -0600
7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
7, 23 -- Content-class: urn:content-classes:message
7, 23 -- MIME-Version: 1.0
7, 23 -- Content-Type: text/plain;
7, 23 -- charset="us-ascii"
7, 23 -- Subject: Spark emission test?
7, 23 -- Date: Thu, 18 Jan 2007 10:12:18 -0600
7, 23 -- Message-ID:
{91108EF9255B394CBF8B7E3789814A41C68D0E-at-UM-XMAIL08.um.umsystem.edu}
7, 23 -- X-MS-Has-Attach:
7, 23 -- X-MS-TNEF-Correlator:
7, 23 -- Thread-Topic: Spark emission test?
7, 23 -- Thread-Index: Acc7G207OlT1I53OTyafRC/FuYiYDg==
7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
7, 23 -- To: {microscopy-at-microscopy.com}
7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 16:12:19.0651 (UTC)
FILETIME=[6DDA0930:01C73B1B]
7, 23 -- Content-Transfer-Encoding: 8bit
7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0IGCJAl020222
==============================End of -
Headers==============================


==============================Original Headers==============================
18, 27 -- From nwwhite-at-bwxt.com Thu Jan 18 10:24:47 2007
18, 27 -- Received: from bwxt.com (lynsmtp1.bwxt.com [131.184.146.8])
18, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IGOkGM031352
18, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 18 Jan 2007 10:24:46 -0600
18, 27 -- Received: from ([131.184.13.224])
18, 27 -- by lynsmtp1.bwxt.com with ESMTP id 5202488.3675195;
18, 27 -- Thu, 18 Jan 2007 11:24:20 -0500
18, 27 -- Received: from BWXSPO01.BWXS.BWXTECH.NET ([131.184.13.31]) by bwxslynbh01.BWXS.BWXTECH.NET with Microsoft SMTPSVC(6.0.3790.1830);
18, 27 -- Thu, 18 Jan 2007 11:24:20 -0500
18, 27 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
18, 27 -- Content-class: urn:content-classes:message
18, 27 -- MIME-Version: 1.0
18, 27 -- Content-Type: text/plain;
18, 27 -- charset="us-ascii"
18, 27 -- Subject: RE: [Microscopy] Spark emission test?
18, 27 -- Date: Thu, 18 Jan 2007 11:24:20 -0500
18, 27 -- Message-ID: {360DDAABED436B49BA397357CCEE8BB703D110-at-BWXSPO01.BWXS.BWXTECH.NET}
18, 27 -- X-MS-Has-Attach:
18, 27 -- X-MS-TNEF-Correlator:
18, 27 -- Thread-Topic: [Microscopy] Spark emission test?
18, 27 -- Thread-Index: Acc7G5uvDzuNapjlSNqW5XCH7jMkNAAACgIg
18, 27 -- From: "White, Woody N." {NWWhite-at-bwxt.com}
18, 27 -- To: {TindallR-at-missouri.edu} ,
18, 27 -- "MicroscopyListServer" {Microscopy-at-msa.microscopy.com}
18, 27 -- X-OriginalArrivalTime: 18 Jan 2007 16:24:20.0707 (UTC) FILETIME=[1BA27330:01C73B1D]
18, 27 -- Content-Transfer-Encoding: 8bit
18, 27 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0IGOkGM031352
==============================End of - Headers==============================




From: frank.karl-at-degussa.com
Date: Thu, 18 Jan 2007 10:33:13 -0600
Subject: [Microscopy] Re: Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Randy,
While this could be some type of laser emission spectroscopy test, I have
also read about spark test for metals. As I remember it different steels
and other metals will shoot out different colored, shaped, morphology ,
behavior (you catch the drift ?) sparks when touch to a grind wheel. It's
been reported that an experienced operator can tell what grade of stainless
steel based on the properties of sparks.

Never tried this myself.......... Too hard to get the grinding wheel under
my stereomicroscope......




TindallR-at-missouri
.edu To: frank.karl-at-degussa.com
cc:
01/18/2007 11:14 Subject: [Microscopy] Spark emission test?
AM
Please respond to
TindallR








----------------------------------------------------------------------------

The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


And along a completely different line, we've been asked if we can do a
"spark emission test". Since I don't know what that is, and Google
hasn't helped, I don't know if we can or not.

Has anybody heard of this test? Seems to be a way of checking sample
composition at concentrations below those attainable by EDS, but that's
about all I can tell.

Sorry if this isn't microscopy-related, but I don't even know that, at
this point. We'll see what the collective comes up with on this one!

Thanks to all.

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original
Headers==============================
7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 10:12:19 2007
7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu
(um-tsmtpout1.um.umsystem.edu [209.106.228.23])
7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0IGCJAl020222
7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
10:12:19 -0600
7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by
um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
7, 23 -- Thu, 18 Jan 2007 10:12:19 -0600
7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
7, 23 -- Content-class: urn:content-classes:message
7, 23 -- MIME-Version: 1.0
7, 23 -- Content-Type: text/plain;
7, 23 -- charset="us-ascii"
7, 23 -- Subject: Spark emission test?
7, 23 -- Date: Thu, 18 Jan 2007 10:12:18 -0600
7, 23 -- Message-ID:
{91108EF9255B394CBF8B7E3789814A41C68D0E-at-UM-XMAIL08.um.umsystem.edu}
7, 23 -- X-MS-Has-Attach:
7, 23 -- X-MS-TNEF-Correlator:
7, 23 -- Thread-Topic: Spark emission test?
7, 23 -- Thread-Index: Acc7G207OlT1I53OTyafRC/FuYiYDg==
7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
7, 23 -- To: {microscopy-at-microscopy.com}
7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 16:12:19.0651 (UTC)
FILETIME=[6DDA0930:01C73B1B]
7, 23 -- Content-Transfer-Encoding: 8bit
7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0IGCJAl020222
==============================End of -
Headers==============================




==============================Original Headers==============================
25, 18 -- From frank.karl-at-degussa.com Thu Jan 18 10:33:12 2007
25, 18 -- Received: from framailout1.rz.itson.com (mailout2.degussa.com [149.216.91.173])
25, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IGXCkx009902
25, 18 -- for {microscopy-at-msa.microscopy.com} ; Thu, 18 Jan 2007 10:33:12 -0600
25, 18 -- Received: from mobuscomm01.mail.degussa.com (uss1026.applications.degussanet.com [10.88.88.98])
25, 18 -- by framailout1.rz.itson.com (8.13.4/8.13.4/Debian-3sarge3) with SMTP id l0IGX9ae009353
25, 18 -- for {microscopy-at-msa.microscopy.com} ; Thu, 18 Jan 2007 17:33:09 +0100
25, 18 -- In-Reply-To: {200701181614.l0IGET2g022368-at-ns.microscopy.com}
25, 18 -- Subject: Re: [Microscopy] Spark emission test?
25, 18 -- To: TindallR-at-missouri.edu, microscopy-at-msa.microscopy.com
25, 18 -- X-Mailer: Lotus Notes Release 6.5.5 November 30, 2005
25, 18 -- Message-ID: {OFFEA8A4ED.CFFC3950-ON86257267.005A4DCA-85257267.005AEB26-at-degussa.com}
25, 18 -- From: frank.karl-at-degussa.com
25, 18 -- Date: Thu, 18 Jan 2007 11:33:04 -0500
25, 18 -- X-MIMETrack: Serialize by Router on MOBUSComm01/DHexternal/US(Release 6.5.1|January 21, 2004) at
25, 18 -- 01/18/2007 10:33:11 AM
25, 18 -- MIME-Version: 1.0
25, 18 -- Content-type: text/plain; charset=US-ASCII
==============================End of - Headers==============================




From: dljones-at-bestweb.net
Date: Thu, 18 Jan 2007 11:15:32 -0600
Subject: [Microscopy] Re: Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Randy,

Your question of "spark emission test" sounds familiar but I can't put my
finger on it. I can think of three techniques that may be what is being
asked: Spark Testing, Spark Source Mass Spectrometery, or Optical Emission
Spectroscopy.

Spark Testing - this is a technique that simply uses a grinder. A piece of
metal is pushed against the grinder. The sparks that fly off are
indicative of what the alloy is. I only know of it being used in metal
alloys. It would be the kind of test someone dealing with a scrap yard
would find useful. It can basically tell you what grade of metal you are
dealing with. It takes a fair amount of practice to be able to visually
know what alloy the spark pattern is indicating. People that do this test
frequently can do quite well with it as far as getting a good handle on
what alloy they have in their hand.

Spark Source Mass Spectrometery - This is used for a wide range of
samples, but tends to do better with samples that are conductive, although
that is not necessarily a limitation. There are two other techniques that
are related to this: Laser Ionization Mass Spectrometery and Inductively
Coupled Plasma Atomic Emission.

Optical Emission Spectroscopy - I'm not too familiar with this technique,
but I seem to recall they will often use a spark source in the process.
When they do that, I think this is then called Spark Emission
Spectroscopy. You might want to check out Bruker AXS Microanalysis, I
believe they make some of these instruments.

Perhaps if you post what they are testing and what they want from the
analysis, that may help. I hope I've given you at least some food for
thought.

Good luck,

dj



On Thu, 18 Jan 2007, TindallR-at-missouri.edu wrote:

}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} And along a completely different line, we've been asked if we can do a
} "spark emission test". Since I don't know what that is, and Google
} hasn't helped, I don't know if we can or not.
}
} Has anybody heard of this test? Seems to be a way of checking sample
} composition at concentrations below those attainable by EDS, but that's
} about all I can tell.
}
} Sorry if this isn't microscopy-related, but I don't even know that, at
} this point. We'll see what the collective comes up with on this one!
}
} Thanks to all.
}
} Randy
}
} Randy Tindall
} Senior EM Specialist
} Electron Microscopy Core Facility---We Do Small Well!
} W125 Veterinary Medicine
} University of Missouri
} Columbia, MO 65211
} Tel: (573) 882-8304
} Fax: (573) 884-2227
} Email: tindallr-at-missouri.edu
} Web: http://www.emc.missouri.edu
} On-line calendar:
} http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
} Week&NavType=Both&Type=TimePlan
}
}
} ==============================Original Headers==============================
} 7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 10:12:19 2007
} 7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
} 7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IGCJAl020222
} 7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 10:12:19 -0600
} 7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
} 7, 23 -- Thu, 18 Jan 2007 10:12:19 -0600
} 7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
} 7, 23 -- Content-class: urn:content-classes:message
} 7, 23 -- MIME-Version: 1.0
} 7, 23 -- Content-Type: text/plain;
} 7, 23 -- charset="us-ascii"
} 7, 23 -- Subject: Spark emission test?
} 7, 23 -- Date: Thu, 18 Jan 2007 10:12:18 -0600
} 7, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D0E-at-UM-XMAIL08.um.umsystem.edu}
} 7, 23 -- X-MS-Has-Attach:
} 7, 23 -- X-MS-TNEF-Correlator:
} 7, 23 -- Thread-Topic: Spark emission test?
} 7, 23 -- Thread-Index: Acc7G207OlT1I53OTyafRC/FuYiYDg==
} 7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
} 7, 23 -- To: {microscopy-at-microscopy.com}
} 7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 16:12:19.0651 (UTC) FILETIME=[6DDA0930:01C73B1B]
} 7, 23 -- Content-Transfer-Encoding: 8bit
} 7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0IGCJAl020222
} ==============================End of - Headers==============================
}


==============================Original Headers==============================
13, 19 -- From dljones-at-bestweb.net Thu Jan 18 11:15:31 2007
13, 19 -- Received: from vms044pub.verizon.net (vms044pub.verizon.net [206.46.252.44])
13, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IHFULY021977
13, 19 -- for {Microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 11:15:31 -0600
13, 19 -- Received: from localhost ([71.247.249.221])
13, 19 -- by vms044.mailsrvcs.net (Sun Java System Messaging Server 6.2-6.01 (built Apr
13, 19 -- 3 2006)) with ESMTPA id {0JC200NM7QL94A26-at-vms044.mailsrvcs.net} for
13, 19 -- Microscopy-at-microscopy.com; Thu, 18 Jan 2007 11:15:14 -0600 (CST)
13, 19 -- Date: Thu, 18 Jan 2007 12:14:16 -0500 (Eastern Standard Time)
13, 19 -- From: "David L. Jones" {dljones-at-bestweb.net}
13, 19 -- Subject: Re: [Microscopy] Spark emission test?
13, 19 -- In-reply-to: {200701181618.l0IGIW0e026577-at-ns.microscopy.com}
13, 19 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
13, 19 -- To: TindallR-at-missouri.edu
13, 19 -- Cc: Microscopy-at-microscopy.com
13, 19 -- Message-id: {Pine.WNT.4.64.0701181137340.3320-at-H-F1}
13, 19 -- MIME-version: 1.0
13, 19 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
13, 19 -- References: {200701181618.l0IGIW0e026577-at-ns.microscopy.com}
==============================End of - Headers==============================




From: werner-at-rosharon.oilfield.slb.com
Date: Thu, 18 Jan 2007 13:55:03 -0600
Subject: [Microscopy] Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

dj is right, when metallurgists say "spark test" they generally mean
optical emission spectroscopy (OES). OES can give reliable sulfur and
phosphorous (for instance) readings at low concentrations I would never
have the chutzpa to report from EDS.

The grinding wheel test can distinguish high-carbon (short sparks, in dense
clusters) from low-carbon steels (few sparks, long, and not in
clusters). I do not think it can distinguish whether any other alloying
elements are present or not. There is a great ASTM document - STP 550
"Nondestructive Rapid Identification of Metals and Alloys by Spot Test"
from 1974 that covers chemical spot tests for qualitative alloy i.d. but
that is 'way off-topic.

Regards,
Andrew

At 11:16 AM 1/18/2007, you wrote:



} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
8, 36 -- From werner-at-rosharon.oilfield.slb.com Thu Jan 18 13:55:02 2007
8, 36 -- Received: from mail.slb.com (nammta01.sugar-land.nam.slb.com [163.188.150.130])
8, 36 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IJt2Wl003343
8, 36 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 13:55:02 -0600
8, 36 -- Received: from pmxchannel_int-daemon.nammta01.sugar-land.nam.slb.com by
8, 36 -- nammta01.sugar-land.nam.slb.com
8, 36 -- (iPlanet Messaging Server 5.2 Patch 2 (built Jul 14 2004))
8, 36 -- id {0JC200301XENHX-at-nammta01.sugar-land.nam.slb.com} for
8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:23 +0000 (GMT)
8, 36 -- Received: from usxsl052.slb.atosorigin-asp.com
8, 36 -- (usxsl052.slb.atosorigin-asp.com [199.6.136.167])
8, 36 -- by nammta01.sugar-land.nam.slb.com
8, 36 -- (iPlanet Messaging Server 5.2 Patch 2 (built Jul 14 2004))
8, 36 -- with ESMTP id {0JC200G99XEN8P-at-nammta01.sugar-land.nam.slb.com} for
8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:23 +0000 (GMT)
8, 36 -- Received: from pmxchannel-daemon.us085mbx01.slb.atosorigin-asp.com by
8, 36 -- us085mbx01.slb.atosorigin-asp.com
8, 36 -- (Sun Java System Messaging Server 6.2-1 (built Feb 24 2005))
8, 36 -- id {0JC20092VXEJW500-at-us085mbx01.slb.atosorigin-asp.com} for
8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:19 +0000 (GMT)
8, 36 -- Received: from WERNER1-OFS.rosharon.oilfield.slb.com ([163.188.46.5])
8, 36 -- by us085mbx01.slb.atosorigin-asp.com
8, 36 -- (Sun Java System Messaging Server 6.2-1 (built Feb 24 2005))
8, 36 -- with ESMTPSA id {0JC2003Z9XEJQX50-at-us085mbx01.slb.atosorigin-asp.com} for
8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:19 +0000 (GMT)
8, 36 -- Date: Thu, 18 Jan 2007 13:42:18 -0600
8, 36 -- From: Andrew Werner {werner-at-rosharon.oilfield.slb.com}
8, 36 -- Subject: Re: [Microscopy] Re: Spark emission test?
8, 36 -- In-reply-to: {200701181716.l0IHGH9F023234-at-ns.microscopy.com}
8, 36 -- To: microscopy-at-microscopy.com
8, 36 -- Message-id: {6.2.1.2.2.20070118133232.03439d28-at-us1061-pop3.mail.slb.com}
8, 36 -- MIME-version: 1.0
8, 36 -- X-Mailer: QUALCOMM Windows Eudora Version 6.2.1.2
8, 36 -- Content-type: text/plain; charset=us-ascii; format=flowed
8, 36 -- Content-transfer-encoding: 7BIT
8, 36 -- References: {200701181716.l0IHGH9F023234-at-ns.microscopy.com}
==============================End of - Headers==============================




From: bigelow-at-engin.umich.edu
Date: Thu, 18 Jan 2007 13:59:24 -0600
Subject: [Microscopy] RE: Etching gold

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

In an ancient edition of The Metals Handbook (Amer Soc for Materials,
ASM) I found the following :


"The most widely used nonelectrolytic etch is a mixture of equal
volumes of 10% ammonium persulfate in water with 10% KCN in water.
Although the separate solutions are stable in air, the mixture must
be used within a few minutes of mixing. Since this solution and the
fumes from it are poisonous, it should be used under a hood (and
with other relevant precautions) . . . . . . When swabbed on
specimens of the usual gold alloys and palladium alloys the mixture
acts rapidly and smoothly. It may also be used on silver and certain
nickle alloys. For more resistant alloys, 20% solutions may be
employed"

"The same metals can be etched when made the anode in a 5% KCN
solution" (I'd suggest a voltage of about 5 V). Also mentioned is
electrolytic etching using an a.c. voltage (you can probably use a
variac as a source for this, again about 5 volts) with a 20% solution
of hydrochloric acid saturated with sodium chloride.

The great advantage of Gold and the other noble metals is their
chemical stabilite, and so your task is a non-trivial one.

Good luck,
--
Wilbur C. Bigelow, Professor Emeritus
Materials Sci. & Engr., Univ. of Michigan
Ann Arbor, Michigan 48109-2136
e-mail: bigelow-at-engin.umich.edu;
Fx:734-763-4788; Ph:734-975-0858
Address mail to: 2911 Whittier Court
Ann Arbor, MI 48104-6731

==============================Original Headers==============================
6, 14 -- From bigelow-at-engin.umich.edu Thu Jan 18 13:59:24 2007
6, 14 -- Received: from srvr20.engin.umich.edu (srvr20.engin.umich.edu [141.213.75.22])
6, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IJxNTZ009333
6, 14 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 13:59:23 -0600
6, 14 -- Received: from [141.212.131.221] (bigelow-g4.engin.umich.edu [141.212.131.221])
6, 14 -- by srvr20.engin.umich.edu (8.13.6/8.13.6) with ESMTP id l0IJxNcC020136
6, 14 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 14:59:23 -0500 (EST)
6, 14 -- Mime-Version: 1.0
6, 14 -- Message-Id: {p06210200c1d5793b7bb7-at-[141.212.131.221]}
6, 14 -- Date: Thu, 18 Jan 2007 14:59:22 -0500
6, 14 -- To: Microscopy Listserver {microscopy-at-microscopy.com}
6, 14 -- From: Wil Bigelow {bigelow-at-engin.umich.edu}
6, 14 -- Subject: [Microscopy] RE: Etching gold
6, 14 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
==============================End of - Headers==============================




From: smalinskas-at-yahoo.com
Date: Thu, 18 Jan 2007 14:22:27 -0600
Subject: [Microscopy] Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Randy,

This sounds like optical emission arc-spectrometry.
It sparks a metal sample and the different light
wavelengths are channeled and read to quantify the
elements. We use it in our lab for accurate
measurements of elemental content in different metal
alloys. This technique is considered very accurate,
considerably better and more reliable than EDS
techniques. Some elements can be detected in the ppm
range.

Stu Smalinskas, P.E.
Metallurgist
SKF
Plymouth, Michigan
(734) 414-6862

--- frank.karl-at-degussa.com wrote:

} Randy,
} While this could be some type of laser emission
} spectroscopy test, I have
} also read about spark test for metals. As I
} remember it different steels
} and other metals will shoot out different colored,
} shaped, morphology ,
} behavior (you catch the drift ?) sparks when touch
} to a grind wheel. It's
} been reported that an experienced operator can tell
} what grade of stainless
} steel based on the properties of sparks.
}
} Never tried this myself.......... Too hard to get
} the grinding wheel under
} my stereomicroscope......
}
}
}
}
}
--- Randy wrote:
}
} And along a completely different line, we've been
} asked if we can do a
} "spark emission test". Since I don't know what that
} is, and Google
} hasn't helped, I don't know if we can or not.
}
} Has anybody heard of this test? Seems to be a way
} of checking sample
} composition at concentrations below those attainable
} by EDS, but that's
} about all I can tell.
}
} Sorry if this isn't microscopy-related, but I don't
} even know that, at
} this point. We'll see what the collective comes up
} with on this one!
}
} Thanks to all.
}
} Randy
}
} Randy Tindall
} Senior EM Specialist
} Electron Microscopy Core Facility---We Do Small
} Well!
} W125 Veterinary Medicine
} University of Missouri
} Columbia, MO 65211
} Tel: (573) 882-8304
} Fax: (573) 884-2227
} Email: tindallr-at-missouri.edu
} Web: http://www.emc.missouri.edu
} On-line calendar:
}
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
} Week&NavType=Both&Type=TimePlan
}



____________________________________________________________________________________
Do you Yahoo!?
Everyone is raving about the all-new Yahoo! Mail beta.
http://new.mail.yahoo.com

==============================Original Headers==============================
7, 20 -- From smalinskas-at-yahoo.com Thu Jan 18 14:22:27 2007
7, 20 -- Received: from web34413.mail.mud.yahoo.com (web34413.mail.mud.yahoo.com [66.163.178.162])
7, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0IKMR5h025518
7, 20 -- for {microscopy-at-sparc5.microscopy.com} ; Thu, 18 Jan 2007 14:22:27 -0600
7, 20 -- Received: (qmail 39218 invoked by uid 60001); 18 Jan 2007 20:22:26 -0000
7, 20 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
7, 20 -- s=s1024; d=yahoo.com;
7, 20 -- h=X-YMail-OSG:Received:Date:From:Subject:To:In-Reply-To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
7, 20 -- b=JZCRqLAkID61w/jC3In0bdqAW1jS4FD/LY3jw0bryfCxQF+q2vP9FngOg80uM5vKgEnp8Q9++rmbWpADMESwX3ks63T3RBAwdQehrREbEgjl+6MCtcyl0vSQg7XQnVAOAZCkdoks5Aybt+5jFGcuiUGjOogDL50jj/kdRlmyeJo=;
7, 20 -- X-YMail-OSG: VRGraHoVM1mlDkQhOovEDc2WkHKrPO363WKRTxaY
7, 20 -- Received: from [141.151.33.213] by web34413.mail.mud.yahoo.com via HTTP; Thu, 18 Jan 2007 12:22:26 PST
7, 20 -- Date: Thu, 18 Jan 2007 12:22:26 -0800 (PST)
7, 20 -- From: Kestutis Smalinskas {smalinskas-at-yahoo.com}
7, 20 -- Subject: Re: [Microscopy] Re: Spark emission test?
7, 20 -- To: microscopy-at-ns.microscopy.com, tindallr-at-missouri.edu
7, 20 -- In-Reply-To: {200701181634.l0IGYKL8012393-at-ns.microscopy.com}
7, 20 -- MIME-Version: 1.0
7, 20 -- Content-Type: text/plain; charset=iso-8859-1
7, 20 -- Content-Transfer-Encoding: 8bit
7, 20 -- Message-ID: {748483.38927.qm-at-web34413.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Thu, 18 Jan 2007 14:34:50 -0600
Subject: [Microscopy] Spark emission test

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Many thanks to all who responded to my query about spark emission
testing. I have passed on the information to the client who requested
it, and even found a local source to do the test----right here in the UM
system.

Sorry for cluttering up the list with what turned out to be a
non-microscopy issue, but I learned something new.

Next question: is there anything the listserve denizens can't answer?

Thanks again!

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original Headers==============================
7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 14:34:50 2007
7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IKYoNV004299
7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 14:34:50 -0600
7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
7, 23 -- Thu, 18 Jan 2007 14:34:49 -0600
7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
7, 23 -- Content-class: urn:content-classes:message
7, 23 -- MIME-Version: 1.0
7, 23 -- Content-Type: text/plain;
7, 23 -- charset="us-ascii"
7, 23 -- Subject: Spark emission test
7, 23 -- Date: Thu, 18 Jan 2007 14:34:49 -0600
7, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D1F-at-UM-XMAIL08.um.umsystem.edu}
7, 23 -- X-MS-Has-Attach:
7, 23 -- X-MS-TNEF-Correlator:
7, 23 -- Thread-Topic: Spark emission test
7, 23 -- Thread-Index: Acc7QBmlkLDae5eJSMug3056bJpIBw==
7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
7, 23 -- To: {microscopy-at-microscopy.com}
7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 20:34:50.0159 (UTC) FILETIME=[19E2D3F0:01C73B40]
7, 23 -- Content-Transfer-Encoding: 8bit
7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0IKYoNV004299
==============================End of - Headers==============================




From: dljones-at-bestweb.net
Date: Thu, 18 Jan 2007 15:57:08 -0600
Subject: [Microscopy] Re: Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Andrew,

Actually, the grinding wheel test can tell a lot more than one would
think. You can tell stainless steels, high speed steels, if the alloy
contains cobalt, and more. I used to have an old text that listed the
spark patterns for a wide range of alloys. But I haven't been able to find
it in years. It was handed out in a technical course on welding I took
many years ago at a community college. This was an excellent reference
that had several pages of spark patterns in color identifying the various
alloys that each spark test indicated. It has to be in color as color is
one of the identifiers. If anyone knows of a reference that has an
extensive color chart like this, can you please send me the reference?
I've been looking for a long time to find this again.

Randy,

You didn't say what the specific test was you have found to be it. Is it
indeed OES?

dj


On Thu, 18 Jan 2007, werner-at-rosharon.oilfield.slb.com wrote:

} Date: Thu, 18 Jan 2007 13:59:42 -0600
} From: werner-at-rosharon.oilfield.slb.com
} To: dljones-at-bestweb.net
} Subject: [Microscopy] Spark emission test?
}
}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} dj is right, when metallurgists say "spark test" they generally mean
} optical emission spectroscopy (OES). OES can give reliable sulfur and
} phosphorous (for instance) readings at low concentrations I would never
} have the chutzpa to report from EDS.
}
} The grinding wheel test can distinguish high-carbon (short sparks, in dense
} clusters) from low-carbon steels (few sparks, long, and not in
} clusters). I do not think it can distinguish whether any other alloying
} elements are present or not. There is a great ASTM document - STP 550
} "Nondestructive Rapid Identification of Metals and Alloys by Spot Test"
} from 1974 that covers chemical spot tests for qualitative alloy i.d. but
} that is 'way off-topic.
}
} Regards,
} Andrew
}
} At 11:16 AM 1/18/2007, you wrote:
}
}
}
} } ----------------------------------------------------------------------------
} } The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} } To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} } On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} } ----------------------------------------------------------------------------
} }
} } Randy,
} }
} } Your question of "spark emission test" sounds familiar but I can't put my
} } finger on it. I can think of three techniques that may be what is being
} } asked: Spark Testing, Spark Source Mass Spectrometery, or Optical Emission
} } Spectroscopy.
} }
} } Spark Testing - this is a technique that simply uses a grinder. A piece of
} } metal is pushed against the grinder. The sparks that fly off are
} } indicative of what the alloy is. I only know of it being used in metal
} } alloys. It would be the kind of test someone dealing with a scrap yard
} } would find useful. It can basically tell you what grade of metal you are
} } dealing with. It takes a fair amount of practice to be able to visually
} } know what alloy the spark pattern is indicating. People that do this test
} } frequently can do quite well with it as far as getting a good handle on
} } what alloy they have in their hand.
} }
} } Spark Source Mass Spectrometery - This is used for a wide range of
} } samples, but tends to do better with samples that are conductive, although
} } that is not necessarily a limitation. There are two other techniques that
} } are related to this: Laser Ionization Mass Spectrometery and Inductively
} } Coupled Plasma Atomic Emission.
} }
} } Optical Emission Spectroscopy - I'm not too familiar with this technique,
} } but I seem to recall they will often use a spark source in the process.
} } When they do that, I think this is then called Spark Emission
} } Spectroscopy. You might want to check out Bruker AXS Microanalysis, I
} } believe they make some of these instruments.
} }
} } Perhaps if you post what they are testing and what they want from the
} } analysis, that may help. I hope I've given you at least some food for
} } thought.
} }
} } Good luck,
} }
} } dj
} }
} }
} }
} } On Thu, 18 Jan 2007, TindallR-at-missouri.edu wrote:
} }
} } }
} } }
} } }
} } }
} } ----------------------------------------------------------------------------
} } } The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} } } To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} } } On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} } }
} } ----------------------------------------------------------------------------
} } }
} } } And along a completely different line, we've been asked if we can do a
} } } "spark emission test". Since I don't know what that is, and Google
} } } hasn't helped, I don't know if we can or not.
} } }
} } } Has anybody heard of this test? Seems to be a way of checking sample
} } } composition at concentrations below those attainable by EDS, but that's
} } } about all I can tell.
} } }
} } } Sorry if this isn't microscopy-related, but I don't even know that, at
} } } this point. We'll see what the collective comes up with on this one!
} } }
} } } Thanks to all.
} } }
} } } Randy
} } }
} } } Randy Tindall
} } } Senior EM Specialist
} } } Electron Microscopy Core Facility---We Do Small Well!
} } } W125 Veterinary Medicine
} } } University of Missouri
} } } Columbia, MO 65211
} } } Tel: (573) 882-8304
} } } Fax: (573) 884-2227
} } } Email: tindallr-at-missouri.edu
} } } Web: http://www.emc.missouri.edu
} } } On-line calendar:
} } } http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
} } } Week&NavType=Both&Type=TimePlan
} } }
} } }
} } } ==============================Original
} } Headers==============================
} } } 7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 10:12:19 2007
} } } 7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu
} } (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
} } } 7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP
} } id l0IGCJAl020222
} } } 7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
} } 10:12:19 -0600
} } } 7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34])
} } by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
} } } 7, 23 -- Thu, 18 Jan 2007 10:12:19 -0600
} } } 7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
} } } 7, 23 -- Content-class: urn:content-classes:message
} } } 7, 23 -- MIME-Version: 1.0
} } } 7, 23 -- Content-Type: text/plain;
} } } 7, 23 -- charset="us-ascii"
} } } 7, 23 -- Subject: Spark emission test?
} } } 7, 23 -- Date: Thu, 18 Jan 2007 10:12:18 -0600
} } } 7, 23 -- Message-ID:
} } {91108EF9255B394CBF8B7E3789814A41C68D0E-at-UM-XMAIL08.um.umsystem.edu}
} } } 7, 23 -- X-MS-Has-Attach:
} } } 7, 23 -- X-MS-TNEF-Correlator:
} } } 7, 23 -- Thread-Topic: Spark emission test?
} } } 7, 23 -- Thread-Index: Acc7G207OlT1I53OTyafRC/FuYiYDg==
} } } 7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
} } } 7, 23 -- To: {microscopy-at-microscopy.com}
} } } 7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 16:12:19.0651 (UTC)
} } FILETIME=[6DDA0930:01C73B1B]
} } } 7, 23 -- Content-Transfer-Encoding: 8bit
} } } 7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
} } ns.microscopy.com id l0IGCJAl020222
} } } ==============================End of -
} } Headers==============================
} } }
} }
} }
} } ==============================Original Headers==============================
} } 13, 19 -- From dljones-at-bestweb.net Thu Jan 18 11:15:31 2007
} } 13, 19 -- Received: from vms044pub.verizon.net (vms044pub.verizon.net
} } [206.46.252.44])
} } 13, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP
} } id l0IHFULY021977
} } 13, 19 -- for {Microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 11:15:31
} } -0600
} } 13, 19 -- Received: from localhost ([71.247.249.221])
} } 13, 19 -- by vms044.mailsrvcs.net (Sun Java System Messaging Server
} } 6.2-6.01 (built Apr
} } 13, 19 -- 3 2006)) with ESMTPA id {0JC200NM7QL94A26-at-vms044.mailsrvcs.net} for
} } 13, 19 -- Microscopy-at-microscopy.com; Thu, 18 Jan 2007 11:15:14 -0600 (CST)
} } 13, 19 -- Date: Thu, 18 Jan 2007 12:14:16 -0500 (Eastern Standard Time)
} } 13, 19 -- From: "David L. Jones" {dljones-at-bestweb.net}
} } 13, 19 -- Subject: Re: [Microscopy] Spark emission test?
} } 13, 19 -- In-reply-to: {200701181618.l0IGIW0e026577-at-ns.microscopy.com}
} } 13, 19 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
} } 13, 19 -- To: TindallR-at-missouri.edu
} } 13, 19 -- Cc: Microscopy-at-microscopy.com
} } 13, 19 -- Message-id: {Pine.WNT.4.64.0701181137340.3320-at-H-F1}
} } 13, 19 -- MIME-version: 1.0
} } 13, 19 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
} } 13, 19 -- References: {200701181618.l0IGIW0e026577-at-ns.microscopy.com}
} } ==============================End of - Headers==============================
}
}
} ==============================Original Headers==============================
} 8, 36 -- From werner-at-rosharon.oilfield.slb.com Thu Jan 18 13:55:02 2007
} 8, 36 -- Received: from mail.slb.com (nammta01.sugar-land.nam.slb.com [163.188.150.130])
} 8, 36 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0IJt2Wl003343
} 8, 36 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 13:55:02 -0600
} 8, 36 -- Received: from pmxchannel_int-daemon.nammta01.sugar-land.nam.slb.com by
} 8, 36 -- nammta01.sugar-land.nam.slb.com
} 8, 36 -- (iPlanet Messaging Server 5.2 Patch 2 (built Jul 14 2004))
} 8, 36 -- id {0JC200301XENHX-at-nammta01.sugar-land.nam.slb.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:23 +0000 (GMT)
} 8, 36 -- Received: from usxsl052.slb.atosorigin-asp.com
} 8, 36 -- (usxsl052.slb.atosorigin-asp.com [199.6.136.167])
} 8, 36 -- by nammta01.sugar-land.nam.slb.com
} 8, 36 -- (iPlanet Messaging Server 5.2 Patch 2 (built Jul 14 2004))
} 8, 36 -- with ESMTP id {0JC200G99XEN8P-at-nammta01.sugar-land.nam.slb.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:23 +0000 (GMT)
} 8, 36 -- Received: from pmxchannel-daemon.us085mbx01.slb.atosorigin-asp.com by
} 8, 36 -- us085mbx01.slb.atosorigin-asp.com
} 8, 36 -- (Sun Java System Messaging Server 6.2-1 (built Feb 24 2005))
} 8, 36 -- id {0JC20092VXEJW500-at-us085mbx01.slb.atosorigin-asp.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:19 +0000 (GMT)
} 8, 36 -- Received: from WERNER1-OFS.rosharon.oilfield.slb.com ([163.188.46.5])
} 8, 36 -- by us085mbx01.slb.atosorigin-asp.com
} 8, 36 -- (Sun Java System Messaging Server 6.2-1 (built Feb 24 2005))
} 8, 36 -- with ESMTPSA id {0JC2003Z9XEJQX50-at-us085mbx01.slb.atosorigin-asp.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:19 +0000 (GMT)
} 8, 36 -- Date: Thu, 18 Jan 2007 13:42:18 -0600
} 8, 36 -- From: Andrew Werner {werner-at-rosharon.oilfield.slb.com}
} 8, 36 -- Subject: Re: [Microscopy] Re: Spark emission test?
} 8, 36 -- In-reply-to: {200701181716.l0IHGH9F023234-at-ns.microscopy.com}
} 8, 36 -- To: microscopy-at-microscopy.com
} 8, 36 -- Message-id: {6.2.1.2.2.20070118133232.03439d28-at-us1061-pop3.mail.slb.com}
} 8, 36 -- MIME-version: 1.0
} 8, 36 -- X-Mailer: QUALCOMM Windows Eudora Version 6.2.1.2
} 8, 36 -- Content-type: text/plain; charset=us-ascii; format=flowed
} 8, 36 -- Content-transfer-encoding: 7BIT
} 8, 36 -- References: {200701181716.l0IHGH9F023234-at-ns.microscopy.com}
} ==============================End of - Headers==============================
}


==============================Original Headers==============================
9, 20 -- From "dljones-at-bestweb.net"-at-bestweb.net Thu Jan 18 15:57:08 2007
9, 20 -- Received: from mta5.srv.hcvlny.cv.net (mta5.srv.hcvlny.cv.net [167.206.4.200])
9, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0ILv7Cf016738
9, 20 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 15:57:08 -0600
9, 20 -- Received: from localhost (ool-44c62883.dyn.optonline.net [68.198.40.131])
9, 20 -- by mta5.srv.hcvlny.cv.net
9, 20 -- (Sun Java System Messaging Server 6.2-6.01 (built Apr 3 2006))
9, 20 -- with ESMTP id {0JC300E9X3MIVOF0-at-mta5.srv.hcvlny.cv.net} for
9, 20 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 16:56:59 -0500 (EST)
9, 20 -- Date: Thu, 18 Jan 2007 16:48:42 -0500 (Eastern Standard Time)
9, 20 -- From: "David L. Jones" {"dljones-at-bestweb.net"-at-bestweb.net}
9, 20 -- Subject: Re: [Microscopy] Spark emission test?
9, 20 -- In-reply-to: {200701181959.l0IJxg9t010372-at-ns.microscopy.com}
9, 20 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
9, 20 -- Cc: microscopy-at-microscopy.com
9, 20 -- Message-id: {Pine.WNT.4.64.0701181516340.176-at-dljtoshiba}
9, 20 -- MIME-version: 1.0
9, 20 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
9, 20 -- Content-transfer-encoding: 7BIT
9, 20 -- References: {200701181959.l0IJxg9t010372-at-ns.microscopy.com}
==============================End of - Headers==============================




From: gregm-at-andersonlabs.com
Date: Thu, 18 Jan 2007 17:54:18 -0600
Subject: [Microscopy] viaWWW: Spark Spectroscopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both gregm-at-andersonlabs.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: gregm-at-andersonlabs.com
Name: Greg Mann

Title-Subject: [Filtered] Spark Spectroscopy

Question: Optical Emission Spectroscopy (OES) commonly uses an AC spark source to excite a small volume of metallic sample in a coronal discharge. In one application a high frequency AC source strikes an arc with a tungsten electrode. Basically the excited electrons have jumped an electron shell, and on return to their relaxed state, they emit a photon of light characteristic of their source. The light generated is directed into an evacuated spectral box, passed through a grating and separated into its spectral components. Each wavelength of interest (i.e. element of interest) is detected. Solid state detectors have can be nearly continuous in wavelength detection, while traditional photo multipliers are discreet in location, often with greater sensitivity. Other OES techniques may employ a DC arc, glow discharge source, or Inductively coupled plasma (ICP). Not to be confused the older generation spark station software (I believe it was Leica Cambridge). Or the grinding wheel test that looks at the color of the spark, and the lengths of the fingers on the generated spark to estimate manganese and carbon contents.

For more information, contact me offline, as we do quite a bit of these bulk analysis metals.

Greg M



---------------------------------------------------------------------------


==============================Original Headers==============================
10, 14 -- From zaluzec-at-microscopy.com Thu Jan 18 17:54:18 2007
10, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
10, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0INsHGm029719
10, 14 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 17:54:17 -0600
10, 14 -- Mime-Version: 1.0
10, 14 -- X-Sender: (Unverified)
10, 14 -- Message-Id: {p06110402c1d5b918fea0-at-[206.69.208.22]}
10, 14 -- Date: Thu, 18 Jan 2007 17:54:15 -0600
10, 14 -- To: microscopy-at-microscopy.com
10, 14 -- From: gregm-at-andersonlabs.com (by way of MicroscopyListserver)
10, 14 -- Subject: viaWWW: Spark Spectroscopy
10, 14 -- Content-Type: text/plain; charset="us-ascii"
10, 14 -- Content-Transfer-Encoding: 8bit
10, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0INsHGm029719
==============================End of - Headers==============================




From: dainis-at-red5wood.com
Date: Thu, 18 Jan 2007 17:54:46 -0600
Subject: [Microscopy] viaWWW: Leica 1400 microtome manual etc

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both dainis-at-red5wood.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: dainis-at-red5wood.com
Name: Dainis

Title-Subject: [Filtered] Leica 1400 microtome manual etc

Question: Dear Microscopists
Does anyone have a spare (or is prepared to copy) manual for the Leica 1400 microtome please?
Also has anyone used ImageJ with the CRI RGB Micro Color filter?
Thanks
Dainis

---------------------------------------------------------------------------

==============================Original Headers==============================
5, 12 -- From zaluzec-at-microscopy.com Thu Jan 18 17:54:46 2007
5, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
5, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0INsgU5030282
5, 12 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007 17:54:44 -0600
5, 12 -- Mime-Version: 1.0
5, 12 -- X-Sender: (Unverified)
5, 12 -- Message-Id: {p06110403c1d5b9310469-at-[206.69.208.22]}
5, 12 -- Date: Thu, 18 Jan 2007 17:54:41 -0600
5, 12 -- To: microscopy-at-microscopy.com
5, 12 -- From: dainis-at-red5wood.com (by way of MicroscopyListserver)
5, 12 -- Subject: viaWWW: Leica 1400 microtome manual etc
5, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: mjaffer-at-science.uct.ac.za
Date: Fri, 19 Jan 2007 01:21:29 -0600
Subject: [Microscopy] Disposal of JEOL 200CX

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear All

We need to dispose our working 1979 vintage JEOL 200CX to make way for
a new instrument. The microscope will be taken out of service by of
February 2007. The complete instrument or parts will be available to
anyone who is interested. We have the manuals, service diagrams, a
variety of holders as well as a few spare parts for this instrument.
Please contact me offline if you are interested.

Thanks in advance.

Mohamed

************************************
Mohamed Jaffer
Electron Microscope Unit
University of Cape Town
Private Bag
Rondebosch, 7701
South Africa

Tel: +27 21 6503354
Fax: + 27 21 6891528

Email: mjaffer-at-science.uct.ac.za

************************************






==============================Original Headers==============================
13, 20 -- From mjaffer-at-science.uct.ac.za Fri Jan 19 01:21:29 2007
13, 20 -- Received: from mail.uct.ac.za (mail.uct.ac.za [137.158.128.3])
13, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0J7LSSJ025726
13, 20 -- for {Microscopy-at-microscopy.com} ; Fri, 19 Jan 2007 01:21:29 -0600
13, 20 -- Received: from [137.158.37.13] (helo=science.uct.ac.za)
13, 20 -- by mail.uct.ac.za with esmtp (Exim 4.44 (FreeBSD))
13, 20 -- id 1H7o3z-000ANv-VP
13, 20 -- for Microscopy-at-microscopy.com; Fri, 19 Jan 2007 09:21:28 +0200
13, 20 -- Message-ID: {45B07177.A37A6686-at-science.uct.ac.za}
13, 20 -- Date: Fri, 19 Jan 2007 09:21:27 +0200
13, 20 -- From: Mohamed A Jaffer {mjaffer-at-science.uct.ac.za}
13, 20 -- Reply-To: mjaffer-at-science.uct.ac.za
13, 20 -- Organization: University of Cape Town
13, 20 -- X-Mailer: Mozilla 4.75 [en] (Windows NT 5.0; U)
13, 20 -- X-Accept-Language: en
13, 20 -- MIME-Version: 1.0
13, 20 -- To: "Microscopy-at-microscopy.com" {Microscopy-at-microscopy.com}
13, 20 -- Subject: Disposal of JEOL 200CX
13, 20 -- Content-Type: text/plain; charset=us-ascii
13, 20 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: nmedvitz-at-bostwicklaboratories.com
Date: Fri, 19 Jan 2007 08:47:04 -0600
Subject: [Microscopy] viaWWW: TEM job opening

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both nmedvitz-at-bostwicklaboratories.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: nmedvitz-at-bostwicklaboratories.com
Name: Neil Medvitz

Organization: Bostwick Laboratories

Title-Subject: [Filtered] TEM job opening

Question: I am looking to hire a couple people to do EM in a renal pathology lab in the Richmond Virginia area. Please contact mr via e-mail if interested.

Neil Medvitz
Electron Microscopist
Bostwick Laboratoriesô
For Absolute ConfidenceÆ

4355 Innslake Drive
Glen Allen, Virginia 23060
Phone: (804) 967-9225 x 1488
Toll Free: (800) 214-6628
Email: nmedvitz-at-bostwicklaboratories.com




---------------------------------------------------------------------------


==============================Original Headers==============================
12, 14 -- From zaluzec-at-microscopy.com Fri Jan 19 08:47:04 2007
12, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
12, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0JEl2na012200
12, 14 -- for {microscopy-at-microscopy.com} ; Fri, 19 Jan 2007 08:47:04 -0600
12, 14 -- Mime-Version: 1.0
12, 14 -- X-Sender: (Unverified)
12, 14 -- Message-Id: {p06110401c1d68a4e0854-at-[206.69.208.22]}
12, 14 -- Date: Fri, 19 Jan 2007 08:47:00 -0600
12, 14 -- To: microscopy-at-microscopy.com
12, 14 -- From: nmedvitz-at-bostwicklaboratories.com (by way of MicroscopyListserver)
12, 14 -- Subject: viaWWW: TEM job opening
12, 14 -- Content-Type: text/plain; charset="iso-8859-1"
12, 14 -- Content-Transfer-Encoding: 8bit
12, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0JEl2na012200
==============================End of - Headers==============================




From: fschamber-at-aspexcorp.com
Date: Fri, 19 Jan 2007 10:36:01 -0600
Subject: [Microscopy] Re: Spark emission test?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html



-----Original Message-----
X-from: dljones-at-bestweb.net [mailto:dljones-at-bestweb.net]
Sent: Thursday, January 18, 2007 5:05 PM
To: Fred Schamber

Andrew,

Actually, the grinding wheel test can tell a lot more than one would
think. You can tell stainless steels, high speed steels, if the alloy
contains cobalt, and more. I used to have an old text that listed the
spark patterns for a wide range of alloys. But I haven't been able to
find
it in years. It was handed out in a technical course on welding I took
many years ago at a community college. This was an excellent reference
that had several pages of spark patterns in color identifying the
various
alloys that each spark test indicated. It has to be in color as color is

one of the identifiers. If anyone knows of a reference that has an
extensive color chart like this, can you please send me the reference?
I've been looking for a long time to find this again.

Randy,

You didn't say what the specific test was you have found to be it. Is it

indeed OES?

dj


On Thu, 18 Jan 2007, werner-at-rosharon.oilfield.slb.com wrote:

} Date: Thu, 18 Jan 2007 13:59:42 -0600
} From: werner-at-rosharon.oilfield.slb.com
} To: dljones-at-bestweb.net
} Subject: [Microscopy] Spark emission test?
}
}
}
}
}
------------------------------------------------------------------------
----
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of
America
} To Subscribe/Unsubscribe --
http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
}
------------------------------------------------------------------------
----
}
} dj is right, when metallurgists say "spark test" they generally mean
} optical emission spectroscopy (OES). OES can give reliable sulfur and
} phosphorous (for instance) readings at low concentrations I would
never
} have the chutzpa to report from EDS.
}
} The grinding wheel test can distinguish high-carbon (short sparks, in
dense
} clusters) from low-carbon steels (few sparks, long, and not in
} clusters). I do not think it can distinguish whether any other
alloying
} elements are present or not. There is a great ASTM document - STP 550
} "Nondestructive Rapid Identification of Metals and Alloys by Spot
Test"
} from 1974 that covers chemical spot tests for qualitative alloy i.d.
but
} that is 'way off-topic.
}
} Regards,
} Andrew
}
} At 11:16 AM 1/18/2007, you wrote:
}
}
}
} }
------------------------------------------------------------------------
----
} } The Microscopy ListServer -- CoSponsor: The Microscopy Society of
America
} } To Subscribe/Unsubscribe --
http://www.microscopy.com/MicroscopyListserver
} } On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} }
------------------------------------------------------------------------
----
} }
} } Randy,
} }
} } Your question of "spark emission test" sounds familiar but I can't
put my
} } finger on it. I can think of three techniques that may be what is
being
} } asked: Spark Testing, Spark Source Mass Spectrometery, or Optical
Emission
} } Spectroscopy.
} }
} } Spark Testing - this is a technique that simply uses a grinder. A
piece of
} } metal is pushed against the grinder. The sparks that fly off are
} } indicative of what the alloy is. I only know of it being used in
metal
} } alloys. It would be the kind of test someone dealing with a scrap
yard
} } would find useful. It can basically tell you what grade of metal you
are
} } dealing with. It takes a fair amount of practice to be able to
visually
} } know what alloy the spark pattern is indicating. People that do this
test
} } frequently can do quite well with it as far as getting a good handle
on
} } what alloy they have in their hand.
} }
} } Spark Source Mass Spectrometery - This is used for a wide range of
} } samples, but tends to do better with samples that are conductive,
although
} } that is not necessarily a limitation. There are two other techniques
that
} } are related to this: Laser Ionization Mass Spectrometery and
Inductively
} } Coupled Plasma Atomic Emission.
} }
} } Optical Emission Spectroscopy - I'm not too familiar with this
technique,
} } but I seem to recall they will often use a spark source in the
process.
} } When they do that, I think this is then called Spark Emission
} } Spectroscopy. You might want to check out Bruker AXS Microanalysis, I
} } believe they make some of these instruments.
} }
} } Perhaps if you post what they are testing and what they want from the
} } analysis, that may help. I hope I've given you at least some food for
} } thought.
} }
} } Good luck,
} }
} } dj
} }
} }
} }
} } On Thu, 18 Jan 2007, TindallR-at-missouri.edu wrote:
} }
} } }
} } }
} } }
} } }
} }
------------------------------------------------------------------------
----
} } } The Microscopy ListServer -- CoSponsor: The Microscopy Society of
America
} } } To Subscribe/Unsubscribe --
http://www.microscopy.com/MicroscopyListserver
} } } On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} } }
} }
------------------------------------------------------------------------
----
} } }
} } } And along a completely different line, we've been asked if we can do
a
} } } "spark emission test". Since I don't know what that is, and Google
} } } hasn't helped, I don't know if we can or not.
} } }
} } } Has anybody heard of this test? Seems to be a way of checking
sample
} } } composition at concentrations below those attainable by EDS, but
that's
} } } about all I can tell.
} } }
} } } Sorry if this isn't microscopy-related, but I don't even know that,
at
} } } this point. We'll see what the collective comes up with on this
one!
} } }
} } } Thanks to all.
} } }
} } } Randy
} } }
} } } Randy Tindall
} } } Senior EM Specialist
} } } Electron Microscopy Core Facility---We Do Small Well!
} } } W125 Veterinary Medicine
} } } University of Missouri
} } } Columbia, MO 65211
} } } Tel: (573) 882-8304
} } } Fax: (573) 884-2227
} } } Email: tindallr-at-missouri.edu
} } } Web: http://www.emc.missouri.edu
} } } On-line calendar:
} } }
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
} } } Week&NavType=Both&Type=TimePlan
} } }
} } }
} } } ==============================Original
} } Headers==============================
} } } 7, 23 -- From TindallR-at-missouri.edu Thu Jan 18 10:12:19 2007
} } } 7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu
} } (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
} } } 7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP
} } id l0IGCJAl020222
} } } 7, 23 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
} } 10:12:19 -0600
} } } 7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu
([209.106.228.34])
} } by um-tsmtpout1.um.umsystem.edu with Microsoft
SMTPSVC(6.0.3790.1830);
} } } 7, 23 -- Thu, 18 Jan 2007 10:12:19 -0600
} } } 7, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
} } } 7, 23 -- Content-class: urn:content-classes:message
} } } 7, 23 -- MIME-Version: 1.0
} } } 7, 23 -- Content-Type: text/plain;
} } } 7, 23 -- charset="us-ascii"
} } } 7, 23 -- Subject: Spark emission test?
} } } 7, 23 -- Date: Thu, 18 Jan 2007 10:12:18 -0600
} } } 7, 23 -- Message-ID:
} } {91108EF9255B394CBF8B7E3789814A41C68D0E-at-UM-XMAIL08.um.umsystem.edu}
} } } 7, 23 -- X-MS-Has-Attach:
} } } 7, 23 -- X-MS-TNEF-Correlator:
} } } 7, 23 -- Thread-Topic: Spark emission test?
} } } 7, 23 -- Thread-Index: Acc7G207OlT1I53OTyafRC/FuYiYDg==
} } } 7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
} } } 7, 23 -- To: {microscopy-at-microscopy.com}
} } } 7, 23 -- X-OriginalArrivalTime: 18 Jan 2007 16:12:19.0651 (UTC)
} } FILETIME=[6DDA0930:01C73B1B]
} } } 7, 23 -- Content-Transfer-Encoding: 8bit
} } } 7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
} } ns.microscopy.com id l0IGCJAl020222
} } } ==============================End of -
} } Headers==============================
} } }
} }
} }
} } ==============================Original
Headers==============================
} } 13, 19 -- From dljones-at-bestweb.net Thu Jan 18 11:15:31 2007
} } 13, 19 -- Received: from vms044pub.verizon.net (vms044pub.verizon.net
} } [206.46.252.44])
} } 13, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP
} } id l0IHFULY021977
} } 13, 19 -- for {Microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
11:15:31
} } -0600
} } 13, 19 -- Received: from localhost ([71.247.249.221])
} } 13, 19 -- by vms044.mailsrvcs.net (Sun Java System Messaging Server
} } 6.2-6.01 (built Apr
} } 13, 19 -- 3 2006)) with ESMTPA id
{0JC200NM7QL94A26-at-vms044.mailsrvcs.net} for
} } 13, 19 -- Microscopy-at-microscopy.com; Thu, 18 Jan 2007 11:15:14 -0600
(CST)
} } 13, 19 -- Date: Thu, 18 Jan 2007 12:14:16 -0500 (Eastern Standard
Time)
} } 13, 19 -- From: "David L. Jones" {dljones-at-bestweb.net}
} } 13, 19 -- Subject: Re: [Microscopy] Spark emission test?
} } 13, 19 -- In-reply-to:
{200701181618.l0IGIW0e026577-at-ns.microscopy.com}
} } 13, 19 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
} } 13, 19 -- To: TindallR-at-missouri.edu
} } 13, 19 -- Cc: Microscopy-at-microscopy.com
} } 13, 19 -- Message-id: {Pine.WNT.4.64.0701181137340.3320-at-H-F1}
} } 13, 19 -- MIME-version: 1.0
} } 13, 19 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
} } 13, 19 -- References: {200701181618.l0IGIW0e026577-at-ns.microscopy.com}
} } ==============================End of -
Headers==============================
}
}
} ==============================Original
Headers==============================
} 8, 36 -- From werner-at-rosharon.oilfield.slb.com Thu Jan 18 13:55:02
2007
} 8, 36 -- Received: from mail.slb.com (nammta01.sugar-land.nam.slb.com
[163.188.150.130])
} 8, 36 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0IJt2Wl003343
} 8, 36 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
13:55:02 -0600
} 8, 36 -- Received: from
pmxchannel_int-daemon.nammta01.sugar-land.nam.slb.com by
} 8, 36 -- nammta01.sugar-land.nam.slb.com
} 8, 36 -- (iPlanet Messaging Server 5.2 Patch 2 (built Jul 14 2004))
} 8, 36 -- id {0JC200301XENHX-at-nammta01.sugar-land.nam.slb.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:23 +0000
(GMT)
} 8, 36 -- Received: from usxsl052.slb.atosorigin-asp.com
} 8, 36 -- (usxsl052.slb.atosorigin-asp.com [199.6.136.167])
} 8, 36 -- by nammta01.sugar-land.nam.slb.com
} 8, 36 -- (iPlanet Messaging Server 5.2 Patch 2 (built Jul 14 2004))
} 8, 36 -- with ESMTP id
{0JC200G99XEN8P-at-nammta01.sugar-land.nam.slb.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:23 +0000
(GMT)
} 8, 36 -- Received: from
pmxchannel-daemon.us085mbx01.slb.atosorigin-asp.com by
} 8, 36 -- us085mbx01.slb.atosorigin-asp.com
} 8, 36 -- (Sun Java System Messaging Server 6.2-1 (built Feb 24 2005))
} 8, 36 -- id {0JC20092VXEJW500-at-us085mbx01.slb.atosorigin-asp.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:19 +0000
(GMT)
} 8, 36 -- Received: from WERNER1-OFS.rosharon.oilfield.slb.com
([163.188.46.5])
} 8, 36 -- by us085mbx01.slb.atosorigin-asp.com
} 8, 36 -- (Sun Java System Messaging Server 6.2-1 (built Feb 24 2005))
} 8, 36 -- with ESMTPSA id
{0JC2003Z9XEJQX50-at-us085mbx01.slb.atosorigin-asp.com} for
} 8, 36 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 19:42:19 +0000
(GMT)
} 8, 36 -- Date: Thu, 18 Jan 2007 13:42:18 -0600
} 8, 36 -- From: Andrew Werner {werner-at-rosharon.oilfield.slb.com}
} 8, 36 -- Subject: Re: [Microscopy] Re: Spark emission test?
} 8, 36 -- In-reply-to: {200701181716.l0IHGH9F023234-at-ns.microscopy.com}
} 8, 36 -- To: microscopy-at-microscopy.com
} 8, 36 -- Message-id:
{6.2.1.2.2.20070118133232.03439d28-at-us1061-pop3.mail.slb.com}
} 8, 36 -- MIME-version: 1.0
} 8, 36 -- X-Mailer: QUALCOMM Windows Eudora Version 6.2.1.2
} 8, 36 -- Content-type: text/plain; charset=us-ascii; format=flowed
} 8, 36 -- Content-transfer-encoding: 7BIT
} 8, 36 -- References: {200701181716.l0IHGH9F023234-at-ns.microscopy.com}
} ==============================End of -
Headers==============================
}


==============================Original
Headers==============================
9, 20 -- From "dljones-at-bestweb.net"-at-bestweb.net Thu Jan 18 15:57:08 2007
9, 20 -- Received: from mta5.srv.hcvlny.cv.net (mta5.srv.hcvlny.cv.net
[167.206.4.200])
9, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0ILv7Cf016738
9, 20 -- for {microscopy-at-microscopy.com} ; Thu, 18 Jan 2007
15:57:08 -0600
9, 20 -- Received: from localhost (ool-44c62883.dyn.optonline.net
[68.198.40.131])
9, 20 -- by mta5.srv.hcvlny.cv.net
9, 20 -- (Sun Java System Messaging Server 6.2-6.01 (built Apr 3
2006))
9, 20 -- with ESMTP id {0JC300E9X3MIVOF0-at-mta5.srv.hcvlny.cv.net} for
9, 20 -- microscopy-at-microscopy.com; Thu, 18 Jan 2007 16:56:59 -0500
(EST)
9, 20 -- Date: Thu, 18 Jan 2007 16:48:42 -0500 (Eastern Standard Time)
9, 20 -- From: "David L. Jones" {"dljones-at-bestweb.net"-at-bestweb.net}
9, 20 -- Subject: Re: [Microscopy] Spark emission test?
9, 20 -- In-reply-to: {200701181959.l0IJxg9t010372-at-ns.microscopy.com}
9, 20 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
9, 20 -- Cc: microscopy-at-microscopy.com
9, 20 -- Message-id: {Pine.WNT.4.64.0701181516340.176-at-dljtoshiba}
9, 20 -- MIME-version: 1.0
9, 20 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
9, 20 -- Content-transfer-encoding: 7BIT
9, 20 -- References: {200701181959.l0IJxg9t010372-at-ns.microscopy.com}
==============================End of -
Headers==============================


==============================Original Headers==============================
19, 20 -- From fschamber-at-aspexcorp.com Fri Jan 19 10:36:01 2007
19, 20 -- Received: from aspexcorp2k6.local (mail.aspexcorp.com [67.141.199.81])
19, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0JGa0R0025169
19, 20 -- for {Microscopy-at-microscopy.com} ; Fri, 19 Jan 2007 10:36:00 -0600
19, 20 -- Content-class: urn:content-classes:message
19, 20 -- MIME-Version: 1.0
19, 20 -- Content-Type: text/plain;
19, 20 -- charset="us-ascii"
19, 20 -- Subject: RE: [Microscopy] Re: Spark emission test?
19, 20 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
19, 20 -- Date: Fri, 19 Jan 2007 11:35:50 -0500
19, 20 -- Message-ID: {65657329C810FE4B96B17390F9BE463E06BF3E-at-ASPEXCORP2K3.aspexcorp2k6.local}
19, 20 -- X-MS-Has-Attach:
19, 20 -- X-MS-TNEF-Correlator:
19, 20 -- Thread-Topic: [Microscopy] Re: Spark emission test?
19, 20 -- Thread-Index: Acc7TLT9vKeUQ6lMRFO7a5jCaV9I6QAmY4Pg
19, 20 -- From: "Fred Schamber" {fschamber-at-aspexcorp.com}
19, 20 -- To: {Microscopy-at-microscopy.com}
19, 20 -- Content-Transfer-Encoding: 8bit
19, 20 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0JGa0R0025169
==============================End of - Headers==============================




From: r-holdford-at-ti.com
Date: Fri, 19 Jan 2007 17:40:59 -0600
Subject: [Microscopy] Re: viaWWW: EBL/ PMMA Mask

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Kurt: Since my company banned cyanide-based etches some years back. one
of my favorite gold etches is as follows:

4.6 grams potassium iodide
1.3 grams Iodine
100 milliliters of DI water
(ratios can be increased/decreased as needed)

Use at room temp. This removes 2-4 microns of Au in around 5 minutes,
according to my notes. You might want to test it on something
expendable before you use it on the real deal. It will store a long time.

klangwor-at-uoregon.edu wrote:
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} This Question/Comment was submitted to the Microscopy Listserver
} using the WWW based Form at http://www.microscopy.com/MicroscopyListserver/MLFormMail.html
} ---------------------------------------------------------------------------
} Remember this posting is most likely not from a Subscriber, so when replying
} please copy both klangwor-at-uoregon.edu as well as the MIcroscopy Listserver
} ---------------------------------------------------------------------------
}
} Email: klangwor-at-uoregon.edu
} Name: Kurt Langworthy
}
} Organization: University of Oregon
}
} Title-Subject: [Filtered] EBL/ PMMA Mask
}
} Question: All,
} I'm looking for a gold etching recipt that will allow me to use PMMA as a mask. KCN looks promising, but is highly toxic. Does anyone have any better ideas?
} Thanks,
} Kurt
}
} ---------------------------------------------------------------------------
}
} ==============================Original Headers==============================
} 6, 12 -- From zaluzec-at-microscopy.com Wed Jan 17 18:34:22 2007
} 6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
} 6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0I0YJlb010881
} 6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 17 Jan 2007 18:34:21 -0600
} 6, 12 -- Mime-Version: 1.0
} 6, 12 -- X-Sender: (Unverified)
} 6, 12 -- Message-Id: {p06110407c1d470f44a74-at-[206.69.208.22]}
} 6, 12 -- Date: Wed, 17 Jan 2007 18:34:19 -0600
} 6, 12 -- To: microscopy-at-microscopy.com
} 6, 12 -- From: klangwor-at-uoregon.edu (by way of MicroscopyListserver)
} 6, 12 -- Subject: viaWWW: EBL/ PMMA Mask
} 6, 12 -- Content-Type: text/plain; charset="us-ascii"
} ==============================End of - Headers==============================
}
}

--
~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
Becky Holdford (r-holdford-at-ti.com)
972-995-2360
972-648-8743 (pager)
SC Packaging FA Development
Texas Instruments, Inc.
Dallas, TX
~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~


==============================Original Headers==============================
6, 23 -- From r-holdford-at-ti.com Fri Jan 19 17:40:58 2007
6, 23 -- Received: from calf.ext.ti.com (calf.ext.ti.com [198.47.26.144])
6, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0JNew1s012201
6, 23 -- for {Microscopy-at-MSA.Microscopy.Com} ; Fri, 19 Jan 2007 17:40:58 -0600
6, 23 -- Received: from dlep33.itg.ti.com ([157.170.170.112])
6, 23 -- by calf.ext.ti.com (8.13.7/8.13.7) with ESMTP id l0JNeRND006255
6, 23 -- (version=TLSv1/SSLv3 cipher=DHE-RSA-AES256-SHA bits=256 verify=NO);
6, 23 -- Fri, 19 Jan 2007 17:40:38 -0600
6, 23 -- Received: from [156.117.194.120] (localhost [127.0.0.1])
6, 23 -- by dlep33.itg.ti.com (8.13.7/8.13.7) with ESMTP id l0JNeMol019754;
6, 23 -- Fri, 19 Jan 2007 17:40:22 -0600 (CST)
6, 23 -- Message-ID: {45B156E5.4050106-at-ti.com}
6, 23 -- Date: Fri, 19 Jan 2007 17:40:21 -0600
6, 23 -- From: Becky Holdford {r-holdford-at-ti.com}
6, 23 -- Organization: SC Packaging Development -- FA Development
6, 23 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
6, 23 -- MIME-Version: 1.0
6, 23 -- To: klangwor-at-uoregon.edu, MSA ListServer {Microscopy-at-MSA.Microscopy.Com}
6, 23 -- Subject: Re: [Microscopy] viaWWW: EBL/ PMMA Mask
6, 23 -- References: {200701180034.l0I0YWPP011425-at-ns.microscopy.com}
6, 23 -- In-Reply-To: {200701180034.l0I0YWPP011425-at-ns.microscopy.com}
6, 23 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
6, 23 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: jeremy.wilburn-at-gmail.com
Date: Sat, 20 Jan 2007 09:27:38 -0600
Subject: [Microscopy] viaWWW: biological microscopy question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both jeremy.wilburn-at-gmail.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: jeremy.wilburn-at-gmail.com
Name: Jeremy Wilburn

Organization: Vanderbilt University

Title-Subject: [Filtered] biological microscopy question

Question: I am a chemistry grad student working with both live cells and cultured pancreatic islets. More specifically I do electrochemical (electrophysiology) measurements with ultramicroelectrodes, and am needing to put together an instrument similar to a patch-clamp on an inverted microscope. I am in the market for an inverted microscope now, but am limited in funding (~$8-10k).

My problem is the following. Since I use a capillary electrode body (which must remain completely vertical, unlike a patch pipette), I am unable to use the top condenser for imaging (due to both space restraints and shadowing from the electrode body). I was thinking to get a fluorescence-type inverted microscope and remove the filter, making a bottom-illumination similar to a metallurgical microscope, but would still let me do fluorescence at some point...would this work?

My second option would be a more expensive scope (like Olympus IX71) that has multiple light ports...

Any other suggestions (including a good scope to look for) would be greatly appreciated.

---------------------------------------------------------------------------

==============================Original Headers==============================
9, 12 -- From zaluzec-at-microscopy.com Sat Jan 20 09:27:37 2007
9, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
9, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFRbth004257
9, 12 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:27:37 -0600
9, 12 -- Mime-Version: 1.0
9, 12 -- X-Sender: (Unverified)
9, 12 -- Message-Id: {p06110404c1d7e559607a-at-[206.69.208.22]}
9, 12 -- Date: Sat, 20 Jan 2007 09:27:35 -0600
9, 12 -- To: microscopy-at-microscopy.com
9, 12 -- From: jeremy.wilburn-at-gmail.com (by way of MicroscopyListserver)
9, 12 -- Subject: viaWWW: biological microscopy question
9, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: Remi.Delville-at-ua.ac.be
Date: Sat, 20 Jan 2007 09:28:00 -0600
Subject: [Microscopy] viaWWW: Electropolishing TiNiAu

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both Remi.Delville-at-ua.ac.be as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: Remi.Delville-at-ua.ac.be
Name: RÈmi DELVILLE

Organization: EMAT, University of Antwerp, Belgium

Title-Subject: [Filtered] Electropolishing TiNiAu

Question: Dear microscopists,

I am currently looking for a suitable electropolishing solution for the ternary shape-memory alloys TiNiAu in order to have a thinned sample for electron microscopy.

I already tried several solutions that are known to work well with TiNi. Unfortunately, the added gold at a concentration around 15 at% seems to severely hamper the electro-polishing: the same conditions as for Ni-Ti give etched specimens with bad edges of the hole.
We've played with tension, temperature, fluid speed,etc.. but without success.


Solutions tried that work with NiTi but not with my sample:

80% methanol + 20% sulphuric acid (-17ƒC)
93% acetic acid + 7% perchloric acid (5ƒC)
93% methanol + 7% perchloric acid (-10ƒC)
75% methanol + 25% nitric acid (-22ƒC)

I already looked up in the literature and couldn't find anything close to my alloy. So I was wondering if you might have a suggestion of an alternative, eg, adding an extra acid to one of the above baths or maybe something completely different.

Many thanks in advance,


RÈmi Delville

EMAT - Electron Microscopy for Material Science - University of Antwerp, Belgium



---------------------------------------------------------------------------


==============================Original Headers==============================
19, 14 -- From zaluzec-at-microscopy.com Sat Jan 20 09:27:59 2007
19, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
19, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFRwJL004565
19, 14 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:27:59 -0600
19, 14 -- Mime-Version: 1.0
19, 14 -- X-Sender: (Unverified)
19, 14 -- Message-Id: {p06110405c1d7e57165f2-at-[206.69.208.22]}
19, 14 -- Date: Sat, 20 Jan 2007 09:27:58 -0600
19, 14 -- To: microscopy-at-microscopy.com
19, 14 -- From: Remi.Delville-at-ua.ac.be (by way of MicroscopyListserver)
19, 14 -- Subject: viaWWW: Electropolishing TiNiAu
19, 14 -- Content-Type: text/plain; charset="iso-8859-1"
19, 14 -- Content-Transfer-Encoding: 8bit
19, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0KFRwJL004565
==============================End of - Headers==============================




From: peter.tomic-at-renwireless.com
Date: Sat, 20 Jan 2007 09:28:30 -0600
Subject: [Microscopy] viaWWW: FEI XL-50 Experience

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both peter.tomic-at-renwireless.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: peter.tomic-at-renwireless.com
Name: Peter Tomic

Organization: Renaissance Wireless

Title-Subject: [Filtered] FEI XL-50 Experience

Question: Listers,

I'd like to hear from anyone that has owned/operated an FEI XL50 SEM. I have one we are bringing up in our MEMS lab. There are some unusual aspects to this instrument such as the exchange port.

My understanding is that there were not very many XL-50's made. Does anyone currently own one?

You may respond on-list or off-list if you care to.

Peter Tomic
Renaissance Wireless Corp.
Somerset, New Jersey, USA


---------------------------------------------------------------------------

==============================Original Headers==============================
11, 12 -- From zaluzec-at-microscopy.com Sat Jan 20 09:28:29 2007
11, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
11, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFSSDY005532
11, 12 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:28:29 -0600
11, 12 -- Mime-Version: 1.0
11, 12 -- X-Sender: (Unverified)
11, 12 -- Message-Id: {p06110406c1d7e58f6d02-at-[206.69.208.22]}
11, 12 -- Date: Sat, 20 Jan 2007 09:28:28 -0600
11, 12 -- To: microscopy-at-microscopy.com
11, 12 -- From: peter.tomic-at-renwireless.com (by way of MicroscopyListserver)
11, 12 -- Subject: viaWWW: FEI XL-50 Experience
11, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: liang-at-saturn.med.nyu.edu
Date: Sat, 20 Jan 2007 09:29:10 -0600
Subject: [Microscopy] viaWWW: Job Opening Research Technician in EM core facility

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both liang-at-saturn.med.nyu.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: liang-at-saturn.med.nyu.edu
Name: Alice Liang

Organization: NYU school of Medicine

Title-Subject: [Filtered] Research Technician in EM core facility

Question: Research Technician ñ Electron Microscopy
The Image Core Facility of Skirball Institute, NYU School of Medicine

A full - time research technician position is currently available in the
Image Core Facility at Skirball Institute of NYU School of Medicine. The
facility provides essential electron microscopic and structural analysis
support for school of medicine and local community. The successful candidate
must have at least a B.S. degree, appropriate training and significant
biomedical research experience in histology, electron microscopy (TEM) and
tissue sample preparation. Candidates with an excellent communication skill
and experience in instrument maintenance are encouraged to apply. Skirball
is an academic organization and offers competitive salaries and excellent
fringe benefits. Interested individuals should send resume and three
references to:

Alice Liang, Ph.D.
Director of Image Core Facility
Skirball Institute of Biomolecular Medicine
Skirball 2nd floor, EM Suite
NYU School of Medicine
New York, NY 10016
Tel: 212-263-7644 (o); 212-263-7099 (Lab)
Fax: 212-263-7643
E-mail: liang-at-saturn.med.nyu.edu
http://saturn.med.nyu.edu/facilities/imagecore/

---------------------------------------------------------------------------


==============================Original Headers==============================
8, 14 -- From zaluzec-at-microscopy.com Sat Jan 20 09:29:10 2007
8, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFT9Ya008001
8, 14 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:29:10 -0600
8, 14 -- Mime-Version: 1.0
8, 14 -- X-Sender: (Unverified)
8, 14 -- Message-Id: {p06110407c1d7e5a8730b-at-[206.69.208.22]}
8, 14 -- Date: Sat, 20 Jan 2007 09:29:08 -0600
8, 14 -- To: microscopy-at-microscopy.com
8, 14 -- From: liang-at-saturn.med.nyu.edu (by way of MicroscopyListserver)
8, 14 -- Subject: viaWWW: Job Opening Research Technician in EM core facility
8, 14 -- Content-Type: text/plain; charset="iso-8859-1"
8, 14 -- Content-Transfer-Encoding: 8bit
8, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0KFT9Ya008001
==============================End of - Headers==============================




From: rafaldb-at-gmail.com
Date: Sat, 20 Jan 2007 09:30:34 -0600
Subject: [Microscopy] viaWWW: Positions available: Technical University of Denmark

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both rafaldb-at-gmail.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: rafaldb-at-gmail.com
Name: Rafal Dunin-Borkowski

Organization: Technical University of Denmark

Title-Subject: [Filtered] Positions available: Technical University of Denmark

Question: Immediate opening:

-----------------------------------------------------------------
Senior scientists in Electron Microscopy at the Center for Electron Nanoscopy, Technical University of Denmark

Apply before: 12.02.2007 at 12:00 (noon)

The Center for Electron Nanoscopy (CEN) at the Technical University of Denmark (DTU), which has been made possible by a generous donation from the A.P. Moller Foundation, seeks to appoint one or more senior scientists to permanent positions in advanced electron microscopy for materials science and nanotechnology. The number of appointments will depend on the quality and experience of the applicants.

CEN-DTU will be fully operational in 2007 and will contain seven new electron microscopes. There will be two aberration-corrected monochromated TEMs, one of which will be equipped with a gas reaction capability.

Applicants must have outstanding international research reputations and hands-on expertise in a wide range of advanced electron microscopy techniques. They will be expected to perform experimental work with users of the facility, and to assist CEN's director in raising research funding and in developing an internationally leading program of advanced research.

Expertise in one or more of the following areas is essential: gas reaction TEM; aberration-correction; monochromated EELS; electron tomography; electron holography; direct methods; advanced image analysis and simulation for quantitative electron microscopy.

Positions are to be filled in 2007, on dates to be agreed with the successful applicants.

The salary and appointment terms will be based on the current collective agreement for Danish University faculty members.

All interested candidates, irrespective of age, gender, race, religion and ethnic background, are invited to apply.

Further information can be obtained from CEN's director Rafal Dunin-Borkowski (rafaldb-at-gmail.com).

Applications, which should include a resumÈ, list of publications and statement of research interests, should be sent to

The Rector
Technical University of Denmark
Building 101 A
DK-2800 Lyngby, Denmark

and must be received before 12:00 on 12 February 2007.

-----------------------------------------------------------------


---------------------------------------------------------------------------


==============================Original Headers==============================
22, 14 -- From zaluzec-at-microscopy.com Sat Jan 20 09:30:34 2007
22, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
22, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFUXDh012623
22, 14 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:30:34 -0600
22, 14 -- Mime-Version: 1.0
22, 14 -- X-Sender: (Unverified)
22, 14 -- Message-Id: {p06110408c1d7e5d07c40-at-[206.69.208.22]}
22, 14 -- Date: Sat, 20 Jan 2007 09:30:32 -0600
22, 14 -- To: microscopy-at-microscopy.com
22, 14 -- From: rafaldb-at-gmail.com (by way of MicroscopyListserver)
22, 14 -- Subject: viaWWW: Positions available: Technical University of Denmark
22, 14 -- Content-Type: text/plain; charset="iso-8859-1"
22, 14 -- Content-Transfer-Encoding: 8bit
22, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0KFUXDh012623
==============================End of - Headers==============================




From: dac-at-research.umass.edu
Date: Sat, 20 Jan 2007 09:32:13 -0600
Subject: [Microscopy] viaWWW: Sputter coating details

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both dac-at-research.umass.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: dac-at-research.umass.edu
Name: Dale Callaham

Organization: University of Massachusetts

Title-Subject: [Filtered] Sputter coating details

Question: Hi listees,

This message is in response to the thread about Polaron E5100 sputter coaters and operating conditions: HV knob setting and voltages to be used. I saw mention that finer grain size of the sputtered film is achieved when a lower voltage (I think 700V was mentioned) is used, and I have seen this elsewhere. I'm trying to come to terms with this and sometimes it is worth going back to basics.

Several sources show graphs of the Voltage and Current relationships in a DC plasma:

http://science-education.pppl.gov/SummerInst/SGershman/Structure_of_Glow_Discharge.pdf
(Sophia Gershman must be one heck of a High School teacher!)

http://www.emitech.co.uk/sputter-coating-brief2.htm

These data show that for the range of conditions typically used in a sputter coater the actual voltage that exists in the plasma is clamped by the physics of the plasma and while the knob may set 700V or 2.2kV, the difference between the set voltage and the voltage across the plasma must be accounted for by IR drop in the winding resistance and any ballasting/limiting resistors in the circuit (the E5100 has resistance in the primary leads, 5k ohm in the secondary lead, plus the considerable winding resistance of the secondary HV winding). The description of the behaviour of the plasma is that the current increases by increasing the volume of the plasma - any you can see this for yourself - but as best I can judge by the graphs in the referenced papers, the voltage may increase only very slightly, and it is difficult for me to see that it would change the energy of electrons bombarding the target significantly.

Is the "finer grain with lower voltage" real? If it is real, what is the basis for it? Have I completely misinterpreted the data?

Sincerely,

Dale Callaham
Umass, Amherst
dac-at-research.umass.edu



---------------------------------------------------------------------------

==============================Original Headers==============================
16, 12 -- From zaluzec-at-microscopy.com Sat Jan 20 09:32:12 2007
16, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
16, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFW96E018638
16, 12 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:32:11 -0600
16, 12 -- Mime-Version: 1.0
16, 12 -- X-Sender: (Unverified)
16, 12 -- Message-Id: {p0611040ac1d7e65f9dde-at-[206.69.208.22]}
16, 12 -- Date: Sat, 20 Jan 2007 09:32:09 -0600
16, 12 -- To: microscopy-at-microscopy.com
16, 12 -- From: dac-at-research.umass.edu (by way of MicroscopyListserver)
16, 12 -- Subject: viaWWW: Sputter coating details
16, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: edelmare-at-muohio.edu
Date: Sat, 20 Jan 2007 14:54:54 -0600
Subject: [Microscopy] Re: viaWWW: biological microscopy question

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Jeremy:

The reflected light imaging will work, but your biggest issue will
be contrast in the biological samples. And for reflected light
imaging you do NOT need or want a high brightness light source like
those used for florescence work. Cost savings ($100 vs $5000).

As for contrast reflected light DIC might be ideal however since it
has depth of focus issues seeing the electrode as it approaches the
cell would be problematic. And it is costly. I suggest you look at
reflected light dark-field. A rare beast but might be what you need.

Another approach would be oblique illumination. I seem to remember
that some inverted scopes had/have tilting lamp/condensors so you can
tilt them off axis 35? 45? 60-degrees or so. . . . in fact I just
went and checked my IX-81 illumination column tilts backwards, and
the manual states "Even with the illumination column tilted the
specimen surface will be illuminated, which is convient for rough
confirmation of the specimen location on initial positioning of the
specimen." My guess would be that setting Köhler illumination for
highest resolution would be an issue but would still allow very
reasonable bright-field imaging.

good luck.




On 20 Jan 2007 at 9:29, jeremy.wilburn-at-gmail.com wrote:

}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} This Question/Comment was submitted to the Microscopy Listserver
} using the WWW based Form at http://www.microscopy.com/MLFormMail.html
} ---------------------------------------------------------------------------
} Remember this posting is most likely not from a Subscriber, so when replying
} please copy both jeremy.wilburn-at-gmail.com as well as the MIcroscopy Listserver
} ---------------------------------------------------------------------------
}
} Email: jeremy.wilburn-at-gmail.com
} Name: Jeremy Wilburn
}
} Organization: Vanderbilt University
}
} Title-Subject: [Filtered] biological microscopy question
}
} Question: I am a chemistry grad student working with both live cells and cultured pancreatic islets. More specifically I do electrochemical (electrophysiology) measurements with ultramicroelectrodes, and am needing to put together an instrument similar to a patch-clamp on an inverted microscope. I am in the market for an inverted microscope now, but am limited in funding (~$8-10k).
}
} My problem is the following. Since I use a capillary electrode body (which must remain completely vertical, unlike a patch pipette), I am unable to use the top condenser for imaging (due to both space restraints and shadowing from the electrode body). I was thinking to get a fluorescence-type inverted microscope and remove the filter, making a bottom-illumination similar to a metallurgical microscope, but would still let me do fluorescence at some point...would this work?
}
} My second option would be a more expensive scope (like Olympus IX71) that has multiple light ports...
}
} Any other suggestions (including a good scope to look for) would be greatly appreciated.
}
} ---------------------------------------------------------------------------
}
} ==============================Original Headers==============================
} 9, 12 -- From zaluzec-at-microscopy.com Sat Jan 20 09:27:37 2007
} 9, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
} 9, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KFRbth004257
} 9, 12 -- for {microscopy-at-microscopy.com} ; Sat, 20 Jan 2007 09:27:37 -0600
} 9, 12 -- Mime-Version: 1.0
} 9, 12 -- X-Sender: (Unverified)
} 9, 12 -- Message-Id: {p06110404c1d7e559607a-at-[206.69.208.22]}
} 9, 12 -- Date: Sat, 20 Jan 2007 09:27:35 -0600
} 9, 12 -- To: microscopy-at-microscopy.com
} 9, 12 -- From: jeremy.wilburn-at-gmail.com (by way of MicroscopyListserver)
} 9, 12 -- Subject: viaWWW: biological microscopy question
} 9, 12 -- Content-Type: text/plain; charset="us-ascii"
} ==============================End of - Headers==============================


Richard E. Edelmann, Ph.D.
Electron Microscopy Facility Director
364 Pearson Hall
Miami University, Oxford, OH 45056
Ph: 513.529.5712 Fax: 513.529.4243
E-mail: edelmare-at-muohio.edu
http://www.emf.muohio.edu

"A main-frame: The biggest PC peripheral available."



==============================Original Headers==============================
15, 33 -- From edelmare-at-muohio.edu Sat Jan 20 14:54:54 2007
15, 33 -- Received: from spamfirewall.muohio.edu (walrus.mcs.muohio.edu [134.53.6.27])
15, 33 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0KKsrKX009982
15, 33 -- for {microscopy-at-Microscopy.com} ; Sat, 20 Jan 2007 14:54:53 -0600
15, 33 -- X-ASG-Debug-ID: 1169326488-3fd400980000-Dem1zR
15, 33 -- X-Barracuda-URL: http://spamfirewall.muohio.edu:80/cgi-bin/mark.cgi
15, 33 -- X-Barracuda-Connect: mulnx24.mcs.muohio.edu[134.53.6.11]
15, 33 -- X-Barracuda-Start-Time: 1169326488
15, 33 -- Received: from mulnx24.mcs.muohio.edu (mulnx24.mcs.muohio.edu [134.53.6.11])
15, 33 -- by spamfirewall.muohio.edu (Spam Firewall) with ESMTP
15, 33 -- id A4F9033C62D; Sat, 20 Jan 2007 15:54:48 -0500 (EST)
15, 33 -- Received: from [192.168.1.23] ([134.53.14.105])
15, 33 -- by mulnx24.mcs.muohio.edu (Switch-3.1.8/Switch-3.1.7) with ESMTP id l0KKsmU4020217;
15, 33 -- Sat, 20 Jan 2007 15:54:48 -0500
15, 33 -- From: "Richard E. Edelmann" {edelmare-at-muohio.edu}
15, 33 -- To: jeremy.wilburn-at-gmail.com
15, 33 -- Date: Sat, 20 Jan 2007 15:54:46 -0500
15, 33 -- MIME-Version: 1.0
15, 33 -- X-ASG-Orig-Subj: Re: [Microscopy] viaWWW: biological microscopy question
15, 33 -- Subject: Re: [Microscopy] viaWWW: biological microscopy question
15, 33 -- CC: microscopy-at-Microscopy.com
15, 33 -- Message-ID: {45B23B46.5131.11055C7D-at-edelmare.muohio.edu}
15, 33 -- Priority: normal
15, 33 -- In-reply-to: {200701201529.l0KFTDqB008077-at-ns.microscopy.com}
15, 33 -- References: {200701201529.l0KFTDqB008077-at-ns.microscopy.com}
15, 33 -- X-mailer: Pegasus Mail for Windows (4.41)
15, 33 -- Content-type: text/plain; charset=ISO-8859-1
15, 33 -- Content-description: Mail message body
15, 33 -- X-Barracuda-Virus-Scanned: by Barracuda Spam Firewall at muohio.edu
15, 33 -- X-Barracuda-Spam-Score: 0.00
15, 33 -- X-Barracuda-Spam-Status: No, SCORE=0.00 using per-user scores of TAG_LEVEL=3.5 QUARANTINE_LEVEL=1000.0 KILL_LEVEL=1000.0
15, 33 -- Content-Transfer-Encoding: 8bit
15, 33 -- X-MIME-Autoconverted: from Quoted-printable to 8bit by ns.microscopy.com id l0KKsrKX009982
==============================End of - Headers==============================




From: mgb-at-ansto.gov.au
Date: Mon, 22 Jan 2007 08:03:07 -0600
Subject: [Microscopy] viaWWW: Wanted: anode chamber for JEOL 2000fxII TEM

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both mgb-at-ansto.gov.au as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: mgb-at-ansto.gov.au
Name: Mark Blackford

Organization: Australian Nuclear Science and Technology Organisation

Title-Subject: [Filtered] Wanted: anode chamber for JEOL 2000fxII TEM

Question: Hi All,

I'm hoping someone may be able to help locate a working anode chamber (electron gun) for a JEOL 2000fxII TEM. Our microscope can only be operated at 100kV due to electrical discharging in the anode chamber at higher accelerating voltages. Several attempts at cleaning the internal components have not been entirely successful.

A fully refurbished unit can be supplied by JEOL, but not cheaply.

I would like to contact anyone who has a suitable fully functional anode chamber, capable of working at 200kV, that they don't happen to need anymore. Please contact me off-line. Cheers,

Mark Blackford

---------------------------------------------------------------------------

==============================Original Headers==============================
10, 12 -- From zaluzec-at-microscopy.com Mon Jan 22 08:03:07 2007
10, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
10, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0ME37Dh025551
10, 12 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007 08:03:07 -0600
10, 12 -- Mime-Version: 1.0
10, 12 -- X-Sender: (Unverified)
10, 12 -- Message-Id: {p06110405c1da748cf355-at-[206.69.208.22]}
10, 12 -- Date: Mon, 22 Jan 2007 08:03:05 -0600
10, 12 -- To: microscopy-at-microscopy.com
10, 12 -- From: mgb-at-ansto.gov.au (by way of MicroscopyListserver)
10, 12 -- Subject: viaWWW: Wanted: anode chamber for JEOL 2000fxII TEM
10, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: dmcdaniel-at-usuhs.mil
Date: Mon, 22 Jan 2007 08:03:29 -0600
Subject: [Microscopy] viaWWW: LR White polymerization problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both dmcdaniel-at-usuhs.mil as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: dmcdaniel-at-usuhs.mil
Name: Dennis McDaniel

Title-Subject: [Filtered] LR White polymerization problem

Question: I am having difficulty polymerizing a sample in LR White. The cells (macropahges and dendritic cells) were grown on a piece of aclar, fixed with formaldehyde/glutaraldehyde, post-fixed with uranyl acetate, dehydrated in a graduated series of EtOH, and infiltrated with LR White. I then sandwiched the aclar film containing the cells between two larger pieces of aclar (since the edges don't polymerize due to air exposure) and put the sample in the oven at 55C. After 48h, the resin immediately around the piece of aclar containing the cells appears very fragmented and cracked and is obviously unsuitable for sectioning, while the resin further from the cells appears to have polymerized properly. At first I thought it was an infiltration problem, but I did several changes of 100% resin and even left the samples overnight on a rotator at room temperature in 100% resin to ensure complete infiltration. Does anyone have any idea what they problem might be? Thanks.

---------------------------------------------------------------------------

==============================Original Headers==============================
5, 12 -- From zaluzec-at-microscopy.com Mon Jan 22 08:03:29 2007
5, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
5, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0ME3SfO025798
5, 12 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007 08:03:29 -0600
5, 12 -- Mime-Version: 1.0
5, 12 -- X-Sender: (Unverified)
5, 12 -- Message-Id: {p06110406c1da74a6f98c-at-[206.69.208.22]}
5, 12 -- Date: Mon, 22 Jan 2007 08:03:27 -0600
5, 12 -- To: microscopy-at-microscopy.com
5, 12 -- From: dmcdaniel-at-usuhs.mil (by way of MicroscopyListserver)
5, 12 -- Subject: viaWWW: LR White polymerization problem
5, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: krensing-at-ucalgary.ca
Date: Mon, 22 Jan 2007 09:55:24 -0600
Subject: [Microscopy] Re: viaWWW: LR White polymerization problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dennis,
I think that your problem is a result of incomplete dehydration. LR
White is supposed to handle some water without problem, but when you
sandwich the pieces, you create a limited volume for diffusion and the
water becomes too great locally. Make sure your ethanol is dry (on
molecular sieve or freshly opened).
That's my theory.
Kim

dmcdaniel-at-usuhs.mil wrote:

} Email: dmcdaniel-at-usuhs.mil
} Name: Dennis McDaniel
}
} Title-Subject: [Filtered] LR White polymerization problem
}
} Question: I am having difficulty polymerizing a sample in LR White. The cells (macropahges and dendritic cells) were grown on a piece of aclar, fixed with formaldehyde/glutaraldehyde, post-fixed with uranyl acetate, dehydrated in a graduated series of EtOH, and infiltrated with LR White. I then sandwiched the aclar film containing the cells between two larger pieces of aclar (since the edges don't polymerize due to air exposure) and put the sample in the oven at 55C. After 48h, the resin immediately around the piece of aclar containing the cells appears very fragmented and cracked and is obviously unsuitable for sectioning, while the resin further from the cells appears to have polymerized properly. At first I thought it was an infiltration problem, but I did several changes of 100% resin and even left the samples overnight on a rotator at room temperature in 100% resin to ensure complete infiltration. Does anyone have any idea what they problem might be? Thanks.
}
} ---------------------------------------------------------------------------


--
Kim Rensing Ph.D.
Manager,
Microscopy and Imaging Facility

University of Calgary,
Health Sciences Centre
B129 - 3330 Hospital Drive NW
Calgary, AB, Canada T2N 4N1
403-220-3488
krensing-at-ucalgary.ca

==============================Original Headers==============================
6, 22 -- From krensing-at-ucalgary.ca Mon Jan 22 09:55:23 2007
6, 22 -- Received: from smtp2.ucalgary.ca (smtp2.ucalgary.ca [136.159.34.223])
6, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0MFtNkD018643
6, 22 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007 09:55:23 -0600
6, 22 -- Received: from [136.159.164.163] (unknown [136.159.164.163])
6, 22 -- by smtp2.ucalgary.ca (Postfix) with ESMTP id 501331004B;
6, 22 -- Mon, 22 Jan 2007 08:55:22 -0700 (MST)
6, 22 -- Message-ID: {45B4DE64.1000509-at-ucalgary.ca}
6, 22 -- Date: Mon, 22 Jan 2007 08:55:16 -0700
6, 22 -- From: Kim Rensing {krensing-at-ucalgary.ca}
6, 22 -- Organization: Microscopy and Imaging Facility, U. of Calgary
6, 22 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
6, 22 -- MIME-Version: 1.0
6, 22 -- To: dmcdaniel-at-usuhs.mil, microscopy-at-microscopy.com
6, 22 -- Subject: Re: [Microscopy] viaWWW: LR White polymerization problem
6, 22 -- References: {200701221414.l0MEEtPY012668-at-ns.microscopy.com}
6, 22 -- In-Reply-To: {200701221414.l0MEEtPY012668-at-ns.microscopy.com}
6, 22 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
6, 22 -- Content-Transfer-Encoding: 7bit
6, 22 -- X-UCalgary-MailScanner-Information: Please contact IT Help Desk at (403) 220-5555 for more information
6, 22 -- X-UCalgary-MailScanner: Found to be clean
6, 22 -- X-UCalgary-MailScanner-From: krensing-at-ucalgary.ca
==============================End of - Headers==============================




From: gary.m.brown-at-exxonmobil.com
Date: Mon, 22 Jan 2007 11:22:02 -0600
Subject: [Microscopy] Re: viaWWW: LR White polymerization problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dennis,

I suspect that your polymerization problems arise from entrapped air. I
eliminate that problem by purging my oven with nitrogen, thus alleviating
other more tedious approaches. Works every time.

Regards,

Gary M. Brown
Microscopy Specialist
ExxonMobil Chemical Company
Baytown Technology & Engineering - West
5200 Bayway Drive
Baytown, Texas 77520-2101
phone: 281 834 2387
fax: 281 834 2395
e-mail: Gary.M.Brown-at-ExxonMobil.com



dmcdaniel-at-usuh
s.mil
To
gary.m.brown-at-exxonmobil.com
01/22/07 08:08 cc
AM
Subject
[Microscopy] viaWWW: LR White
Please respond polymerization problem
to
dmcdaniel-at-usuh
s.mil










----------------------------------------------------------------------------

The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
http://microscopy.com/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when
replying
please copy both dmcdaniel-at-usuhs.mil as well as the MIcroscopy
Listserver
---------------------------------------------------------------------------

Email: dmcdaniel-at-usuhs.mil
Name: Dennis McDaniel

Title-Subject: [Filtered] LR White polymerization problem

Question: I am having difficulty polymerizing a sample in LR White. The
cells (macropahges and dendritic cells) were grown on a piece of aclar,
fixed with formaldehyde/glutaraldehyde, post-fixed with uranyl acetate,
dehydrated in a graduated series of EtOH, and infiltrated with LR White. I
then sandwiched the aclar film containing the cells between two larger
pieces of aclar (since the edges don't polymerize due to air exposure) and
put the sample in the oven at 55C. After 48h, the resin immediately around
the piece of aclar containing the cells appears very fragmented and cracked
and is obviously unsuitable for sectioning, while the resin further from
the cells appears to have polymerized properly. At first I thought it was
an infiltration problem, but I did several changes of 100% resin and even
left the samples overnight on a rotator at room temperature in 100% resin
to ensure complete infiltration. Does anyone have any idea what they
problem might be? Thanks.

---------------------------------------------------------------------------

==============================Original
Headers==============================
5, 12 -- From zaluzec-at-microscopy.com Mon Jan 22 08:03:29 2007
5, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
5, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0ME3SfO025798
5, 12 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007
08:03:29 -0600
5, 12 -- Mime-Version: 1.0
5, 12 -- X-Sender: (Unverified)
5, 12 -- Message-Id: {p06110406c1da74a6f98c-at-[206.69.208.22]}
5, 12 -- Date: Mon, 22 Jan 2007 08:03:27 -0600
5, 12 -- To: microscopy-at-microscopy.com
5, 12 -- From: dmcdaniel-at-usuhs.mil (by way of MicroscopyListserver)
5, 12 -- Subject: viaWWW: LR White polymerization problem
5, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of -
Headers==============================



==============================Original Headers==============================
23, 19 -- From gary.m.brown-at-exxonmobil.com Mon Jan 22 11:22:01 2007
23, 19 -- Received: from hoespc02.exxonmobil.com (hoespc02.exxonmobil.com [192.67.48.39])
23, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0MHM1TQ031633
23, 19 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007 11:22:01 -0600
23, 19 -- Received: from dalnmg03.na.xom.com (dalnmg03.na.xom.com [131.126.97.122])
23, 19 -- by hoespc02.exxonmobil.com (Switch-3.1.11/Switch-3.1.10) with ESMTP id l0MHLsWc015677;
23, 19 -- Mon, 22 Jan 2007 11:21:56 -0600 (CST)
23, 19 -- In-Reply-To: {200701221408.l0ME81Zl032651-at-ns.microscopy.com}
23, 19 -- Subject: Re: [Microscopy] viaWWW: LR White polymerization problem
23, 19 -- Importance:
23, 19 -- To: dmcdaniel-at-usuhs.mil, microscopy-at-microscopy.com
23, 19 -- X-Mailer: Lotus Notes 652HF109 January 14, 2005
23, 19 -- Message-ID: {OF1C37A9C1.76CA9570-ON8625726B.005F1376-8625726B.005F661F-at-exxonmobil.com}
23, 19 -- From: gary.m.brown-at-exxonmobil.com
23, 19 -- Date: Mon, 22 Jan 2007 11:21:54 -0600
23, 19 -- X-MIMETrack: Serialize by Router on Dalnmg03.na.xom.com/S/ExxonMobil(652FP1HF193|March
23, 19 -- 02, 2006) at 01/22/2007 11:21:59 AM
23, 19 -- MIME-Version: 1.0
23, 19 -- Content-type: text/plain; charset=US-ASCII
==============================End of - Headers==============================




From: zerfasp-at-ors.od.nih.gov
Date: Mon, 22 Jan 2007 19:24:30 -0600
Subject: [Microscopy] viaWWW: Part for print processor

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both zerfasp-at-ors.od.nih.gov as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: zerfasp-at-ors.od.nih.gov
Name: Patricia Zerfas

Organization: NIH

Title-Subject: [Filtered] Part for print processor

Question: Does anyone have a MohrPro8, model ME-42 print processor they are no longer using? I need to obtain the roller assembly that processes the print through the water and drying. My roller assembly is broken beyond repair and this model is no longer made. I am willing to pay for shipment.

Thanks,
Patricia Zerfas
National Institute of Health
Building 28A, Room 112
28 Library Drive
Bethesda, MD 20892 USA
ph # (301) 496-4464

---------------------------------------------------------------------------

==============================Original Headers==============================
7, 12 -- From zaluzec-at-microscopy.com Mon Jan 22 19:24:29 2007
7, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
7, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0N1OSAo022259
7, 12 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007 19:24:29 -0600
7, 12 -- Mime-Version: 1.0
7, 12 -- X-Sender: (Unverified)
7, 12 -- Message-Id: {p06110400c1db143d61b4-at-[206.69.208.22]}
7, 12 -- Date: Mon, 22 Jan 2007 19:24:26 -0600
7, 12 -- To: microscopy-at-microscopy.com
7, 12 -- From: zerfasp-at-ors.od.nih.gov (by way of MicroscopyListserver)
7, 12 -- Subject: viaWWW: Part for print processor
7, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: ptarq-at-evex.com
Date: Mon, 22 Jan 2007 19:35:08 -0600
Subject: [Microscopy] viaWWW: In memory of James Hillier

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Colleagues:


I regret to inform you of the passing of fellow electron microscopist, friend, neighbor and mentor, James Hillier. James Hillier of Princeton, NJ developed the first operational electron microscope in 1938, died Monday January 15, 2007 at University Medical Center at Princeton. He was 91. He was a senior scientist and vice president at the Radio Corporation of America and a director of RCA's David Sarnoff Research Center.



As a graduate student at the University of Toronto in 1938, under Professor Eli Franklin Burton, Mr. Hillier and fellow student Albert Prebus designed and built the first north American commercial electron microscope prototype.



After obtaining a master's degree and a doctorate in 1941 from the University of Toronto, he was immediately hired by The Radio Corporation of America in Camden, where he developed the first commercially available electron microscope, later developing RCA's videodisc, a precursor to the DVD. He rose through the corporate ranks to become executive vice president for research and engineering and senior scientist at RCA in 1969. He also served as the director of the David Sarnoff Research Center in West Windsor.



Mr. Hillier held 41 patents for devices and processes, the most significant for the electron microscope. After receiving a joint award from the American Public Health Association and the Albert and Mary Lasker Foundation for medical research in 1960, Mr. Hillier said to a reporter from Time magazine, "The electron microscope is like the monkey wrench on the garage wall; what you do with it is the important thing."



After retiring from RCA in 1977, he focused his attention on the National Inventors Hall of Fame in Akron, Ohio, and later The James Hillier Foundation for Science Education, dedicated to funding the college education of bright scientifically oriented students. In 1993, he established the James Hillier Foundation, which awards scholarships each year to science students from Brant County, Ontario.



Dr. Hillier received honorary degrees from New Jersey Institute of Technology and the University of Toronto. A public school in Brantford, Ontario, his birthplace, has been named in his honor. He was one of the first scientists elected to the National Inventors Hall of Fame in 1980. In 1997, he was made officer of the Order of Canada, the country's highest civilian honor.



His late wife owned and managed the Flower Basket in Princeton and later owned two additional Princeton flower shops. Both artists, Mr. Hillier made photo-realistic pastels and drawings while his wife was an abstract artist and an award-winning flower arranger. Son of the late James and Ethel Cooke Hillier, husband of the late Florence M. Bell Hillier, who died in 1992 after 55 years of marriage, father of the late William W. Hillier, who died in 2002, he is survived by his son, architect James Robert Hillier of New Hope, Pa.; sisters May Hillier of Brantford, Ontario, and Thelma Henshaw of Naples, Fla.; three grandsons; a granddaughter; and three great-grandchildren.



In lieu of flowers, memorial contributions may be made to the James Hillier Foundation, 34 Hill Ave., Brantford, Ontario, Canada, N3R 4HI. {http://comdir.bfree.on.ca/hillier/index.html} http://comdir.bfree.on.ca/hillier/index.html





Sincerely,

Peter P. Tarquinio











Evex Inc.,
Peter Tarquinio
857 State Road
Princeton, NJ 08540

Telephone - 609-252-9192
Fax - 609-252-9091
E-mail - {mailto:ptarq-at-evex.com} ptarq-at-evex.com
{http://www.evex.com} http://www.evex.com









==============================Original Headers==============================
30, 12 -- From zaluzec-at-microscopy.com Mon Jan 22 19:35:08 2007
30, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
30, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0N1Z6V3000858
30, 12 -- for {microscopy-at-microscopy.com} ; Mon, 22 Jan 2007 19:35:08 -0600
30, 12 -- Mime-Version: 1.0
30, 12 -- X-Sender: (Unverified)
30, 12 -- Message-Id: {p06110402c1db15429ed4-at-[206.69.208.22]}
30, 12 -- Date: Mon, 22 Jan 2007 19:35:06 -0600
30, 12 -- To: microscopy-at-microscopy.com
30, 12 -- From: {ptarq-at-evex.com} (by way of MicroscopyListserver)
30, 12 -- Subject: viaWWW: In memory of James Hillier
30, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: LarryZ-at-fai.us
Date: Tue, 23 Jan 2007 13:04:08 -0600
Subject: [Microscopy] Rebuilding Amray 1850 FE Gun

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

We are re-building and replacing the gun FE tip on an Amray 1850 FE.
Question: What is the tolerance, i.e., the distance from the suppressor
to the extractor? We have a sketch showing 0.765mm, but no plus or minus
tolerance. Does anyone know what the tolerance should be?

Thank you for your help.

Larry Zagorski
Sr. Metallurgist
FAI Materials Testing Laboratory
825 Chance Road
Marietta, GA 30066
770-928-1930
larryz-at-fai.us



==============================Original Headers==============================
5, 17 -- From LarryZ-at-fai.us Tue Jan 23 13:04:08 2007
5, 17 -- Received: from mx.cbeyond.com (mx.cbeyond.com [66.180.96.58])
5, 17 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0NJ478I032500
5, 17 -- for {Microscopy-at-microscopy.com} ; Tue, 23 Jan 2007 13:04:08 -0600
5, 17 -- Received: from [69.15.9.254] (port=3292 helo=[127.0.0.1])
5, 17 -- by mx.cbeyond.com with asmtp (Exim 4.34)
5, 17 -- id 1H9QwB-0006SF-EU
5, 17 -- for Microscopy-at-microscopy.com; Tue, 23 Jan 2007 14:04:07 -0500
5, 17 -- Message-ID: {45B660B7.4080308-at-fai.us}
5, 17 -- Date: Tue, 23 Jan 2007 14:23:35 -0500
5, 17 -- From: Larry Zagorski {LarryZ-at-fai.us}
5, 17 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
5, 17 -- MIME-Version: 1.0
5, 17 -- To: Microscopy-at-microscopy.com
5, 17 -- Subject: Rebuilding Amray 1850 FE Gun
5, 17 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
5, 17 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: jehrman-at-mta.ca
Date: Wed, 24 Jan 2007 12:26:27 -0600
Subject: [Microscopy] instructions: Vigor TW-1000 tweezer sharpener

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Greetings listers,

I was going through some cabinets and came across a Vigor TW-1000
tweezer sharpener (sold by Pella, apparently, a long time ago, judging
by the yellowing of the box) sans instructions. I think I've figured out
how the thing works, but would like a copy of the instructions to see if
I have all the parts and all those parts in the correct location. Is
there anyone out there in the ether who has a copy they could scan,
xerox, or otherwise send to me? The beer is on me and I'll show you the
fantastically high tides if you ever find yourself in Sackville, New
Brunswick (doesn't everybody wish they were here?).

Thank in advance,

Jim

--

James M. Ehrman
Digital Microscopy Facility
Mount Allison University
Sackville, NB E4L 1G7
CANADA

phone: 506-364-2519
fax: 506-364-2505
email: jehrman-at-mta.ca
www: http://www.mta.ca/dmf


==============================Original Headers==============================
8, 19 -- From jehrman-at-mta.ca Wed Jan 24 12:26:27 2007
8, 19 -- Received: from mailserv.mta.ca (mailserv.mta.ca [138.73.1.1])
8, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0OIQRTG004348
8, 19 -- for {Microscopy-at-MSA.Microscopy.com} ; Wed, 24 Jan 2007 12:26:27 -0600
8, 19 -- Received: from host-22-245.mta.ca ([138.73.22.245])
8, 19 -- by mailserv.mta.ca with esmtp (Exim 4.61)
8, 19 -- (envelope-from {jehrman-at-mta.ca} )
8, 19 -- id 1H9mpG-0000gG-0f
8, 19 -- for Microscopy-at-MSA.Microscopy.com; Wed, 24 Jan 2007 14:26:26 -0400
8, 19 -- Message-ID: {45B7A4C5.2090605-at-mta.ca}
8, 19 -- Date: Wed, 24 Jan 2007 14:26:13 -0400
8, 19 -- From: "James M. Ehrman" {jehrman-at-mta.ca}
8, 19 -- Organization: Mount Allison University
8, 19 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
8, 19 -- MIME-Version: 1.0
8, 19 -- To: Microscopy Listserv {Microscopy-at-MSA.Microscopy.com}
8, 19 -- Subject: instructions: Vigor TW-1000 tweezer sharpener
8, 19 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
8, 19 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: msteglic-at-mdanderson.org
Date: Wed, 24 Jan 2007 13:14:16 -0600
Subject:

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Did you try contacting Ted Pella? They may be able to help.

Mannie Steglich





jehrman-at-mta.ca

01/24/2007 12:30 PM
Please respond to jehrman




To:
msteglic-at-mdanderson.org
cc:






Greetings listers,

I was going through some cabinets and came across a Vigor TW-1000
tweezer sharpener (sold by Pella, apparently, a long time ago, judging
by the yellowing of the box) sans instructions. I think I've figured out
how the thing works, but would like a copy of the instructions to see if
I have all the parts and all those parts in the correct location. Is
there anyone out there in the ether who has a copy they could scan,
xerox, or otherwise send to me? The beer is on me and I'll show you the
fantastically high tides if you ever find yourself in Sackville, New
Brunswick (doesn't everybody wish they were here?).

Thank in advance,

Jim

--

James M. Ehrman
Digital Microscopy Facility
Mount Allison University
Sackville, NB E4L 1G7
CANADA

phone: 506-364-2519
fax: 506-364-2505
email: jehrman-at-mta.ca
www: http://www.mta.ca/dmf


==============================Original
Headers==============================
8, 19 -- From jehrman-at-mta.ca Wed Jan 24 12:26:27 2007
8, 19 -- Received: from mailserv.mta.ca (mailserv.mta.ca [138.73.1.1])
8, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8)
with ESMTP id l0OIQRTG004348
8, 19 -- for {Microscopy-at-MSA.Microscopy.com} ; Wed, 24 Jan
2007 12:26:27 -0600
8, 19 -- Received: from host-22-245.mta.ca ([138.73.22.245])
8, 19 -- by mailserv.mta.ca with esmtp (Exim 4.61)
8, 19 -- (envelope-from {jehrman-at-mta.ca} )
8, 19 -- id 1H9mpG-0000gG-0f
8, 19 -- for Microscopy-at-MSA.Microscopy.com; Wed, 24 Jan
2007 14:26:26 -0400
8, 19 -- Message-ID: {45B7A4C5.2090605-at-mta.ca}
8, 19 -- Date: Wed, 24 Jan 2007 14:26:13 -0400
8, 19 -- From: "James M. Ehrman" {jehrman-at-mta.ca}
8, 19 -- Organization: Mount Allison University
8, 19 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
8, 19 -- MIME-Version: 1.0
8, 19 -- To: Microscopy Listserv {Microscopy-at-MSA.Microscopy.com}
8, 19 -- Subject: instructions: Vigor TW-1000 tweezer sharpener
8, 19 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
8, 19 -- Content-Transfer-Encoding: 7bit
==============================End of -
Headers==============================




==============================Original Headers==============================
33, 19 -- From msteglic-at-mdanderson.org Wed Jan 24 13:14:16 2007
33, 19 -- Received: from mdairnmail2.mdacc.tmc.edu (mdairnmail2.mdacc.tmc.edu [143.111.251.124])
33, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0OJECol016834
33, 19 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 13:14:14 -0600
33, 19 -- Received: from ([143.111.84.152])
33, 19 -- by mdairnmail2.mdacc.tmc.edu with ESMTP id KP-BRCRK.60446901;
33, 19 -- Wed, 24 Jan 2007 13:13:36 -0600
33, 19 -- To: jehrman-at-mta.ca
33, 19 -- Cc: Microscopy-at-msa.microscopy.com
33, 19 -- Subject: Re: [Microscopy] instructions: Vigor TW-1000 tweezer sharpener
33, 19 -- MIME-Version: 1.0
33, 19 -- X-Mailer: Lotus Notes Release 5.0.11 July 24, 2002
33, 19 -- Message-ID: {OFE888E453.442A3D95-ON8625726D.006984DE-8625726D.00696FE3-at-mdacc.tmc.edu}
33, 19 -- From: msteglic-at-mdanderson.org
33, 19 -- Date: Wed, 24 Jan 2007 13:13:34 -0600
33, 19 -- X-MIMETrack: Serialize by Router on UTM-MAIL04A/HOU/UTMDACC(Release 5.0.13a |April 8, 2004) at
33, 19 -- 01/24/2007 01:13:36 PM,
33, 19 -- Serialize complete at 01/24/2007 01:13:36 PM
33, 19 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: jehrman-at-mta.ca
Date: Wed, 24 Jan 2007 13:14:33 -0600
Subject: [Microscopy] Thanks: instructions: Vigor TW-1000 tweezer sharpener

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Listers,

After the usual glut of "Out of Office" replies (gotta love those
people -- can't shoot 'em) I have a
contact who'll send me a copy of the instructions. Thanks for the other
offers.

Cheers,

Jim

--

James M. Ehrman
Digital Microscopy Facility
Mount Allison University
Sackville, NB E4L 1G7
CANADA

phone: 506-364-2519
fax: 506-364-2505
email: jehrman-at-mta.ca
www: http://www.mta.ca/dmf


==============================Original Headers==============================
8, 19 -- From jehrman-at-mta.ca Wed Jan 24 13:14:33 2007
8, 19 -- Received: from mailserv.mta.ca (mailserv.mta.ca [138.73.1.1])
8, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0OJEUbs016931
8, 19 -- for {Microscopy-at-MSA.Microscopy.com} ; Wed, 24 Jan 2007 13:14:32 -0600
8, 19 -- Received: from host-22-245.mta.ca ([138.73.22.245])
8, 19 -- by mailserv.mta.ca with esmtp (Exim 4.61)
8, 19 -- (envelope-from {jehrman-at-mta.ca} )
8, 19 -- id 1H9nZl-0006pj-14
8, 19 -- for Microscopy-at-MSA.Microscopy.com; Wed, 24 Jan 2007 15:14:29 -0400
8, 19 -- Message-ID: {45B7B008.5010209-at-mta.ca}
8, 19 -- Date: Wed, 24 Jan 2007 15:14:16 -0400
8, 19 -- From: "James M. Ehrman" {jehrman-at-mta.ca}
8, 19 -- Organization: Mount Allison University
8, 19 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
8, 19 -- MIME-Version: 1.0
8, 19 -- To: Microscopy Listserv {Microscopy-at-MSA.Microscopy.com}
8, 19 -- Subject: Thanks: instructions: Vigor TW-1000 tweezer sharpener
8, 19 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
8, 19 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Wed, 24 Jan 2007 13:44:05 -0600
Subject: [Microscopy] TEM: Fixation of mouse brain tissue

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Listers,

Does anyone have any pet techniques/fixatives for TEM processing of
brain tissue? Specifically mouse brain, if it matters. We just did a
run with less than stellar results----poor membranes, etc. We are now
trying a couple different things, but if anyone has some tried-and-true
tips, it might save us a ton of time. No immunolabeling, just
ultrastructure.

Some recipes call for picric acid. Can anyone tell me what purpose this
serves?

Thanks a heap, as usual!

Cheers,
Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original Headers==============================
7, 23 -- From TindallR-at-missouri.edu Wed Jan 24 13:44:05 2007
7, 23 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
7, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0OJi4V8007057
7, 23 -- for {microscopy-at-microscopy.com} ; Wed, 24 Jan 2007 13:44:05 -0600
7, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
7, 23 -- Wed, 24 Jan 2007 13:44:04 -0600
7, 23 -- x-mimeole: Produced By Microsoft Exchange V6.5
7, 23 -- Content-class: urn:content-classes:message
7, 23 -- MIME-Version: 1.0
7, 23 -- Content-Type: text/plain;
7, 23 -- charset="us-ascii"
7, 23 -- Subject: TEM: Fixation of mouse brain tissue
7, 23 -- Date: Wed, 24 Jan 2007 13:44:04 -0600
7, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D40-at-UM-XMAIL08.um.umsystem.edu}
7, 23 -- X-MS-Has-Attach:
7, 23 -- X-MS-TNEF-Correlator:
7, 23 -- Thread-Topic: TEM: Fixation of mouse brain tissue
7, 23 -- Thread-Index: Acc/8AJnCy/SuSXRTS2WyasNFvjxlQ==
7, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
7, 23 -- To: {microscopy-at-microscopy.com}
7, 23 -- X-OriginalArrivalTime: 24 Jan 2007 19:44:04.0365 (UTC) FILETIME=[00EE0BD0:01C73FF0]
7, 23 -- Content-Transfer-Encoding: 8bit
7, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0OJi4V8007057
==============================End of - Headers==============================




From: bozzola-at-siu.edu
Date: Wed, 24 Jan 2007 14:09:41 -0600
Subject: [Microscopy] Re: TEM: Fixation of mouse brain tissue

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

CNS tissue is best fixed by perfusion, without a doubt. Conventional
procedures (excision) hardly ever give very good results with CNS.

So, my first question is: did you use perfusion?

JB


} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


--
##############################################################
John J. Bozzola, Ph.D., Director
I.M.A.G.E. (Integrated Microscopy & Graphics Expertise)
750 Communications Drive - MC 4402
Southern Illinois University
Carbondale, IL 62901 U.S.A.
Phone: 618-453-3730
Email: bozzola-at-siu.edu
##############################################################

==============================Original Headers==============================
7, 20 -- From bozzola-at-siu.edu Wed Jan 24 14:09:41 2007
7, 20 -- Received: from abbmta1.siu.edu (abbmta1.siu.edu [131.230.254.205])
7, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0OK9e4L018531
7, 20 -- for {Microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 14:09:41 -0600
7, 20 -- Received: from [131.230.177.142] (ws177142.microscope.siu.edu [131.230.177.142])
7, 20 -- by abbmta1.siu.edu (Switch-3.1.11/Switch-3.1.11) with ESMTP id l0OK9cYV006561
7, 20 -- for {Microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 14:09:39 -0600 (CST)
7, 20 -- Mime-Version: 1.0
7, 20 -- X-Sender: bozzola-at-saluki-mail.siu.edu
7, 20 -- Message-Id: {p06110403c1dd6c9bbb5d-at-[131.230.177.142]}
7, 20 -- In-Reply-To: {200701241945.l0OJjCBs008740-at-ns.microscopy.com}
7, 20 -- References: {200701241945.l0OJjCBs008740-at-ns.microscopy.com}
7, 20 -- Date: Wed, 24 Jan 2007 14:09:37 -0600
7, 20 -- To: Microscopy-at-msa.microscopy.com
7, 20 -- From: "John J. Bozzola" {bozzola-at-siu.edu}
7, 20 -- Subject: Re: [Microscopy] TEM: Fixation of mouse brain tissue
7, 20 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
7, 20 -- X-Spam-Score: 0.00%
7, 20 -- X-MASF: 0.00%
7, 20 -- X-Whitelist: 0.00%
==============================End of - Headers==============================




From: mcauliff-at-umdnj.edu
Date: Wed, 24 Jan 2007 14:30:54 -0600
Subject: [Microscopy] Re: TEM: Fixation of mouse brain tissue

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Randy:

I use 2-3% glutaraldehyde in 0.1M phosphate buffer at room temp., pH
7.3-7.4, perfused via the left ventricle. While I have a small
peristaltic pump I have used gravity with excellent results. While some
will argue otherwise, one does need a fancy apparatus or high pressure,
I don't even measure the pressure. I use just a few ml of the same
buffer (without fancy additives) as a washout and at least 50-75 ml of
fix perfused in 5 minutes. In my experience most labs with less than
optimal results use 1. a tiny little needle in the ventricle 2. an
inadquate rate of flow and 3. way too much washout. Use an 18-20 g
needle (same inside diameter as the aorta!) inserted into the vent. then
have the assistant just nick the r. atrium, being careful not to go too
deep and cut the aorta .......-at-#$%^!! Going too far into the vent. can
be a problem so put a piece of cork or tygon tubing aroung the shaft of
the needle so you don't go too deep. Make everything fresh. Fix for a
few hours at room temp., multiple rinses in buffer, osmicate (reasonably
fresh osmium) thin slices/small pieces in the same buffer for several
hours, multiple buffer rinses, store in buffer if needed. If you must
keep orientation cut 200 micron Vibratome sections (before osmication)
so you can see the antomy after osmium turns everything black.
Dehydrate rapidly in graded ethanols (too much time in ethanols will
remove cytoplasm, Hyatt's book has the illustrations and refs. to prove
this), clear, infiltrate and embed. Uranyl acetate (after multiple
washes in sodium acetate to remove phosphates) will improve contrast. I
have not used K-ferricyanide to improve osmium staining.
Some labs use a little bit of picric acid as a protein coagulator I
suppose. Plenty of people get good results without it.

Geoff


TindallR-at-missouri.edu wrote:

} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


--
--
**********************************************
Geoff McAuliffe, Ph.D.
Neuroscience and Cell Biology
Robert Wood Johnson Medical School
675 Hoes Lane, Piscataway, NJ 08854
voice: (732)-235-4583; fax: -4029
mcauliff-at-umdnj.edu
**********************************************



==============================Original Headers==============================
10, 35 -- From mcauliff-at-umdnj.edu Wed Jan 24 14:30:54 2007
10, 35 -- Received: from zix01.umdnj.edu (zix01.UMDNJ.EDU [130.219.34.124])
10, 35 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0OKUrTU029852
10, 35 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 14:30:54 -0600
10, 35 -- Received: from zix01.umdnj.edu (ZixVPM [127.0.0.1])
10, 35 -- by Outbound.umdnj.edu (Proprietary) with ESMTP id 6A61E1C3348
10, 35 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 15:30:52 -0500 (EST)
10, 35 -- Received: from umdnj.edu (imail2.UMDNJ.EDU [130.219.34.139])
10, 35 -- by zix01.umdnj.edu (Proprietary) with ESMTP id AD722A7B9B
10, 35 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 15:30:50 -0500 (EST)
10, 35 -- Received: from ([130.219.34.131])
10, 35 -- by imail2.umdnj.edu with ESMTP id KP-BRCRF.70631915;
10, 35 -- Wed, 24 Jan 2007 15:30:29 -0500
10, 35 -- Received: from conversion-daemon.Polaris.umdnj.edu by Polaris.umdnj.edu
10, 35 -- (iPlanet Messaging Server 5.2 HotFix 2.02 (built Oct 21 2004))
10, 35 -- id {0JCE00I013A2PW-at-Polaris.umdnj.edu} (original mail from mcauliff-at-umdnj.edu)
10, 35 -- for microscopy-at-msa.microscopy.com; Wed, 24 Jan 2007 15:30:29 -0500 (EST)
10, 35 -- Received: from [127.0.0.1] ([10.138.2.123])
10, 35 -- by Polaris.umdnj.edu (iPlanet Messaging Server 5.2 HotFix 2.02 (built Oct 21
10, 35 -- 2004)) with ESMTP id {0JCE00AMD3ML7H-at-Polaris.umdnj.edu} ; Wed,
10, 35 -- 24 Jan 2007 15:30:21 -0500 (EST)
10, 35 -- Date: Wed, 24 Jan 2007 15:31:45 -0500
10, 35 -- From: Geoff McAuliffe {mcauliff-at-umdnj.edu}
10, 35 -- Subject: Re: [Microscopy] TEM: Fixation of mouse brain tissue
10, 35 -- In-reply-to: {200701241945.l0OJjGp8008888-at-ns.microscopy.com}
10, 35 -- To: TindallR-at-missouri.edu,
10, 35 -- MicroscopyListserver {microscopy-at-msa.microscopy.com}
10, 35 -- Message-id: {45B7C231.2020903-at-umdnj.edu}
10, 35 -- MIME-version: 1.0
10, 35 -- Content-type: text/plain; format=flowed; charset=ISO-8859-1
10, 35 -- Content-transfer-encoding: 7BIT
10, 35 -- X-Accept-Language: en-us, en
10, 35 -- User-Agent: Mozilla/5.0 (Windows; U; Windows NT 5.1; en-US; rv:1.7.2)
10, 35 -- Gecko/20040804 Netscape/7.2 (ax)
10, 35 -- References: {200701241945.l0OJjGp8008888-at-ns.microscopy.com}
==============================End of - Headers==============================




From: smythen-at-musc.edu
Date: Wed, 24 Jan 2007 18:40:45 -0600
Subject: [Microscopy] viaWWW: vigor tweezer sharpener

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both smythen-at-musc.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: smythen-at-musc.edu
Name: Nancy Smythe

Organization: Medical University of South Carolina

Title-Subject: [Filtered] vigor tweezer sharpener

Question: Good luck. I bought one about 6 years ago without instructions. But I do get results, so if you don't get the instructions let me know and I will tell you my techniques and you can share what you know and between the two of us we can maybe get it straight!

---------------------------------------------------------------------------

==============================Original Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Wed Jan 24 18:40:45 2007
6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0P0ehBW013013
6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 24 Jan 2007 18:40:45 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110400c1ddacf99ca5-at-[206.69.208.22]}
6, 12 -- Date: Wed, 24 Jan 2007 18:40:42 -0600
6, 12 -- To: microscopy-at-microscopy.com
6, 12 -- From: smythen-at-musc.edu (by way of MicroscopyListserver)
6, 12 -- Subject: viaWWW: vigor tweezer sharpener
6, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: tivol-at-caltech.edu
Date: Wed, 24 Jan 2007 19:02:21 -0600
Subject: [Microscopy] Re: viaWWW: vigor tweezer sharpener

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


On Jan 24, 2007, at 4:40 PM, smythen-at-musc.edu wrote:

} Good luck. I bought one about 6 years ago without instructions. But
} I do get results, so if you don't get the instructions let me know and
} I will tell you my techniques and you can share what you know and
} between the two of us we can maybe get it straight!
}
Dear Jim & Nancy,
Could you please post your techniques to the list? TIA.
Yours,
Bill Tivol, PhD
EM Scientist and Manager
Cryo-Electron Microscopy Facility
Broad Center, Mail Code 114-96
California Institute of Technology
Pasadena CA 91125
(626) 395-8833
tivol-at-caltech.edu


==============================Original Headers==============================
4, 24 -- From tivol-at-caltech.edu Wed Jan 24 19:02:20 2007
4, 24 -- Received: from outgoing-mail.its.caltech.edu (outgoing-mail.its.caltech.edu [131.215.239.19])
4, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0P12K85024375
4, 24 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 19:02:20 -0600
4, 24 -- Received: from earth-dog.caltech.edu (earth-dog [192.168.1.3])
4, 24 -- by earth-ox-postvirus (Postfix) with ESMTP
4, 24 -- id EEBAF10A099; Wed, 24 Jan 2007 17:02:19 -0800 (PST)
4, 24 -- Received: from [192.168.159.233] (jpix-01.caltech.edu [131.215.2.133])
4, 24 -- (using TLSv1 with cipher RC4-SHA (128/128 bits))
4, 24 -- (No client certificate requested)
4, 24 -- by water-ox.its.caltech.edu (Postfix) with ESMTP
4, 24 -- id 093E52EEEE; Wed, 24 Jan 2007 17:02:19 -0800 (PST)
4, 24 -- Mime-Version: 1.0 (Apple Message framework v624)
4, 24 -- In-Reply-To: {200701250040.l0P0et4L013176-at-ns.microscopy.com}
4, 24 -- References: {200701250040.l0P0et4L013176-at-ns.microscopy.com}
4, 24 -- Content-Type: text/plain; charset=US-ASCII; format=flowed
4, 24 -- Message-Id: {c986f3c20b4cc3e981ac867fdf0820cb-at-caltech.edu}
4, 24 -- Content-Transfer-Encoding: 7bit
4, 24 -- From: Bill Tivol {tivol-at-caltech.edu}
4, 24 -- Subject: Re: [Microscopy] viaWWW: vigor tweezer sharpener
4, 24 -- Date: Wed, 24 Jan 2007 17:10:13 -0800
4, 24 -- To: microscopy-at-msa.microscopy.com, smythen-at-musc.edu
4, 24 -- X-Mailer: Apple Mail (2.624)
4, 24 -- X-Spam-Scanned: at Caltech-ITS on earth-dog by amavisd-2.3.3
==============================End of - Headers==============================




From: r.sims-at-auckland.ac.nz
Date: Wed, 24 Jan 2007 19:30:07 -0600
Subject: [Microscopy] 840 vacuum problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


Here's one for all the 840 afficianados:

In the 'ready' state all is well, with all the valve lights as they should be, and a
pressure of around 10 - 5 in the chamber (there are 3 WDS spectros leaking away
into the chamber) and around 5 x 10 - 6 in the gun.

When I push the GUN button to vent the gun, the lights do what they should, including
the opening of LV1, but the gun pressure comes not up to atmospheric but to about
100 mbar, and the pressure in the vacuum ballast tank starts climbing, until the PiG3
light comes on and the system devotes itself, as it should, to re-evacuating the ballast
tank.

Am I correct in guessing that V1B, the gun pumpdown valve, is not closing properly
and is allowing the atmosphere to leak via LV1, the gun, and V1B, into the ballast
tank?

I would welcome support (or, I suppose, otherwise) in this guess before I start ripping
things apart.

There is a time constraint here, as we have the NZ EM conference in 10 days time
and there's to be a teaching workshop involving the 840.

Help!

cheers

rtch

--
Ritchie Sims Ph D Phone : 64 9 3737599 ext 87713
Microanalyst Fax : 64 9 3737435
Department of Geology email : r.sims-at-auckland.ac.nz
The University of Auckland
Private Bag 92019
Auckland
New Zealand


==============================Original Headers==============================
12, 27 -- From r.sims-at-auckland.ac.nz Wed Jan 24 19:30:06 2007
12, 27 -- Received: from harpo.itss.auckland.ac.nz (harpo.itss.auckland.ac.nz [130.216.190.13])
12, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0P1U513003330
12, 27 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 19:30:06 -0600
12, 27 -- Received: from localhost (localhost.localdomain [127.0.0.1])
12, 27 -- by harpo.itss.auckland.ac.nz (Postfix) with ESMTP id 5A6193485B
12, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 25 Jan 2007 14:30:04 +1300 (NZDT)
12, 27 -- Received: from harpo.itss.auckland.ac.nz ([127.0.0.1])
12, 27 -- by localhost (smtpc.itss.auckland.ac.nz [127.0.0.1]) (amavisd-new, port 10024)
12, 27 -- with ESMTP id 23181-05 for {microscopy-at-msa.microscopy.com} ;
12, 27 -- Thu, 25 Jan 2007 14:30:04 +1300 (NZDT)
12, 27 -- Received: from rs (r.sims.glg.auckland.ac.nz [130.216.59.18])
12, 27 -- by harpo.itss.auckland.ac.nz (Postfix) with ESMTP id 317AE3485A
12, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 25 Jan 2007 14:30:03 +1300 (NZDT)
12, 27 -- From: "Ritchie Sims" {r.sims-at-auckland.ac.nz}
12, 27 -- To: microscopy-at-msa.microscopy.com
12, 27 -- Date: Thu, 25 Jan 2007 14:30:24 +1300
12, 27 -- MIME-Version: 1.0
12, 27 -- Subject: 840 vacuum problem
12, 27 -- Message-ID: {45B8BF00.12939.1379784-at-localhost}
12, 27 -- Priority: normal
12, 27 -- In-reply-to: {200701250103.l0P135c6026032-at-ns.microscopy.com}
12, 27 -- X-mailer: Pegasus Mail for Windows (4.21c)
12, 27 -- Content-type: text/plain; charset=US-ASCII
12, 27 -- Content-transfer-encoding: 7BIT
12, 27 -- Content-description: Mail message body
12, 27 -- X-Virus-Scanned: by amavisd-new at mailhost.auckland.ac.nz
==============================End of - Headers==============================




From: jzheng-at-uci.edu
Date: Wed, 24 Jan 2007 20:38:58 -0600
Subject: [Microscopy] Gatan ion mill

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi,
I need an easy and quick method to clean the anode in Gatan dual mill.
Could you pleasse help? Thanks, Jeffrey



==============================Original Headers==============================
3, 20 -- From jzheng-at-uci.edu Wed Jan 24 20:38:58 2007
3, 20 -- Received: from relay2.es.uci.edu (relay2.es.uci.edu [128.200.80.28])
3, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0P2cv4C015265
3, 20 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 20:38:57 -0600
3, 20 -- Received: from webmail.uci.edu (webmail-http3.es.uci.edu [128.200.80.166])
3, 20 -- by relay2.es.uci.edu (8.13.1/8.13.1) with ESMTP id l0P2cuTO018259
3, 20 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 18:38:57 -0800
3, 20 -- Received: from 128.195.177.193
3, 20 -- (SquirrelMail authenticated user jzheng)
3, 20 -- by webmail.uci.edu with HTTP;
3, 20 -- Wed, 24 Jan 2007 18:38:57 -0800 (PST)
3, 20 -- Message-ID: {4052.128.195.177.193.1169692737.squirrel-at-webmail.uci.edu}
3, 20 -- Date: Wed, 24 Jan 2007 18:38:57 -0800 (PST)
3, 20 -- Subject: Gatan ion mill
3, 20 -- From: jzheng-at-uci.edu
3, 20 -- To: microscopy-at-msa.microscopy.com
3, 20 -- User-Agent: SquirrelMail/1.4.9a
3, 20 -- MIME-Version: 1.0
3, 20 -- Content-Type: text/plain;charset=iso-8859-1
3, 20 -- Content-Transfer-Encoding: 8bit
==============================End of - Headers==============================




From: peter.heimann-at-uni-bielefeld.de
Date: Thu, 25 Jan 2007 02:46:56 -0600
Subject: [Microscopy] Re: TEM: Fixation of mouse brain tissue

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Randy,
Geoff McCauliff's tips are my experiences too.
a minute comment: use gravity perfusion; a wonderful & detailed
instruction is given in:

Forssmann, W.G., Ito, S., Weihe, E., Aoki, A., Dym, M. and Fawcett,
D.W. (1977) Anat Rec 188, 307-14.

a small but important tip: Filtrate ALL solutions used for perfusion
through a 1 (or 2) mikrometer filter before use!

If you ever have seen a capillary in cross section and measured its
diameter, you know why.

Use the two resp. tree-step technique described in the paper, in my
experience perfusion with the "rinse" solution for 20-30 seconds
followed by a single or two-step perfusion (3-5 minutes) with a mixture
of freshly (!) prepared form- and glutaraldehyde (the latter prepared
by the manufacturer according to the Anderson-technique) is sufficient
for an adult wildtype mouse.

good luck,
peter heimann


==============================Original Headers==============================
7, 31 -- From peter.heimann-at-uni-bielefeld.de Thu Jan 25 02:46:56 2007
7, 31 -- Received: from smtp-relay.uni-bielefeld.de (mux2-unibi-smtp.hrz.uni-bielefeld.de [129.70.204.73])
7, 31 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0P8ktmQ031325
7, 31 -- for {microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 02:46:56 -0600
7, 31 -- Received: from pmxchannel-daemon.mux2-unibi-smtp.hrz.uni-bielefeld.de by
7, 31 -- mux2-unibi-smtp.hrz.uni-bielefeld.de
7, 31 -- (Sun Java System Messaging Server 6.2-8.02 (built Dec 21 2006))
7, 31 -- id {0JCF003011Q6FR00-at-mux2-unibi-smtp.hrz.uni-bielefeld.de} for
7, 31 -- microscopy-at-microscopy.com; Thu, 25 Jan 2007 09:46:54 +0100 (CET)
7, 31 -- Received: from [129.70.82.133] by mux2-unibi-smtp.hrz.uni-bielefeld.de
7, 31 -- (Sun Java System Messaging Server 6.2-8.02 (built Dec 21 2006))
7, 31 -- with ESMTP id {0JCF000A41Q5BH80-at-mux2-unibi-smtp.hrz.uni-bielefeld.de} for
7, 31 -- microscopy-at-microscopy.com; Thu, 25 Jan 2007 09:46:54 +0100 (CET)
7, 31 -- Date: Thu, 25 Jan 2007 09:46:52 +0100
7, 31 -- From: PETER HEIMANN {peter.heimann-at-uni-bielefeld.de}
7, 31 -- Subject: Re: TEM: Fixation of mouse brain tissue
7, 31 -- In-reply-to: {200701241944.l0OJiRh8007561-at-ns.microscopy.com}
7, 31 -- To: microscopy-at-microscopy.com
7, 31 -- Reply-to: peter.heimann-at-uni-bielefeld.de
7, 31 -- Message-id: {45B86E7C.20404-at-UNI-BIELEFELD.DE}
7, 31 -- Organization: Uni Bielefeld
7, 31 -- MIME-version: 1.0
7, 31 -- Content-type: text/plain; charset=ISO-8859-1; format=flowed
7, 31 -- Content-transfer-encoding: 7BIT
7, 31 -- X-Accept-Language: en-us, en
7, 31 -- X-EnvFrom: peter.heimann-at-uni-bielefeld.de
7, 31 -- X-PMX-Version: 5.2.1.279297, Antispam-Engine: 2.5.0.283055,
7, 31 -- Antispam-Data: 2007.1.25.3433, pmx1.hrz.uni-bielefeld.de
7, 31 -- X-Connecting-IP: 127.0.0.1
7, 31 -- References: {200701241944.l0OJiRh8007561-at-ns.microscopy.com}
7, 31 -- User-Agent: Mozilla Thunderbird 1.0.2 (Windows/20050317)
==============================End of - Headers==============================




From: cammer-at-aecom.yu.edu
Date: Thu, 25 Jan 2007 10:17:50 -0600
Subject: [Microscopy] color CCD Zeiss AxioCam MRc problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi

Vacuum systems have high and low speed pumping areas their problems show up
as set out below, high (1) and low (2).

1. If a reasonable vacuum is held in the high vacuum area but the rotary
pump or the backing tank show a considerable decrease in their vacuum level,
the problem is in an area with close proximity to a main pumping line

2. If a poor vacuum is held in the high vacuum area and the rotary pump or
the backing tank also show a considerable decrease in its vacuum level, the
problem is in an area with some distance (vacuum wise) to a main pumping
line.

Always ask what did I do last and it seems that here lies the problem, first
reverse the action does the system recover? It would suggest a leak
between the gun, trying to reach atmospheric, and the column trying to
retain its high vacuum.

Good luck

Steve Chapman
Senior Consultant Protrain
For electron microscopy consultancy and training world wide
Tel +44 1280 816512 Fax +44 1280 814007
Mobile +44 7711 606967 Web www.emcourses.com


----- Original Message -----
X-from: {r.sims-at-auckland.ac.nz}
To: {protrain-at-emcourses.com}
Sent: Thursday, January 25, 2007 1:30 AM

We found that the red (typically rhodamine) images taken with our
Zeiss Axiocam MRc are fuzzy. We looked at the blue and found this a
problem too, but since in fluorescence people typically use the blue
just as a marker for nucleii rather than as stain for fine structure,
they didn't notice this problem. (Illustrations at
http://www.aecom.yu.edu/aif/temp/axiocamproblem/ ).

This made us worry about the sharpness of images of histological
stained samples, such as H&E.

First we checked that this isn't a problem due to the fluorescent
filter sets. To do this, we filtered the light to be red, green or
blue at the halogen lamp using RGB filters removed from an old 35mm
film recorder and imaged the sample with no filters in the light path
between the sample and the camera. The result was that the red and
blue images were very fuzzy.

Shooting in the b/w mode solves this problem, but then the images are
not color. Also, this may solve the problem for fluorescence, but
what about brightfield or other techniques that involve colors?

Can anyone explain why this happens and how to fix it? Is this a
problem with our sytem uniquely or is this a problem with the Axiocam
in general?

Thanks!

-Michael
____________________________________________________________________________
Michael Cammer Analytical Imaging Facility Albert Einstein Coll. of Med.
URL: http://www.aecom.yu.edu/aif/


==============================Original Headers==============================
8, 29 -- From cammer-at-aecom.yu.edu Thu Jan 25 10:17:49 2007
8, 29 -- Received: from mx1.aecom.yu.edu (mx1.aecom.yu.edu [129.98.1.51])
8, 29 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PGHnp3030006
8, 29 -- for {microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 10:17:49 -0600
8, 29 -- Received: from draco.aecom.yu.edu (draco.aecom.yu.edu [129.98.1.160])
8, 29 -- by mx1.aecom.yu.edu (Postfix) with ESMTP id 5A9FA9F00EF;
8, 29 -- Thu, 25 Jan 2007 11:17:49 -0500 (EST)
8, 29 -- Received: from draco.aecom.yu.edu (unknown [127.0.0.1])
8, 29 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id 8D5C38B4006;
8, 29 -- Thu, 25 Jan 2007 11:06:04 -0500 (EST)
8, 29 -- X-AuditID: 816201a0-a13c8bb000007022-82-45b8d56c712a
8, 29 -- Received: from post.aecom.yu.edu (post.aecom.yu.edu [129.98.1.100])
8, 29 -- by draco.aecom.yu.edu (Symantec Mail Security) with ESMTP id 4E7FE718002;
8, 29 -- Thu, 25 Jan 2007 11:06:04 -0500 (EST)
8, 29 -- Received: from AIF3.aecom.yu.edu (aif3.aif.aecom.yu.edu [129.98.80.70])
8, 29 -- (using TLSv1 with cipher DHE-RSA-AES256-SHA (256/256 bits))
8, 29 -- (No client certificate requested)
8, 29 -- by post.aecom.yu.edu (Postfix) with ESMTP id C3E7E26;
8, 29 -- Thu, 25 Jan 2007 11:17:43 -0500 (EST)
8, 29 -- Message-Id: {7.0.1.0.2.20070125111540.04f87920-at-aecom.yu.edu}
8, 29 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
8, 29 -- Date: Thu, 25 Jan 2007 11:17:43 -0500
8, 29 -- To: microscopy-at-microscopy.com
8, 29 -- From: Michael Cammer {cammer-at-aecom.yu.edu}
8, 29 -- Subject: color CCD Zeiss AxioCam MRc problem
8, 29 -- Cc: yadeng-at-aecom.yu.edu, rleonard-at-aecom.yu.edu, jmorreale-at-zeiss.com
8, 29 -- Mime-Version: 1.0
8, 29 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
8, 29 -- X-Brightmail-Tracker: AAAAAA==
==============================End of - Headers==============================




From: wim5-at-lehigh.edu
Date: Thu, 25 Jan 2007 11:05:58 -0600
Subject: [Microscopy] Re: 840 vacuum problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

If you use the "VENT" button rather than the "GUN VENT" to vent the
whole column and the ballast pressure holds up, your problem is most
likely that the V7 valve O-ring is leaking between the gun and
specimen chambers. This would also eliminate V1B as a possibility.

Unfortunately, you will have to tear down the column to get to the V7
O-ring. The good news is, this valve should only present a problem
when doing a gun vent. To avoid introducing more serious problems I
would just use the VENT button temporarily when needed and postpone
the repair until after your conference.

Regards,
Bill

On 1/24/07, r.sims-at-auckland.ac.nz {r.sims-at-auckland.ac.nz} wrote:
}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
}
} Here's one for all the 840 afficianados:
}
} In the 'ready' state all is well, with all the valve lights as they should be, and a
} pressure of around 10 - 5 in the chamber (there are 3 WDS spectros leaking away
} into the chamber) and around 5 x 10 - 6 in the gun.
}
} When I push the GUN button to vent the gun, the lights do what they should, including
} the opening of LV1, but the gun pressure comes not up to atmospheric but to about
} 100 mbar, and the pressure in the vacuum ballast tank starts climbing, until the PiG3
} light comes on and the system devotes itself, as it should, to re-evacuating the ballast
} tank.
}
} Am I correct in guessing that V1B, the gun pumpdown valve, is not closing properly
} and is allowing the atmosphere to leak via LV1, the gun, and V1B, into the ballast
} tank?
}
} I would welcome support (or, I suppose, otherwise) in this guess before I start ripping
} things apart.
}
} There is a time constraint here, as we have the NZ EM conference in 10 days time
} and there's to be a teaching workshop involving the 840.
}
} Help!
}
} cheers
}
} rtch
}
} --
} Ritchie Sims Ph D Phone : 64 9 3737599 ext 87713
} Microanalyst Fax : 64 9 3737435
} Department of Geology email : r.sims-at-auckland.ac.nz
} The University of Auckland
} Private Bag 92019
} Auckland
} New Zealand
}
}
} ==============================Original Headers==============================
} 12, 27 -- From r.sims-at-auckland.ac.nz Wed Jan 24 19:30:06 2007
} 12, 27 -- Received: from harpo.itss.auckland.ac.nz (harpo.itss.auckland.ac.nz [130.216.190.13])
} 12, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0P1U513003330
} 12, 27 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007 19:30:06 -0600
} 12, 27 -- Received: from localhost (localhost.localdomain [127.0.0.1])
} 12, 27 -- by harpo.itss.auckland.ac.nz (Postfix) with ESMTP id 5A6193485B
} 12, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 25 Jan 2007 14:30:04 +1300 (NZDT)
} 12, 27 -- Received: from harpo.itss.auckland.ac.nz ([127.0.0.1])
} 12, 27 -- by localhost (smtpc.itss.auckland.ac.nz [127.0.0.1]) (amavisd-new, port 10024)
} 12, 27 -- with ESMTP id 23181-05 for {microscopy-at-msa.microscopy.com} ;
} 12, 27 -- Thu, 25 Jan 2007 14:30:04 +1300 (NZDT)
} 12, 27 -- Received: from rs (r.sims.glg.auckland.ac.nz [130.216.59.18])
} 12, 27 -- by harpo.itss.auckland.ac.nz (Postfix) with ESMTP id 317AE3485A
} 12, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 25 Jan 2007 14:30:03 +1300 (NZDT)
} 12, 27 -- From: "Ritchie Sims" {r.sims-at-auckland.ac.nz}
} 12, 27 -- To: microscopy-at-msa.microscopy.com
} 12, 27 -- Date: Thu, 25 Jan 2007 14:30:24 +1300
} 12, 27 -- MIME-Version: 1.0
} 12, 27 -- Subject: 840 vacuum problem
} 12, 27 -- Message-ID: {45B8BF00.12939.1379784-at-localhost}
} 12, 27 -- Priority: normal
} 12, 27 -- In-reply-to: {200701250103.l0P135c6026032-at-ns.microscopy.com}
} 12, 27 -- X-mailer: Pegasus Mail for Windows (4.21c)
} 12, 27 -- Content-type: text/plain; charset=US-ASCII
} 12, 27 -- Content-transfer-encoding: 7BIT
} 12, 27 -- Content-description: Mail message body
} 12, 27 -- X-Virus-Scanned: by amavisd-new at mailhost.auckland.ac.nz
} ==============================End of - Headers==============================
}


--
William J Mushock
Lehigh University
Whitaker Laboratory
5 East Packer Ave.
Bethlehem, PA 18015
(610)-758-4283

==============================Original Headers==============================
6, 26 -- From bill.mushock-at-gmail.com Thu Jan 25 11:05:58 2007
6, 26 -- Received: from wx-out-0506.google.com (wx-out-0506.google.com [66.249.82.238])
6, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PH5vu1009442
6, 26 -- for {Microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 11:05:58 -0600
6, 26 -- Received: by wx-out-0506.google.com with SMTP id s16so560115wxc
6, 26 -- for {Microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 09:05:57 -0800 (PST)
6, 26 -- DomainKey-Signature: a=rsa-sha1; c=nofws;
6, 26 -- d=gmail.com; s=beta;
6, 26 -- h=received:message-id:date:from:sender:to:subject:in-reply-to:mime-version:content-type:content-transfer-encoding:content-disposition:references:x-google-sender-auth;
6, 26 -- b=h0j2UNl6VAYGJsJMfcp6v4S6qjd3GX2nVyxcRD1aMv1iWfUnmyVZqRueusr+RCakUcYzcwEYfi47DV+xGNb0YL9S6D4qj9Few4Tc2PwMrJ9s2u727BsUP1Ii9OjoJiXwOW3INQKMRfxu+cPJrH6LXK2Vc1mvqQLQbs4W96Sxerc=
6, 26 -- Received: by 10.70.52.5 with SMTP id z5mr4060864wxz.1169744757551;
6, 26 -- Thu, 25 Jan 2007 09:05:57 -0800 (PST)
6, 26 -- Received: by 10.70.36.7 with HTTP; Thu, 25 Jan 2007 09:05:57 -0800 (PST)
6, 26 -- Message-ID: {61452d440701250905w34ff2e77j4a56c18c6f15e33c-at-mail.gmail.com}
6, 26 -- Date: Thu, 25 Jan 2007 12:05:57 -0500
6, 26 -- From: "William J Mushock" {wim5-at-lehigh.edu}
6, 26 -- Sender: bill.mushock-at-gmail.com
6, 26 -- To: Microscopy-at-microscopy.com
6, 26 -- Subject: Re: [Microscopy] 840 vacuum problem
6, 26 -- In-Reply-To: {200701250132.l0P1Whdp007019-at-ns.microscopy.com}
6, 26 -- MIME-Version: 1.0
6, 26 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
6, 26 -- Content-Transfer-Encoding: 7bit
6, 26 -- Content-Disposition: inline
6, 26 -- References: {200701250132.l0P1Whdp007019-at-ns.microscopy.com}
6, 26 -- X-Google-Sender-Auth: 7b3ead8b693005a4
==============================End of - Headers==============================




From: ani.issaian-at-csun.edu
Date: Thu, 25 Jan 2007 11:14:16 -0600
Subject: [Microscopy]

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi everyone,

I am trying to order photographic paper racks which are used inside a dryer box after printing. If anyone knows were I can buy them, please let me know.
Thanks,


ANI M ISSAIAN
Electron Microscopy Facility Manager
California State University, Northridge
18111 Nordhoff street, Northridge, CA 91330-8303
Biology dept., MC 8303, Room CS2205
Phone: (818) 677-3383
Fax (818) 677-2034

==============================Original Headers==============================
4, 28 -- From ani.issaian-at-csun.edu Thu Jan 25 11:14:15 2007
4, 28 -- Received: from plover.csun.edu (plover.csun.edu [130.166.1.24])
4, 28 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PHEDfV020498
4, 28 -- for {Microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 11:14:14 -0600
4, 28 -- Received: from puffin.csun.edu (puffin.csun.edu [130.166.1.21])
4, 28 -- by plover.csun.edu (MOS 3.8.2-GA)
4, 28 -- with ESMTP id DMG61952
4, 28 -- for {Microscopy-at-microscopy.com} ;
4, 28 -- Thu, 25 Jan 2007 09:14:11 -0800 (PST)
4, 28 -- Received: from cuckoo.csun.edu (cuckoo.csun.edu [130.166.1.230])
4, 28 -- by puffin.csun.edu (MOS 3.7.5a-GA)
4, 28 -- with ESMTP id FJW26576
4, 28 -- for {Microscopy-at-microscopy.com} ;
4, 28 -- Thu, 25 Jan 2007 09:14:10 -0800 (PST)
4, 28 -- Received: (from cuckoo.csun.edu [130.166.114.202])
4, 28 -- by cuckoo.csun.edu (MOS 3.8.2-GA)
4, 28 -- with HTTPS/1.1 id ARZ65214 (AUTH ami24015);
4, 28 -- Thu, 25 Jan 2007 09:14:10 -0800 (PST)
4, 28 -- From: Ani M Issaian {ani.issaian-at-csun.edu}
4, 28 -- To: Microscopy-at-microscopy.com
4, 28 -- Reply-To: ani.issaian-at-csun.edu
4, 28 -- X-Mailer: Mirapoint Webmail Direct 3.8.2-GA
4, 28 -- MIME-Version: 1.0
4, 28 -- Content-Type: text/plain; charset=us-ascii
4, 28 -- Content-Transfer-Encoding: 7bit
4, 28 -- Message-Id: {20070125091410.ARZ65214-at-cuckoo.csun.edu}
4, 28 -- Date: Thu, 25 Jan 2007 09:14:10 -0800 (PST)
4, 28 -- X-Junkmail-Whitelist: YES (by domain whitelist at plover.csun.edu)
==============================End of - Headers==============================




From: donovan-at-uoregon.edu
Date: Thu, 25 Jan 2007 11:24:37 -0600
Subject: [Microscopy] Nitrogen free epoxy?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

All,
Can anyone tell me of a source for a nitrogen free epoxy? I realize
that may be a contradiction in terms, but I need to find a mounting
media that can hold a water soluble polymer coated wire for cross
sectioning for subsequent nitrogen analysis by EPMA.

Thanks!
john

John J. Donovan donovan-at-uoregon.edu
University of Oregon (541) 346-4632 (office)
1260 Franklin Blvd (541) 346-4655 (probe)
Eugene, OR (541) 346-4778 (SEM)
97403-1272 (541) 346-4692 (FAX)

Lab Web: http://epmalab.uoregon.edu/
EPMA (SX100)Schedule: http://sweetwater.uoregon.edu/sx100
EPMA (SX50) Schedule: http://sweetwater.uoregon.edu/epma
SEM (Ultra) Schedule: http://sweetwater.uoregon.edu/zeiss
SEM (Quanta) Schedule: http://sweetwater.uoregon.edu/sem
Remote Access: http://epmalab.uoregon.edu/howto.htm
Personal: http://www.uoregon.edu/~donovan/

"Aristotle presented with a moon rock would have no difficulty in
discerning it as an object not fundamentally different from
terrestrial materials. So much for Thomas Kuhn."

"I'd rather be uncertain and wrong, than be certain and wrong."




==============================Original Headers==============================
9, 20 -- From donovan-at-uoregon.edu Thu Jan 25 11:24:37 2007
9, 20 -- Received: from smtp.uoregon.edu (mserv2.uoregon.edu [128.223.142.41])
9, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PHOadg031693
9, 20 -- for {Microscopy-at-Microscopy.Com} ; Thu, 25 Jan 2007 11:24:36 -0600
9, 20 -- Received: from Probev.uoregon.edu (probev.uoregon.edu [128.223.92.75])
9, 20 -- (authenticated bits=0)
9, 20 -- by smtp.uoregon.edu (8.13.8/8.13.8) with ESMTP id l0PHOahW007001
9, 20 -- (version=TLSv1/SSLv3 cipher=DHE-RSA-AES256-SHA bits=256 verify=NOT)
9, 20 -- for {Microscopy-at-Microscopy.Com} ; Thu, 25 Jan 2007 09:24:36 -0800
9, 20 -- Message-Id: {7.0.1.0.0.20070125092420.039399b8-at-uoregon.edu}
9, 20 -- Message-Id: {7.0.1.0.0.20070125091904.038e7cd0-at-uoregon.edu}
9, 20 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
9, 20 -- Date: Thu, 25 Jan 2007 09:24:32 -0800
9, 20 -- To: Microscopy-at-Microscopy.Com
9, 20 -- From: John Donovan {donovan-at-uoregon.edu}
9, 20 -- Subject: Nitrogen free epoxy?
9, 20 -- Mime-Version: 1.0
9, 20 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
9, 20 -- X-Virus-Scanned: ClamAV 0.88.7/2489/Thu Jan 25 04:35:24 2007 on mserv2
9, 20 -- X-Virus-Status: Clean
==============================End of - Headers==============================




From: pli-at-dal.ca
Date: Thu, 25 Jan 2007 12:48:53 -0600
Subject: [Microscopy] WTB: an used Acer Altos computer for TEM

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Our current computer (Acer Altos 11000) that controls FEI's Tecnai 12 TEM is
having serious problems. I am looking for an used Acer Altos 11000, or
12000, or 21000 to replace our current one. If you have one for sale or if
you would like to upgrade yours, could you please let me know? I will pay
for the computer, handling and shipping cost.

Thank you,
Ping

--
Ping Li, Ph.D.
Director, Scientific Imaging Suite
Department of Biology
Dalhousie University
Halifax, NS B3H 4J1
Canada

Tel: 902-494-3309
Fax: 902-494-3736
E-mail: Ping.Li-at-Dal.Ca



==============================Original Headers==============================
6, 22 -- From pli-at-dal.ca Thu Jan 25 12:48:52 2007
6, 22 -- Received: from SXT-SM-1.UCIS.Dal.Ca (SXT-SM-1.UCIS.Dal.Ca [129.173.5.72])
6, 22 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PImow6012020
6, 22 -- for {Microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 12:48:51 -0600
6, 22 -- Received: from SISPL (PLi-7.Biology.Dal.Ca [129.173.34.142])
6, 22 -- by SXT-SM-1.UCIS.Dal.Ca (8.12.11/8.12.10) with SMTP id l0PImit1011138
6, 22 -- for {Microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 14:48:49 -0400
6, 22 -- Message-ID: {006d01c740b1$80797120$8e22ad81-at-SISPL}
6, 22 -- From: "Ping Li" {pli-at-dal.ca}
6, 22 -- To: {Microscopy-at-microscopy.com}
6, 22 -- Subject: WTB: an used Acer Altos computer for TEM
6, 22 -- Date: Thu, 25 Jan 2007 14:49:05 -0400
6, 22 -- MIME-Version: 1.0
6, 22 -- Content-Type: text/plain;
6, 22 -- format=flowed;
6, 22 -- charset="iso-8859-1";
6, 22 -- reply-type=original
6, 22 -- Content-Transfer-Encoding: 7bit
6, 22 -- X-Priority: 3
6, 22 -- X-MSMail-Priority: Normal
6, 22 -- X-Mailer: Microsoft Outlook Express 6.00.2900.3028
6, 22 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: jehrman-at-mta.ca
Date: Thu, 25 Jan 2007 13:25:22 -0600
Subject: [Microscopy] JEOL 5xxx photo CRT

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi all,

With somewhat mixed emotions, we finally decommissioned our darkroom and
transferred
the (nearly new) 4x5 setup to the Fine Arts Department. This move was
prompted by
the upgrade of our JEOL 5600 computer system, which essentially disabled
the photo system
(not that we ever used it). On a recent service visit I had the engineer
remove the system entirely.
JEOL Canada doesn't have any interest in retrieving it, so on the off
chance that somebody desperately
needs a replacement, I'm offering it to anyone who will pay shipping for
the thing. Let me know
relatively quickly - it's probably destined for the dumpster if I can't
find someone to give it to.
There's a new Polaroid back on it, if that makes it any more attractive.

Cheers,

Jim

--

James M. Ehrman
Digital Microscopy Facility
Mount Allison University
Sackville, NB E4L 1G7
CANADA

phone: 506-364-2519
fax: 506-364-2505
email: jehrman-at-mta.ca
www: http://www.mta.ca/dmf


==============================Original Headers==============================
8, 19 -- From jehrman-at-mta.ca Thu Jan 25 13:25:22 2007
8, 19 -- Received: from mailserv.mta.ca (mailserv.mta.ca [138.73.1.1])
8, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PJPM7Q023863
8, 19 -- for {Microscopy-at-MSA.Microscopy.com} ; Thu, 25 Jan 2007 13:25:22 -0600
8, 19 -- Received: from host-22-245.mta.ca ([138.73.22.245])
8, 19 -- by mailserv.mta.ca with esmtp (Exim 4.61)
8, 19 -- (envelope-from {jehrman-at-mta.ca} )
8, 19 -- id 1HAADk-0002D3-GG
8, 19 -- for Microscopy-at-MSA.Microscopy.com; Thu, 25 Jan 2007 15:25:17 -0400
8, 19 -- Message-ID: {45B9040E.20507-at-mta.ca}
8, 19 -- Date: Thu, 25 Jan 2007 15:25:02 -0400
8, 19 -- From: "James M. Ehrman" {jehrman-at-mta.ca}
8, 19 -- Organization: Mount Allison University
8, 19 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
8, 19 -- MIME-Version: 1.0
8, 19 -- To: Microscopy Listserv {Microscopy-at-MSA.Microscopy.com}
8, 19 -- Subject: JEOL 5xxx photo CRT
8, 19 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
8, 19 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: lamoreaux-at-mail.csi.cuny.edu
Date: Thu, 25 Jan 2007 14:40:10 -0600
Subject: [Microscopy] research associate postion

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

The College of Staten Island of The City University of New York has an
opening for a Research Associate to serve as a Facilities Manager for the
Program in Neuroscience and Advanced Imaging Facility. A description of the
position can be found at:

http://www.csi.cuny.edu/administration/jobs/mp12932.pdf

Applications may sent to my attention:

William L'Amoreaux, Ph.D.
Associate Professor and Deputy Chair
Director, Advanced Imaging Facility
CUNY College of Staten Island
2800 Victory Blvd.
Staten Island, NY 10314
(718) 982-3864 - office
(718) 982-3852



==============================Original Headers==============================
5, 20 -- From lamoreaux-at-mail.csi.cuny.edu Thu Jan 25 14:40:10 2007
5, 20 -- Received: from mail.csi.cuny.edu (mail.csi.cuny.edu [163.238.8.100])
5, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PKe9JI003792
5, 20 -- for {Microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 14:40:09 -0600
5, 20 -- Received: from D7K4H8C1 [163.238.22.9] by mail.csi.cuny.edu with ESMTP
5, 20 -- (SMTPD-8.22) id A5A90440; Thu, 25 Jan 2007 15:40:09 -0500
5, 20 -- Reply-To: {lamoreaux-at-mail.csi.cuny.edu}
5, 20 -- From: "William L'Amoreaux" {lamoreaux-at-mail.csi.cuny.edu}
5, 20 -- To: {Microscopy-at-microscopy.com}
5, 20 -- Subject: research associate postion
5, 20 -- Date: Thu, 25 Jan 2007 15:40:05 -0500
5, 20 -- Organization: College of Staten Island
5, 20 -- Message-ID: {001c01c740c0$febe95b0$0916eea3-at-D7K4H8C1}
5, 20 -- MIME-Version: 1.0
5, 20 -- Content-Type: text/plain;
5, 20 -- charset="us-ascii"
5, 20 -- Content-Transfer-Encoding: 7bit
5, 20 -- X-Mailer: Microsoft Office Outlook 11
5, 20 -- thread-index: AcdAwP6ki/NffyhDTIecNLGTCYQiEg==
5, 20 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: jbpawley-at-wisc.edu
Date: Thu, 25 Jan 2007 19:18:50 -0600
Subject: [Microscopy] Twelfth International UBC 3D LiveCell Course

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


Twelfth Annual INTERNATIONAL 12-Day Short Course on 3D Microscopy of Living Cells, June 17 - 28, 2007 (Pre-course: June 16)

Eleventh, Post-course Workshop on 3D Image Processing, June 30 -July 2

Organized by Prof. James Pawley, (University of Wisconsin-Madison)

in association with the Departments of Pharmacology and Physiology and the Brain Research Centre,
University of British Columbia, Vancouver, BC, Canada

DATES

Applications must be received by Tuesday, March 15, 2007
Deposit due Friday, April 15, 2007
Registration 5:00 - 7:00 PM Saturday, June 16, 2007
First Lecture 7:30 PM Saturday, June 16, 2007
Live-cell Course ends, noon Thursday, June 28, 2007
3D Image Processing Course, Saturday, June 30 - Monday, July 2, 2007


APPLICATIONS DUE BY MARCH 15, 2007

APPLICATIONS
Applicants must complete a questionnaire on the web to assess knowledge level, field of interest and proposed personal, live-cell, project. But don't let this put you off: if you plan to use 3D microscopy on living cells, we can usually find a way to make it work.

Enrollment will be limited to about 32 participants (exact number depends on number of 3D Systems available). Selection will be made on the basis of background and perceived need. Those without previous LM experience will be provided with access to basic texts to read before the course begins. Application forms may be down-loaded from the WWW site at

http://www.3dcourse.ubc.ca/application.htm , or obtained from:

Prof. James B. Pawley, Ph. 608-263-3147
Room 223, Zoology Research Building, FAX 608-265-5315
250 N. Mills Street, Madison, WI, 53706 JBPAWLEY-at-WISC.EDU

Additional information is available from:
http://www.3dcourse.ubc.ca/brochure.htm and links.

We expect to have at least 13, 3D microscope workstations for student use and there will be an international faculty of 22.

Application deadlines:

Application forms should be received for screening by March 15, 2007. Successful applicants will be notified by April 1, and a deposit of 50% must be received by April 15, 2007. In general, refunds of the deposit will only be possible if someone on the waiting list can take your place but this has not been a problem in previous years. The remaining balance is due before Registration.


Pre-Course Tuition (1/2 day Basic Optical principles) $150 (US)
3D Live-cell Course Tuition (includes lunches, 2 big snacks, 3 dinners, incl. the
NEW Third Edition of the Handbook of Biological Confocal Microscopy): $2,850 (US)
Workshop Tuition (includes lunches, snacks, final dinner): $1,200 (US)

Room/board about $40/day (US) depending on room type.

I hope that this includes all of the information that you need, but if not, please get back to me.

Jim Pawley

--
****************************************
Prof. James B. Pawley, Ph. 608-263-3147
Room 223, Zoology Research Building, FAX 608-265-5315
1117 Johnson Ave., Madison, WI, 53706 JBPAWLEY-at-WISC.EDU
"A scientist is not one who can answer questions but one who can
question answers." Theodore Schick Jr., Skeptical Enquirer, 21-2:39

==============================Original Headers==============================
23, 12 -- From zaluzec-at-microscopy.com Thu Jan 25 19:18:49 2007
23, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
23, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0Q1ImSV019471
23, 12 -- for {microscopy-at-microscopy.com} ; Thu, 25 Jan 2007 19:18:49 -0600
23, 12 -- Mime-Version: 1.0
23, 12 -- X-Sender: (Unverified)
23, 12 -- Message-Id: {p06110400c1df0711bdc4-at-[206.69.208.22]}
23, 12 -- Date: Thu, 25 Jan 2007 19:18:48 -0600
23, 12 -- To: microscopy-at-microscopy.com
23, 12 -- From: James Pawley {jbpawley-at-wisc.edu} (by way of MicroscopyListserver)
23, 12 -- Subject: Twelfth International UBC 3D LiveCell Course
23, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: kenconverse-at-qualityimages.biz
Date: Thu, 25 Jan 2007 19:43:36 -0600
Subject: [Microscopy] 840 vacuum problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi rtch,
The problem could well be V1B or V7 (column isolation). I haven't had much
opportunity to work on JEOL spectrometers, but the question arises: is
there a thin window detector and could its vacuum manifold be part of the
problem?

The part that bothers me the most is that the gun is being held to about
100mB (75 Torr) while being vented (this is a huge gas load to handle and
maintain that kind of pressure) and PiG 3 is ***not*** being dumped
instantaneously along with a big blast of DP oil being sent into the
specimen chamber.

I would be looking for a combination of a leak in V1B or V7 and either a
restriction in the nitrogen line to LV1 or perhaps its timer is shutting it
very quickly.

Btw what additional gauging have you got that would let you know that the
gun is at about 100mB? Do you have a convection gauge or Bourdon Tube gauge
attached to the gun?

I also strongly second Jim's comment about "what was the last thing you did
before you had the problem?", but I bet you already know that.

A very interesting problem. Keep us posted on what you find.

Ken Converse
owner

QUALITY IMAGES
Servicing Scanning Electron Microscopes
Since 1981
474 So. Bridgton Rd.
Bridgton, ME 04009
207-647-4348
Fax 207-647-2688
kenconverse-at-qualityimages.biz
qualityimages.biz



-----Original Message-----
X-from: r.sims-at-auckland.ac.nz [mailto:r.sims-at-auckland.ac.nz]
Sent: Wednesday, January 24, 2007 8:33 PM
To: kenconverse-at-qualityimages.biz


Here's one for all the 840 afficianados:

In the 'ready' state all is well, with all the valve lights as they should
be, and a
pressure of around 10 - 5 in the chamber (there are 3 WDS spectros leaking
away
into the chamber) and around 5 x 10 - 6 in the gun.

When I push the GUN button to vent the gun, the lights do what they should,
including
the opening of LV1, but the gun pressure comes not up to atmospheric but to
about
100 mbar, and the pressure in the vacuum ballast tank starts climbing, until
the PiG3
light comes on and the system devotes itself, as it should, to re-evacuating
the ballast
tank.

Am I correct in guessing that V1B, the gun pumpdown valve, is not closing
properly
and is allowing the atmosphere to leak via LV1, the gun, and V1B, into the
ballast
tank?

I would welcome support (or, I suppose, otherwise) in this guess before I
start ripping
things apart.

There is a time constraint here, as we have the NZ EM conference in 10 days
time
and there's to be a teaching workshop involving the 840.

Help!

cheers

rtch

--
Ritchie Sims Ph D Phone : 64 9 3737599 ext
87713
Microanalyst Fax : 64 9 3737435
Department of Geology email :
r.sims-at-auckland.ac.nz
The University of Auckland
Private Bag 92019
Auckland
New Zealand


==============================Original Headers==============================
12, 27 -- From r.sims-at-auckland.ac.nz Wed Jan 24 19:30:06 2007 12, 27 --
Received: from harpo.itss.auckland.ac.nz (harpo.itss.auckland.ac.nz
[130.216.190.13])
12, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id
l0P1U513003330
12, 27 -- for {microscopy-at-msa.microscopy.com} ; Wed, 24 Jan 2007
19:30:06 -0600
12, 27 -- Received: from localhost (localhost.localdomain [127.0.0.1])
12, 27 -- by harpo.itss.auckland.ac.nz (Postfix) with ESMTP id
5A6193485B
12, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 25 Jan 2007
14:30:04 +1300 (NZDT)
12, 27 -- Received: from harpo.itss.auckland.ac.nz ([127.0.0.1]) 12, 27 --
by localhost (smtpc.itss.auckland.ac.nz [127.0.0.1]) (amavisd-new, port
10024) 12, 27 -- with ESMTP id 23181-05 for
{microscopy-at-msa.microscopy.com} ; 12, 27 -- Thu, 25 Jan 2007 14:30:04 +1300
(NZDT) 12, 27 -- Received: from rs (r.sims.glg.auckland.ac.nz
[130.216.59.18])
12, 27 -- by harpo.itss.auckland.ac.nz (Postfix) with ESMTP id
317AE3485A
12, 27 -- for {microscopy-at-msa.microscopy.com} ; Thu, 25 Jan 2007
14:30:03 +1300 (NZDT)
12, 27 -- From: "Ritchie Sims" {r.sims-at-auckland.ac.nz}
12, 27 -- To: microscopy-at-msa.microscopy.com
12, 27 -- Date: Thu, 25 Jan 2007 14:30:24 +1300
12, 27 -- MIME-Version: 1.0
12, 27 -- Subject: 840 vacuum problem
12, 27 -- Message-ID: {45B8BF00.12939.1379784-at-localhost}
12, 27 -- Priority: normal
12, 27 -- In-reply-to: {200701250103.l0P135c6026032-at-ns.microscopy.com}
12, 27 -- X-mailer: Pegasus Mail for Windows (4.21c)
12, 27 -- Content-type: text/plain; charset=US-ASCII
12, 27 -- Content-transfer-encoding: 7BIT
12, 27 -- Content-description: Mail message body
12, 27 -- X-Virus-Scanned: by amavisd-new at mailhost.auckland.ac.nz
==============================End of - Headers==============================





_________________________________________________________________
Need personalized email and website? Look no further. It's easy
with Doteasy $0 Web Hosting! Learn more at www.doteasy.com


==============================Original Headers==============================
34, 28 -- From kenconverse-at-qualityimages.biz Thu Jan 25 19:43:35 2007
34, 28 -- Received: from dpmailmta02.doteasy.com (dpmailmta02.doteasy.com [65.61.209.88])
34, 28 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0Q1hZAY030854
34, 28 -- for {Microscopy-at-MSA.Microscopy.Com} ; Thu, 25 Jan 2007 19:43:35 -0600
34, 28 -- Received: from qualityimages.biz (unverified [192.168.101.16])
34, 28 -- by dpmailmta02.doteasy.com (DEO) with ESMTP id 52497752-1814644
34, 28 -- for multiple; Thu, 25 Jan 2007 18:16:15 -0800
34, 28 -- Received: from Ken [76.179.237.214] by qualityimages.biz with ESMTP
34, 28 -- (SMTPD32-8.05) id AD2972C0138; Thu, 25 Jan 2007 17:45:13 -0800
34, 28 -- From: "Ken Converse" {kenconverse-at-qualityimages.biz}
34, 28 -- To: {r.sims-at-auckland.ac.nz} , "MSA Listserver" {Microscopy-at-MSA.Microscopy.Com}
34, 28 -- Subject: RE: [Microscopy] 840 vacuum problem
34, 28 -- Date: Thu, 25 Jan 2007 20:43:06 -0500
34, 28 -- Message-ID: {002201c740eb$55a753b0$6401a8c0-at-Ken}
34, 28 -- MIME-Version: 1.0
34, 28 -- Content-Type: text/plain;
34, 28 -- charset="us-ascii"
34, 28 -- X-Priority: 3 (Normal)
34, 28 -- X-MSMail-Priority: Normal
34, 28 -- X-Mailer: Microsoft Outlook, Build 10.0.6626
34, 28 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
34, 28 -- Importance: Normal
34, 28 -- In-Reply-To: {200701250133.l0P1XAjA008071-at-ns.microscopy.com}
34, 28 -- X-IMSTrailer: __IMail_7__
34, 28 -- X-IP-stats: Incoming Last 0, First 120, in=2905538, out=0, spam=0 ip=192.168.101.16
34, 28 -- X-External-IP: 192.168.101.16
34, 28 -- Content-Transfer-Encoding: 8bit
34, 28 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0Q1hZAY030854
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Fri, 26 Jan 2007 05:18:28 -0600
Subject: [Microscopy] Re: color CCD Zeiss AxioCam MRc problem

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi!

We have the same camera and I must say that I never
noticed that. We have the B/W axiocam too and since
this one is more sensitive than the color one, I
usually use it for fluorescence work. Of course each
wavelength having a specific convergence angle it is
necessary to refocus each time you change the filter,
however this does not explain why you get a sharp
image in BW mode (except if you refocused in this
mode).
The color camera is used for more "classical"
stainings like HE.
In general, I find the image pretty grainy. For
example the digital zoom is absolutely useless because
the image becomes too "pixely". Perhaps it is becaused
I am used to the high definition of TEM images,
perhaps it is because I do most of the LM work on
single cell imaging (reaching the physical limits of
LM). To answer this question rationally I should try a
higher resolution camera, however I am pretty sure I
won't have the credits just to answer this question
;-)
I questioned the Zeiss team about the quality of my
images and they say it is normal, so I have to accept
it somehow.

Regards,

Stephane

--- cammer-at-aecom.yu.edu wrote:

}
}
}
}
----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The
} Microscopy Society of America
} To Subscribe/Unsubscribe --
} http://www.microscopy.com/MicroscopyListserver
} On-Line Help
}
http://www.microscopy.com/MicroscopyListserver/FAQ.html
}
----------------------------------------------------------------------------
}
} We found that the red (typically rhodamine) images
} taken with our
} Zeiss Axiocam MRc are fuzzy. We looked at the blue
} and found this a
} problem too, but since in fluorescence people
} typically use the blue
} just as a marker for nucleii rather than as stain
} for fine structure,
} they didn't notice this problem. (Illustrations at
} http://www.aecom.yu.edu/aif/temp/axiocamproblem/ ).
}
} This made us worry about the sharpness of images of
} histological
} stained samples, such as H&E.
}
} First we checked that this isn't a problem due to
} the fluorescent
} filter sets. To do this, we filtered the light to
} be red, green or
} blue at the halogen lamp using RGB filters removed
} from an old 35mm
} film recorder and imaged the sample with no filters
} in the light path
} between the sample and the camera. The result was
} that the red and
} blue images were very fuzzy.
}
} Shooting in the b/w mode solves this problem, but
} then the images are
} not color. Also, this may solve the problem for
} fluorescence, but
} what about brightfield or other techniques that
} involve colors?
}
} Can anyone explain why this happens and how to fix
} it? Is this a
} problem with our sytem uniquely or is this a problem
} with the Axiocam
} in general?
}
} Thanks!
}
} -Michael
}
____________________________________________________________________________
} Michael Cammer Analytical Imaging Facility
} Albert Einstein Coll. of Med.
} URL: http://www.aecom.yu.edu/aif/
}
}
} ==============================Original
} Headers==============================
} 8, 29 -- From cammer-at-aecom.yu.edu Thu Jan 25
} 10:17:49 2007
} 8, 29 -- Received: from mx1.aecom.yu.edu
} (mx1.aecom.yu.edu [129.98.1.51])
} 8, 29 -- by ns.microscopy.com
} (8.12.11.20060308/8.12.8) with ESMTP id
} l0PGHnp3030006
} 8, 29 -- for {microscopy-at-microscopy.com} ; Thu, 25
} Jan 2007 10:17:49 -0600
} 8, 29 -- Received: from draco.aecom.yu.edu
} (draco.aecom.yu.edu [129.98.1.160])
} 8, 29 -- by mx1.aecom.yu.edu (Postfix) with ESMTP
} id 5A9FA9F00EF;
} 8, 29 -- Thu, 25 Jan 2007 11:17:49 -0500 (EST)
} 8, 29 -- Received: from draco.aecom.yu.edu (unknown
} [127.0.0.1])
} 8, 29 -- by draco.aecom.yu.edu (Symantec Mail
} Security) with ESMTP id 8D5C38B4006;
} 8, 29 -- Thu, 25 Jan 2007 11:06:04 -0500 (EST)
} 8, 29 -- X-AuditID:
} 816201a0-a13c8bb000007022-82-45b8d56c712a
} 8, 29 -- Received: from post.aecom.yu.edu
} (post.aecom.yu.edu [129.98.1.100])
} 8, 29 -- by draco.aecom.yu.edu (Symantec Mail
} Security) with ESMTP id 4E7FE718002;
} 8, 29 -- Thu, 25 Jan 2007 11:06:04 -0500 (EST)
} 8, 29 -- Received: from AIF3.aecom.yu.edu
} (aif3.aif.aecom.yu.edu [129.98.80.70])
} 8, 29 -- (using TLSv1 with cipher
} DHE-RSA-AES256-SHA (256/256 bits))
} 8, 29 -- (No client certificate requested)
} 8, 29 -- by post.aecom.yu.edu (Postfix) with ESMTP
} id C3E7E26;
} 8, 29 -- Thu, 25 Jan 2007 11:17:43 -0500 (EST)
} 8, 29 -- Message-Id:
} {7.0.1.0.2.20070125111540.04f87920-at-aecom.yu.edu}
} 8, 29 -- X-Mailer: QUALCOMM Windows Eudora Version
} 7.0.1.0
} 8, 29 -- Date: Thu, 25 Jan 2007 11:17:43 -0500
} 8, 29 -- To: microscopy-at-microscopy.com
} 8, 29 -- From: Michael Cammer {cammer-at-aecom.yu.edu}
} 8, 29 -- Subject: color CCD Zeiss AxioCam MRc
} problem
} 8, 29 -- Cc: yadeng-at-aecom.yu.edu,
} rleonard-at-aecom.yu.edu, jmorreale-at-zeiss.com
} 8, 29 -- Mime-Version: 1.0
} 8, 29 -- Content-Type: text/plain;
} charset="us-ascii"; format=flowed
} 8, 29 -- X-Brightmail-Tracker: AAAAAA==
} ==============================End of -
} Headers==============================
}




____________________________________________________________________________________
Do you Yahoo!?
Everyone is raving about the all-new Yahoo! Mail beta.
http://new.mail.yahoo.com

==============================Original Headers==============================
9, 21 -- From nizets2-at-yahoo.com Fri Jan 26 05:18:28 2007
9, 21 -- Received: from web37405.mail.mud.yahoo.com (web37405.mail.mud.yahoo.com [209.191.91.137])
9, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0QBIS9A018787
9, 21 -- for {microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 05:18:28 -0600
9, 21 -- Received: (qmail 59506 invoked by uid 60001); 26 Jan 2007 11:18:27 -0000
9, 21 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
9, 21 -- s=s1024; d=yahoo.com;
9, 21 -- h=X-YMail-OSG:Received:Date:From:Subject:To:Cc:In-Reply-To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
9, 21 -- b=5xtcf3ab0P0lvV3K+ARcllmK8q6VYiM3tFHdaaU9YMHNcynCgtt8AtrwkadChoB48V9fHRqgnjT2hFc80sg7QCTCUxdGGuVZ2dieoeTuh512WNx6H6DYINpd2aLwLEXgohXdfvfVvdpop6u5zg9cE2Wwyos45HA5OIej83X74Cw=;
9, 21 -- X-YMail-OSG: SZCGGSkVM1noy50s_7cZ2wnm8LdubtYip1L_5vpDShbX6U7dBg4FRh6O6uKFabbS1HS9sigpJZ_ZeFDNi4ox5DeVVRbvsBNAEV5GY7ehgdUTd2605rU1mIulPbxrFnVU09VtKetok5m6wGA13kX.WhyKaaX3wjd5l1fmckaPYiKo
9, 21 -- Received: from [80.122.101.102] by web37405.mail.mud.yahoo.com via HTTP; Fri, 26 Jan 2007 03:18:27 PST
9, 21 -- Date: Fri, 26 Jan 2007 03:18:27 -0800 (PST)
9, 21 -- From: Stephane Nizet {nizets2-at-yahoo.com}
9, 21 -- Subject: Re: [Microscopy] color CCD Zeiss AxioCam MRc problem
9, 21 -- To: cammer-at-aecom.yu.edu
9, 21 -- Cc: microscopy-at-microscopy.com
9, 21 -- In-Reply-To: {200701251623.l0PGNVuH005535-at-ns.microscopy.com}
9, 21 -- MIME-Version: 1.0
9, 21 -- Content-Type: text/plain; charset=iso-8859-1
9, 21 -- Content-Transfer-Encoding: 8bit
9, 21 -- Message-ID: {648803.48936.qm-at-web37405.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: dljones-at-bestweb.net
Date: Fri, 26 Jan 2007 08:21:46 -0600
Subject: [Microscopy] Re: Nitrogen free epoxy?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

John,

I really know little of what you are doing, but I haven't seen any
answers to your question so I thought I'd throw something back at you.

Can you use something other than epoxy? For example a polyester mounting
material?

The polyester poly(ethyene terephthalate) doesn't have any nitrogen in it,
for example. I think several other forms of polyesters also do not contain
nitrogen, but I am no expert in that field.

There is a commercial product from Extec that is a polyester mount
material. I don't know if it's nitrogen free, but you could ask them for
the MSDS of the product and see. A link to their web page where they have
their cold mounting supplies is:

http://www.extec.com/coldmounting.htm

Disclaimer: I have no commercial interest in this company or it's
products.

I do use their polyester mounting material when I have samples I need to
look at close to the edge with the mount and I don't want all the added
elements usually found in epoxies. But I've not yet been concerned with
nitrogen as one of those elements...

Good luck, and I would like to know what you do end up finding if you
wouldn't mind posting your results...

dj

On Thu, 25 Jan 2007, donovan-at-uoregon.edu wrote:

}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} All,
} Can anyone tell me of a source for a nitrogen free epoxy? I realize
} that may be a contradiction in terms, but I need to find a mounting
} media that can hold a water soluble polymer coated wire for cross
} sectioning for subsequent nitrogen analysis by EPMA.
}
} Thanks!
} john
}
} John J. Donovan donovan-at-uoregon.edu
} University of Oregon (541) 346-4632 (office)
} 1260 Franklin Blvd (541) 346-4655 (probe)
} Eugene, OR (541) 346-4778 (SEM)
} 97403-1272 (541) 346-4692 (FAX)
}
} Lab Web: http://epmalab.uoregon.edu/
} EPMA (SX100)Schedule: http://sweetwater.uoregon.edu/sx100
} EPMA (SX50) Schedule: http://sweetwater.uoregon.edu/epma
} SEM (Ultra) Schedule: http://sweetwater.uoregon.edu/zeiss
} SEM (Quanta) Schedule: http://sweetwater.uoregon.edu/sem
} Remote Access: http://epmalab.uoregon.edu/howto.htm
} Personal: http://www.uoregon.edu/~donovan/
}
} "Aristotle presented with a moon rock would have no difficulty in
} discerning it as an object not fundamentally different from
} terrestrial materials. So much for Thomas Kuhn."
}
} "I'd rather be uncertain and wrong, than be certain and wrong."
}
}
}
}
} ==============================Original Headers==============================
} 9, 20 -- From donovan-at-uoregon.edu Thu Jan 25 11:24:37 2007
} 9, 20 -- Received: from smtp.uoregon.edu (mserv2.uoregon.edu [128.223.142.41])
} 9, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0PHOadg031693
} 9, 20 -- for {Microscopy-at-Microscopy.Com} ; Thu, 25 Jan 2007 11:24:36 -0600
} 9, 20 -- Received: from Probev.uoregon.edu (probev.uoregon.edu [128.223.92.75])
} 9, 20 -- (authenticated bits=0)
} 9, 20 -- by smtp.uoregon.edu (8.13.8/8.13.8) with ESMTP id l0PHOahW007001
} 9, 20 -- (version=TLSv1/SSLv3 cipher=DHE-RSA-AES256-SHA bits=256 verify=NOT)
} 9, 20 -- for {Microscopy-at-Microscopy.Com} ; Thu, 25 Jan 2007 09:24:36 -0800
} 9, 20 -- Message-Id: {7.0.1.0.0.20070125092420.039399b8-at-uoregon.edu}
} 9, 20 -- Message-Id: {7.0.1.0.0.20070125091904.038e7cd0-at-uoregon.edu}
} 9, 20 -- X-Mailer: QUALCOMM Windows Eudora Version 7.0.1.0
} 9, 20 -- Date: Thu, 25 Jan 2007 09:24:32 -0800
} 9, 20 -- To: Microscopy-at-Microscopy.Com
} 9, 20 -- From: John Donovan {donovan-at-uoregon.edu}
} 9, 20 -- Subject: Nitrogen free epoxy?
} 9, 20 -- Mime-Version: 1.0
} 9, 20 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
} 9, 20 -- X-Virus-Scanned: ClamAV 0.88.7/2489/Thu Jan 25 04:35:24 2007 on mserv2
} 9, 20 -- X-Virus-Status: Clean
} ==============================End of - Headers==============================
}


==============================Original Headers==============================
13, 19 -- From dljones-at-bestweb.net Fri Jan 26 08:21:45 2007
13, 19 -- Received: from vms046pub.verizon.net (vms046pub.verizon.net [206.46.252.46])
13, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QELjiD000634
13, 19 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 08:21:45 -0600
13, 19 -- Received: from localhost ([71.247.249.221])
13, 19 -- by vms046.mailsrvcs.net (Sun Java System Messaging Server 6.2-6.01 (built Apr
13, 19 -- 3 2006)) with ESMTPA id {0JCH00AC2BVXV1B7-at-vms046.mailsrvcs.net} for
13, 19 -- Microscopy-at-microscopy.com; Fri, 26 Jan 2007 08:21:39 -0600 (CST)
13, 19 -- Date: Fri, 26 Jan 2007 09:20:48 -0500 (Eastern Standard Time)
13, 19 -- From: "David L. Jones" {dljones-at-bestweb.net}
13, 19 -- Subject: Re: [Microscopy] Nitrogen free epoxy?
13, 19 -- In-reply-to: {200701251728.l0PHSolQ005552-at-ns.microscopy.com}
13, 19 -- X-X-Sender: dljones-at-pop-croton.bestweb.net
13, 19 -- To: donovan-at-uoregon.edu
13, 19 -- Cc: Microscopy-at-microscopy.com
13, 19 -- Message-id: {Pine.WNT.4.64.0701260828200.4048-at-H-F1}
13, 19 -- MIME-version: 1.0
13, 19 -- Content-type: TEXT/PLAIN; charset=US-ASCII; format=flowed
13, 19 -- References: {200701251728.l0PHSolQ005552-at-ns.microscopy.com}
==============================End of - Headers==============================




From: hikonishi-at-gmail.com
Date: Fri, 26 Jan 2007 12:35:31 -0600
Subject: [Microscopy] Ultramicrotomy testing

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hello:

We are planning to buy an ultramicrotome at a high probability, and we would
like to microtome some materials for testing. If some labs in the
manufacturers can help us, please contact me by email. Our materials are
metal and perhaps ceramics.



Hiromi Konishi, Ph.D.
Laboratory Manager
The S.W. Bailey X-ray Diffraction Laboratory
Room A353 Weeks Hall
Voice: (608) 262-9784, (608) 262-0915
Fax: (608) 262-0693
Department of Geology and Geophysics
University of Wisconsin-Madison
1215 W Dayton St. Madison WI 53706


==============================Original Headers==============================
6, 30 -- From hikonishi-at-gmail.com Fri Jan 26 12:35:31 2007
6, 30 -- Received: from py-out-1112.google.com (py-out-1112.google.com [64.233.166.178])
6, 30 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QIZVU7018415
6, 30 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 12:35:31 -0600
6, 30 -- Received: by py-out-1112.google.com with SMTP id b50so462752pyh
6, 30 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 10:35:29 -0800 (PST)
6, 30 -- DomainKey-Signature: a=rsa-sha1; c=nofws;
6, 30 -- d=gmail.com; s=beta;
6, 30 -- h=received:message-id:from:to:subject:date:mime-version:content-type:content-transfer-encoding:x-priority:x-msmail-priority:x-mailer:x-mimeole;
6, 30 -- b=IjEz3uvHL3yMtLR2aQgYhYeow1xdq1tuZWN/O3m2vm2bY4s2STDk26QWQYb5deLstS3b1VIuiWSX6CZLdqS57JePH8+w5Sz2HnT7kMIIEDxGrS++WwNcifsqytrVTfbhJXSW9MAG8DcA0cvztNFlkzcF7KBmV7EqV0U98x3iiJI=
6, 30 -- Received: by 10.35.96.11 with SMTP id y11mr4404012pyl.1169836529656;
6, 30 -- Fri, 26 Jan 2007 10:35:29 -0800 (PST)
6, 30 -- Received: from HKONISHI ( [144.92.207.106])
6, 30 -- by mx.google.com with ESMTP id z52sm2803277pyg.2007.01.26.10.35.28;
6, 30 -- Fri, 26 Jan 2007 10:35:28 -0800 (PST)
6, 30 -- Message-ID: {000501c74178$bd28b0f0$6acf5c90-at-HKONISHI}
6, 30 -- From: "Hiromi Konishi" {hikonishi-at-gmail.com}
6, 30 -- To: {Microscopy-at-microscopy.com}
6, 30 -- Subject: Ultramicrotomy testing
6, 30 -- Date: Fri, 26 Jan 2007 12:35:05 -0600
6, 30 -- MIME-Version: 1.0
6, 30 -- Content-Type: text/plain;
6, 30 -- format=flowed;
6, 30 -- charset="iso-2022-jp";
6, 30 -- reply-type=original
6, 30 -- Content-Transfer-Encoding: 7bit
6, 30 -- X-Priority: 3
6, 30 -- X-MSMail-Priority: Normal
6, 30 -- X-Mailer: Microsoft Outlook Express 6.00.2900.2180
6, 30 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.2180
==============================End of - Headers==============================




From: Jessica.Wagner-at-childrens.harvard.edu
Date: Fri, 26 Jan 2007 14:24:29 -0600
Subject: [Microscopy] imaging the occular view through another port?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hello list. I have a question that's just for fun. I'm using a Nikon TE-2000 wth all 4 ports and a beam splitter that can direct light 50/50 split between two ports. Is there a way to set up a camera to image not a sample but specifically the image that I'm seeing? I have a minor defect in my eye, a wrinkle caused by slight detachment of the vitreous (I think due to having LASIK done). When I look through the scope at a bright field, I can see an image of the wrinkle. I'm curious to know if I can capture it, but my guess is that the image only exists in the occulars? Maybe I could somehow use a dichroic mirror?

Thanks,
Jessica



==============================Original Headers==============================
4, 37 -- From Jessica.Wagner-at-childrens.harvard.edu Fri Jan 26 14:24:29 2007
4, 37 -- Received: from mail2.childrenshospital.org (mail2.childrenshospital.org [134.174.20.64])
4, 37 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QKOTAx003925
4, 37 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 14:24:29 -0600
4, 37 -- Received: from tumsmtp1.CHBOSTON.ORG (tumsmtp1.tch.harvard.edu [10.1.101.46])
4, 37 -- by mail2.childrenshospital.org (Postfix) with ESMTP id B770480F1
4, 37 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 15:24:26 -0500 (EST)
4, 37 -- Received: from 10.1.102.175 by tumsmtp1.CHBOSTON.ORG with ESMTP (MMS
4, 37 -- SMTP Relay (Email Firewall v6.3.0)); Fri, 26 Jan 2007 15:24:13 -0500
4, 37 -- X-Server-Uuid: 1F337096-A893-456A-BE4C-C6341410F3EE
4, 37 -- Received: from CHEXV4.CHBOSTON.ORG ([10.1.102.188]) by
4, 37 -- chexsmtp2.CHBOSTON.ORG with Microsoft SMTPSVC(6.0.3790.1830); Fri, 26
4, 37 -- Jan 2007 15:24:13 -0500
4, 37 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
4, 37 -- Content-class: urn:content-classes:message
4, 37 -- MIME-Version: 1.0
4, 37 -- Subject: imaging the occular view through another port?
4, 37 -- Date: Fri, 26 Jan 2007 15:24:12 -0500
4, 37 -- Message-ID: {6B335BE3804E0A40978750E51EFC076108B83B-at-CHEXV4.CHBOSTON.ORG}
4, 37 -- X-MS-Has-Attach:
4, 37 -- X-MS-TNEF-Correlator:
4, 37 -- Thread-Topic: imaging the occular view through another port?
4, 37 -- Thread-Index: AcdBh/FJTzzdmMO7TlybuNWpOytJvw==
4, 37 -- From: "Wagner, Jessica" {Jessica.Wagner-at-childrens.harvard.edu}
4, 37 -- To: Microscopy-at-microscopy.com
4, 37 -- X-OriginalArrivalTime: 26 Jan 2007 20:24:13.0614 (UTC)
4, 37 -- FILETIME=[F1C7C4E0:01C74187]
4, 37 -- X-TMWD-Spam-Summary: TS=20070126202415; SEV=2.2.0; DFV=B2007012608;
4, 37 -- IFV=2.0.4,4.0-9; AIF=B2007012608; RPD=5.02.0004; ENG=IBF;
4, 37 -- RPDID=7374723D303030312E30413031303230332E34354241363336462E303032442C73733D312C6667733D30;
4, 37 -- CAT=NONE; CON=NONE
4, 37 -- X-MMS-Spam-Filter-ID: B2007012608_5.02.0004_4.0-9
4, 37 -- X-WSS-ID: 69A4BCE73B47225184-01-01
4, 37 -- Content-Type: text/plain;
4, 37 -- charset=iso-8859-1
4, 37 -- Content-Transfer-Encoding: 8bit
4, 37 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0QKOTAx003925
==============================End of - Headers==============================




From: jquinn-at-www.matscieng.sunysb.edu
Date: Fri, 26 Jan 2007 14:28:55 -0600
Subject: [Microscopy] re: SEM for 12" or 300mm sample

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


Folks -

I am looking for a SEM with EDS to check out
a 12" or 300mm Si sample/disk/target.

We would like to scan around the
surface of the silicon target.

Please respond "off-listserv".

kind regards,

Jim






==============================Original Headers==============================
12, 13 -- From jquinn-at-www.matscieng.sunysb.edu Fri Jan 26 14:28:54 2007
12, 13 -- Received: from www.matscieng.sunysb.edu (www.matscieng.sunysb.edu [129.49.36.33])
12, 13 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QKSsHi010007
12, 13 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 14:28:54 -0600
12, 13 -- Received: (from jquinn-at-localhost)
12, 13 -- by www.matscieng.sunysb.edu (8.11.6/8.11.6) id l0QKNkj14934;
12, 13 -- Fri, 26 Jan 2007 15:23:46 -0500
12, 13 -- Date: Fri, 26 Jan 2007 15:23:46 -0500
12, 13 -- From: Jim Quinn {jquinn-at-www.matscieng.sunysb.edu}
12, 13 -- Message-Id: {200701262023.l0QKNkj14934-at-www.matscieng.sunysb.edu}
12, 13 -- To: Microscopy-at-microscopy.com
12, 13 -- Subject: re: SEM for 12" or 300mm sample
12, 13 -- Cc: waldvogel-at-prodigy.net
==============================End of - Headers==============================




From: bozzola-at-siu.edu
Date: Fri, 26 Jan 2007 15:01:39 -0600
Subject: [Microscopy] Re: imaging the occular view through another port?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hello,

I have experienced this same phenomenon with "floaters" in my eyes.
Under the right conditions, I can see the floaters beautifully and
sometimes even the corneal surface. As explained to me by an
ophthomologist, this is due to the projection of a shadow (of the
objects) onto the retina. Therefore, you would not be able to "go the
other way" and see the image in a viewing port. I believe that the
only way to image this would be with a slit lamp (as in an
ophthomological examination).

Maybe others on the list would care to comment since I am certainly
no expert but speaking only from personal experience.

JB

} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America

==============================Original Headers==============================
5, 20 -- From bozzola-at-siu.edu Fri Jan 26 15:01:39 2007
5, 20 -- Received: from abbmta1.siu.edu (abbmta1.siu.edu [131.230.254.205])
5, 20 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QL1dYK003114
5, 20 -- for {Microscopy-at-msa.microscopy.com} ; Fri, 26 Jan 2007 15:01:39 -0600
5, 20 -- Received: from [131.230.177.142] (ws177142.microscope.siu.edu [131.230.177.142])
5, 20 -- by abbmta1.siu.edu (Switch-3.1.11/Switch-3.1.11) with ESMTP id l0QL15wG014885
5, 20 -- for {Microscopy-at-msa.microscopy.com} ; Fri, 26 Jan 2007 15:01:16 -0600 (CST)
5, 20 -- Mime-Version: 1.0
5, 20 -- X-Sender: bozzola-at-saluki-mail.siu.edu
5, 20 -- Message-Id: {p06110401c1e01ae71b78-at-[131.230.177.142]}
5, 20 -- In-Reply-To: {200701262026.l0QKQHhM006040-at-ns.microscopy.com}
5, 20 -- References: {200701262026.l0QKQHhM006040-at-ns.microscopy.com}
5, 20 -- Date: Fri, 26 Jan 2007 15:01:04 -0600
5, 20 -- To: Microscopy-at-msa.microscopy.com
5, 20 -- From: "John J. Bozzola" {bozzola-at-siu.edu}
5, 20 -- Subject: Re: [Microscopy] imaging the occular view through another port?
5, 20 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
5, 20 -- X-Spam-Score: 0.00%
5, 20 -- X-MASF: 0.00%
5, 20 -- X-Whitelist: 0.00%
==============================End of - Headers==============================




From: tivol-at-caltech.edu
Date: Fri, 26 Jan 2007 15:39:41 -0600
Subject: [Microscopy] Re: imaging the occular view through another port?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


On Jan 26, 2007, at 12:24 PM, Jessica.Wagner-at-childrens.harvard.edu
wrote:

} } Hello list. I have a question that's just for fun. I'm using a
} } Nikon TE-2000 wth all 4 ports and a beam splitter that can direct
} } light 50/50 split between two ports. Is there a way to set up a
} } camera to image not a sample but specifically the image that I'm
} } seeing? I have a minor defect in my eye, a wrinkle caused by slight
} } detachment of the vitreous (I think due to having LASIK done). When
} } I look through the scope at a bright field, I can see an image of the
} } wrinkle. I'm curious to know if I can capture it, but my guess is
} } that the image only exists in the occulars? Maybe I could somehow
} } use a dichroic mirror?
} }
} I have experienced this same phenomenon with "floaters" in my eyes.
} Under the right conditions, I can see the floaters beautifully and
} sometimes even the corneal surface. As explained to me by an
} ophthomologist, this is due to the projection of a shadow (of the
} objects) onto the retina. Therefore, you would not be able to "go the
} other way" and see the image in a viewing port. I believe that the
} only way to image this would be with a slit lamp (as in an
} ophthomological examination).
}
} Maybe others on the list would care to comment since I am certainly
} no expert but speaking only from personal experience.

Dear Jessica and John,
The image "that I'm seeing" can be thought of as the image that a user
with normal sight would see transformed by the effect of the defect, so
although it is easy to see the "normal" image, and as John suggests,
you could see the defect with a slit lamp, you could not see the
perturbation caused by the defect (although there are ways in theory to
calculate it from the structure of the defect). Of course, you could
look at a grid or some such test object and draw what you see, so the
distortions noted in the grid will tell you what distortions to expect
in a more complex image. You could even look at the grid separately
with each eye and compare the image from the undamaged eye to that from
the eye with the defect, and, if your brain has not already adapted to
seeing the distorted image, looking at a grid that has been distorted
in one eye could be interpreted by your brain as a 3D image of the grid
that appears to be closer to you in some places than in others, like
the effect of viewing stereo images, but much more subtle.
Yours,
Bill Tivol, PhD
EM Scientist and Manager
Cryo-Electron Microscopy Facility
Broad Center, Mail Code 114-96
California Institute of Technology
Pasadena CA 91125
(626) 395-8833
tivol-at-caltech.edu


==============================Original Headers==============================
5, 24 -- From tivol-at-caltech.edu Fri Jan 26 15:39:40 2007
5, 24 -- Received: from outgoing-mail.its.caltech.edu (outgoing-mail.its.caltech.edu [131.215.239.19])
5, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QLdexu017495
5, 24 -- for {microscopy-at-msa.microscopy.com} ; Fri, 26 Jan 2007 15:39:40 -0600
5, 24 -- Received: from water-dog.caltech.edu (water-dog [192.168.1.26])
5, 24 -- by water-ox-postvirus (Postfix) with ESMTP id D8F352EF8F
5, 24 -- for {microscopy-at-msa.microscopy.com} ; Fri, 26 Jan 2007 13:39:35 -0800 (PST)
5, 24 -- Received: from [192.168.159.233] (jpix-01.caltech.edu [131.215.2.133])
5, 24 -- (using TLSv1 with cipher RC4-SHA (128/128 bits))
5, 24 -- (No client certificate requested)
5, 24 -- by earth-ox.its.caltech.edu (Postfix) with ESMTP id 9BF4810A109
5, 24 -- for {microscopy-at-msa.microscopy.com} ; Fri, 26 Jan 2007 13:39:34 -0800 (PST)
5, 24 -- Mime-Version: 1.0 (Apple Message framework v624)
5, 24 -- In-Reply-To: {200701262024.l0QKOc5u004144-at-ns.microscopy.com}
5, 24 -- References: {200701262024.l0QKOc5u004144-at-ns.microscopy.com}
5, 24 -- Content-Type: text/plain; charset=US-ASCII; format=flowed
5, 24 -- Message-Id: {9a126178c30930f80e890b1ee25dc61d-at-caltech.edu}
5, 24 -- Content-Transfer-Encoding: 7bit
5, 24 -- From: Bill Tivol {tivol-at-caltech.edu}
5, 24 -- Subject: Re: [Microscopy] imaging the occular view through another port?
5, 24 -- Date: Fri, 26 Jan 2007 13:47:32 -0800
5, 24 -- To: microscopy-at-msa.microscopy.com
5, 24 -- X-Mailer: Apple Mail (2.624)
5, 24 -- X-Spam-Scanned: at Caltech-ITS on water-dog by amavisd-2.3.3
==============================End of - Headers==============================




From: ahlst007-at-umn.edu
Date: Fri, 26 Jan 2007 15:46:29 -0600
Subject: [Microscopy] Re: Nitrogen free epoxy?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

John,

Unless you will be looking for trace amounts of
nitrogen in your polymer, you might be able to
get away with using one of the Epon epoxy
substitutes, like EMbed 812.

Of its four components, only the accelerator
DMP-30 has any N, 3 atoms per molecule, only a
small amount is used in the mix, about 1.25%
v/v, and the mass fraction of N in that 1.25% is
even less. Another accelerator that can be used
with EMbed812, BDMA, has only one N atom per
molecule.

As it happened, I was scheduled to do some
SEM-EDS this morning, so I took a look at a
sample of an EMbed 812 block, hardened with
DMP-30. I could see no detectable N peak in the
spectra. I'll send you a JPG of a spectra off-List.

As you will be doing EPMA, your sensitivity for
N maybe be greater. But you could just check
some epoxy resins to see if there is enough N
present to be a problem for your polymer work.

But of course, there is nothing sweeter than a
nice, clean elemental analysis!

Good luck in your search for an N-free medium.

Gib

Gilbert (Gib) Ahlstrand, Electron Microscopist,
Imaging Center, University of Minnesota
123 Snyder Hall
St. Paul, MN 55108
http://www.cbs.umn.edu/ic

donovan-at-uoregon.edu wrote:
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} All,
} Can anyone tell me of a source for a nitrogen free epoxy? I realize
} that may be a contradiction in terms, but I need to find a mounting
} media that can hold a water soluble polymer coated wire for cross
} sectioning for subsequent nitrogen analysis by EPMA.
}
} Thanks!
} john
}
} John J. Donovan donovan-at-uoregon.edu
} University of Oregon (541) 346-4632 (office)
} 1260 Franklin Blvd (541) 346-4655 (probe)
} Eugene, OR (541) 346-4778 (SEM)
} 97403-1272 (541) 346-4692 (FAX)

==============================Original Headers==============================
10, 21 -- From ahlst007-at-umn.edu Fri Jan 26 15:46:29 2007
10, 21 -- Received: from mta-m2.tc.umn.edu (mta-m2.tc.umn.edu [160.94.23.21])
10, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QLkTOO028495
10, 21 -- for {Microscopy-at-Microscopy.com} ; Fri, 26 Jan 2007 15:46:29 -0600
10, 21 -- Received: from [160.94.39.28] (x160-94-39-28.cbs.umn.edu [160.94.39.28])
10, 21 -- by mta-m2.tc.umn.edu (UMN smtpd) with ESMTP
10, 21 -- for {Microscopy-at-Microscopy.com} ; Fri, 26 Jan 2007 15:46:28 -0600 (CST)
10, 21 -- X-Umn-Remote-Mta: [N] x160-94-39-28.cbs.umn.edu [160.94.39.28] #+LO+TS+AU+HN
10, 21 -- Message-ID: {45BA766B.7040308-at-umn.edu}
10, 21 -- Date: Fri, 26 Jan 2007 15:45:15 -0600
10, 21 -- From: Gib Ahlstrand {ahlst007-at-umn.edu}
10, 21 -- Reply-To: ahlst007-at-umn.edu
10, 21 -- Organization: Imaging Center UM
10, 21 -- User-Agent: Thunderbird 1.5.0.9 (Macintosh/20061207)
10, 21 -- MIME-Version: 1.0
10, 21 -- To: Microscopy-at-Microscopy.com
10, 21 -- Subject: Re: [Microscopy] Nitrogen free epoxy?
10, 21 -- References: {200701251728.l0PHSYC9005007-at-ns.microscopy.com}
10, 21 -- In-Reply-To: {200701251728.l0PHSYC9005007-at-ns.microscopy.com}
10, 21 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
10, 21 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: rcommon-at-msu.edu
Date: Fri, 26 Jan 2007 15:58:38 -0600
Subject: [Microscopy] imaging the occular view through another port?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html


You probably would not like one possible approach to solving this imaging
problem. I am thinking of Kevin Kline's solution to a similar problem in
the movie "Wild Wild West".

Ralph Common


==============================Original Headers==============================
4, 24 -- From rcommon-at-msu.edu Fri Jan 26 15:58:38 2007
4, 24 -- Received: from sys28.mail.msu.edu (sys28.mail.msu.edu [35.9.75.128])
4, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QLwcbH007237
4, 24 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 15:58:38 -0600
4, 24 -- Received: from [35.9.122.125] (helo=emlab)
4, 24 -- by sys28.mail.msu.edu with esmtpsa (Exim 4.52 #1)
4, 24 -- (TLSv1:RC4-MD5:128)
4, 24 -- id 1HAZ5i-00010e-9d
4, 24 -- for Microscopy-at-microscopy.com; Fri, 26 Jan 2007 16:58:38 -0500
4, 24 -- From: "Ralph Common" {rcommon-at-msu.edu}
4, 24 -- To: {Microscopy-at-microscopy.com}
4, 24 -- Subject: imaging the occular view through another port?
4, 24 -- Date: Fri, 26 Jan 2007 17:00:10 -0500
4, 24 -- Message-ID: {006801c74195$5abe6c30$7d7a0923-at-msu.edu}
4, 24 -- MIME-Version: 1.0
4, 24 -- Content-Type: text/plain;
4, 24 -- charset="iso-8859-1"
4, 24 -- Content-Transfer-Encoding: 7bit
4, 24 -- X-Priority: 3 (Normal)
4, 24 -- X-MSMail-Priority: Normal
4, 24 -- X-Mailer: Microsoft Outlook CWS, Build 9.0.2416 (9.0.2911.0)
4, 24 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2800.1896
4, 24 -- Importance: Normal
4, 24 -- X-Virus: None found by Clam AV
==============================End of - Headers==============================




From: CGoldsmith-at-cdc.gov
Date: Fri, 26 Jan 2007 17:33:12 -0600
Subject: [Microscopy] viaWWW: Southeastern Microscopy Society annual meetin

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both CGoldsmith-at-cdc.gov as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: CGoldsmith-at-cdc.gov
Name: Cynthia Goldsmith

Organization: Southeastern Microscopy Society

Title-Subject: [Filtered] Southeastern Microscopy Society annual meetin

Question: The 2007 annual meeting for the Southeastern Microscopy Society (SEMS) will be held April 11-13 in historic Decatur, Georgia, a suburb of Atlanta. We have an exciting program planned, including pre-meeting workshops on Bio-TEM Imaging, QuantomiX Wet SEM, Cryoultramicrotomy, and Deconvolution Microscopy. In addition, invited speakers will present information on such topics as the examination of the Gospel of Judas, World Trade Center dust, and macrophage cholesterol metabolism.

Deadlines for registration and abstract submission are February 23, 2007. There is additional meeting information available at the SEMS website at www.southeasternmicroscopy.org.


Cynthia S. Goldsmith
Secretary, Southeastern Microscopy Society (SEMS)



---------------------------------------------------------------------------

==============================Original Headers==============================
11, 12 -- From zaluzec-at-microscopy.com Fri Jan 26 17:33:11 2007
11, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
11, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0QNX93x019671
11, 12 -- for {microscopy-at-microscopy.com} ; Fri, 26 Jan 2007 17:33:11 -0600
11, 12 -- Mime-Version: 1.0
11, 12 -- X-Sender: (Unverified)
11, 12 -- Message-Id: {p06110402c1e040131ba2-at-[206.69.208.22]}
11, 12 -- Date: Fri, 26 Jan 2007 17:33:09 -0600
11, 12 -- To: microscopy-at-microscopy.com
11, 12 -- From: CGoldsmith-at-cdc.gov (by way of MicroscopyListserver)
11, 12 -- Subject: viaWWW: Southeastern Microscopy Society annual meetin
11, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Mon, 29 Jan 2007 09:38:24 -0600
Subject: [Microscopy] TEM:Maleic vs. maleate vs. malic acid buffers

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Listers,

I guess this is my month for desperate listserve queries, so here I go
again.

We are trying to follow a not-very-detailed protocol for brain tissue
which utilizes maleic acid buffer for wash steps before and after the
"uranyl acetate colorization solution". One would presume that this
refers to en bloc staining also using this buffer, but no details are
given as to molarity, buffer composition, etc. We have no experience
with this buffer, although I've heard of it, and are having a terrible
time finding good references regarding its use and formulation.

My research in our reference library and online shows many instances of
en bloc staining using maleic acid buffer, tris-maleate buffer, or
simply maleic acid, so my questions are:

1) Are these terms pretty much interchangeable (i.e., is tris-maleate
buffer just maleic acid buffer with the addition of tris)?

2) Do you all have any preferred formulations for this buffer in the UA
step? Most of the refs I've seen use 0.05M - 0.2M buffer with 0.5-1%
UA.

3) Is this buffer pretty much restricted to en bloc staining, at least
for use in TEM studies?

4) What are the advantages of this buffer over aqueous en bloc staining
with UA?

Thanks to all.

Cheers,
Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


==============================Original Headers==============================
12, 23 -- From TindallR-at-missouri.edu Mon Jan 29 09:38:23 2007
12, 23 -- Received: from um-nsmtpout1.um.umsystem.edu (um-nsmtpout1.um.umsystem.edu [209.106.228.53])
12, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0TFcNVY008362
12, 23 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 09:38:23 -0600
12, 23 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-nsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
12, 23 -- Mon, 29 Jan 2007 09:38:23 -0600
12, 23 -- x-mimeole: Produced By Microsoft Exchange V6.5
12, 23 -- Content-class: urn:content-classes:message
12, 23 -- MIME-Version: 1.0
12, 23 -- Content-Type: text/plain;
12, 23 -- charset="us-ascii"
12, 23 -- Subject: TEM:Maleic vs. maleate vs. malic acid buffers
12, 23 -- Date: Mon, 29 Jan 2007 09:38:22 -0600
12, 23 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D7A-at-UM-XMAIL08.um.umsystem.edu}
12, 23 -- X-MS-Has-Attach:
12, 23 -- X-MS-TNEF-Correlator:
12, 23 -- Thread-Topic: TEM:Maleic vs. maleate vs. malic acid buffers
12, 23 -- Thread-Index: AcdDu4WL6a4MycTjQd+Res2DXMOOtQ==
12, 23 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
12, 23 -- To: {microscopy-at-microscopy.com}
12, 23 -- X-OriginalArrivalTime: 29 Jan 2007 15:38:23.0068 (UTC) FILETIME=[827F61C0:01C743BB]
12, 23 -- Content-Transfer-Encoding: 8bit
12, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0TFcNVY008362
==============================End of - Headers==============================




From: mconstan-at-princeton.edu
Date: Mon, 29 Jan 2007 10:04:18 -0600
Subject: [Microscopy] AskAMicroscopist: SEM beam penetration

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question was submitted to Ask-A-Microscopist by (mconstan-at-princeton.edu)
from
http://www.msa.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html
on Monday, January 29, 2007 at 02:58:11
Remember to consider the Grade/Age of the student when considering the Question
---------------------------------------------------------------------------
Please reply to both mconstan-at-princeton.edu as well as to the
Microscopy Listserver
---------------------------------------------------------------------------

Email: mconstan-at-princeton.edu
Name: Michael

Organization: Princeton University

Education: Undergraduate College

Location: Princeton, NJ, USA

Title: SEM beam penetration

Question: Hi. I'm an undergrad working in a lab and I would like to
use SEM to view gold nanoparticles embedded in agarose gel. About
how far could the SEM beam penetrate into 1% agarose (I'm guessing I
would have to let the water in the gel evaporate)? Would it be
better to use SE or BSE? What kV should I try? Thanks for any help
you can give me!

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-microscopy.com Mon Jan 29 10:04:17 2007
8, 12 -- Received: from [207.42.54.42] (msdvpn8.msd.anl.gov [130.202.238.72])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0TG4GSB020035
8, 12 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 10:04:17 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: (Unverified)
8, 12 -- Message-Id: {p06020400c1e3ccfb5db3-at-[207.42.54.42]}
8, 12 -- Date: Mon, 29 Jan 2007 10:10:52 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: mconstan-at-princeton.edu (by way of Ask-A-Microscopist)
8, 12 -- Subject: AskAMicroscopist: SEM beam penetration
8, 12 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
==============================End of - Headers==============================




From: phillipst-at-missouri.edu
Date: Mon, 29 Jan 2007 10:17:23 -0600
Subject: [Microscopy] Re: Maleate buffer

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Here is what we use with success:

Sodium Maleate (Maleic acid, monosodium salt)
MW = 138.1
50 mM = 691 mg/ 100 mls

Assume uranyl acetate stock is 5%. Dilute 1 part + 9 parts of 50 mM sodium
maleate

En bloc staining protocol:

After osmium, rinse tissue 3x 5 min each in dH2O

rinse tissue 3x 5 min each in 50 mM sodium maleate, pH 5.2

incubate tissue for 1 hr to overnight in 0.5% - 1.0% uranyl acetate in 50
mM sodium maleate, pH 5.2 (for overnight, probably should use 0.5% UA, for
1 hr probably should use 1% UA).

rinse 3x dH2O for 10 min

50% ethanol 10 min

70% ethanol 10 min

95% ethanol 10 min

3x 100% ethanol 10 min

start resin infiltration



==============================Original Headers==============================
15, 19 -- From PhillipsT-at-missouri.edu Mon Jan 29 10:17:23 2007
15, 19 -- Received: from um-nsmtpout1.um.umsystem.edu (um-nsmtpout1.um.umsystem.edu [209.106.228.53])
15, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0TGHMKd031221
15, 19 -- for {Microscopy-at-msa.microscopy.com} ; Mon, 29 Jan 2007 10:17:23 -0600
15, 19 -- Received: from um-tsmtpout1.um.umsystem.edu ([209.106.228.23]) by um-nsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
15, 19 -- Mon, 29 Jan 2007 10:17:22 -0600
15, 19 -- Received: from phillips-dell.missouri.edu ([128.206.81.132]) by um-tsmtpout1.um.umsystem.edu over TLS secured channel with Microsoft SMTPSVC(6.0.3790.1830);
15, 19 -- Mon, 29 Jan 2007 10:17:21 -0600
15, 19 -- Message-Id: {6.0.0.22.2.20070129101755.04c27060-at-pop.missouri.edu}
15, 19 -- X-Sender: phillipst-at-pop.missouri.edu
15, 19 -- X-Mailer: QUALCOMM Windows Eudora Version 6.0.0.22
15, 19 -- Date: Mon, 29 Jan 2007 10:19:18 -0600
15, 19 -- To: Microscopy-at-msa.microscopy.com
15, 19 -- From: Tom Phillips {phillipst-at-missouri.edu}
15, 19 -- Subject: Re: Maleate buffer
15, 19 -- Cc: tindallr-at-missouri.edu
15, 19 -- Mime-Version: 1.0
15, 19 -- Content-Type: text/plain; charset="us-ascii"; format=flowed
15, 19 -- X-OriginalArrivalTime: 29 Jan 2007 16:17:22.0173 (UTC) FILETIME=[F4B67AD0:01C743C0]
==============================End of - Headers==============================




From: john.mardinly-at-intel.com
Date: Mon, 29 Jan 2007 11:51:40 -0600
Subject: [Microscopy] imaging the occular view through another port?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

What kind of detail do you need? That will probably determine your
conditions.

Understand that the beam will start scattering and interacting once it
encounters the first surface which will be agarose. Secondary electrons
are limited to escaping from the first few nm of surface, so even if
they are generated at some depth, they will not escape to be detected. I
expect you would see primarily the agarose surface.

There is also the question of imaging the agarose at all. I presume it
would charge without coating and would need to be examined in variable
pressure or environmental mode. That would eliminate the choice of SE
unless you have a gaseous SE detector, in which case the previous
comments still apply.

BSE could give you some information from some depth below the surface.
It will still be difficult and there will be some scattering.

Also, what happens to the structure of the agarose under vacuum? I
suppose there will be appreciable loss of moisture, so what you see may
not be anything like what you had. I would check the mass loss as a
result of exposure to vacuum. It might be too extreme.

I would probably start at 20kV for voltage and set the current to give a
decent BSE image. You will probably have to experiment from there.

Warren Straszheim

-----Original Message-----
X-from: mconstan-at-princeton.edu [mailto:mconstan-at-princeton.edu]
Sent: Monday, January 29, 2007 10:05 AM
To: wesaia-at-iastate.edu

Jessica;
Some time ago, there was a thread about LASIK and microscopists.
I do not recall any of the reports mentioning this sort of post-surgical
vision difficulty. Is this perhaps a subject that should be re-visited?

John Mardinly
Intel

Disclaimer: The opinions of this author do not represent the opinions of
Intel Corporation.

-----Original Message-----
X-from: Jessica.Wagner-at-childrens.harvard.edu
[mailto:Jessica.Wagner-at-childrens.harvard.edu]
Sent: Friday, January 26, 2007 12:25 PM
To: Mardinly, John

Hello list. I have a question that's just for fun. I'm using a Nikon
TE-2000 wth all 4 ports and a beam splitter that can direct light 50/50
split between two ports. Is there a way to set up a camera to image not
a sample but specifically the image that I'm seeing? I have a minor
defect in my eye, a wrinkle caused by slight detachment of the vitreous
(I think due to having LASIK done). When I look through the scope at a
bright field, I can see an image of the wrinkle. I'm curious to know if
I can capture it, but my guess is that the image only exists in the
occulars? Maybe I could somehow use a dichroic mirror?

Thanks,
Jessica



==============================Original
Headers==============================
4, 37 -- From Jessica.Wagner-at-childrens.harvard.edu Fri Jan 26 14:24:29
2007
4, 37 -- Received: from mail2.childrenshospital.org
(mail2.childrenshospital.org [134.174.20.64])
4, 37 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0QKOTAx003925
4, 37 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007
14:24:29 -0600
4, 37 -- Received: from tumsmtp1.CHBOSTON.ORG (tumsmtp1.tch.harvard.edu
[10.1.101.46])
4, 37 -- by mail2.childrenshospital.org (Postfix) with ESMTP id
B770480F1
4, 37 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007
15:24:26 -0500 (EST)
4, 37 -- Received: from 10.1.102.175 by tumsmtp1.CHBOSTON.ORG with ESMTP
(MMS
4, 37 -- SMTP Relay (Email Firewall v6.3.0)); Fri, 26 Jan 2007 15:24:13
-0500
4, 37 -- X-Server-Uuid: 1F337096-A893-456A-BE4C-C6341410F3EE
4, 37 -- Received: from CHEXV4.CHBOSTON.ORG ([10.1.102.188]) by
4, 37 -- chexsmtp2.CHBOSTON.ORG with Microsoft SMTPSVC(6.0.3790.1830);
Fri, 26
4, 37 -- Jan 2007 15:24:13 -0500
4, 37 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
4, 37 -- Content-class: urn:content-classes:message
4, 37 -- MIME-Version: 1.0
4, 37 -- Subject: imaging the occular view through another port?
4, 37 -- Date: Fri, 26 Jan 2007 15:24:12 -0500
4, 37 -- Message-ID:
{6B335BE3804E0A40978750E51EFC076108B83B-at-CHEXV4.CHBOSTON.ORG}
4, 37 -- X-MS-Has-Attach:
4, 37 -- X-MS-TNEF-Correlator:
4, 37 -- Thread-Topic: imaging the occular view through another port?
4, 37 -- Thread-Index: AcdBh/FJTzzdmMO7TlybuNWpOytJvw==
4, 37 -- From: "Wagner, Jessica" {Jessica.Wagner-at-childrens.harvard.edu}
4, 37 -- To: Microscopy-at-microscopy.com
4, 37 -- X-OriginalArrivalTime: 26 Jan 2007 20:24:13.0614 (UTC)
4, 37 -- FILETIME=[F1C7C4E0:01C74187]
4, 37 -- X-TMWD-Spam-Summary: TS=20070126202415; SEV=2.2.0;
DFV=B2007012608;
4, 37 -- IFV=2.0.4,4.0-9; AIF=B2007012608; RPD=5.02.0004; ENG=IBF;
4, 37 --
RPDID=7374723D303030312E30413031303230332E34354241363336462E303032442C73
733D312C6667733D30;
4, 37 -- CAT=NONE; CON=NONE
4, 37 -- X-MMS-Spam-Filter-ID: B2007012608_5.02.0004_4.0-9
4, 37 -- X-WSS-ID: 69A4BCE73B47225184-01-01
4, 37 -- Content-Type: text/plain;
4, 37 -- charset=iso-8859-1
4, 37 -- Content-Transfer-Encoding: 8bit
4, 37 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0QKOTAx003925
==============================End of -
Headers==============================


==============================Original Headers==============================
13, 34 -- From john.mardinly-at-intel.com Mon Jan 29 11:51:39 2007
13, 34 -- Received: from mga02.intel.com (mga02.intel.com [134.134.136.20])
13, 34 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0THpd49013923
13, 34 -- for {Microscopy-at-msa.microscopy.com} ; Mon, 29 Jan 2007 11:51:39 -0600
13, 34 -- Received: from orsmga001.jf.intel.com ([10.7.209.18])
13, 34 -- by mga02.intel.com with ESMTP; 29 Jan 2007 09:51:39 -0800
13, 34 -- Received: from fmsmsx333.amr.corp.intel.com ([132.233.42.2])
13, 34 -- by orsmga001.jf.intel.com with ESMTP; 29 Jan 2007 09:51:38 -0800
13, 34 -- X-ExtLoop1: 1
13, 34 -- X-IronPort-AV: i="4.13,253,1167638400";
13, 34 -- d="scan'208"; a="190293831:sNHT77980763"
13, 34 -- Received: from scsmsx411.amr.corp.intel.com ([10.3.90.30]) by fmsmsx333.amr.corp.intel.com with Microsoft SMTPSVC(6.0.3790.1830);
13, 34 -- Mon, 29 Jan 2007 09:51:31 -0800
13, 34 -- Received: from scsmsx413.amr.corp.intel.com ([10.3.90.32]) by scsmsx411.amr.corp.intel.com with Microsoft SMTPSVC(6.0.3790.1830);
13, 34 -- Mon, 29 Jan 2007 09:51:30 -0800
13, 34 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
13, 34 -- Content-class: urn:content-classes:message
13, 34 -- MIME-Version: 1.0
13, 34 -- Content-Type: text/plain;
13, 34 -- charset="us-ascii"
13, 34 -- Subject: RE: [Microscopy] imaging the occular view through another port?
13, 34 -- Date: Mon, 29 Jan 2007 09:51:29 -0800
13, 34 -- Message-ID: {1DF4C4D62339DB4C9C98DF04213995B60263F0B5-at-scsmsx413.amr.corp.intel.com}
13, 34 -- In-Reply-To: {200701262024.l0QKObkm004099-at-ns.microscopy.com}
13, 34 -- X-MS-Has-Attach:
13, 34 -- X-MS-TNEF-Correlator:
13, 34 -- Thread-Topic: [Microscopy] imaging the occular view through another port?
13, 34 -- Thread-Index: AcdBiAPf8S8hIZO1T+u/pIAfVESIzACRKOmg
13, 34 -- From: "Mardinly, John" {john.mardinly-at-intel.com}
13, 34 -- To: {Jessica.Wagner-at-childrens.harvard.edu}
13, 34 -- Cc: {Microscopy-at-msa.microscopy.com}
13, 34 -- X-OriginalArrivalTime: 29 Jan 2007 17:51:30.0799 (UTC) FILETIME=[1B8E67F0:01C743CE]
13, 34 -- Content-Transfer-Encoding: 8bit
13, 34 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0THpd49013923
==============================End of - Headers==============================




From: VGriffiths-at-mtech.edu
Date: Mon, 29 Jan 2007 12:00:34 -0600
Subject: [Microscopy] Epoxy free mounting material

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Another possibility is to use a mounting material from long,long ago.
This is Wood's metal, a quaternary eutectic of Bi/50%, Pb/25%, Sn12.5%,
Cd12.5%. It's used for sprinkler plugs in automatic systems.
If your polymer can stand 70'C, (melting point of the alloy) it.s a good
way to go since you be sure there's no nitrogen in it. The alloy is
fairly soft and it wet's well.

Vern Griffiths
Montana Tech


==============================Original Headers==============================
3, 23 -- From VGriffiths-at-mtech.edu Mon Jan 29 12:00:34 2007
3, 23 -- Received: from cug1.umt.edu (cug1.umt.edu [150.131.192.195])
3, 23 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0TI0XPf001959
3, 23 -- for {Microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 12:00:33 -0600
3, 23 -- Received: from mtmailr3.butte.campus (mtmailr3.mtech.edu [10.34.34.114])
3, 23 -- by cug1.umt.edu (Switch-3.1.8/Switch-3.1.7) with ESMTP id l0TI0WJh012580
3, 23 -- for {Microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 11:00:32 -0700 (MST)
3, 23 -- Content-class: urn:content-classes:message
3, 23 -- MIME-Version: 1.0
3, 23 -- Content-Type: text/plain;
3, 23 -- charset="us-ascii"
3, 23 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
3, 23 -- Subject: Epoxy free mounting material
3, 23 -- Date: Mon, 29 Jan 2007 11:00:32 -0700
3, 23 -- Message-ID: {137E940998125E4FBEE1F53EBA11D800E5BB5E-at-mtmailr3.butte.campus}
3, 23 -- X-MS-Has-Attach:
3, 23 -- X-MS-TNEF-Correlator:
3, 23 -- Thread-Topic: Epoxy free mounting material
3, 23 -- Thread-Index: AcdDz15U04uodbytS1WUWB3KRqeNbw==
3, 23 -- From: "Griffiths, Vern" {VGriffiths-at-mtech.edu}
3, 23 -- To: {Microscopy-at-microscopy.com}
3, 23 -- Content-Transfer-Encoding: 8bit
3, 23 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0TI0XPf001959
==============================End of - Headers==============================




From: gradice-at-richmond.edu
Date: Mon, 29 Jan 2007 14:05:41 -0600
Subject: [Microscopy] real microscopes vs lens equations

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

I'm teaching a bioimaging course for undergraduates. We are now
discussing the lens equations that relate object and image distance
to focal length:

1/p + 1/q = 1/f

and magnification:

M = q/p

where p is the object distance and q is the image distance.

We've demonstrated these relationships using lenses on an optical
bench, and now I want to transfer to real microscopes.

The students have access to Nikon Alphaphot microscopes with a fixed
tube length of 160 mm, which I understand is measured from the
nosepiece to the eyepiece mount. The eyepiece mount is approximately
the same as the primary image plane on these 'scopes. So does that
mean that the image distance for each lens should be 160 mm plus the
distance to the middle of the objective lens? Or is it more
complicated than that, because the lens equations don't work out very
well using those values for image distance.

For example, the 10x objective should then have an image distance of
about 160 + 38 mm = 198 mm. Which would then predict an object
distance of 19.8 mm and a focal length of 18 mm. All of that makes
sense since the object should be just beyond the front focal point of
the objective to make a real inverted image at the primary image plane.

But in practice, the working distance of this lens is 6.1 mm, which
puts the actual object distance at around 6.3 mm, well inside the
presumed focal point, and puts the image distance at around 63 mm for
a 10 magnification

Something doesn't add up. What am I missing?





Gary P. Radice gradice-at-richmond.edu
Department of Biology 804-289-8107 (voice)
University of Richmond 804-289-8233 (FAX)
Richmond VA 23173 http://www.richmond.edu/~gradice



==============================Original Headers==============================
17, 21 -- From gradice-at-richmond.edu Mon Jan 29 14:05:38 2007
17, 21 -- Received: from spandex.richmond.edu (spandex.richmond.edu [141.166.24.58])
17, 21 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0TK5XPQ015998
17, 21 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 14:05:36 -0600
17, 21 -- Received: from rayon.richmond.edu (rayon.richmond.edu [141.166.30.10])
17, 21 -- by spandex.richmond.edu (8.13.1/8.13.1) with ESMTP id l0TK5WbO021856
17, 21 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 15:05:32 -0500
17, 21 -- Received: from [141.166.185.175] (ffa185175.richmond.edu [141.166.185.175])
17, 21 -- by rayon.richmond.edu (8.12.11.20060308/8.12.11) with ESMTP id l0TK5WUW007000
17, 21 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 15:05:32 -0500
17, 21 -- Mime-Version: 1.0 (Apple Message framework v752.2)
17, 21 -- Content-Transfer-Encoding: 7bit
17, 21 -- Message-Id: {38ED7F44-618C-4970-9B9E-C5BC26FAE970-at-richmond.edu}
17, 21 -- Content-Type: text/plain; charset=US-ASCII; delsp=yes; format=flowed
17, 21 -- To: microscopy-at-microscopy.com
17, 21 -- From: gradice {gradice-at-richmond.edu}
17, 21 -- Subject: real microscopes vs lens equations
17, 21 -- Date: Mon, 29 Jan 2007 15:04:25 -0500
17, 21 -- X-Mailer: Apple Mail (2.752.2)
17, 21 -- X-Spam-Detail: -1.44 ALL_TRUSTED
17, 21 -- X-Scanned-By: MIMEDefang 2.49 on 141.166.30.10
==============================End of - Headers==============================




From: Jessica.Wagner-at-childrens.harvard.edu
Date: Mon, 29 Jan 2007 16:18:54 -0600
Subject: [Microscopy] imaging the occular view through another port?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

John,

I missed that thread, but I'll be happy to add my two cents now.
Assuming the PVD (posterior vitreous detachment) doesn't get any worse,
for me the benefits of LASIK still outweigh this minor negative. The
'floater' that I see is an annoyance, but it doesn't truly inhibit my
vision through the scope (or outside the scope for that matter).

But I had LASIK only a year ago, and was never warned PVD could be a
complication, so this says to me that there is still much unknown about
the procedure and it's results. Unfortunately, complications of laser
eye surgeries really aren't tracked all that well; doctors are not
required to report them, unless they are related to a device, but even
then, many doctors are not aware of FDA regulations about device event
reporting or how/to whom they should report adverse events.

Here is an article about PVD's and LASIK:
http://www.springerlink.com/content/j3100858467pu5k1/

And thanks to everyone who offered imaging advice!

Jessica


-----Original Message-----
X-from: john.mardinly-at-intel.com [mailto:john.mardinly-at-intel.com]
Sent: Monday, January 29, 2007 12:56 PM
To: Wagner, Jessica

Jessica;
Some time ago, there was a thread about LASIK and microscopists.
I do not recall any of the reports mentioning this sort of post-surgical
vision difficulty. Is this perhaps a subject that should be re-visited?

John Mardinly
Intel

Disclaimer: The opinions of this author do not represent the opinions of
Intel Corporation.

-----Original Message-----
X-from: Jessica.Wagner-at-childrens.harvard.edu
[mailto:Jessica.Wagner-at-childrens.harvard.edu]
Sent: Friday, January 26, 2007 12:25 PM
To: Mardinly, John

Hello list. I have a question that's just for fun. I'm using a Nikon
TE-2000 wth all 4 ports and a beam splitter that can direct light 50/50
split between two ports. Is there a way to set up a camera to image not
a sample but specifically the image that I'm seeing? I have a minor
defect in my eye, a wrinkle caused by slight detachment of the vitreous
(I think due to having LASIK done). When I look through the scope at a
bright field, I can see an image of the wrinkle. I'm curious to know if
I can capture it, but my guess is that the image only exists in the
occulars? Maybe I could somehow use a dichroic mirror?

Thanks,
Jessica



==============================Original
Headers==============================
4, 37 -- From Jessica.Wagner-at-childrens.harvard.edu Fri Jan 26 14:24:29
2007 4, 37 -- Received: from mail2.childrenshospital.org
(mail2.childrenshospital.org [134.174.20.64])
4, 37 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0QKOTAx003925
4, 37 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007
14:24:29 -0600
4, 37 -- Received: from tumsmtp1.CHBOSTON.ORG (tumsmtp1.tch.harvard.edu
[10.1.101.46])
4, 37 -- by mail2.childrenshospital.org (Postfix) with ESMTP id
B770480F1
4, 37 -- for {Microscopy-at-microscopy.com} ; Fri, 26 Jan 2007
15:24:26 -0500 (EST)
4, 37 -- Received: from 10.1.102.175 by tumsmtp1.CHBOSTON.ORG with ESMTP
(MMS 4, 37 -- SMTP Relay (Email Firewall v6.3.0)); Fri, 26 Jan 2007
15:24:13 -0500 4, 37 -- X-Server-Uuid:
1F337096-A893-456A-BE4C-C6341410F3EE
4, 37 -- Received: from CHEXV4.CHBOSTON.ORG ([10.1.102.188]) by 4, 37 --
chexsmtp2.CHBOSTON.ORG with Microsoft SMTPSVC(6.0.3790.1830); Fri, 26 4,
37 -- Jan 2007 15:24:13 -0500 4, 37 -- X-MimeOLE: Produced By Microsoft
Exchange V6.5 4, 37 -- Content-class: urn:content-classes:message 4, 37
-- MIME-Version: 1.0 4, 37 -- Subject: imaging the occular view through
another port? 4, 37 -- Date: Fri, 26 Jan 2007 15:24:12 -0500 4, 37 --
Message-ID: {6B335BE3804E0A40978750E51EFC076108B83B-at-CHEXV4.CHBOSTON.ORG}
4, 37 -- X-MS-Has-Attach:
4, 37 -- X-MS-TNEF-Correlator:
4, 37 -- Thread-Topic: imaging the occular view through another port? 4,
37 -- Thread-Index: AcdBh/FJTzzdmMO7TlybuNWpOytJvw== 4, 37 -- From:
"Wagner, Jessica" {Jessica.Wagner-at-childrens.harvard.edu}
4, 37 -- To: Microscopy-at-microscopy.com
4, 37 -- X-OriginalArrivalTime: 26 Jan 2007 20:24:13.0614 (UTC) 4, 37 --
FILETIME=[F1C7C4E0:01C74187] 4, 37 -- X-TMWD-Spam-Summary:
TS=20070126202415; SEV=2.2.0; DFV=B2007012608; 4, 37 --
IFV=2.0.4,4.0-9; AIF=B2007012608; RPD=5.02.0004; ENG=IBF; 4, 37 --
RPDID=7374723D303030312E30413031303230332E34354241363336462E303032442C73
733D312C6667733D30;
4, 37 -- CAT=NONE; CON=NONE
4, 37 -- X-MMS-Spam-Filter-ID: B2007012608_5.02.0004_4.0-9
4, 37 -- X-WSS-ID: 69A4BCE73B47225184-01-01
4, 37 -- Content-Type: text/plain;
4, 37 -- charset=iso-8859-1
4, 37 -- Content-Transfer-Encoding: 8bit
4, 37 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0QKOTAx003925 ==============================End of
- Headers==============================


==============================Original
Headers==============================
13, 34 -- From john.mardinly-at-intel.com Mon Jan 29 11:51:39 2007 13, 34
-- Received: from mga02.intel.com (mga02.intel.com [134.134.136.20])
13, 34 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0THpd49013923
13, 34 -- for {Microscopy-at-msa.microscopy.com} ; Mon, 29 Jan 2007
11:51:39 -0600
13, 34 -- Received: from orsmga001.jf.intel.com ([10.7.209.18])
13, 34 -- by mga02.intel.com with ESMTP; 29 Jan 2007 09:51:39 -0800
13, 34 -- Received: from fmsmsx333.amr.corp.intel.com ([132.233.42.2])
13, 34 -- by orsmga001.jf.intel.com with ESMTP; 29 Jan 2007 09:51:38
-0800
13, 34 -- X-ExtLoop1: 1
13, 34 -- X-IronPort-AV: i="4.13,253,1167638400";
13, 34 -- d="scan'208"; a="190293831:sNHT77980763"
13, 34 -- Received: from scsmsx411.amr.corp.intel.com ([10.3.90.30]) by
fmsmsx333.amr.corp.intel.com with Microsoft SMTPSVC(6.0.3790.1830);
13, 34 -- Mon, 29 Jan 2007 09:51:31 -0800
13, 34 -- Received: from scsmsx413.amr.corp.intel.com ([10.3.90.32]) by
scsmsx411.amr.corp.intel.com with Microsoft SMTPSVC(6.0.3790.1830);
13, 34 -- Mon, 29 Jan 2007 09:51:30 -0800
13, 34 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
13, 34 -- Content-class: urn:content-classes:message
13, 34 -- MIME-Version: 1.0
13, 34 -- Content-Type: text/plain;
13, 34 -- charset="us-ascii"
13, 34 -- Subject: RE: [Microscopy] imaging the occular view through
another port? 13, 34 -- Date: Mon, 29 Jan 2007 09:51:29 -0800 13, 34 --
Message-ID:
{1DF4C4D62339DB4C9C98DF04213995B60263F0B5-at-scsmsx413.amr.corp.intel.com}
13, 34 -- In-Reply-To: {200701262024.l0QKObkm004099-at-ns.microscopy.com}
13, 34 -- X-MS-Has-Attach:
13, 34 -- X-MS-TNEF-Correlator:
13, 34 -- Thread-Topic: [Microscopy] imaging the occular view through
another port? 13, 34 -- Thread-Index:
AcdBiAPf8S8hIZO1T+u/pIAfVESIzACRKOmg
13, 34 -- From: "Mardinly, John" {john.mardinly-at-intel.com}
13, 34 -- To: {Jessica.Wagner-at-childrens.harvard.edu}
13, 34 -- Cc: {Microscopy-at-msa.microscopy.com}
13, 34 -- X-OriginalArrivalTime: 29 Jan 2007 17:51:30.0799 (UTC)
FILETIME=[1B8E67F0:01C743CE] 13, 34 -- Content-Transfer-Encoding: 8bit
13, 34 -- X-MIME-Autoconverted: from quoted-printable to 8bit by
ns.microscopy.com id l0THpd49013923 ==============================End of
- Headers==============================



==============================Original Headers==============================
28, 37 -- From Jessica.Wagner-at-childrens.harvard.edu Mon Jan 29 16:18:54 2007
28, 37 -- Received: from mail2.childrenshospital.org (mail2.childrenshospital.org [134.174.20.64])
28, 37 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0TMIrbi030608
28, 37 -- for {Microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 16:18:53 -0600
28, 37 -- Received: from tumsmtp1.CHBOSTON.ORG (tumsmtp1.tch.harvard.edu [10.1.101.46])
28, 37 -- by mail2.childrenshospital.org (Postfix) with ESMTP id C3D458082
28, 37 -- for {Microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 17:18:50 -0500 (EST)
28, 37 -- Received: from 10.1.102.174 by tumsmtp1.CHBOSTON.ORG with ESMTP (MMS
28, 37 -- SMTP Relay (Email Firewall v6.3.0)); Mon, 29 Jan 2007 17:18:43 -0500
28, 37 -- X-Server-Uuid: 1F337096-A893-456A-BE4C-C6341410F3EE
28, 37 -- Received: from CHEXV4.CHBOSTON.ORG ([10.1.102.188]) by
28, 37 -- chexsmtp1.CHBOSTON.ORG with Microsoft SMTPSVC(6.0.3790.1830); Mon, 29
28, 37 -- Jan 2007 17:18:43 -0500
28, 37 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
28, 37 -- Content-class: urn:content-classes:message
28, 37 -- MIME-Version: 1.0
28, 37 -- Subject: LASIK and microscopists
28, 37 -- Date: Mon, 29 Jan 2007 17:18:42 -0500
28, 37 -- Message-ID: {6B335BE3804E0A40978750E51EFC076107BF91-at-CHEXV4.CHBOSTON.ORG}
28, 37 -- X-MS-Has-Attach:
28, 37 -- X-MS-TNEF-Correlator:
28, 37 -- Thread-Topic: LASIK and microscopists
28, 37 -- Thread-Index: AcdD829TWE3SHiDVT4OxJPpkd6QQAg==
28, 37 -- From: "Wagner, Jessica" {Jessica.Wagner-at-childrens.harvard.edu}
28, 37 -- To: Microscopy-at-microscopy.com
28, 37 -- X-OriginalArrivalTime: 29 Jan 2007 22:18:43.0039 (UTC)
28, 37 -- FILETIME=[6F8416F0:01C743F3]
28, 37 -- X-TMWD-Spam-Summary: TS=20070129221845; SEV=2.2.0; DFV=B2007012910;
28, 37 -- IFV=2.0.4,4.0-9; AIF=B2007012910; RPD=5.02.0004; ENG=IBF;
28, 37 -- RPDID=7374723D303030312E30413031303230312E34354245373243342E303046372C73733D312C6667733D30;
28, 37 -- CAT=NONE; CON=NONE
28, 37 -- X-MMS-Spam-Filter-ID: B2007012910_5.02.0004_4.0-9
28, 37 -- X-WSS-ID: 69A0AD493B47707430-01-01
28, 37 -- Content-Type: text/plain;
28, 37 -- charset=us-ascii
28, 37 -- Content-Transfer-Encoding: 8bit
28, 37 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0TMIrbi030608
==============================End of - Headers==============================




From: sbarlow-at-sunstroke.sdsu.edu
Date: Mon, 29 Jan 2007 20:07:58 -0600
Subject: [Microscopy] negative staining viruses in cesium chloride

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

One of my users is trying to visualize virus particles by negative
stain, normally a routine procedure here. However, her sample is
straight off a cesium chloride purification. Will the CsCl solution
inhibit binding of the virus particles to coated grids, or should she
be able to use standard neg. staining procedures to see her
particles. She didn't see any particles in her first attempts.
While trying to calculate virus concentrations to determine if her
sample is concentrated enough, I wondered if the CsCl will inhibit
virus sticking to the grids--I have no experience with virus samples
that have not been purified into some other buffer system.

Anyone have a protocol we can use for such a sample, or does she need
to get rid of the CsCl before she tries to do any negative stain
imaging.

Your advice is appreciated.

Steve
--
Dr. Steven Barlow
EM Facility/Biology Dept.
San Diego State University
5500 Campanile Drive
San Diego CA 92182-4614
phone: (619) 594-4523
fax: (619) 594-5676

email: sbarlow-at-sunstroke.sdsu.edu
http://www.sci.sdsu.edu/emfacility

==============================Original Headers==============================
5, 18 -- From sbarlow-at-sunstroke.sdsu.edu Mon Jan 29 20:07:58 2007
5, 18 -- Received: from sciences.sdsu.edu (sciences.sdsu.edu [130.191.140.33])
5, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0U27vjw013631
5, 18 -- for {Microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 20:07:58 -0600
5, 18 -- Received: from [146.244.225.211] ([146.244.225.211])
5, 18 -- by sciences.sdsu.edu (8.12.11.20060614/8.12.10/SCEC-032005-9-50) with ESMTP id l0U27pVC000828
5, 18 -- for {Microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 18:07:51 -0800 (PST)
5, 18 -- Mime-Version: 1.0
5, 18 -- Message-Id: {p06230908c1e4573227d1-at-[146.244.225.211]}
5, 18 -- Date: Mon, 29 Jan 2007 18:07:42 -0800
5, 18 -- To: Microscopy-at-microscopy.com
5, 18 -- From: Steve Barlow {sbarlow-at-sunstroke.sdsu.edu}
5, 18 -- Subject: negative staining viruses in cesium chloride
5, 18 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
5, 18 -- X-Greylist: Sender IP whitelisted, not delayed by milter-greylist-1.6 (sciences.sdsu.edu [130.191.140.35]); Mon, 29 Jan 2007 18:07:52 -0800 (PST)
5, 18 -- X-COS-MailScanner: Found to be clean
5, 18 -- X-MailScanner-SpamCheck: not spam, SpamAssassin (score=0, required 8)
5, 18 -- X-MailScanner-From: sbarlow-at-sunstroke.sdsu.edu
==============================End of - Headers==============================




From: skuo-at-jhu.edu
Date: Mon, 29 Jan 2007 21:06:16 -0600
Subject: [Microscopy] viaWWW: Job Opening: Microscopist

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both skuo-at-jhu.edu as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: skuo-at-jhu.edu
Name: Scot Kuo

Organization: Johns Hopkins University

Title-Subject: [Filtered] Job Opening: Microscopist, try#2

Question: ...Try#2: Prior attempt lost line-breaks...
{br /}

With light, fluorescence or confocal microscopy experience,
a microscopy specialist is sought to provide user support
in Johns Hopkins School of Medicine Microscope Facility
(http://www.hopkinsmedicine.org/micfac/). This facility
provides light, fluorescence and electron microscopy
services to more than 250 users throughout Johns Hopkins.
As part of a team with other full-time staff, the
candidateís main duties are to help assist in training and
supervising new users, as well as help users troubleshoot
experiments. Because of the diversity of equipment within
the facility, we will provide initial training as necessary.
Consequently, the candidate must display resourceful
independence, a willingness to learn and have strong
analytical and computer skills.
{br /}

Please refer to JHUjobs (http://jobs.jhu.edu/) posting #26703
(https://hrnt.jhu.edu/jhujobs/job_view.cfm?view_req_id=26703)
for details about minimal qualifications and application
process. More qualified individuals will be considered for
appropriate compensation.
{br /}

===============================================================================
{br /}
Scot C. Kuo, skuo-at-jhu.edu, www.jhu.edu/cmml/; www.hopkinsmedicine.org/micfac/
{br /}
Director, Imaging for Basic Biomedical Sciences, (410)955-4536
{br /}
Wood Basic Science Bldg, Room G10, 725 N Wolfe St, Baltimore MD 21205
{br /}
Associate Professor, Biomedical Engineering & Cell Biology
{br /}
===============================================================================
{br /}


---------------------------------------------------------------------------


==============================Original Headers==============================
11, 14 -- From zaluzec-at-microscopy.com Mon Jan 29 21:06:16 2007
11, 14 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
11, 14 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0U36FbG025864
11, 14 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 21:06:16 -0600
11, 14 -- Mime-Version: 1.0
11, 14 -- X-Sender: (Unverified)
11, 14 -- Message-Id: {p06110400c1e466962733-at-[206.69.208.22]}
11, 14 -- Date: Mon, 29 Jan 2007 21:06:14 -0600
11, 14 -- To: microscopy-at-microscopy.com
11, 14 -- From: skuo-at-jhu.edu (by way of MicroscopyListserver)
11, 14 -- Subject: viaWWW: Job Opening: Microscopist
11, 14 -- Content-Type: text/plain; charset="iso-8859-1"
11, 14 -- Content-Transfer-Encoding: 8bit
11, 14 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0U36FbG025864
==============================End of - Headers==============================




From: swtkeller-at-yahoo.com
Date: Mon, 29 Jan 2007 21:06:45 -0600
Subject: [Microscopy] viaWWW: TEM-Defect density Threshold

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both swtkeller-at-yahoo.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: swtkeller-at-yahoo.com
Name: Sandra Keller

Organization: TA/SICCO

Title-Subject: [Filtered] TEM-Defect density Threshold

Question: Hi All:
I was asked by a colleague of mine to inquire from the Listserver: What is the approximate defect density in silicon, one can easily visualize using a 120-200 KV TEM (plane view and cross-sectional view)?

Thanks in advance for being such a valuable resource,
Sandra Keller

---------------------------------------------------------------------------

==============================Original Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Mon Jan 29 21:06:45 2007
6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0U36i9C026687
6, 12 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 21:06:44 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110401c1e466b52e6d-at-[206.69.208.22]}
6, 12 -- Date: Mon, 29 Jan 2007 21:06:42 -0600
6, 12 -- To: microscopy-at-microscopy.com
6, 12 -- From: swtkeller-at-yahoo.com (by way of MicroscopyListserver)
6, 12 -- Subject: viaWWW: TEM-Defect density Threshold
6, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: mconstan-at-princeton.edu
Date: Mon, 29 Jan 2007 21:08:55 -0600
Subject: [Microscopy] AskAMicroscopist SEM beam penetration

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question was submitted to Ask-A-Microscopist by (mconstan-at-princeton.edu)
from http://www.msa.microscopy.org/Ask-A-Microscopist/Ask-A-Microscopist.html on Monday, January 29, 2007 at 02:58:11
Remember to consider the Grade/Age of the student when considering the Question
---------------------------------------------------------------------------
Please reply to both mconstan-at-princeton.edu as well as to the Microscopy Listserver
---------------------------------------------------------------------------

Email: mconstan-at-princeton.edu
Name: Michael

Organization: Princeton University

Education: Undergraduate College

Location: Princeton, NJ, USA

Title: SEM beam penetration

Question: Hi. I'm an undergrad working in a lab and I would like to use SEM to view gold nanoparticles embedded in agarose gel. About how far could the SEM beam penetrate into 1% agarose (I'm guessing I would have to let the water in the gel evaporate)? Would it be better to use SE or BSE? What kV should I try? Thanks for any help you can give me!

---------------------------------------------------------------------------

==============================Original Headers==============================
8, 12 -- From zaluzec-at-ultra5.microscopy.com Mon Jan 29 21:08:54 2007
8, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
8, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0U38rHV001437
8, 12 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007 21:08:54 -0600
8, 12 -- Mime-Version: 1.0
8, 12 -- X-Sender: zaluzec-at-ultra5.microscopy.com (Unverified)
8, 12 -- Message-Id: {p06110402c1e467344c2e-at-[206.69.208.22]}
8, 12 -- Date: Mon, 29 Jan 2007 21:08:52 -0600
8, 12 -- To: microscopy-at-microscopy.com
8, 12 -- From: mconstan-at-princeton.edu (by way of Ask-A-Microscopist)
8, 12 -- Subject: AskAMicroscopist SEM beam penetration
8, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: richard.beanland-at-bookham.com
Date: Tue, 30 Jan 2007 03:10:23 -0600
Subject: [Microscopy] viaWWW: TEM-Defect density Threshold

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Sandra,
It's an easy calculation to do if you make some rough
assumptions about the amount of thin material in your TEM specimen and
the visibility of defects. For a standard cross section which has
roughly 100um of electron transparent material away from the edges of
the hole and is on average 1um thick, you have 0.01 x 0.0001 = 0.000001
cm^2 of the original wafer surface in your cross-section specimen. So
if you see just one defect you have a density of 10^6 defects cm^-2. If
your defects are only visible in thin material then revise the number
upwards. If you have a FIB section (say 30um wide and 0.3 um thick) you
need over 10^8 defects cm^-2 to catch one in your specimen.

For a plan view specimen which has the same 100um electron transparent
area and average thickness of 1um, say around a hole 50um wide then the
amount of the original wafer surface you can see is
pi*(150um)^2-pi*(50um)^2 = 0.0006 cm^2. So if you see one defect you
have a density of about 1600 defects cm^-2.

Having said that I know from experience it can take some time to find
defects in plan view specimens even when the density as high as 10^5
defects cm^-2, simply because the specimen is always bent and you have
to tilt the specimen through a diffraction condition to see them.

Defect etching is very useful for seeing low densities of defects if
processing allows. You can even do a light defect etch and then make a
TEM specimen - in which case you will find that not all of the defects
actually produce an etch pit, which just goes to show that there is no
perfect technique for measuring defect densities.

All the best

Richard



________________________________________
Richard Beanland
Materials Analysis
Bookham
Caswell
Towcester
Northants
NN12 8EQ
United Kingdom
Tel. +44 1327 356362
Fax. +44 1327 356775
http://www.bookham.com
________________________________________


-----Original Message-----
X-from: swtkeller-at-yahoo.com [mailto:swtkeller-at-yahoo.com]
Sent: 30 January 2007 03:08
To: Richard Beanland

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://microscopy.com/MLFormMail.html
------------------------------------------------------------------------
---
Remember this posting is most likely not from a Subscriber, so when
replying
please copy both swtkeller-at-yahoo.com as well as the MIcroscopy
Listserver
------------------------------------------------------------------------
---

Email: swtkeller-at-yahoo.com
Name: Sandra Keller

Organization: TA/SICCO

Title-Subject: [Filtered] TEM-Defect density Threshold

Question: Hi All:
I was asked by a colleague of mine to inquire from the Listserver: What
is the approximate defect density in silicon, one can easily visualize
using a 120-200 KV TEM (plane view and cross-sectional view)?

Thanks in advance for being such a valuable resource,
Sandra Keller

------------------------------------------------------------------------
---

==============================Original
Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Mon Jan 29 21:06:45 2007
6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com
[206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0U36i9C026687
6, 12 -- for {microscopy-at-microscopy.com} ; Mon, 29 Jan 2007
21:06:44 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110401c1e466b52e6d-at-[206.69.208.22]}
6, 12 -- Date: Mon, 29 Jan 2007 21:06:42 -0600
6, 12 -- To: microscopy-at-microscopy.com
6, 12 -- From: swtkeller-at-yahoo.com (by way of MicroscopyListserver)
6, 12 -- Subject: viaWWW: TEM-Defect density Threshold
6, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of -
Headers==============================

=======================================================================
This e-mail is intended for the person it is addressed to only. The
information contained in it may be confidential and/or protected by
law. If you are not the intended recipient of this message, you must
not make any use of this information, or copy or show it to any
person. Please contact us immediately to tell us that you have
received this e-mail, and return the original to us. Any use,
forwarding, printing or copying of this message is strictly prohibited.
No part of this message can be considered a request for goods or
services.
=======================================================================


==============================Original Headers==============================
22, 32 -- From richard.beanland-at-bookham.com Tue Jan 30 03:10:23 2007
22, 32 -- Received: from mail82.messagelabs.com (mail82.messagelabs.com [195.245.231.67])
22, 32 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0U9AMfR031960
22, 32 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 03:10:22 -0600
22, 32 -- X-VirusChecked: Checked
22, 32 -- X-Env-Sender: richard.beanland-at-bookham.com
22, 32 -- X-Msg-Ref: server-6.tower-82.messagelabs.com!1170148220!1610727!3
22, 32 -- X-StarScan-Version: 5.5.10.7.1; banners=bookham.com,-,-
22, 32 -- X-Originating-IP: [213.249.209.179]
22, 32 -- Received: (qmail 23692 invoked from network); 30 Jan 2007 09:10:20 -0000
22, 32 -- Received: from unknown (HELO cas-smx-brdg-02.BOOKHAM.ENTERPRISE.PRI) (213.249.209.179)
22, 32 -- by server-6.tower-82.messagelabs.com with SMTP; 30 Jan 2007 09:10:20 -0000
22, 32 -- Received: from cas-smx-02.BOOKHAM.ENTERPRISE.PRI ([10.4.0.223]) by cas-smx-brdg-02.BOOKHAM.ENTERPRISE.PRI with Microsoft SMTPSVC(6.0.3790.1830);
22, 32 -- Tue, 30 Jan 2007 09:11:38 +0000
22, 32 -- Content-class: urn:content-classes:message
22, 32 -- MIME-Version: 1.0
22, 32 -- Content-Type: text/plain;
22, 32 -- charset="us-ascii"
22, 32 -- X-MimeOLE: Produced By Microsoft Exchange V6.5
22, 32 -- Subject: RE: [Microscopy] viaWWW: TEM-Defect density Threshold
22, 32 -- Date: Tue, 30 Jan 2007 09:11:35 -0000
22, 32 -- Message-ID: {9645D3E33E4C6548B12A7B25F611533E353709-at-cas-smx-02.BOOKHAM.ENTERPRISE.PRI}
22, 32 -- X-MS-Has-Attach:
22, 32 -- X-MS-TNEF-Correlator:
22, 32 -- Thread-Topic: [Microscopy] viaWWW: TEM-Defect density Threshold
22, 32 -- Thread-Index: AcdEG/1pUaZIhLZCRG2NwtjBllRCOwAL5JIw
22, 32 -- From: "Richard Beanland" {richard.beanland-at-bookham.com}
22, 32 -- To: {swtkeller-at-yahoo.com}
22, 32 -- Cc: {microscopy-at-microscopy.com}
22, 32 -- X-OriginalArrivalTime: 30 Jan 2007 09:11:38.0341 (UTC) FILETIME=[A5D0C950:01C7444E]
22, 32 -- Content-Transfer-Encoding: 8bit
22, 32 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0U9AMfR031960
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Tue, 30 Jan 2007 04:07:41 -0600
Subject: [Microscopy] basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi

The simplest solution to your problem is to down load one of the basic Monte
Carlo simulations and consider your material to be Carbon. The Joy-Nockolds
version is on our web site under hints and tips.

Depending upon the microscope that you are using, the magnification, the kV
and the working distance, you could find the so called "secondary electron
signal" (everhart-Thornley detector) carries sufficient backscattered
information to provide the data that you require. Be aware of the fall off
in resolution when using dedicated backscatter as the performance may fall
below that required to visualise your particles. You should also consider
that depending on the microscope and your own skill the size of particles
may be too small to visualise?

You should try the complete range of kV as the information will always be
better at one particular kV. It is impossible perhaps foolish to try and
guess what that may be. You may need higher kVs for penetration but too
much will fog the image with unimportant sub surface data, whilst too low a
kV may not provide the resolution that you require. Microscopists are
scientists we should experiment!

Good luck

Steve Chapman
Protrain for EM training & consultancy world wide
www.emcourses.com
Tel +44 1280 816512 Fax +44 1280 814007
Mobile +44 7711 606967


----- Original Message -----
X-from: {mconstan-at-princeton.edu}
To: {protrain-at-emcourses.com}
Sent: Tuesday, January 30, 2007 3:10 AM

Hello colleagues,

I have (again) a pretty basic, non-life-threatening
question about ultramicrotomy (with a Leica EM UC6 in
my case).
When I am cutting ultrathin sections it may happen I
have to pause for a while (because I have to sign
autographs, or to pick up sections from water). After
the pause when I start the cutting cycle again, it
very often does not cut immediately but I have to cut
500nm or more to touch the block again.
I wondered if this was a sort of automatic protection
from the machine, which retracts a little after some
time of inactivity to avoid damage to the knife, or if
this a the normal consequence of an effect of physics.
Any comment on this?

Stephane




____________________________________________________________________________________
It's here! Your new message!
Get new email alerts with the free Yahoo! Toolbar.
http://tools.search.yahoo.com/toolbar/features/mail/

==============================Original Headers==============================
6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007
6, 19 -- Received: from web37407.mail.mud.yahoo.com (web37407.mail.mud.yahoo.com [209.191.91.139])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0UA7eBg022570
6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 04:07:40 -0600
6, 19 -- Received: (qmail 41697 invoked by uid 60001); 30 Jan 2007 10:07:40 -0000
6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 -- h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
6, 19 -- b=1+gBHYOW2Ee1SsEBAqxWuYWO5/nL8nn468i6i/JHanUAFCuYaxMvYlJD69RYqw8wN8Ck9AxdtzwfdWjcEPbq8VT8OKfknMmjNnA6uBWbdWqpeBvEE2CUqDpIOntBmBc9KIeJGijGR+OPFEeRXti883DSnoTmaxRoxUIQ3t718kI=;
6, 19 -- X-YMail-OSG: F4N2dwUVM1kyUlEszoDiJrQ93JszCzNI5jTxknPqdu3C.NSUiFGAGTXveQLO_7tHUG3Bz6CcIkJT4uNj48uba2WMjJiJYFTgohvdDVQ.1dKt8T8WSpk-
6, 19 -- Received: from [80.122.101.102] by web37407.mail.mud.yahoo.com via HTTP; Tue, 30 Jan 2007 02:07:40 PST
6, 19 -- Date: Tue, 30 Jan 2007 02:07:40 -0800 (PST)
6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: basic question ultramicrotopy
6, 19 -- To: microscopy-at-microscopy.com
6, 19 -- MIME-Version: 1.0
6, 19 -- Content-Type: text/plain; charset=iso-8859-1
6, 19 -- Content-Transfer-Encoding: 8bit
6, 19 -- Message-ID: {389124.40167.qm-at-web37407.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: David.Patton-at-uwe.ac.uk
Date: Tue, 30 Jan 2007 05:57:17 -0600
Subject: [Microscopy] basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Stephane, you are lucky - no accidental 1000nm sections! When I stop to
sign autographs or explain to new users what is going on, the block in
my Ultracut E sneaks forward and I demonstrate how to ruin the knife
with a thick section. I accept this is probably due to operator error
although an expert suggested that the block is compressed during
sectioning and if left for a while will expand forward.

Dave

-----Original Message-----
X-from: nizets2-at-yahoo.com [mailto:nizets2-at-yahoo.com]
Sent: 30 January 2007 10:12
To: David Patton

Hello colleagues,

I have (again) a pretty basic, non-life-threatening
question about ultramicrotomy (with a Leica EM UC6 in
my case).
When I am cutting ultrathin sections it may happen I
have to pause for a while (because I have to sign
autographs, or to pick up sections from water). After
the pause when I start the cutting cycle again, it
very often does not cut immediately but I have to cut
500nm or more to touch the block again.
I wondered if this was a sort of automatic protection
from the machine, which retracts a little after some
time of inactivity to avoid damage to the knife, or if
this a the normal consequence of an effect of physics.
Any comment on this?

Stephane




________________________________________________________________________
____________
It's here! Your new message!
Get new email alerts with the free Yahoo! Toolbar.
http://tools.search.yahoo.com/toolbar/features/mail/

==============================Original
Headers==============================
6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007
6, 19 -- Received: from web37407.mail.mud.yahoo.com
(web37407.mail.mud.yahoo.com [209.191.91.139])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP
id l0UA7eBg022570
6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007
04:07:40 -0600
6, 19 -- Received: (qmail 41697 invoked by uid 60001); 30 Jan 2007
10:07:40 -0000
6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 --
h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Co
ntent-Transfer-Encoding:Message-ID;
6, 19 --
b=1+gBHYOW2Ee1SsEBAqxWuYWO5/nL8nn468i6i/JHanUAFCuYaxMvYlJD69RYqw8wN8Ck9A
xdtzwfdWjcEPbq8VT8OKfknMmjNnA6uBWbdWqpeBvEE2CUqDpIOntBmBc9KIeJGijGR+OPFE
eRXti883DSnoTmaxRoxUIQ3t718kI=;
6, 19 -- X-YMail-OSG:
F4N2dwUVM1kyUlEszoDiJrQ93JszCzNI5jTxknPqdu3C.NSUiFGAGTXveQLO_7tHUG3Bz6Cc
IkJT4uNj48uba2WMjJiJYFTgohvdDVQ.1dKt8T8WSpk-
6, 19 -- Received: from [80.122.101.102] by web37407.mail.mud.yahoo.com
via HTTP; Tue, 30 Jan 2007 02:07:40 PST
6, 19 -- Date: Tue, 30 Jan 2007 02:07:40 -0800 (PST)
6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: basic question ultramicrotopy
6, 19 -- To: microscopy-at-microscopy.com
6, 19 -- MIME-Version: 1.0
6, 19 -- Content-Type: text/plain; charset=iso-8859-1
6, 19 -- Content-Transfer-Encoding: 8bit
6, 19 -- Message-ID: {389124.40167.qm-at-web37407.mail.mud.yahoo.com}
==============================End of -
Headers==============================


This incoming email to UWE has been independently scanned for viruses by
McAfee anti-virus software and none were detected


This email was independently scanned for viruses by McAfee anti-virus software and none were found


==============================Original Headers==============================
18, 34 -- From David.Patton-at-uwe.ac.uk Tue Jan 30 05:57:16 2007
18, 34 -- Received: from mailapp03.uwe.ac.uk (mailapp03.uwe.ac.uk [164.11.132.65])
18, 34 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0UBvFst002960
18, 34 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 05:57:16 -0600
18, 34 -- Received: from (mta02.uwe.ac.uk [164.11.132.62]) by mailapp03.uwe.ac.uk with smtp
18, 34 -- id 6e3e_2833e114_b058_11db_8e32_00142221cca9;
18, 34 -- Tue, 30 Jan 2007 11:51:15 +0000
18, 34 -- Received: from egen-uwe01.campus.ads.uwe.ac.uk
18, 34 -- (egen-uwe01.uwe.ac.uk [164.11.249.121])
18, 34 -- by mta02.uwe.ac.uk (iPlanet Messaging Server 5.2 HotFix 2.07 (built Jun 24
18, 34 -- 2005)) with ESMTP id {0JCO00GXCJV0WN-at-mta02.uwe.ac.uk} for
18, 34 -- microscopy-at-microscopy.com; Tue, 30 Jan 2007 11:57:01 +0000 (GMT)
18, 34 -- Date: Tue, 30 Jan 2007 11:51:13 +0000
18, 34 -- From: David Patton {David.Patton-at-uwe.ac.uk}
18, 34 -- Subject: RE: [Microscopy] basic question ultramicrotopy
18, 34 -- In-reply-to: {200701301011.l0UABfML029533-at-ns.microscopy.com}
18, 34 -- To: nizets2-at-yahoo.com
18, 34 -- Cc: microscopy-at-microscopy.com
18, 34 -- Message-id: {F247F674896BE243AD8263C5280E2BDB02B0B1E5-at-egen-uwe01}
18, 34 -- MIME-version: 1.0
18, 34 -- X-MIMEOLE: Produced By Microsoft Exchange V6.5
18, 34 -- Content-type: text/plain; charset=us-ascii
18, 34 -- Content-class: urn:content-classes:message
18, 34 -- Thread-topic: [Microscopy] basic question ultramicrotopy
18, 34 -- Thread-index: AcdEVxCgRwxzSHS8QpGuRKg0YxRSOwACYQBA
18, 34 -- X-MS-Has-Attach:
18, 34 -- X-MS-TNEF-Correlator:
18, 34 -- X-NAIMIME-Disclaimer: 1
18, 34 -- X-NAIMIME-Modified: 1
18, 34 -- X-NAI-Spam-Score: -2.5
18, 34 -- X-NAI-Spam-Rules: 1 Rules triggered
18, 34 -- BAYES_00=-2.5
18, 34 -- Content-Transfer-Encoding: 8bit
18, 34 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0UBvFst002960
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Tue, 30 Jan 2007 08:19:52 -0600
Subject: [Microscopy] basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Stephane,

I have the opposite problem, maybe because I'm usually asking for
autographs, instead of signing them. If I take a break, I learned the
hard way that the first swipe of the block over the knife will usually
cut off a section about as thick as a slice of bread. I've often
wondered why----especially after one time losing a single layer of cells
which floated off into the sunset. (I had spares, so no worries, mate.)
We have Leica UCTs.

Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


-----Original Message-----
X-from: nizets2-at-yahoo.com [mailto:nizets2-at-yahoo.com]
Sent: Tuesday, January 30, 2007 4:08 AM
To: Tindall, Randy D.

Hello colleagues,

I have (again) a pretty basic, non-life-threatening question about
ultramicrotomy (with a Leica EM UC6 in my case).
When I am cutting ultrathin sections it may happen I have to pause for a
while (because I have to sign autographs, or to pick up sections from
water). After the pause when I start the cutting cycle again, it very
often does not cut immediately but I have to cut 500nm or more to touch
the block again.
I wondered if this was a sort of automatic protection from the machine,
which retracts a little after some time of inactivity to avoid damage to
the knife, or if this a the normal consequence of an effect of physics.
Any comment on this?

Stephane




________________________________________________________________________
____________
It's here! Your new message!
Get new email alerts with the free Yahoo! Toolbar.
http://tools.search.yahoo.com/toolbar/features/mail/

==============================Original
Headers==============================
6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007 6, 19 --
Received: from web37407.mail.mud.yahoo.com (web37407.mail.mud.yahoo.com
[209.191.91.139])
6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP
id l0UA7eBg022570
6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007
04:07:40 -0600
6, 19 -- Received: (qmail 41697 invoked by uid 60001); 30 Jan 2007
10:07:40 -0000 6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
6, 19 -- s=s1024; d=yahoo.com;
6, 19 --
h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Co
ntent-Transfer-Encoding:Message-ID;
6, 19 --
b=1+gBHYOW2Ee1SsEBAqxWuYWO5/nL8nn468i6i/JHanUAFCuYaxMvYlJD69RYqw8wN8Ck9A
xdtzwfdWjcEPbq8VT8OKfknMmjNnA6uBWbdWqpeBvEE2CUqDpIOntBmBc9KIeJGijGR+OPFE
eRXti883DSnoTmaxRoxUIQ3t718kI=;
6, 19 -- X-YMail-OSG:
F4N2dwUVM1kyUlEszoDiJrQ93JszCzNI5jTxknPqdu3C.NSUiFGAGTXveQLO_7tHUG3Bz6Cc
IkJT4uNj48uba2WMjJiJYFTgohvdDVQ.1dKt8T8WSpk-
6, 19 -- Received: from [80.122.101.102] by web37407.mail.mud.yahoo.com
via HTTP; Tue, 30 Jan 2007 02:07:40 PST 6, 19 -- Date: Tue, 30 Jan 2007
02:07:40 -0800 (PST) 6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
6, 19 -- Subject: basic question ultramicrotopy 6, 19 -- To:
microscopy-at-microscopy.com 6, 19 -- MIME-Version: 1.0 6, 19 --
Content-Type: text/plain; charset=iso-8859-1 6, 19 --
Content-Transfer-Encoding: 8bit 6, 19 -- Message-ID:
{389124.40167.qm-at-web37407.mail.mud.yahoo.com}
==============================End of -
Headers==============================



==============================Original Headers==============================
17, 26 -- From TindallR-at-missouri.edu Tue Jan 30 08:19:51 2007
17, 26 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
17, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UEJpVw016138
17, 26 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 08:19:51 -0600
17, 26 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
17, 26 -- Tue, 30 Jan 2007 08:19:50 -0600
17, 26 -- x-mimeole: Produced By Microsoft Exchange V6.5
17, 26 -- Content-class: urn:content-classes:message
17, 26 -- MIME-Version: 1.0
17, 26 -- Content-Type: text/plain;
17, 26 -- charset="us-ascii"
17, 26 -- Subject: RE: [Microscopy] basic question ultramicrotopy
17, 26 -- Date: Tue, 30 Jan 2007 08:19:51 -0600
17, 26 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D85-at-UM-XMAIL08.um.umsystem.edu}
17, 26 -- In-Reply-To: {200701301008.l0UA8QDp024543-at-ns.microscopy.com}
17, 26 -- X-MS-Has-Attach:
17, 26 -- X-MS-TNEF-Correlator:
17, 26 -- Thread-Topic: [Microscopy] basic question ultramicrotopy
17, 26 -- Thread-Index: AcdEVpXqhzJvu355StiQwG7GHLeYTAAIprEg
17, 26 -- References: {200701301008.l0UA8QDp024543-at-ns.microscopy.com}
17, 26 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
17, 26 -- To: {nizets2-at-yahoo.com}
17, 26 -- Cc: {microscopy-at-microscopy.com}
17, 26 -- X-OriginalArrivalTime: 30 Jan 2007 14:19:50.0827 (UTC) FILETIME=[B43237B0:01C74479]
17, 26 -- Content-Transfer-Encoding: 8bit
17, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0UEJpVw016138
==============================End of - Headers==============================




From: Michal.Jarnik-at-fccc.edu
Date: Tue, 30 Jan 2007 08:40:45 -0600
Subject: [Microscopy] Re: basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

I don't really think this is an operator error, it does happen all the
time, standard or cryo, I am not sure why. After breaking a couple of
cryoblocks (not ruining any diamond on plastic, so far), I always go
200-400 nm back (using coarse advance) when pausing for section
retrieval. Usually I have to wait for 3-4 cycles for the microtome to
start cutting again, but it solves the problem.

Michal

nizets2-at-yahoo.com wrote:
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} Hello colleagues,
}
} I have (again) a pretty basic, non-life-threatening
} question about ultramicrotomy (with a Leica EM UC6 in
} my case).
} When I am cutting ultrathin sections it may happen I
} have to pause for a while (because I have to sign
} autographs, or to pick up sections from water). After
} the pause when I start the cutting cycle again, it
} very often does not cut immediately but I have to cut
} 500nm or more to touch the block again.
} I wondered if this was a sort of automatic protection
} from the machine, which retracts a little after some
} time of inactivity to avoid damage to the knife, or if
} this a the normal consequence of an effect of physics.
} Any comment on this?
}
} Stephane
}
}
}
}
} ____________________________________________________________________________________
} It's here! Your new message!
} Get new email alerts with the free Yahoo! Toolbar.
} http://tools.search.yahoo.com/toolbar/features/mail/
}
} ==============================Original Headers==============================
} 6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007
} 6, 19 -- Received: from web37407.mail.mud.yahoo.com (web37407.mail.mud.yahoo.com [209.191.91.139])
} 6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0UA7eBg022570
} 6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 04:07:40 -0600
} 6, 19 -- Received: (qmail 41697 invoked by uid 60001); 30 Jan 2007 10:07:40 -0000
} 6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
} 6, 19 -- s=s1024; d=yahoo.com;
} 6, 19 -- h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
} 6, 19 -- b=1+gBHYOW2Ee1SsEBAqxWuYWO5/nL8nn468i6i/JHanUAFCuYaxMvYlJD69RYqw8wN8Ck9AxdtzwfdWjcEPbq8VT8OKfknMmjNnA6uBWbdWqpeBvEE2CUqDpIOntBmBc9KIeJGijGR+OPFEeRXti883DSnoTmaxRoxUIQ3t718kI=;
} 6, 19 -- X-YMail-OSG: F4N2dwUVM1kyUlEszoDiJrQ93JszCzNI5jTxknPqdu3C.NSUiFGAGTXveQLO_7tHUG3Bz6CcIkJT4uNj48uba2WMjJiJYFTgohvdDVQ.1dKt8T8WSpk-
} 6, 19 -- Received: from [80.122.101.102] by web37407.mail.mud.yahoo.com via HTTP; Tue, 30 Jan 2007 02:07:40 PST
} 6, 19 -- Date: Tue, 30 Jan 2007 02:07:40 -0800 (PST)
} 6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
} 6, 19 -- Subject: basic question ultramicrotopy
} 6, 19 -- To: microscopy-at-microscopy.com
} 6, 19 -- MIME-Version: 1.0
} 6, 19 -- Content-Type: text/plain; charset=iso-8859-1
} 6, 19 -- Content-Transfer-Encoding: 8bit
} 6, 19 -- Message-ID: {389124.40167.qm-at-web37407.mail.mud.yahoo.com}
} ==============================End of - Headers==============================
}

==============================Original Headers==============================
3, 19 -- From Michal.Jarnik-at-fccc.edu Tue Jan 30 08:40:45 2007
3, 19 -- Received: from azah.fccc.edu (azah.fccc.edu [131.249.4.237])
3, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UEejgq027634
3, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 08:40:45 -0600
3, 19 -- Received: from [131.249.6.85] (emf1.fccc.edu [131.249.6.85])
3, 19 -- (authenticated bits=0)
3, 19 -- by azah.fccc.edu (8.13.6/8.13.6) with ESMTP id l0UEejK7011668
3, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 09:40:45 -0500 (EST)
3, 19 -- Message-ID: {45BF58E9.4050809-at-fccc.edu}
3, 19 -- Date: Tue, 30 Jan 2007 09:40:41 -0500
3, 19 -- From: Michal Jarnik {Michal.Jarnik-at-fccc.edu}
3, 19 -- User-Agent: Thunderbird 1.5.0.5 (Macintosh/20060719)
3, 19 -- MIME-Version: 1.0
3, 19 -- To: microscopy-at-microscopy.com
3, 19 -- Subject: Re: [Microscopy] basic question ultramicrotopy
3, 19 -- References: {200701301013.l0UADfhx032601-at-ns.microscopy.com}
3, 19 -- In-Reply-To: {200701301013.l0UADfhx032601-at-ns.microscopy.com}
3, 19 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
3, 19 -- Content-Transfer-Encoding: 7bit
==============================End of - Headers==============================




From: thoward-at-unm.edu
Date: Tue, 30 Jan 2007 09:01:49 -0600
Subject: [Microscopy] Re: basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

I must be doing something very wrong - I'm only called
away from the microtome to be informed of malfunctioning
microscopes/gel boxes/processors/etc. - never to sign
autographs.
When I return from the crisis, sometimes the block is
further forward, sometimes back - I always retract the
knife out of paranoia. I think the block moving away from
the knife is due to thermal effects; you may generate
enough friction during cutting to cause thermal expansion
of the block, which then shrinks back when you stop
cutting. At least, that is what I was told when I was
learning to cut. I think it moves forward out of pure
contrariness. I know on the old thermal advance microtomes
you were pretty much guaranteed a trashed block/knife if
you didn't retract when re-starting...there were all kinds
of odd behaviors associated with those machines.

Tamara


On Tue, 30 Jan 2007 04:08:34 -0600
nizets2-at-yahoo.com wrote:
}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy
} Society of America
} To Subscribe/Unsubscribe --
} http://www.microscopy.com/MicroscopyListserver
} On-Line Help
} http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} Hello colleagues,
}
} I have (again) a pretty basic, non-life-threatening
} question about ultramicrotomy (with a Leica EM UC6 in
} my case).
} When I am cutting ultrathin sections it may happen I
} have to pause for a while (because I have to sign
} autographs, or to pick up sections from water). After
} the pause when I start the cutting cycle again, it
} very often does not cut immediately but I have to cut
} 500nm or more to touch the block again.
} I wondered if this was a sort of automatic protection
} from the machine, which retracts a little after some
} time of inactivity to avoid damage to the knife, or if
} this a the normal consequence of an effect of physics.
} Any comment on this?
}
} Stephane
}
}
}
}
} ____________________________________________________________________________________
} It's here! Your new message!
} Get new email alerts with the free Yahoo! Toolbar.
} http://tools.search.yahoo.com/toolbar/features/mail/
}
} ==============================Original
} Headers==============================
} 6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007
} 6, 19 -- Received: from web37407.mail.mud.yahoo.com
} (web37407.mail.mud.yahoo.com [209.191.91.139])
} 6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8)



From: David.Patton-at-uwe.ac.uk
Date: Tue, 30 Jan 2007 09:15:00 -0600
Subject: [Microscopy] Re: basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

On a related theme. Without pausing to sign autographs etc has anyone
demonstrated that turning the "wheel" only moves the block up and down,
only to have puzzled students say "But it is cutting sections."?

Dave

-----Original Message-----
X-from: thoward-at-unm.edu [mailto:thoward-at-unm.edu]
Sent: 30 January 2007 15:06
To: David Patton

I must be doing something very wrong - I'm only called
away from the microtome to be informed of malfunctioning
microscopes/gel boxes/processors/etc. - never to sign
autographs.
When I return from the crisis, sometimes the block is
further forward, sometimes back - I always retract the
knife out of paranoia. I think the block moving away from
the knife is due to thermal effects; you may generate
enough friction during cutting to cause thermal expansion
of the block, which then shrinks back when you stop
cutting. At least, that is what I was told when I was
learning to cut. I think it moves forward out of pure
contrariness. I know on the old thermal advance microtomes
you were pretty much guaranteed a trashed block/knife if
you didn't retract when re-starting...there were all kinds
of odd behaviors associated with those machines.

Tamara


On Tue, 30 Jan 2007 04:08:34 -0600
nizets2-at-yahoo.com wrote:
}
}
}
}
------------------------------------------------------------------------
----
} The Microscopy ListServer -- CoSponsor: The Microscopy
} Society of America
} To Subscribe/Unsubscribe --
} http://www.microscopy.com/MicroscopyListserver
} On-Line Help
} http://www.microscopy.com/MicroscopyListserver/FAQ.html
}
------------------------------------------------------------------------
----
}
} Hello colleagues,
}
} I have (again) a pretty basic, non-life-threatening
} question about ultramicrotomy (with a Leica EM UC6 in
} my case).
} When I am cutting ultrathin sections it may happen I
} have to pause for a while (because I have to sign
} autographs, or to pick up sections from water). After
} the pause when I start the cutting cycle again, it
} very often does not cut immediately but I have to cut
} 500nm or more to touch the block again.
} I wondered if this was a sort of automatic protection
} from the machine, which retracts a little after some
} time of inactivity to avoid damage to the knife, or if
} this a the normal consequence of an effect of physics.
} Any comment on this?
}
} Stephane
}
}
}
}
}
________________________________________________________________________
____________
} It's here! Your new message!
} Get new email alerts with the free Yahoo! Toolbar.
} http://tools.search.yahoo.com/toolbar/features/mail/
}
} ==============================Original
} Headers==============================
} 6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007
} 6, 19 -- Received: from web37407.mail.mud.yahoo.com
} (web37407.mail.mud.yahoo.com [209.191.91.139])
} 6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8)



From: rblyston-at-trinity.edu
Date: Tue, 30 Jan 2007 09:31:11 -0600
Subject: [Microscopy] Differential Ultramicrotomy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

To all:

The chucked block in the unattended ultramicrotome obeys the
Hehesenberg Uncertainty Principle. As I trust you remember the
position of a chucked block whose momentary momentum is zero will
come to rest at a less accurate location. When the circular motion
of the microtome wheel approaches zero then the product of the
position and momentum approaches the Kerplank's constant. Thus using
more mathematical rigor the conjugate quantity of the sections
becomes the standard deviation from the property of being silver.
Associated with this pause in the angular momentum of the microtome
will be section chatter due to wave-particle duality. Ultramicrotome
manufactures have tried to address this problem with an adaptor known
as the von Neumann measurement, which according to some is better
than the Landau calculation.

I hope recalling this information lays to rest in no uncertain terms
the phenomenon of zero momentum block creep or in some cases inverse
block creep.

Bob Blystone



==============================Original Headers==============================
6, 17 -- From rblyston-at-trinity.edu Tue Jan 30 09:31:11 2007
6, 17 -- Received: from ms-smtp-04.texas.rr.com (ms-smtp-04.texas.rr.com [24.93.47.43])
6, 17 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UFVBCS029656
6, 17 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 09:31:11 -0600
6, 17 -- Received: from [192.168.0.2] (cpe-66-69-93-33.satx.res.rr.com [66.69.93.33])
6, 17 -- by ms-smtp-04.texas.rr.com (8.13.6/8.13.6) with ESMTP id l0UFV7IW003436
6, 17 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 09:31:08 -0600 (CST)
6, 17 -- Mime-Version: 1.0 (Apple Message framework v752.2)
6, 17 -- Content-Transfer-Encoding: 7bit
6, 17 -- Message-Id: {32791431-6E70-43AE-BC6B-BCEF6BEFCD19-at-trinity.edu}
6, 17 -- Content-Type: text/plain; charset=US-ASCII; delsp=yes; format=flowed
6, 17 -- To: Microscopy-at-microscopy.com
6, 17 -- From: Robert Blystone {rblyston-at-trinity.edu}
6, 17 -- Subject: Differential Ultramicrotomy
6, 17 -- Date: Tue, 30 Jan 2007 09:31:04 -0600
6, 17 -- X-Mailer: Apple Mail (2.752.2)
6, 17 -- X-Virus-Scanned: Symantec AntiVirus Scan Engine
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Tue, 30 Jan 2007 09:36:46 -0600
Subject: [Microscopy] Differential Ultramicrotomy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Zowie, Bob! Can I have your autograph?

Randy

-----Original Message-----
X-from: rblyston-at-trinity.edu [mailto:rblyston-at-trinity.edu]
Sent: Tuesday, January 30, 2007 9:32 AM
To: Tindall, Randy D.

To all:

The chucked block in the unattended ultramicrotome obeys the Hehesenberg
Uncertainty Principle. As I trust you remember the position of a
chucked block whose momentary momentum is zero will come to rest at a
less accurate location. When the circular motion of the microtome wheel
approaches zero then the product of the position and momentum approaches
the Kerplank's constant. Thus using more mathematical rigor the
conjugate quantity of the sections
becomes the standard deviation from the property of being silver.
Associated with this pause in the angular momentum of the microtome will
be section chatter due to wave-particle duality. Ultramicrotome
manufactures have tried to address this problem with an adaptor known as
the von Neumann measurement, which according to some is better than the
Landau calculation.

I hope recalling this information lays to rest in no uncertain terms the
phenomenon of zero momentum block creep or in some cases inverse block
creep.

Bob Blystone



==============================Original
Headers==============================
6, 17 -- From rblyston-at-trinity.edu Tue Jan 30 09:31:11 2007 6, 17 --
Received: from ms-smtp-04.texas.rr.com (ms-smtp-04.texas.rr.com
[24.93.47.43])
6, 17 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0UFVBCS029656
6, 17 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007
09:31:11 -0600
6, 17 -- Received: from [192.168.0.2] (cpe-66-69-93-33.satx.res.rr.com
[66.69.93.33])
6, 17 -- by ms-smtp-04.texas.rr.com (8.13.6/8.13.6) with ESMTP id
l0UFV7IW003436
6, 17 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007
09:31:08 -0600 (CST)
6, 17 -- Mime-Version: 1.0 (Apple Message framework v752.2) 6, 17 --
Content-Transfer-Encoding: 7bit 6, 17 -- Message-Id:
{32791431-6E70-43AE-BC6B-BCEF6BEFCD19-at-trinity.edu}
6, 17 -- Content-Type: text/plain; charset=US-ASCII; delsp=yes;
format=flowed 6, 17 -- To: Microscopy-at-microscopy.com 6, 17 -- From:
Robert Blystone {rblyston-at-trinity.edu} 6, 17 -- Subject: Differential
Ultramicrotomy 6, 17 -- Date: Tue, 30 Jan 2007 09:31:04 -0600 6, 17 --
X-Mailer: Apple Mail (2.752.2) 6, 17 -- X-Virus-Scanned: Symantec
AntiVirus Scan Engine ==============================End of -
Headers==============================



==============================Original Headers==============================
15, 26 -- From TindallR-at-missouri.edu Tue Jan 30 09:36:46 2007
15, 26 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
15, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UFakrK007626
15, 26 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 09:36:46 -0600
15, 26 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
15, 26 -- Tue, 30 Jan 2007 09:36:46 -0600
15, 26 -- x-mimeole: Produced By Microsoft Exchange V6.5
15, 26 -- Content-class: urn:content-classes:message
15, 26 -- MIME-Version: 1.0
15, 26 -- Content-Type: text/plain;
15, 26 -- charset="us-ascii"
15, 26 -- Subject: RE: [Microscopy] Differential Ultramicrotomy
15, 26 -- Date: Tue, 30 Jan 2007 09:36:46 -0600
15, 26 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D88-at-UM-XMAIL08.um.umsystem.edu}
15, 26 -- In-Reply-To: {200701301532.l0UFWJKo031669-at-ns.microscopy.com}
15, 26 -- X-MS-Has-Attach:
15, 26 -- X-MS-TNEF-Correlator:
15, 26 -- Thread-Topic: [Microscopy] Differential Ultramicrotomy
15, 26 -- Thread-Index: AcdEg9Tipa8emWkETBKffY/6uJLcTQAAF7eA
15, 26 -- References: {200701301532.l0UFWJKo031669-at-ns.microscopy.com}
15, 26 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
15, 26 -- To: {rblyston-at-trinity.edu}
15, 26 -- Cc: {microscopy-at-microscopy.com}
15, 26 -- X-OriginalArrivalTime: 30 Jan 2007 15:36:46.0011 (UTC) FILETIME=[730F68B0:01C74484]
15, 26 -- Content-Transfer-Encoding: 8bit
15, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0UFakrK007626
==============================End of - Headers==============================




From: pbond-at-plymouth.ac.uk
Date: Tue, 30 Jan 2007 10:35:08 -0600
Subject: [Microscopy] RE: Differential Ultramicrotomy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

And I thought it was loads more complicated than that!

Pet




} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


--
Peter Bond
Scientific Officer
Plymouth Electron Microscopy Centre
University of Plymouth
Drake Circus
Plymouth
Devon
UK PL4 8AA
Tel: 44 (0)1752 233092
pbond-at-plymouth.ac.uk
http://www.plymouth.ac.uk/emc

==============================Original Headers==============================
8, 16 -- From P.Bond-at-plymouth.ac.uk Tue Jan 30 10:35:04 2007
8, 16 -- Received: from ils103.uopnet.plymouth.ac.uk (ils103.uopnet.plymouth.ac.uk [141.163.66.106])
8, 16 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UGZ3nf021060
8, 16 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 10:35:04 -0600
8, 16 -- Received: from [141.163.37.171] ([141.163.37.171]) by ils103.uopnet.plymouth.ac.uk with Microsoft SMTPSVC(6.0.3790.1830);
8, 16 -- Tue, 30 Jan 2007 16:35:02 +0000
8, 16 -- Mime-Version: 1.0
8, 16 -- Message-Id: {a06240800c1e523a52209-at-[141.163.37.171]}
8, 16 -- In-Reply-To: {200701301539.l0UFd5DE012347-at-ns.microscopy.com}
8, 16 -- References: {200701301539.l0UFd5DE012347-at-ns.microscopy.com}
8, 16 -- Date: Tue, 30 Jan 2007 16:34:07 +0000
8, 16 -- To: microscopy-at-microscopy.com
8, 16 -- From: Peter Bond {pbond-at-plymouth.ac.uk}
8, 16 -- Subject: [Microscopy] RE: Differential Ultramicrotomy
8, 16 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
8, 16 -- X-OriginalArrivalTime: 30 Jan 2007 16:35:02.0439 (UTC) FILETIME=[9717FB70:01C7448C]
==============================End of - Headers==============================




From: glenmac-at-u.washington.edu
Date: Tue, 30 Jan 2007 11:47:31 -0600
Subject: [Microscopy] Re: basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

During the cutting process, the blade is compressing the remainder of
the block, which then expands afterwards.
My mentors led me to believe the overlap with the knife was due to
expansion of the block whenever it was allowed to 'rest'. The
forces equalize when sectioning steadily. This is one reason
microtomes for ulta-thin or semi-thin ( {3 µm) sections retract
during the return stroke.
I don't know if this is really the case, but it does seem to account
for the fact that we can break knives and blocks if not separated
before resuming sectioning.

Regards,
Glen

Glen MacDonald
Core for Communication Research
Virginia Merrill Bloedel Hearing Research Center
Box 357923
University of Washington
Seattle, WA 98195-7923 USA
(206) 616-4156
glenmac-at-u.washington.edu

************************************************************************
******
The box said "Requires WindowsXP or better", so I bought a Macintosh.
************************************************************************
******


On Jan 30, 2007, at 2:09 AM, nizets2-at-yahoo.com wrote:

}
}
}
} ----------------------------------------------------------------------
} ------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of
} America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/
} MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------
} ------
}
} Hello colleagues,
}
} I have (again) a pretty basic, non-life-threatening
} question about ultramicrotomy (with a Leica EM UC6 in
} my case).
} When I am cutting ultrathin sections it may happen I
} have to pause for a while (because I have to sign
} autographs, or to pick up sections from water). After
} the pause when I start the cutting cycle again, it
} very often does not cut immediately but I have to cut
} 500nm or more to touch the block again.
} I wondered if this was a sort of automatic protection
} from the machine, which retracts a little after some
} time of inactivity to avoid damage to the knife, or if
} this a the normal consequence of an effect of physics.
} Any comment on this?
}
} Stephane
}
}
}
}
} ______________________________________________________________________
} ______________
} It's here! Your new message!
} Get new email alerts with the free Yahoo! Toolbar.
} http://tools.search.yahoo.com/toolbar/features/mail/
}
} ==============================Original
} Headers==============================
} 6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007
} 6, 19 -- Received: from web37407.mail.mud.yahoo.com
} (web37407.mail.mud.yahoo.com [209.191.91.139])
} 6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP
} id l0UA7eBg022570
} 6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007
} 04:07:40 -0600
} 6, 19 -- Received: (qmail 41697 invoked by uid 60001); 30 Jan 2007
} 10:07:40 -0000
} 6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
} 6, 19 -- s=s1024; d=yahoo.com;
} 6, 19 -- h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-
} Version:Content-Type:Content-Transfer-Encoding:Message-ID;
} 6, 19 -- b=1+gBHYOW2Ee1SsEBAqxWuYWO5/nL8nn468i6i/
} JHanUAFCuYaxMvYlJD69RYqw8wN8Ck9AxdtzwfdWjcEPbq8VT8OKfknMmjNnA6uBWbdWqp
} eBvEE2CUqDpIOntBmBc9KIeJGijGR+OPFEeRXti883DSnoTmaxRoxUIQ3t718kI=;
} 6, 19 -- X-YMail-OSG:
} F4N2dwUVM1kyUlEszoDiJrQ93JszCzNI5jTxknPqdu3C.NSUiFGAGTXveQLO_7tHUG3Bz6
} CcIkJT4uNj48uba2WMjJiJYFTgohvdDVQ.1dKt8T8WSpk-
} 6, 19 -- Received: from [80.122.101.102] by
} web37407.mail.mud.yahoo.com via HTTP; Tue, 30 Jan 2007 02:07:40 PST
} 6, 19 -- Date: Tue, 30 Jan 2007 02:07:40 -0800 (PST)
} 6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
} 6, 19 -- Subject: basic question ultramicrotopy
} 6, 19 -- To: microscopy-at-microscopy.com
} 6, 19 -- MIME-Version: 1.0
} 6, 19 -- Content-Type: text/plain; charset=iso-8859-1
} 6, 19 -- Content-Transfer-Encoding: 8bit
} 6, 19 -- Message-ID: {389124.40167.qm-at-web37407.mail.mud.yahoo.com}
} ==============================End of -
} Headers==============================



==============================Original Headers==============================
9, 27 -- From glenmac-at-u.washington.edu Tue Jan 30 11:47:30 2007
9, 27 -- Received: from mxout2.cac.washington.edu (mxout2.cac.washington.edu [140.142.33.4])
9, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UHlUkJ001376
9, 27 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 11:47:30 -0600
9, 27 -- Received: from smtp.washington.edu (smtp.washington.edu [140.142.32.139])
9, 27 -- by mxout2.cac.washington.edu (8.13.7+UW06.06/8.13.7+UW06.09) with ESMTP id l0UHlTxE009278
9, 27 -- (version=TLSv1/SSLv3 cipher=DHE-RSA-AES256-SHA bits=256 verify=OK)
9, 27 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 09:47:29 -0800
9, 27 -- X-Auth-Received: from [128.95.178.71] ([128.95.178.71])
9, 27 -- (authenticated authid=glenmac)
9, 27 -- by smtp.washington.edu (8.13.7+UW06.06/8.13.7+UW06.09) with ESMTP id l0UHlTqn003865
9, 27 -- (version=TLSv1/SSLv3 cipher=AES128-SHA bits=128 verify=NOT)
9, 27 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 09:47:29 -0800
9, 27 -- Mime-Version: 1.0 (Apple Message framework v752.2)
9, 27 -- In-Reply-To: {200701301009.l0UA9Yqu025080-at-ns.microscopy.com}
9, 27 -- References: {200701301009.l0UA9Yqu025080-at-ns.microscopy.com}
9, 27 -- Content-Type: text/plain; charset=ISO-8859-1; delsp=yes; format=flowed
9, 27 -- Message-Id: {F61A3F01-BA57-4E26-8E7C-5FF4E04D8C08-at-u.washington.edu}
9, 27 -- From: Glen MacDonald {glenmac-at-u.washington.edu}
9, 27 -- Subject: Re: [Microscopy] basic question ultramicrotopy
9, 27 -- Date: Tue, 30 Jan 2007 09:47:24 -0800
9, 27 -- To: "ListServer-at-MSA.Microscopy.Com Listserver" {microscopy-at-microscopy.com}
9, 27 -- X-Mailer: Apple Mail (2.752.2)
9, 27 -- X-PMX-Version: 5.3.0.289146, Antispam-Engine: 2.5.0.283055, Antispam-Data: 2007.1.30.93432
9, 27 -- X-Uwash-Spam: Gauge=IIIIIII, Probability=7%, Report='__C230066_P5 0, __CP_URI_IN_BODY 0, __CT 0, __CTE 0, __CT_TEXT_PLAIN 0, __HAS_MSGID 0, __HAS_X_MAILER 0, __MIME_TEXT_ONLY 0, __MIME_VERSION 0, __SANE_MSGID 0, __STOCK_PHRASE_24 0'
9, 27 -- Content-Transfer-Encoding: 8bit
9, 27 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0UHlUkJ001376
==============================End of - Headers==============================




From: opmills-at-mtu.edu
Date: Tue, 30 Jan 2007 12:05:15 -0600
Subject: [Microscopy] poor mans TEM camera

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

All,

Our digital TEM camera is dead, the TEM is on it's last leg so I am
not going to have the camera repaired. We use this TEM for teaching
and it is difficult to teach without a camera. I am thinking that
there may be a way to install a removable low cost video camera on
the binocular eyepiece for some applications - like teaching. Has
anyone done this?

Owen

Owen P. Mills
Director, Materials Characterization & Fabrication Facilities
Electron Optics Engineer, Applied Chemical & Morphological Analysis
Laboratory

Materials Science & Engineering
Michigan Technological University
Rm 512 M&M Bldg.
Houghton, MI 49931
PH 906-369-1875
FAX 906-487-2934
mailto:opmills-at-mtu.edu
http://www.mm.mtu.edu/~opmills



==============================Original Headers==============================
7, 31 -- From opmills-at-mtu.edu Tue Jan 30 12:05:14 2007
7, 31 -- Received: from node8.mtu.edu (node8.mtu.edu [141.219.69.8])
7, 31 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UI5EhD012715
7, 31 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 12:05:14 -0600
7, 31 -- Received: from node5.edge.dcsint.mtu.edu (node5.mtu.edu [141.219.69.5])
7, 31 -- by node8.mtu.edu (8.13.1/8.13.1) with ESMTP id l0UI5Cbh028990
7, 31 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 13:05:12 -0500
7, 31 -- Received: from node34.edge.dcsint.mtu.edu (node34.mtu.edu [141.219.69.34])
7, 31 -- by node5.edge.dcsint.mtu.edu (8.12.11.20060308/8.12.8) with ESMTP id l0UI5CDK016955
7, 31 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 13:05:12 -0500
7, 31 -- Received: from mail.mtu.edu (campus0.mtu.edu [141.219.70.8])
7, 31 -- by node34.edge.dcsint.mtu.edu (8.13.8/8.13.8) with ESMTP id l0UI5B9s005195
7, 31 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 13:05:11 -0500
7, 31 -- (envelope-from opmills-at-mtu.edu)
7, 31 -- Received: from node6.edge.dcsint.mtu.edu (node6.mtu.edu [141.219.69.6])
7, 31 -- by mail.mtu.edu (8.13.6+Sun/8.11.6) with ESMTP id l0UI5BHC004120
7, 31 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 13:05:11 -0500 (EST)
7, 31 -- Received: from [141.219.192.45] (opmills.eecn.sabo.mtu.edu [141.219.192.45])
7, 31 -- by node6.edge.dcsint.mtu.edu (8.12.11.20060308/8.12.3/auth_ssl.mc v1.0) with ESMTP id l0UI5BwO030489
7, 31 -- for {Microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 13:05:11 -0500
7, 31 -- Mime-Version: 1.0 (Apple Message framework v752.3)
7, 31 -- Content-Transfer-Encoding: 7bit
7, 31 -- Message-Id: {248461F6-4D9A-4071-8C6C-8B30D858EE0E-at-mtu.edu}
7, 31 -- Content-Type: text/plain; charset=US-ASCII; delsp=yes; format=flowed
7, 31 -- To: Microscopy Listserver {Microscopy-at-microscopy.com}
7, 31 -- From: "Owen P. Mills" {opmills-at-mtu.edu}
7, 31 -- Subject: poor mans TEM camera
7, 31 -- Date: Tue, 30 Jan 2007 13:05:08 -0500
7, 31 -- X-Mailer: Apple Mail (2.752.3)
7, 31 -- X-PMX-Version: 5.3.0.289146, Antispam-Engine: 2.5.0.283055, Antispam-Data: 2007.1.30.95432
7, 31 -- X-PerlMx-Spam: Gauge=IIIIIII, Probability=7%, Report='__C230066_P5 0, __CP_MEDIA_BODY 0, __CP_URI_IN_BODY 0, __CT 0, __CTE 0, __CT_TEXT_PLAIN 0, __HAS_MSGID 0, __HAS_X_MAILER 0, __MIME_TEXT_ONLY 0, __MIME_VERSION 0, __SANE_MSGID 0'
==============================End of - Headers==============================




From: garsha-at-itg.uiuc.edu
Date: Tue, 30 Jan 2007 14:55:32 -0600
Subject: [Microscopy] tolerance specifications for standard microscope slide?

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Does anybody happen to know of a reference or resource for the size and
tolerance specifications for a conventional 25x75 mm microscope slide?
Thanks in advance.

--
Karl Garsha
Head Applications Scientist
Roper Bioscience Advanced Microimaging Group
3440 E. Brittania Drive
Tucson, AZ 85706
Office: 520-547-2704


==============================Original Headers==============================
3, 27 -- From garsha-at-itg.uiuc.edu Tue Jan 30 14:55:31 2007
3, 27 -- Received: from outmail128166.authsmtp.net (outmail128166.authsmtp.net [62.13.128.166])
3, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UKtUAi028471
3, 27 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 14:55:31 -0600
3, 27 -- Received: from outmail128185.authsmtp.co.uk (outmail128185.authsmtp.co.uk [62.13.128.185])
3, 27 -- by punt2.authsmtp.com (8.13.8/8.13.8/Kp) with ESMTP id l0UKtSv8057612
3, 27 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 20:55:29 GMT
3, 27 -- Received: from [144.22.2.112] (66-162-43-45.static.twtelecom.net [66.162.43.45])
3, 27 -- (authenticated bits=0)
3, 27 -- by mail.authsmtp.com (8.13.8/8.13.8/Kp) with ESMTP id l0UKtM3P044076
3, 27 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 20:55:23 GMT
3, 27 -- Message-ID: {45BFB0B6.709-at-itg.uiuc.edu}
3, 27 -- Date: Tue, 30 Jan 2007 13:55:18 -0700
3, 27 -- From: Karl Garsha {garsha-at-itg.uiuc.edu}
3, 27 -- Reply-To: kgarsha-at-roperscientific.com
3, 27 -- Organization: Roper Scientific
3, 27 -- User-Agent: Thunderbird 1.5.0.9 (Windows/20061207)
3, 27 -- MIME-Version: 1.0
3, 27 -- To: microscopy-at-microscopy.com
3, 27 -- Subject: tolerance specifications for standard microscope slide?
3, 27 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
3, 27 -- Content-Transfer-Encoding: 7bit
3, 27 -- X-Server-Quench: 34aa796d-b0a4-11db-8638-001185d377ca
3, 27 -- X-AuthRoute: OCdyaAgTClZaRR4B CiosDTNPDxgkOxYK DBMeOw5bK0AOTg9W KldyK1tYKloHTlZB SnhYBAkaU11uIzIr dAhRbQJNYEpEWgBg UkwHRFRMFQNqHxgC GBobTRt3dQBZeDAs ZSMZGg1YPER+cUZ6 RABTE25IZmMxOWcW VhRFJQMFdB5Lf0sX d1h5V3sQYWUGZ3Jl E1RsYDs4KylYLClP SwcGIBoJWUsAHzA9 TBkeHDIpBgULQD97 NAAvLFIVBkpZN0Q5 K1w6UlUAPlcJDgxS EhYl
3, 27 -- X-Authentic-SMTP: 61633138303639.squirrel.dmpriest.net.uk:536/Kp
3, 27 -- X-Report-SPAM: If SPAM / abuse - report it at: http://www.authsmtp.com/abuse
3, 27 -- X-Virus-Status: No virus detected - but ensure you scan with your own anti-virus system!
==============================End of - Headers==============================




From: tina-at-pbrc.hawaii.edu
Date: Tue, 30 Jan 2007 14:59:18 -0600
Subject: [Microscopy] basic question ultramicrotopy (fwd)

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi, Stephane-

I've had it both ways; block retracts and when I restart it misses, or
block expands and when I restart it whacks the knife. Depends on the type
of resin, how hard it is, the sample in the resin, temperature, humidity,
all kinds of things. Some of my stuff compresses with lengthy sectioning
and the "relaxes" forward when I stop. Other times the block seems to heat
up and be expanding during sectioning, then cools off and retracts when I
stop. In all cases I have learned to retract the knife before starting
again, just in case. Annoying, yes, but better safe than sorry.

I had one customer who had lots of material embedded in LR White many,
many years ago, and brought them in whenever she wanted to try a new
antibody. I had become familiar with them and knew that over time the
tissue (squid parts) expanded out of the face of the blocks. I guess the
tissue was softer than the surrounding resin. I used this "feature" once
when all the TEMs in the state were down and they were desparate to see
if they had immunogold staining. I put the labeled grids in the SEM and
was able to see the features of the tissue by topography, and the
colloidal gold sitting on top.

Please; no autographs!

Aloha,
Tina

} I have (again) a pretty basic, non-life-threatening question about
} ultramicrotomy (with a Leica EM UC6 in my case).
} When I am cutting ultrathin sections it may happen I have to pause for a
} while (because I have to sign autographs, or to pick up sections from
} water). After the pause when I start the cutting cycle again, it very
} often does not cut immediately but I have to cut 500nm or more to touch
} the block again.
} I wondered if this was a sort of automatic protection from the machine,
} which retracts a little after some time of inactivity to avoid damage to
} the knife, or if this a the normal consequence of an effect of physics.
} Any comment on this?
}
} Stephane
}
}
}
}
} ________________________________________________________________________
} ____________
} It's here! Your new message!
} Get new email alerts with the free Yahoo! Toolbar.
} http://tools.search.yahoo.com/toolbar/features/mail/
}
} ==============================Original
} Headers==============================
} 6, 19 -- From nizets2-at-yahoo.com Tue Jan 30 04:07:40 2007 6, 19 --
} Received: from web37407.mail.mud.yahoo.com (web37407.mail.mud.yahoo.com
} [209.191.91.139])
} 6, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP
} id l0UA7eBg022570
} 6, 19 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007
} 04:07:40 -0600
} 6, 19 -- Received: (qmail 41697 invoked by uid 60001); 30 Jan 2007
} 10:07:40 -0000 6, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
} 6, 19 -- s=s1024; d=yahoo.com;
} 6, 19 --
} h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Co
} ntent-Transfer-Encoding:Message-ID;
} 6, 19 --
} b=1+gBHYOW2Ee1SsEBAqxWuYWO5/nL8nn468i6i/JHanUAFCuYaxMvYlJD69RYqw8wN8Ck9A
} xdtzwfdWjcEPbq8VT8OKfknMmjNnA6uBWbdWqpeBvEE2CUqDpIOntBmBc9KIeJGijGR+OPFE
} eRXti883DSnoTmaxRoxUIQ3t718kI=;
} 6, 19 -- X-YMail-OSG:
} F4N2dwUVM1kyUlEszoDiJrQ93JszCzNI5jTxknPqdu3C.NSUiFGAGTXveQLO_7tHUG3Bz6Cc
} IkJT4uNj48uba2WMjJiJYFTgohvdDVQ.1dKt8T8WSpk-
} 6, 19 -- Received: from [80.122.101.102] by web37407.mail.mud.yahoo.com
} via HTTP; Tue, 30 Jan 2007 02:07:40 PST 6, 19 -- Date: Tue, 30 Jan 2007
} 02:07:40 -0800 (PST) 6, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
} 6, 19 -- Subject: basic question ultramicrotopy 6, 19 -- To:
} microscopy-at-microscopy.com 6, 19 -- MIME-Version: 1.0 6, 19 --
} Content-Type: text/plain; charset=iso-8859-1 6, 19 --
} Content-Transfer-Encoding: 8bit 6, 19 -- Message-ID:
} {389124.40167.qm-at-web37407.mail.mud.yahoo.com}
} ==============================End of -
} Headers==============================
}
}
}
} ==============================Original Headers==============================
} 17, 26 -- From TindallR-at-missouri.edu Tue Jan 30 08:19:51 2007
} 17, 26 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
} 17, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UEJpVw016138
} 17, 26 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 08:19:51 -0600
} 17, 26 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
} 17, 26 -- Tue, 30 Jan 2007 08:19:50 -0600
} 17, 26 -- x-mimeole: Produced By Microsoft Exchange V6.5
} 17, 26 -- Content-class: urn:content-classes:message
} 17, 26 -- MIME-Version: 1.0
} 17, 26 -- Content-Type: text/plain;
} 17, 26 -- charset="us-ascii"
} 17, 26 -- Subject: RE: [Microscopy] basic question ultramicrotopy
} 17, 26 -- Date: Tue, 30 Jan 2007 08:19:51 -0600
} 17, 26 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D85-at-UM-XMAIL08.um.umsystem.edu}
} 17, 26 -- In-Reply-To: {200701301008.l0UA8QDp024543-at-ns.microscopy.com}
} 17, 26 -- X-MS-Has-Attach:
} 17, 26 -- X-MS-TNEF-Correlator:
} 17, 26 -- Thread-Topic: [Microscopy] basic question ultramicrotopy
} 17, 26 -- Thread-Index: AcdEVpXqhzJvu355StiQwG7GHLeYTAAIprEg
} 17, 26 -- References: {200701301008.l0UA8QDp024543-at-ns.microscopy.com}
} 17, 26 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
} 17, 26 -- To: {nizets2-at-yahoo.com}
} 17, 26 -- Cc: {microscopy-at-microscopy.com}
} 17, 26 -- X-OriginalArrivalTime: 30 Jan 2007 14:19:50.0827 (UTC) FILETIME=[B43237B0:01C74479]
} 17, 26 -- Content-Transfer-Encoding: 8bit
} 17, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0UEJpVw016138
} ==============================End of - Headers==============================
}

****************************************************************************
* Tina (Weatherby) Carvalho * tina-at-pbrc.hawaii.edu *
* Biological Electron Microscope Facility * (808) 956-6251 *
* University of Hawaii at Manoa * http://www.pbrc.hawaii.edu/bemf*
****************************************************************************




==============================Original Headers==============================
10, 19 -- From tina-at-pbrc.hawaii.edu Tue Jan 30 14:59:18 2007
10, 19 -- Received: from halia.pbrc.hawaii.edu (halia.pbrc.hawaii.edu [128.171.22.7])
10, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UKxHKw032484
10, 19 -- for {Microscopy-at-MSA.Microscopy.Com} ; Tue, 30 Jan 2007 14:59:18 -0600
10, 19 -- Received: from halia.pbrc.hawaii.edu (localhost [127.0.0.1])
10, 19 -- by halia.pbrc.hawaii.edu (8.12.11/8.12.11) with ESMTP id l0UKxDEw029309
10, 19 -- (version=TLSv1/SSLv3 cipher=DHE-RSA-AES256-SHA bits=256 verify=NO)
10, 19 -- for {Microscopy-at-MSA.Microscopy.Com} ; Tue, 30 Jan 2007 10:59:14 -1000 (HST)
10, 19 -- Received: from localhost by halia.pbrc.hawaii.edu (8.12.11/8.12.11/Submit) with ESMTP id l0UKxAOc029306
10, 19 -- for {Microscopy-at-MSA.Microscopy.Com} ; Tue, 30 Jan 2007 10:59:13 -1000 (HST)
10, 19 -- X-Authentication-Warning: halia.pbrc.hawaii.edu: tina owned process doing -bs
10, 19 -- Date: Tue, 30 Jan 2007 10:59:10 -1000 (HST)
10, 19 -- From: Tina Carvalho {tina-at-pbrc.hawaii.edu}
10, 19 -- X-Sender: tina-at-halia
10, 19 -- To: Microscopy Listserver {Microscopy-at-MSA.Microscopy.Com}
10, 19 -- Subject: Re: [Microscopy] RE: basic question ultramicrotopy (fwd)
10, 19 -- Message-ID: {Pine.GSO.4.21.0701301045190.28756-100000-at-halia}
10, 19 -- MIME-Version: 1.0
10, 19 -- Content-Type: TEXT/PLAIN; charset=US-ASCII
==============================End of - Headers==============================




From: beaurega-at-westol.com
Date: Tue, 30 Jan 2007 17:18:13 -0600
Subject: [Microscopy] Re: basic question ultramicrotomy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Stephane,

Your question has three answers. The block can move away from the knife
edge, towards it, or stay at the same distance. The main governing factors
are the mechanical advance setting, temperature, and other heating (or
cooling) affects.

Most microtomists would like to believe that a mechanical advance setting
on a microtome determines the final thickness of sections. The mechanical
advance is a guideline or course adjustment of how thick your sections are.
The final thickness is the sum of the positive mechanical advance and the
± thermal advance.
The section's interference color tells you how thick the section really is
as it floats on water. This assumes you are not cutting something like
polyurethanes or high index eyewear lenses which "mess up" or shift the
interference colors on a thickness chart.

Your gap is caused by a discontinuous use of the mechanical advance and
block cutting while thermal affects continue to operate and you sign your
notebook.
If the block, arm or stage are warm and cooling, the gap will be seen as
you described. If the block is cold and warming up, then the thin sections
will be too thick.
If all the temperatures of all the equipment, the air, and your body &
fingers are the same; then the mechanical advance setting will probably be
real.

It was not unusual for me to see the block retreat from the knife edge at
first, later stabilize, and finally advance towards the knife edge during a
thin sectioning session on one block. You have to remember that your body
is giving off heat, you are touching adjustments on the chuck or arc
segment holder, the air temp might be varying, you picked up the diamond
knife and mounted it, you put cooler water in the knife boat, the internal
electronics are heating the cutting arm, water from the boat cools the
knife edge and block face, or maybe you just took off a cryo unit and the
arm and stage are still warming up. These thermal conditions have affects
on thickness' (and your gap).

If you are going to stop cutting for a few minutes, you should always back
up the knife about 5-10 microns and place the arm in the recommended
position. The knife should never be parked in front of the block face.
Failure to do so can result in a block of resin being shattered and/or a
knife being ruined.

One time I was cutting a very long and 0.25 mm wide block of an automotive
coating in cross section. My lab was located on bedrock and the microtome
was on an isolation table. My sections were being cut and I was getting
bright gold sections. I started the next section and it was gold. A guy
walked up to me and stood next to the microtome table but off to the side.
He was three feet away. The middle of that section changed to another
interference color. Just his standing there made a difference in the
interference color on the rest of that cut. I usually told these people to
not move and finished the next two sections and stopped.

Once I had stopped cutting a block and backed up the knife. When I
returned, the block face did not face off uniformly. It would cut a wedge
shape from right to left. That meant the block was deforming under just
the pressure of the clamping and I was using a "hard" Epon 812 clone epoxy
formulation.

The number of variables in use during microtoming of blocks for TEM quality
thin sections is high and sometimes subtle.

AO Reichert sold or gave away at their convention booths, a nice booklet on
microtoming. I think it was called "Ultramicrotomy---Faults and Problems".
It was a blue color and old. Maybe someone on the list can tell you how
to get a copy. It's worth getting but it is not a textbook. It was more
like a reprint of an article.

Microtomy is a learned skill just like riding a bicycle, only harder. Some
people just can't do it very well. Others could and we said they had "the
touch".

Paul


At 04:08 AM 1/30/07 -0600, you wrote:
}
}
}
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


==============================Original Headers==============================
13, 28 -- From beaurega-at-westol.com Tue Jan 30 17:18:13 2007
13, 28 -- Received: from smtp-gateway-6.winbeam.com (smtp-gateway-6.winbeam.com [64.84.97.71])
13, 28 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0UNICRq021519
13, 28 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 17:18:13 -0600
13, 28 -- Received: from mail.winbeam.com (mail.winbeam.com [64.84.96.10])
13, 28 -- by smtp-gateway-6.winbeam.com (8.13.1/8.12.8) with SMTP id l0UNHPLb022098
13, 28 -- for {microscopy-at-microscopy.com} ; Tue, 30 Jan 2007 18:17:26 -0500
13, 28 -- Received: (qmail 19678 invoked by uid 89); 30 Jan 2007 23:17:22 -0000
13, 28 -- Received: from pitts-69-72-117-31.dynamic-dialup.coretel.net (HELO millenium) (69.72.117.31)
13, 28 -- by mail.winbeam.com with SMTP; 30 Jan 2007 23:17:22 -0000
13, 28 -- Message-Id: {3.0.6.32.20070130181732.007f8d60-at-pop3.norton.antivirus}
13, 28 -- X-Sender: beaurega/mail.westol.com-at-pop3.norton.antivirus
13, 28 -- X-Mailer: QUALCOMM Windows Eudora Light Version 3.0.6 (32)
13, 28 -- Date: Tue, 30 Jan 2007 18:17:32 -0500
13, 28 -- To: nizets2-at-yahoo.com, microscopy-at-microscopy.com
13, 28 -- From: Beaurega {beaurega-at-westol.com}
13, 28 -- Subject: Re: [Microscopy] basic question ultramicrotomy
13, 28 -- In-Reply-To: {200701301008.l0UA816n023338-at-ns.microscopy.com}
13, 28 -- Mime-Version: 1.0
13, 28 -- Content-Type: text/plain; charset="iso-8859-1"
13, 28 -- X--MailScanner-Information: - Please contact Technical Support for more information
13, 28 -- X--MailScanner: Found to be clean (courtesy of MailScanner)
13, 28 -- X--MailScanner-SpamCheck: not spam, SpamAssassin (not cached, score=2.044,
13, 28 -- required 4, AWL 0.04, BAYES_50 2.00)
13, 28 -- X--MailScanner-SpamScore: ss
13, 28 -- X--MailScanner-From: beaurega-at-westol.com
13, 28 -- Content-Transfer-Encoding: 8bit
13, 28 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0UNICRq021519
==============================End of - Headers==============================




From: cgarber-at-2spi.com
Date: Tue, 30 Jan 2007 21:18:43 -0600
Subject: [Microscopy] Specifications on microscope slides

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Try http://www.supercircuits.com/ and type PC164C in search box. Sensitivity
will be adequate. Resolution is NTSC with composite video out. Can be viewed
on TV security monitor and/or digitized with inexpensive USB digitizer from
a computer store. It will work on a binocular with eyepiece removed. You may
have to use an adapter for an eyepiece (as for microscope/telescope) if you
care about particular field of view.

I would use instead a suitable C-mount lens and a gooseneck support from a
desk lamp - camera weight including lens is only a couple of ounces.
Position camera next to binoculars, and point it at a (tilted forward) large
viewing screen.

Should fit within $350 budget (camera, lens, power supply, USB digitizer or
a TV security monitor). Will work with some limitations...

Vitaly Feingold
SIA
2773 Heath Lane
Duluth GA 30096
Ph. 770-232-7785
Fax 770-232-1791
www.sia-cam.com
----- Original Message -----
X-from: {opmills-at-mtu.edu}
To: {vitalylazar-at-att.net}
Sent: Tuesday, January 30, 2007 1:06 PM

Karl Garsha wrote:
==============================================
Does anybody happen to know of a reference or resource for the size and
tolerance specifications for a conventional 25x75 mm microscope slide?
Thanks in advance.
===============================================
What is "conventional" in one part of the world is not necessarily the same
as what would be "conventional" in some other part of the world. So-called
standard microscope slides are made both as 1" x 3" and in other instances,
25 x 75 mm. It usually does not make a difference but if in your case it
does, it is less a matter of talking about tolerances than it is in terms of
specifying the right size. Probably the 1" x 3" are more commonly
available.

There are other instances in the microscopy world where we have a similar
kind of "situation". For example, the small "JEOL" SEM mounts, in North
America are specified as 3/8" but in other countries, they are specified as
10 mm. In general, the mounts are interchangeable in the various SEMs but
the problem comes when one wants to put these mounts into storage boxes.
Those made for 3/8" won't accommodate a 10 mm mount, and vice versa.

Chuck

===================================================
Charles A. Garber, Ph. D. Ph: 1-(610)-436-5400
President
SPI SUPPLIES FAX: 1-(610)-436-5755
PO BOX 656 e-mail: cgarber-at-2spi.com
West Chester, PA 19381-0656 USA Cust. Service: spi2spi-at-2spi.com


Look for us!
############################
WWW: http://www.2spi.com
############################
==================================================



==============================Original Headers==============================
8, 26 -- From cgarber-at-2spi.com Tue Jan 30 21:18:43 2007
8, 26 -- Received: from diskless5.axs2000.net (mail.axs2000.net [209.120.196.43])
8, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0V3Ih13015242
8, 26 -- for {microscopy-at-msa.microscopy.com} ; Tue, 30 Jan 2007 21:18:43 -0600
8, 26 -- Received: from yourb27fb1c401 (c-71-230-36-118.hsd1.pa.comcast.net [71.230.36.118])
8, 26 -- (authenticated bits=0)
8, 26 -- by diskless5.axs2000.net (8.12.11/8.12.11) with ESMTP id l0V3IZJZ016718
8, 26 -- for {microscopy-at-msa.microscopy.com} ; Tue, 30 Jan 2007 22:18:42 -0500
8, 26 -- X-IDV-FirstRcvd: c-71-230-36-118.hsd1.pa.comcast.net [71.230.36.118]
8, 26 -- X-IDV-HELO: yourb27fb1c401
8, 26 -- X-IDV-Authenticated-User: cgarber
8, 26 -- Message-ID: {027101c744e6$7f447c30$6401a8c0-at-yourb27fb1c401}
8, 26 -- From: "Garber, Charles A" {cgarber-at-2spi.com}
8, 26 -- To: "Microscopy Listserver" {microscopy-at-msa.microscopy.com}
8, 26 -- Subject: Specifications on microscope slides
8, 26 -- Date: Tue, 30 Jan 2007 22:18:35 -0500
8, 26 -- MIME-Version: 1.0
8, 26 -- Content-Type: text/plain;
8, 26 -- format=flowed;
8, 26 -- charset="iso-8859-1";
8, 26 -- reply-type=original
8, 26 -- Content-Transfer-Encoding: 7bit
8, 26 -- X-Priority: 3
8, 26 -- X-MSMail-Priority: Normal
8, 26 -- X-Mailer: Microsoft Outlook Express 6.00.2900.3028
8, 26 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
==============================End of - Headers==============================




From: beadrysdale-at-yahoo.com
Date: Wed, 31 Jan 2007 08:30:49 -0600
Subject: [Microscopy] viaWWW: Outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both beadrysdale-at-yahoo.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: beadrysdale-at-yahoo.com
Name: Bea Drysdale

Organization: unaffiliated

Title-Subject: [Filtered] Job outlook for microscopy technicians

Question: Hello. I am considering undertaking a two-year associate degree program to become a microscopy technician. I was wondering if anyone would be willing to comment on the job prospects I might have as new graduate with no science experience beyond this associate degree. The college's placement stats. look good but I would like a second opinion. Also, I am in my early forties and wonder if that would work against me finding a job (one of the draws of microscopy is that it is such a specialized skill that I would think employers would look beyond age). Thanks for any help you can give me.

---------------------------------------------------------------------------

==============================Original Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Wed Jan 31 08:30:49 2007
6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VEUm8u012460
6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 08:30:49 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110400c1e65873d8f2-at-[206.69.208.22]}
6, 12 -- Date: Wed, 31 Jan 2007 08:30:46 -0600
6, 12 -- To: microscopy-at-microscopy.com
6, 12 -- From: beadrysdale-at-yahoo.com (by way of MicroscopyListserver)
6, 12 -- Subject: viaWWW: Outlook for microscopy technicians
6, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of - Headers==============================




From: lcgould-at-med.cornell.edu
Date: Wed, 31 Jan 2007 08:55:50 -0600
Subject: [Microscopy] Re: viaWWW: Outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Dear Bea,
I can address your concern from my perspective as the director of a
Core microscopy facility at an Ivy League Med school. When I am in
the market for new staff, it is like a breath of fresh air to find
someone with training. In the past 6 years, I've hired 2 techs, both
of whom were fresh out of college (4-year). Each of them had
exposure to, but no experience in microscopy and needed training from
the ground up. I was fortunate that tech #2 started working while
tech #1 was still here, so that #1 could train #2, with less imput
from me. The first time around, I had to do all of the hands-on
training, and it took me away from much of my other work. Finding
someone with a microscopy background would be a godsend.
Good luck,
Lee Cohen-Gould
--
Leona Cohen-Gould, M.S.
Sr. Staff Associate
Director, Electron Microscopy & Histology Core Facility
Manager, Optical Microscopy Core Facility
Weill-Cornell Medical College

voice (212)746-6146
fax (212)746-8175
http://www.cornellcelldevbiology.org
http://www.cornellbiochem.org

==============================Original Headers==============================
2, 24 -- From lcgould-at-med.cornell.edu Wed Jan 31 08:55:50 2007
2, 24 -- Received: from smtp-gw1.med.cornell.edu (smtp-gw1.med.cornell.edu [140.251.3.74])
2, 24 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VEtnNx024230
2, 24 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 08:55:50 -0600
2, 24 -- Received: from mpx3.med.cornell.edu (biglb-vlan511vip.med.cornell.edu [140.251.11.120])
2, 24 -- by smtp-gw1.med.cornell.edu (Switch-3.1.8/Switch-3.1.7) with ESMTP id l0VEsKA0025044;
2, 24 -- Wed, 31 Jan 2007 09:55:49 -0500 (EST)
2, 24 -- Received: from [140.251.48.23] (mac110773.med.cornell.edu [140.251.48.23])
2, 24 -- by mpx3.med.cornell.edu
2, 24 -- (Sun Java System Messaging Server 6.1 HotFix 0.11 (built Jan 28 2005))
2, 24 -- with ESMTPA id {0JCQ00IUIMSS5K40-at-mpx3.med.cornell.edu} ; Wed,
2, 24 -- 31 Jan 2007 09:55:40 -0500 (EST)
2, 24 -- Date: Wed, 31 Jan 2007 09:55:37 -0500
2, 24 -- From: "Leona Cohen-Gould" {lcgould-at-med.cornell.edu}
2, 24 -- Subject: Re: [Microscopy] viaWWW: Outlook for microscopy technicians
2, 24 -- In-reply-to: {200701311431.l0VEVt1V013623-at-ns.microscopy.com}
2, 24 -- Sender: "Leona Cohen-Gould" {lcgould-at-med.cornell.edu}
2, 24 -- To: beadrysdale-at-yahoo.com, Microscopy Listserver {microscopy-at-microscopy.com}
2, 24 -- Message-id: {p06230904c1e65cff831b-at-[140.251.48.23]}
2, 24 -- MIME-version: 1.0
2, 24 -- Content-type: text/plain; charset=us-ascii; format=flowed
2, 24 -- Content-transfer-encoding: 7BIT
2, 24 -- References: {200701311431.l0VEVt1V013623-at-ns.microscopy.com}
2, 24 -- X-PMX-Version: 5.2.2.285561, Antispam-Engine: 2.5.0.283055, Antispam-Data: 2007.1.31.64433
==============================End of - Headers==============================




From: TindallR-at-missouri.edu
Date: Wed, 31 Jan 2007 09:36:37 -0600
Subject: [Microscopy] viaWWW: Outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Bea,

I agree with Leona that the job possibilities in microscopy are good for
people with experience and training, and I suspect that being in your
40's would not be a significant factor. It may even work in your favor,
since people (at least the ones who have never met me) might assume that
age equals maturity, stability, responsibility, etc.

That said, there is a factor that is seldom discussed, but definitely
exists. While virtually all job advertisements ask for experience and
training, it is possible to end up in a situation where procedures and
attitudes in a lab are so set in stone that much of the training you
bring to a job will have to be thrown out the window. This is sometimes
due to good reasons, i.e., standard operating procedures are in place
for tracking, verification, and certification reasons, or a lab does a
certain type of work and has really worked out the best ways to
efficiently get the job done that leave little room for variation.
Other times this can simply be a result of micromanagement and other
less-noble reasons. In cases like these, it may be that what the lab
REALLY wants is to train someone from the ground up, but they didn't
really know that. It can be a shocker when it happens.

My point is that I would encourage you to get your degree and go for it.
The microscopy world is full of wonderful and rewarding careers. But be
prepared to be flexible and open to suggestion when joining a lab. My
experience is that virtually everybody has their own, sometimes
radically different, ways of doing things and many people don't accept
change easily or quickly. Be prepared to blend in and as time goes on
you may find that your ideas and training will be allowed to express
themselves, or you may find that the people that hire you have better
ideas that you can learn from. Or, and this is where it gets good, you
will likely find that your new associates are interested in the new
ideas and attitudes you bring to the lab, as well as passing along their
own hard-won experience to you.

Best of luck,
Randy

Randy Tindall
Senior EM Specialist
Electron Microscopy Core Facility---We Do Small Well!
W125 Veterinary Medicine
University of Missouri
Columbia, MO 65211
Tel: (573) 882-8304
Fax: (573) 884-2227
Email: tindallr-at-missouri.edu
Web: http://www.emc.missouri.edu
On-line calendar:
http://biotech.rnet.missouri.edu/cgi-bin/Calcium310.pl?Op=Splash&Amount=
Week&NavType=Both&Type=TimePlan


-----Original Message-----
X-from: beadrysdale-at-yahoo.com [mailto:beadrysdale-at-yahoo.com]
Sent: Wednesday, January 31, 2007 8:33 AM
To: Tindall, Randy D.

This Question/Comment was submitted to the Microscopy Listserver using
the WWW based Form at
http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
------------------------------------------------------------------------
---
Remember this posting is most likely not from a Subscriber, so when
replying
please copy both beadrysdale-at-yahoo.com as well as the MIcroscopy
Listserver
------------------------------------------------------------------------
---

Email: beadrysdale-at-yahoo.com
Name: Bea Drysdale

Organization: unaffiliated

Title-Subject: [Filtered] Job outlook for microscopy technicians

Question: Hello. I am considering undertaking a two-year associate
degree program to become a microscopy technician. I was wondering if
anyone would be willing to comment on the job prospects I might have as
new graduate with no science experience beyond this associate degree.
The college's placement stats. look good but I would like a second
opinion. Also, I am in my early forties and wonder if that would work
against me finding a job (one of the draws of microscopy is that it is
such a specialized skill that I would think employers would look beyond
age). Thanks for any help you can give me.

------------------------------------------------------------------------
---

==============================Original
Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Wed Jan 31 08:30:49 2007 6, 12 --
Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0VEUm8u012460
6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007
08:30:49 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110400c1e65873d8f2-at-[206.69.208.22]}
6, 12 -- Date: Wed, 31 Jan 2007 08:30:46 -0600 6, 12 -- To:
microscopy-at-microscopy.com 6, 12 -- From: beadrysdale-at-yahoo.com (by way
of MicroscopyListserver) 6, 12 -- Subject: viaWWW: Outlook for
microscopy technicians 6, 12 -- Content-Type: text/plain;
charset="us-ascii"
==============================End of -
Headers==============================



==============================Original Headers==============================
20, 26 -- From TindallR-at-missouri.edu Wed Jan 31 09:36:37 2007
20, 26 -- Received: from um-tsmtpout1.um.umsystem.edu (um-tsmtpout1.um.umsystem.edu [209.106.228.23])
20, 26 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VFaa9H004002
20, 26 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 09:36:37 -0600
20, 26 -- Received: from UM-XMAIL08.um.umsystem.edu ([209.106.228.34]) by um-tsmtpout1.um.umsystem.edu with Microsoft SMTPSVC(6.0.3790.1830);
20, 26 -- Wed, 31 Jan 2007 09:36:36 -0600
20, 26 -- x-mimeole: Produced By Microsoft Exchange V6.5
20, 26 -- Content-class: urn:content-classes:message
20, 26 -- MIME-Version: 1.0
20, 26 -- Content-Type: text/plain;
20, 26 -- charset="us-ascii"
20, 26 -- Subject: RE: [Microscopy] viaWWW: Outlook for microscopy technicians
20, 26 -- Date: Wed, 31 Jan 2007 09:36:36 -0600
20, 26 -- Message-ID: {91108EF9255B394CBF8B7E3789814A41C68D96-at-UM-XMAIL08.um.umsystem.edu}
20, 26 -- In-Reply-To: {200701311432.l0VEWeNd014462-at-ns.microscopy.com}
20, 26 -- X-MS-Has-Attach:
20, 26 -- X-MS-TNEF-Correlator:
20, 26 -- Thread-Topic: [Microscopy] viaWWW: Outlook for microscopy technicians
20, 26 -- Thread-Index: AcdFRKn5I28aOGsCRn+6fFXfJvTPUgABYlcg
20, 26 -- References: {200701311432.l0VEWeNd014462-at-ns.microscopy.com}
20, 26 -- From: "Tindall, Randy D." {TindallR-at-missouri.edu}
20, 26 -- To: {beadrysdale-at-yahoo.com}
20, 26 -- Cc: {microscopy-at-microscopy.com}
20, 26 -- X-OriginalArrivalTime: 31 Jan 2007 15:36:36.0295 (UTC) FILETIME=[97AE9D70:01C7454D]
20, 26 -- Content-Transfer-Encoding: 8bit
20, 26 -- X-MIME-Autoconverted: from quoted-printable to 8bit by ns.microscopy.com id l0VFaa9H004002
==============================End of - Headers==============================




From: frank.karl-at-degussa.com
Date: Wed, 31 Jan 2007 09:58:20 -0600
Subject: [Microscopy] Re: viaWWW: Outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Bea,
I'm always glad to see people go into microscopy, regardless of their age.
And yes I believe employer looks beyond age as it take years to acquire the
experience and background to be a general purpose microscopist. You must
never stop learning. From this point on, is where I differ from most of
the feed back I suspect you're going to get. It's just my experience and
my opinion. Please send death threats off line to prevent soaking up the
bandwidth.

In todays market I am seeing BSs hired to do the work that technicians did.
I see MSs hired to do the job that chemists with BSs did 20 years ago. It
appears that only PhDs count. This seem to be the trend in large
companies, especial those with government contracts or grants. Many
companies will assume that you can teach microscopy to anyone off the
street because you're just lookin' at stuff. SEM/EDS/TEM/STEM/microtome,
well they are computer run aren't they? Ever manager I know thinks they
can interpret photomicrographs, especially after I annotate them.

I have seem incredible bright technicians, masters of their trade, people
capable of good scientific, independent work locked into positions that
they have outgrown and are actively working at a significantly higher level
but frozen to title and pay grade (and privileges) because they don't have
a BS or better.

I am and remain a microscopist because I love the work and windows to the
world it opens for me. I have had the opportunity to meet and hob nob with
some of the best. It's the best job in the world and I really love my
current position, but I think you should realize that some opportunities
and options will remain closed to you with only a two year degree, just as
some remain closed with a BS or MS.

Stay safe............Frank






beadrysdale-at-yahoo
.com To: frank.karl-at-degussa.com
cc:
01/31/2007 09:32 Subject: [Microscopy] viaWWW: Outlook for microscopy technicians
AM
Please respond to
beadrysdale








----------------------------------------------------------------------------

The Microscopy ListServer -- CoSponsor: The Microscopy Society of America


This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at
http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when
replying
please copy both beadrysdale-at-yahoo.com as well as the MIcroscopy
Listserver
---------------------------------------------------------------------------

Email: beadrysdale-at-yahoo.com
Name: Bea Drysdale

Organization: unaffiliated

Title-Subject: [Filtered] Job outlook for microscopy technicians

Question: Hello. I am considering undertaking a two-year associate degree
program to become a microscopy technician. I was wondering if anyone
would be willing to comment on the job prospects I might have as new
graduate with no science experience beyond this associate degree. The
college's placement stats. look good but I would like a second opinion.
Also, I am in my early forties and wonder if that would work against me
finding a job (one of the draws of microscopy is that it is such a
specialized skill that I would think employers would look beyond age).
Thanks for any help you can give me.

---------------------------------------------------------------------------

==============================Original
Headers==============================
6, 12 -- From zaluzec-at-microscopy.com Wed Jan 31 08:30:49 2007
6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with
ESMTP id l0VEUm8u012460
6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007
08:30:49 -0600
6, 12 -- Mime-Version: 1.0
6, 12 -- X-Sender: (Unverified)
6, 12 -- Message-Id: {p06110400c1e65873d8f2-at-[206.69.208.22]}
6, 12 -- Date: Wed, 31 Jan 2007 08:30:46 -0600
6, 12 -- To: microscopy-at-microscopy.com
6, 12 -- From: beadrysdale-at-yahoo.com (by way of MicroscopyListserver)
6, 12 -- Subject: viaWWW: Outlook for microscopy technicians
6, 12 -- Content-Type: text/plain; charset="us-ascii"
==============================End of -
Headers==============================




==============================Original Headers==============================
29, 18 -- From frank.karl-at-degussa.com Wed Jan 31 09:58:20 2007
29, 18 -- Received: from framailout1.rz.itson.com (mailout2.degussa.com [149.216.91.173])
29, 18 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VFwJhd015467
29, 18 -- for {microscopy-at-msa.microscopy.com} ; Wed, 31 Jan 2007 09:58:20 -0600
29, 18 -- Received: from mobuscomm01.mail.degussa.com (uss1026.applications.degussanet.com [10.88.88.98])
29, 18 -- by framailout1.rz.itson.com (8.13.4/8.13.4/Debian-3sarge3) with SMTP id l0VFwGOM001081;
29, 18 -- Wed, 31 Jan 2007 16:58:16 +0100
29, 18 -- In-Reply-To: {200701311432.l0VEWn4Z014600-at-ns.microscopy.com}
29, 18 -- Subject: Re: [Microscopy] viaWWW: Outlook for microscopy technicians
29, 18 -- To: beadrysdale-at-yahoo.com, microscopy-at-msa.microscopy.com
29, 18 -- X-Mailer: Lotus Notes Release 6.5.5 November 30, 2005
29, 18 -- Message-ID: {OF100DBE80.A5EB0432-ON86257274.0055795A-85257274.0057B982-at-degussa.com}
29, 18 -- From: frank.karl-at-degussa.com
29, 18 -- Date: Wed, 31 Jan 2007 10:58:12 -0500
29, 18 -- X-MIMETrack: Serialize by Router on MOBUSComm01/DHexternal/US(Release 6.5.1|January 21, 2004) at
29, 18 -- 01/31/2007 09:58:18 AM
29, 18 -- MIME-Version: 1.0
29, 18 -- Content-type: text/plain; charset=US-ASCII
==============================End of - Headers==============================




From: TrogadisJ-at-smh.toronto.on.ca
Date: Wed, 31 Jan 2007 10:23:30 -0600
Subject: [Microscopy] Re: viaWWW: Outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi, Bea

This reply is dry and not passionate - don't let that fool you, I
really love microscopy and my job.

When you mention a lack of science background - are you talking about
biological or electronics type of science? If you have no biological
science experience, you could look for work in a clinical lab where
protocols are pretty well established and have to be standardized. If
the research field appeals to you, this may prove to be challenging and
you may want to increase your biology background as much as possible -
for example, to be able to interpret what you see in a microscope and
figure out whether an image is 'real' and significant or whether the
equipment needs a tune-up. It helps if you understand whatever you see.

These days of digital imaging and complex microscope systems, an
understanding of the electronics is also an asset - make sure your
course covers that adequately. Another area of microscopy is in the
material sciences for which biology is not essential. This also requires
some knowledge of different areas of science. Also, there are always
jobs working for one of the microscope manufacturers or retailers.

The age factor is not an issue these days, competence is more
important. There is always a need for this specialty and actually older
people have more experience of having come across a particular problem
or have done a specific technique - so keep your eyes open especially
for troubleshooting tricks.

Let me know if you need any information, encouragement, whatever.
Good luck.
Judy



Title-Subject: [Filtered] Job outlook for microscopy technicians

Question: Hello. I am considering undertaking a two-year associate
degree program to become a microscopy technician. I was wondering if
anyone would be willing to comment on the job prospects I might have
as
new graduate with no science experience beyond this associate degree.
The college's placement stats. look good but I would like a second
opinion. Also, I am in my early forties and wonder if that would work
against me finding a job (one of the draws of microscopy is that it is
such a specialized skill that I would think employers would look
beyond
age). Thanks for any help you can give me.



Judy Trogadis
Bio-Imaging Coordinator
St. Michael's Hospital, 7Queen
30 Bond St.
Toronto, ON M5B 1W8, Canada
ph: 416-864-6060 x6337
pager: 416-685-9219
fax: 416-864-6043
trogadisj-at-smh.toronto.on.ca


==============================Original Headers==============================
14, 19 -- From trogadisj-at-smh.toronto.on.ca Wed Jan 31 10:23:30 2007
14, 19 -- Received: from smh.toronto.on.ca (mail.smh.toronto.on.ca [199.71.175.103])
14, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VGNTjs026960
14, 19 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 10:23:30 -0600
14, 19 -- Received: from ([172.27.12.43])
14, 19 -- by beethoven.smh.toronto.on.ca with ESMTP id KP-GCT47.46619052;
14, 19 -- Wed, 31 Jan 2007 11:23:00 -0500
14, 19 -- Received: from SMTPOUT-MTA by beethoven.smh.toronto.on.ca
14, 19 -- with Novell_GroupWise; Wed, 31 Jan 2007 11:23:00 -0500
14, 19 -- Message-Id: {s5c07c14.028-at-beethoven.smh.toronto.on.ca}
14, 19 -- X-Mailer: Novell GroupWise Internet Agent 6.5.2
14, 19 -- Date: Wed, 31 Jan 2007 11:22:26 -0500
14, 19 -- From: "Judy Trogadis" {TrogadisJ-at-smh.toronto.on.ca}
14, 19 -- To: {Microscopy-at-microscopy.com} , {beadrysdale-at-yahoo.com}
14, 19 -- Subject: [Microscopy] Re: viaWWW: Outlook for microscopy technicians
14, 19 -- Mime-Version: 1.0
14, 19 -- Content-Type: text/plain; charset=US-ASCII
14, 19 -- Content-Transfer-Encoding: 7bit
14, 19 -- Content-Disposition: inline
==============================End of - Headers==============================




From: nizets2-at-yahoo.com
Date: Wed, 31 Jan 2007 10:52:54 -0600
Subject: [Microscopy] Re: basic question ultramicrotopy

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Thanks to all for your answers.
I don't know why but the vast majority didn't take
seriously the possibility that I may stop cutting to
sign autographs. I will remember this ;-)

In a room with air conditioning, all doors closed and
where all conditions are there to offer an optimal
stable environment, I couldn't believe that
temperature changes could have such effects.
Apparently I was wrong, as most of you point this as
the main cause, even if other obscure reasons have
been proposed too (and even if I would tend to believe
the second category more than the first one).
It is strange to see how a so common technique which
delivers such high quality data can be subject to the
same common uncontrollable events.
Just to be fair I'll finish by pointing out that my
Leica ultramicrotome is near from perfection, it is a
real delight to use and cannot be taken responsible
for this problem.

Stephane



____________________________________________________________________________________
Yahoo! Music Unlimited
Access over 1 million songs.
http://music.yahoo.com/unlimited

==============================Original Headers==============================
5, 19 -- From nizets2-at-yahoo.com Wed Jan 31 10:52:54 2007
5, 19 -- Received: from web37405.mail.mud.yahoo.com (web37405.mail.mud.yahoo.com [209.191.91.137])
5, 19 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with SMTP id l0VGqssa006324
5, 19 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 10:52:54 -0600
5, 19 -- Received: (qmail 46674 invoked by uid 60001); 31 Jan 2007 16:52:53 -0000
5, 19 -- DomainKey-Signature: a=rsa-sha1; q=dns; c=nofws;
5, 19 -- s=s1024; d=yahoo.com;
5, 19 -- h=X-YMail-OSG:Received:Date:From:Subject:To:MIME-Version:Content-Type:Content-Transfer-Encoding:Message-ID;
5, 19 -- b=PDWe+l91MqThy2Dbq9YTuf7KOZDi9MH5blhu2dBLVQZnBpegUJfzdB6M5SlH/f2a91ioHgeswqSZWidl5KSshAHdteW8akGsCOgMKUabs7OSmVXaxo8d7z3RVgmt1BEJfwLVWCOnjPyJI3rg4AGUOwp0HTNzWBlFHfehPGBWGMA=;
5, 19 -- X-YMail-OSG: JuRCPLQVM1kcgmhCOzaEisUN77kFIdvNNQ9l4mbYKCm_XR6OlhJnq8Q4D6xA0y5k4iUgWYcm5iJT1rlyy.iWF7EInW0jvi6j7rz2ZM9ZbGKHMOV5mfY-
5, 19 -- Received: from [80.122.101.102] by web37405.mail.mud.yahoo.com via HTTP; Wed, 31 Jan 2007 08:52:53 PST
5, 19 -- Date: Wed, 31 Jan 2007 08:52:53 -0800 (PST)
5, 19 -- From: Stephane Nizet {nizets2-at-yahoo.com}
5, 19 -- Subject: Re: basic question ultramicrotopy
5, 19 -- To: microscopy-at-microscopy.com
5, 19 -- MIME-Version: 1.0
5, 19 -- Content-Type: text/plain; charset=iso-8859-1
5, 19 -- Content-Transfer-Encoding: 8bit
5, 19 -- Message-ID: {748264.46614.qm-at-web37405.mail.mud.yahoo.com}
==============================End of - Headers==============================




From: gwe-at-ufl.edu
Date: Wed, 31 Jan 2007 11:30:39 -0600
Subject: [Microscopy] Re: viaWWW: Outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

One thing I might add to the comments you have already gotten is that
you need to be prepared to relocate, Perhaps clear across the country.
My last hire was living in California and came to work here in Florida
for an entry level job.
Second, lack of experience may be to your advantage since many hires
are at entry level and folks with more experience are going to want more
money. You are also going to find that jobs in the engineering area
will pay higher than those in the biological area.

Greg Erdos
University of Florida, Retired.

beadrysdale-at-yahoo.com wrote:
} ----------------------------------------------------------------------------
} The Microscopy ListServer -- CoSponsor: The Microscopy Society of America
} To Subscribe/Unsubscribe -- http://www.microscopy.com/MicroscopyListserver
} On-Line Help http://www.microscopy.com/MicroscopyListserver/FAQ.html
} ----------------------------------------------------------------------------
}
} This Question/Comment was submitted to the Microscopy Listserver
} using the WWW based Form at http://www.microscopy.org/MicroscopyListserver/MLFormMail.html
} ---------------------------------------------------------------------------
} Remember this posting is most likely not from a Subscriber, so when replying
} please copy both beadrysdale-at-yahoo.com as well as the MIcroscopy Listserver
} ---------------------------------------------------------------------------
}
} Email: beadrysdale-at-yahoo.com
} Name: Bea Drysdale
}
} Organization: unaffiliated
}
} Title-Subject: [Filtered] Job outlook for microscopy technicians
}
} Question: Hello. I am considering undertaking a two-year associate degree program to become a microscopy technician. I was wondering if anyone would be willing to comment on the job prospects I might have as new graduate with no science experience beyond this associate degree. The college's placement stats. look good but I would like a second opinion. Also, I am in my early forties and wonder if that would work against me finding a job (one of the draws of microscopy is that it is such a specialized skill that I would think employers would look beyond age). Thanks for any help you can give me.
}
} ---------------------------------------------------------------------------
}
} ==============================Original Headers==============================
} 6, 12 -- From zaluzec-at-microscopy.com Wed Jan 31 08:30:49 2007
} 6, 12 -- Received: from [206.69.208.22] (mac22.zaluzec.com [206.69.208.22])
} 6, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VEUm8u012460
} 6, 12 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 08:30:49 -0600
} 6, 12 -- Mime-Version: 1.0
} 6, 12 -- X-Sender: (Unverified)
} 6, 12 -- Message-Id: {p06110400c1e65873d8f2-at-[206.69.208.22]}
} 6, 12 -- Date: Wed, 31 Jan 2007 08:30:46 -0600
} 6, 12 -- To: microscopy-at-microscopy.com
} 6, 12 -- From: beadrysdale-at-yahoo.com (by way of MicroscopyListserver)
} 6, 12 -- Subject: viaWWW: Outlook for microscopy technicians
} 6, 12 -- Content-Type: text/plain; charset="us-ascii"
} ==============================End of - Headers==============================
}

==============================Original Headers==============================
3, 27 -- From gwe-at-ufl.edu Wed Jan 31 11:30:39 2007
3, 27 -- Received: from smtp.ufl.edu (smtp01.osg.ufl.edu [128.227.74.149])
3, 27 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VHUcUk018130
3, 27 -- for {Microscopy-at-MSA.Microscopy.Com} ; Wed, 31 Jan 2007 11:30:38 -0600
3, 27 -- Received: from [10.228.0.96] (ssrb-vpn1-0-96.vpn.ufl.edu [10.228.0.96])
3, 27 -- (authenticated bits=0)
3, 27 -- by smtp.ufl.edu (8.13.7/8.13.7/2.5.9) with ESMTP id l0VHUatn1224878;
3, 27 -- Wed, 31 Jan 2007 12:30:36 -0500
3, 27 -- Message-ID: {45C0D25C.4050509-at-ufl.edu}
3, 27 -- Date: Wed, 31 Jan 2007 12:31:08 -0500
3, 27 -- From: Greg Erdos {gwe-at-ufl.edu}
3, 27 -- Reply-To: gwe-at-ufl.edu
3, 27 -- User-Agent: Mozilla Thunderbird 1.0.7 (Windows/20050923)
3, 27 -- X-Accept-Language: en-us, en
3, 27 -- MIME-Version: 1.0
3, 27 -- To: beadrysdale-at-yahoo.com,
3, 27 -- "Microscopy-at-sparc5.microscopy.com" {Microscopy-at-MSA.Microscopy.Com}
3, 27 -- Subject: Re: [Microscopy] viaWWW: Outlook for microscopy technicians
3, 27 -- References: {200701311431.l0VEVtwG013608-at-ns.microscopy.com}
3, 27 -- In-Reply-To: {200701311431.l0VEVtwG013608-at-ns.microscopy.com}
3, 27 -- Content-Type: text/plain; charset=ISO-8859-1; format=flowed
3, 27 -- Content-Transfer-Encoding: 7bit
3, 27 -- X-Greylist: Sender succeeded SMTP AUTH authentication, not delayed by milter-greylist-3.0rc3 (smtp.ufl.edu [128.227.74.56]); Wed, 31 Jan 2007 12:30:37 -0500 (EST)
3, 27 -- X-Spam-Status: hits=-1.44, required=5, tests=ALL_TRUSTED
3, 27 -- X-UFL-Spam-Status: hits=-1.44, required=5, tests=ALL_TRUSTED
3, 27 -- X-Scanned-By: CNS Open Systems Group (http://open-systems.ufl.edu/services/smtp-relay/)
3, 27 -- X-UFL-Scanned-By: CNS Open Systems Group (http://open-systems.ufl.edu/services/smtp-relay/)
==============================End of - Headers==============================




From: iancozine-at-gmail.com
Date: Wed, 31 Jan 2007 14:34:42 -0600
Subject: [Microscopy] viaWWW: Job outlook for microscopy technicians

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

This Question/Comment was submitted to the Microscopy Listserver
using the WWW based Form at http://www.microscopy.com/MLFormMail.html
---------------------------------------------------------------------------
Remember this posting is most likely not from a Subscriber, so when replying
please copy both iancozine-at-gmail.com as well as the MIcroscopy Listserver
---------------------------------------------------------------------------

Email: iancozine-at-gmail.com
Name: Ian Cozine

Organization: MATC

Title-Subject: [Filtered] Job outlook for microscopy technicians

Question: Bea,

I am a 2nd year student at the EM program at MATC in Madison,
Wisconsin and I graduate this May. While I can't add many informed
comments about the outlook for technicians yet, I can say that I have
thoroughly enjoyed the program. It is a small program so there is
plenty of one on one instruction. We get to spend several hours a
week in the lab, and from day one we are working with the
instruments. I think this is one advantage over a B.S. degree. Upon
graduating I will have 2 years working with all sorts of equipment
and techniques and I think that most 4 year degrees aren't going to
provide that much hands on experience.

I do know that the placement rates are quite good. If you graduate
from the program and are reasonably proficient you will get a job.
Our instructors do a really great job of locating job leads for us
all over the country. One consideration, however, is that chances are
you will be moving for your first job.

I would be hard-pressed to think of another 2 year program where upon
graduation you will get to do such interesting work. I would not
hesitate recommending a microscopy degree to anyone.


-Ian

Email: beadrysdale-at-yahoo.com
Name: Bea Drysdale

Organization: unaffiliated

Title-Subject: [Filtered] Job outlook for microscopy technicians

Question: Hello. I am considering undertaking a two-year associate
degree program to become a microscopy technician. I was wondering
if anyone would be willing to comment on the job prospects I might
have as new graduate with no science experience beyond this associate
degree. The college's placement stats. look good but I would like a
second opinion. Also, I am in my early forties and wonder if that
would work against me finding a job (one of the draws of microscopy
is that it is such a specialized skill that I would think employers
would look beyond age). Thanks for any help you can give me.

---------------------------------------------------------------------------

==============================Original Headers==============================
15, 12 -- From zaluzec-at-microscopy.com Wed Jan 31 14:34:42 2007
15, 12 -- Received: from [146.139.72.120] (msdvpn8.msd.anl.gov [130.202.238.72])
15, 12 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VKYel2001341
15, 12 -- for {microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 14:34:41 -0600
15, 12 -- Mime-Version: 1.0
15, 12 -- X-Sender: (Unverified)
15, 12 -- Message-Id: {p06020400c1e6aede9b65-at-[207.42.54.42]}
15, 12 -- Date: Wed, 31 Jan 2007 14:41:10 -0600
15, 12 -- To: microscopy-at-microscopy.com
15, 12 -- From: iancozine-at-gmail.com (by way of MicroscopyListserver)
15, 12 -- Subject: viaWWW: Job outlook for microscopy technicians
15, 12 -- Content-Type: text/plain; charset="us-ascii" ; format="flowed"
==============================End of - Headers==============================




From: johnwillis-at-valcor.com
Date: Wed, 31 Jan 2007 14:45:47 -0600
Subject: [Microscopy] Camera

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hello:
I use a 3.2 Megapixel Digital camera PN OMCAM32 from The Microscope Store
for OPTICAL microscopy with great results. The camera mounts thru the
eyepiece and has an LCD so you can see exactly what you are getting. The
brighter the light the better the image but I get images in dimmer light
that I can use. The pictures are stored on an internal memory card which is
then downloaded to a PC via cable. It requires no special software for
download.

I talked to Eric Naff. The OMCAM32 is not listed on the web site, but the
store will reply "yes" to an e-mail request regarding it.

The Microscope Store
316 Windy Pines Lane
Rocky Mount, VA 24151
Voice (877) 409-3556
FAX (540) 489-4785
2rhonda-at-microscope-store.com
www.microscope-store.com

I hope this helps,

John R Willis
Tech Writer
Valcor Eng Corp
2 Lawrence Rd
Springfield, NJ 07081
Voice (973) 467-8400 X 7257
FAX (973) 467-8382
johnwillis-at-valcor.com
www.valcor.com


==============================Original Headers==============================
6, 32 -- From johnwillis-at-valcor.com Wed Jan 31 14:45:47 2007
6, 32 -- Received: from mail.valcor.com (mail.valcor.com [12.30.108.82])
6, 32 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l0VKjlWA011519
6, 32 -- for {Microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 14:45:47 -0600
6, 32 -- Received: from mail.valcor.com (localhost.localdomain [127.0.0.1])
6, 32 -- by mail.valcor.com (8.13.7/8.13.4) with ESMTP id l0VKmG60026624
6, 32 -- for {Microscopy-at-microscopy.com} ; Wed, 31 Jan 2007 15:48:17 -0500
6, 32 -- Received: from JohnWillis ([192.168.1.68])(authenticated bits=0)by
6, 32 -- mail.valcor.com (8.13.7/8.13.4) with ESMTP id l0VKli7W026491;Wed, 31 Jan
6, 32 -- 2007 15:47:55 -0500
6, 32 -- Message-ID: {024501c74578$b585b7e0$4401a8c0-at-valcor.local}
6, 32 -- From: "John Willis" {johnwillis-at-valcor.com}
6, 32 -- To: {Microscopy-at-microscopy.com}
6, 32 -- Cc: {johnwillis-at-valcor.com}
6, 32 -- Subject: Camera
6, 32 -- Date: Wed, 31 Jan 2007 15:45:04 -0500
6, 32 -- MIME-Version: 1.0
6, 32 -- Content-Type: text/plain;
6, 32 -- format=flowed;
6, 32 -- charset=iso-8859-1;
6, 32 -- reply-type=original
6, 32 -- Content-Transfer-Encoding: 7bit
6, 32 -- X-Priority: 3
6, 32 -- X-MSMail-Priority: Normal
6, 32 -- X-Mailer: Microsoft Outlook Express 6.00.2900.3028
6, 32 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2900.3028
6, 32 -- X-imss-version: 2.045
6, 32 -- X-imss-result: Passed
6, 32 -- X-imss-scanInfo: M:P L:E SM:0
6, 32 -- X-imss-tmaseResult: TT:0 TS:0.0000 TC:00 TRN:0 TV:3.6.1039(14870.003)
6, 32 -- X-imss-scores: Clean:57.20883 C:2 M:3 S:5 R:5
6, 32 -- X-imss-settings: Baseline:4 C:3 M:3 S:3 R:3 (1.0000 1.0000)
==============================End of - Headers==============================




From: john.brealey-at-imvs.sa.gov.au
Date: Wed, 31 Jan 2007 19:12:51 -0600
Subject: [Microscopy] Ultratsructure of Dendritic Cells

Contents Retrieved from Microscopy Listserver Archives
http://www.microscopylistserver.com/MicroscopyListserver/MicroscopyArchives.html

Hi Listers,

Are there any ultrastructural differences between monocytes and
monocyte-derived dendritic cells?
I've been looking at marmoset monocytes, immature dendritic cells and
matrure dendritic cells and am having difficulties differentiating between
them at the ultrastructural level.

Are there any references available to view images of these cells?

Regards,

John Brealey
EM Unit
Queen Elizabeth Hospital
Adelaide
South Australia


==============================Original Headers==============================
6, 33 -- From john.brealey-at-imvs.sa.gov.au Wed Jan 31 19:12:51 2007
6, 33 -- Received: from adl0741.systems.sa.gov.au (adl0741.systems.sa.gov.au [143.216.236.20])
6, 33 -- by ns.microscopy.com (8.12.11.20060308/8.12.8) with ESMTP id l111Cnln029344
6, 33 -- for {Microscopy-at-MSA.Microscopy.Com} ; Wed, 31 Jan 2007 19:12:50 -0600
6, 33 -- Received: from adl0741.systems.sa.gov.au (localhost [127.0.0.1])
6, 33 -- by adl0741.systems.sa.gov.au OUTGOING (8.12.10/8.12.10) with ESMTP id l111Clji025278
6, 33 -- for {Microscopy-at-MSA.Microscopy.Com} ; Thu, 1 Feb 2007 11:42:47 +1030 (CST)'
6, 33 -- Received: from ablett.imvs.sa.gov.au (ablett.imvs.sa.gov.au [10.20.98.41])
6, 33 -- by adl0741.systems.sa.gov.au INCOMING (8.12.10/8.12.10) with ESMTP id l111ClR6025275
6, 33 -- for {Microscopy-at-MSA.Microscopy.Com} ; Thu, 1 Feb 2007 11:42:47 +1030 (CST)'
6, 33 -- Received: from localhost (mesh.imvs.sa.gov.au [10.20.98.37])
6, 33 -- by ablett.imvs.sa.gov.au (Postfix) with ESMTP id 9A4AC34DA4
6, 33 -- for {Microscopy-at-MSA.Microscopy.Com} ; Thu, 1 Feb 2007 11:42:47 +1030 (CST)
6, 33 -- X-Virus-Scanned: by amavisd-new at imvs.sa.gov.au
6, 33 -- Received: from ablett.imvs.sa.gov.au ([10.20.98.41])
6, 33 -- by localhost (.imvs.sa.gov.au [10.20.98.37]) (amavisd-new, port 10024)
6, 33 -- with LMTP id 5Q4x0tFg1s63 for {Microscopy-at-MSA.Microscopy.Com} ;
6, 33 -- Thu, 1 Feb 2007 11:42:41 +1030 (CST)
6, 33 -- Received: from iqe36042 (iqepc200.imvs.sa.gov.au [10.20.138.200])
6, 33 -- by ablett.imvs.sa.gov.au (Postfix) with SMTP id B38F234DB4
6, 33 -- for {Microscopy-at-MSA.Microscopy.Com} ; Thu, 1 Feb 2007 11:42:41 +1030 (CST)
6, 33 -- From: "John Brealey" {john.brealey-at-imvs.sa.gov.au}
6, 33 -- To: "Listserver" {Microscopy-at-MSA.Microscopy.Com}
6, 33 -- Subject: Ultratsructure of Dendritic Cells
6, 33 -- Date: Thu, 1 Feb 2007 11:42:40 +1030
6, 33 -- Message-ID: {000001c7459e$125729c0$c88a140a-at-iqe36042}
6, 33 -- MIME-Version: 1.0
6, 33 -- Content-Type: text/plain;
6, 33 -- charset="us-ascii"
6, 33 -- Content-Transfer-Encoding: 7bit
6, 33 -- X-Mailer: Microsoft Office Outlook 11
6, 33 -- Thread-Index: AcdFnflmKAUnplW8Rb6sBxBTxsNV+A==
6, 33 -- X-MimeOLE: Produced By Microsoft MimeOLE V6.00.2800.1807
==============================End of - Headers==============================






MicroscopyListserver Archive Email Extraction Software Version NJZ07060908

Return to Microscopy Listserver Home Page


Return to MSA HomePage